K Number
K151387
Manufacturer
Date Cleared
2015-06-24

(29 days)

Product Code
Regulation Number
882.5360
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Stryker Universal Neuro III System is intended for reconstruction and/or rigid fixation of non loadbearing bony areas subsequent to craniectomy and cranial fractures in adults and adolescents (age 12 and higher).

Device Description

The Stryker Universal Neuro III System (UN III) is intended for reconstruction, stabilization and/or rigid fixation of non load-bearing areas subsequent to craniotomy, craniectomy and cranial fractures in adults and adolescents (age 12 or higher). The UN III System consists of an implant module for the respective anatomical and indicated areas with each containing various screw, and plate versions and shapes. This special 510(k) is being submitted due to dimensional modifications to the screw and screwdriver blade. There have been no modifications to the plates, meshes, or other accessory devices.

AI/ML Overview

This document is a 510(k) premarket notification for the Stryker Universal Neuro III System: UN III AXS Screw and UN III AXS Screwdriver Blade. It focuses on demonstrating substantial equivalence to a predicate device following dimensional modifications. It is therefore a submission for a medical device and not for an AI/ML powered device. Hence, most of the questions relating to AI/ML powered devices are not relevant for this submission, and the information is not present in the document.

Here's the breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

The document states that "The subject device met all pre-defined acceptance criteria." However, the specific quantitative acceptance criteria for each test are not explicitly detailed. Instead, only the types of tests performed are listed.

Test PerformedReported Device Performance
Self-Retention TestMet all pre-defined acceptance criteria.
Screw Insertion TestMet all pre-defined acceptance criteria.
Life Cycle Test Fatigue (Screwdriver Blade)Met all pre-defined acceptance criteria.
Torque to Failure Test Quasi Static (Screwdriver Blade)Met all pre-defined acceptance criteria.
End User / End Product TestMet all pre-defined acceptance criteria.
Biocompatibility testingNot required (no change in material, manufacturing, duration/location of contact).
Sterility testingNot required (no change in material, manufacturing, duration/location of contact).

2. Sample Size Used for the Test Set and Data Provenance

This is not applicable as this is a medical device submission, not an AI/ML device. The document details bench testing, not a dataset for an algorithm.

3. Number of Experts Used to Establish Ground Truth and Qualifications

This is not applicable as this is a medical device submission, not an AI/ML device. Ground truth as typically defined for AI/ML models is not relevant here.

4. Adjudication Method

This is not applicable as this is a medical device submission, not an AI/ML device.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No MRMC study was done, as this is a medical device submission and not an AI/ML powered medical device.

6. Standalone (Algorithm Only) Performance

Not applicable. This is a physical medical device (screws and screwdriver blades), not an algorithm.

7. Type of Ground Truth Used

Not applicable. The performance was assessed through physical bench tests. For these tests, the "ground truth" would be the engineering specifications and performance standards that the device is designed to meet.

8. Sample Size for the Training Set

Not applicable. This is a physical medical device, not an AI/ML powered device, so there is no "training set."

9. How the Ground Truth for the Training Set Was Established

Not applicable. There is no training set for this type of medical device submission.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol of a human figure, represented by three overlapping profiles facing to the right, with flowing lines suggesting movement or connection.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 24, 2015

Stryker Craniomaxillofacial Mr. Jonathan Schell Senior Regulatory Compliance Specialist 750 Trade Centre Way, Suite 200 Portage, Michigan 49002

Re: K151387

Trade/Device Name: Stryker Universal Neuro III System: UN III AXS Screw. UN III AXS Screwdriver Blade Regulation Number: 21 CFR 882.5360 Regulation Name: Cranioplasty Plate Fastener Regulatory Class: Class II Product Code: HBW, GWO, GXR Dated: Mav 22, 2015 Received: May 26, 2015

Dear Mr. Schell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carlos L. Pena -SFD/Δ

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

Device Name

Stryker Universal Neuro III System: UN III AXS Screw, UN III AXS Screwdriver Blade

Indications for Use (Describe)

The Stryker Universal Neuro III System is intended for reconstruction and/or rigid fixation of non loadbearing bony areas subsequent to craniectomy and cranial fractures in adults and adolescents (age 12 and higher).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The word is in black and is set against a white background. A small registered trademark symbol is located to the right of the word.

Section 5. 510(k) Summary

This section provides a summary of 510(k) information in accordance with the requirements of 21 CFR 807.92.

I. SUBMITTER

510(k) Owner:Stryker Leibinger GmbH& Co. KGBoetzinger Strasse 41D-79111 Freiburg, Germany
Submitter/ ContactPerson:Jonathan SchellSr. Regulatory Affairs SpecialistStryker Craniomaxillofacial750 Trade Centre WayPortage, MI 49002Phone: 269-389-5596Fax: 877-648-7114
Date prepared:May 22, 2015
II.DEVICE
Trade Name:Stryker Universal Neuro III System (also referred to as UniversalNeuro 3 System): UN III AXS Screw, UN III AXS Screwdriver Blade
Common or Usualname:Neuro Plating System
Classificationname:Cranioplasty Plate Fastener 21 CFR §882.5360Preformed alterable cranioplasty plate 21 CFR §882.5320Burr hole cover 21 CFR §882.5250
Regulatory Class:Class II
Product Code:HBW, GWO, GXR

{4}------------------------------------------------

III. PREDICATE DEVICE

Predicate: Stryker Universal Neuro III System - K112557

This predicate has not been subject to a design-related recall.

IV. DEVICE DESCRIPTION

The Stryker Universal Neuro III System (UN III) is intended for reconstruction, stabilization and/or rigid fixation of non load-bearing areas subsequent to craniotomy, craniectomy and cranial fractures in adults and adolescents (age 12 or higher). The UN III System consists of an implant module for the respective anatomical and indicated areas with each containing various screw, and plate versions and shapes. This special 510(k) is being submitted due to dimensional modifications to the screw and screwdriver blade. There have been no modifications to the plates, meshes, or other accessory devices. The following Table 5-1 lists the devices subject to this Special 510(k):

Article NumberDescription
62-15035AXS Screwdriver Blade Long
62-15036AXS Screwdriver Blade Short
56-1503456-1503556-1503656-1593356-1593456-1593556-17334AXS Screw
29-5603429-5603529-5693329-5693429-56935Preloaded AXS Screw Disc

TABLE 5-1 - LIST OF SUBJECT DEVICES

{5}------------------------------------------------

V. INDICATIONS FOR USE

The Stryker Universal Neuro III System is intended for reconstruction, stabilization and/or rigid fixation of non load-bearing bony areas subsequent to craniotomy, craniectomy and cranial fractures in adults and adolescents (age 12 and higher).

The proposed modifications do not alter the Indications for Use statement for the proposed device. The Indications for Use is identical to the predicate device Stryker Universal Neuro III System (K112557).

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE

The modified Universal Neuro III System is compared to its predicate device for substantial equivalence of technological characteristics based on the following criteria:

  • A. Principle of Operation
  • B. Technological Characteristics

A. Principle of Operation

The basic operational principle of the Stryker Universal Neuro III System (UN III) remains the same: the operating principle for the UN III system is to reconstruct, stabilize and/or provide rigid fixation of non load-bearing areas. Subsequently, the principle of operation of the modified screw device and the screwdriver blade accessory remain the same as the predicate device.

B. Technological Characteristics

The dimensional modification described in this special 510(k) is to the UN III System screw and screwdriver blade accessory. However, the modifications to the UN III screw and screwdriver blade accessory do not alter the technological and operational characteristics of the screw and the screwdriver blade accessory. The technological characteristics remain the same:

  • Same operating principle: reconstruct, stabilize and/or provide rigid fixation of non । load-bearing bony areas
  • Same mode of fixation: plate fixation with screws -
  • Same area of contact and contact duration: the screws have contact to tissue/bone with a duration greater than 30 days; the blades, at worst, may contact breached or compromised surface for less than 24 hours
  • Same material: Screw made from titanium alloy (ASTM F136) the blades are made from stainless steel (ASTM F899),

{6}------------------------------------------------

  • -Similar design: The modifications to the UN III screw and screwdriver blade accessory are dimensional specifications.

VII. PERFORMANCE DATA

Based on the Risk Analysis performed on the modifications to the UN III screw device and the modification to the screwdriver blade accessory, bench testing was performed in support of the substantial equivalence determination.

Biocompatibility and sterility testing was not required as a basis for substantial equivalence. There is no change in the material, manufacturing process, duration or location of contact, or reprocessing methods.

Performance Bench Testing

The following performance bench tests were completed.

  • -Self-Retention Test
  • Screw Insertion Test -
  • Life Cycle Test Fatigue (Screwdriver Blade) -
  • Torque to Failure Test Quasi Static (Screwdriver Blade) -
  • -End User / End Product Test

The subject device met all pre-defined acceptance criteria. Overall, the results of the performance bench tests support the proposed substantial equivalence of the subject device.

Animal Testing

Animal testing was not required as a basis for substantial equivalence.

Clinical Testing

Clinical testing was not required as a basis for substantial equivalence.

VIII. CONCLUSIONS

The results of the non-clinical data demonstrate that the modified UN III screw and screwdriver blade will perform as intended in the specified use conditions. According to the comparison based on the requirements of 21 CFR 807.87 and the information provided herein, it is concluded that the information included in this submission supports substantial equivalence.

§ 882.5360 Cranioplasty plate fastener.

(a)
Identification. A cranioplasty plate fastener is a screw, wire, or other article made of tantalum, vitallium, or stainless steel used to secure a plate to the patient's skull to repair a skull defect.(b)
Classification. Class II (performance standards).