(233 days)
The Amendia Lumbar Interbody Fusion Devices are indicated for intervertebral body spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radios. These DDD patients may also have up to Grade I spondylolisthesis at the involved level. Amendia Lumbar Interbody Fusion Devices are to be used with autogenous bone graft and supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.
The Amendia Interbody Fusion Devices are used to provide structural stability and maintain disc space distraction in skeletally mature adults requiring intervertebral body fusion. They are designed to be used in conjunction with supplemental spinal fixation instrumentation. The Subject Amendia Interbody Fusion Devices (LLIF and OLLIF) are multiple component systems comprised of sterile, single-use implants, designed to treat the lumbar spine.
The Subject Amendia Interbody Fusion System lumbar (LLIF and OLLIF) implants are fabricated from PEEK (ASTM F2026) with tantalum (ASTM F560) x-ray markers, or Titanium alloy (Ti6Al4V ELI, ASTM F136). The Amendia Interbody Fusion System implants are available in a range of sizes and shapes, and are designed to accommodate variations in surgical approach and patient anatomy. Each cage has a hollow center to allow placement of autograft. Ridges on the superior and inferior surfaces of the device help to grip the endplates and prevent expulsion.
This document describes the Amendia Interbody Fusion Devices and their substantial equivalence to predicate devices, focusing on the mechanical testing performed. However, it does not contain information regarding an AI/ML device study or performance against acceptance criteria in the context of an AI/ML device.
Therefore, I cannot provide the requested information in the format of acceptance criteria for an AI/ML device, as the document is about a physical medical device (intervertebral body fusion device) and its mechanical performance testing.
The document discusses:
- Device Description: Amendia Interbody Fusion Devices, including materials (PEEK, tantalum, Titanium alloy) and design features (hollow center for autograft, ridges for grip).
- Indications for Use: Intervertebral body spinal fusion procedures for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1, with or without Grade I spondylolisthesis/retrolisthesis, used with autogenous bone graft and supplemental fixation, and after at least six months of non-operative treatment.
- Performance Testing: Non-clinical mechanical testing, including static and dynamic axial compression and compression shear (ASTM F2077), subsidence (ASTM F2267), and expulsion testing.
- Conclusion: The device was shown to be substantially equivalent to predicate devices based on design, intended use, material composition, function, and mechanical performance.
To directly answer your numbered points regarding an AI/ML device study, based on the provided text:
- A table of acceptance criteria and the reported device performance: Not applicable. The document describes mechanical test standards (ASTM F2077, ASTM F2267) but does not provide a table of acceptance criteria and detailed performance results for these. It just states that performance testing "demonstrated the Subject Device is substantially equivalent to the predicate device."
- Sample sized used for the test set and the data provenance: Not applicable for an AI/ML device. For the mechanical testing, the document states "Non-clinical mechanical testing for the Subject Device was performed on the worst case subject device," implying a limited number of physical samples for mechanical tests, but specific numbers are not given.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable.
- Adjudication method for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
- If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
- The type of ground truth used: Not applicable for an AI/ML device. For the physical device, "ground truth" would be established by the physical testing results and engineering standards.
- The sample size for the training set: Not applicable.
- How the ground truth for the training set was established: Not applicable.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, stacked on top of each other, representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 6, 2016
Amendia, Incorporated Ms. Kristen Allen Senior Regulatory Affairs Specialist 1755 West Oak Parkway Marietta, Georgia 30062
Re: K151310
Trade/Device Name: Amendia Interbody Fusion Devices Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: December 7, 2015 Received: December 8, 2015
Dear Ms. Allen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known)
K151310
Device Name Amendia Interbody Fusion Devices
Indications for Use (Describe)
The Amendia Lumbar Interbody Fusion Devices are indicated for intervertebral body spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radios. These DDD patients may also have up to Grade I spondylolisthesis at the involved level. Amendia Lumbar Interbody Fusion Devices are to be used with autogenous bone graft and supplemental fixation. Patients should have at least six (6) months of non-operative treatment with an intervertebral cage.
Type of Use (Select one or both, as applicable)
| ý Prescription Use (Part 21 CFR 801 Subpart D) |
|---|
| □ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for Amendia. The logo features the word "AMENDIA" in a bold, sans-serif font, with the "A" in a blue square. Below the word "AMENDIA" is the tagline "creating balanced solutions" in a smaller, lighter font. The overall design is clean and modern.
510(k) Summary
Amendia Interbody Fusion Devices
| Submitter: | Amendia, Inc.1755 W. Oak ParkwayMarietta, GA 30062 | |
|---|---|---|
| Contact Person: | Kristen AllenSr. Regulatory Affairs Specialist910-612-4153 (P), 877-420-1213 (F)kallen@amendia.com (e-mail) | |
| Date Prepared: | January 5, 2016 | |
| Trade Name: | Amendia Interbody Fusion Devices | |
| Common Name: | Intervertebral body fusion device | |
| Device Product Codeand Classification: | MAX, 888.3080, Class II, Intervertebral FusionDevice with Bone Graft, Lumbar | |
| Primary Predicate Device: | Zeus Intervertebral Fusion Devices (K081614) | |
| Additional Predicate Devices: | Nuvasive CoRoent System (K141665, K151472)Amendia IBFD (K151322)Medtronic Perimeter Interbody Fusion Device (K132700)Biomet Lateral Spacer System (K122989)Custom Spine Pathway (K080281) |
Purpose of Submission: This 510(k) is intended to add additional LLIF (lateral) and OLLIF (obligue) implant models to the family of lumbar implants originally cleared in K081614.
Device Description:
The Amendia Interbody Fusion Devices are used to provide structural stability and maintain disc space distraction in skeletally mature adults requiring intervertebral body fusion. They are designed to be used in conjunction with supplemental spinal fixation instrumentation. The Subject Amendia Interbody Fusion Devices (LLIF and OLLIF) are multiple component systems comprised of sterile, single-use implants, designed to treat the lumbar spine.
The Subject Amendia Interbody Fusion System lumbar (LLIF and OLLIF) implants are fabricated from PEEK (ASTM F2026) with tantalum (ASTM F560) x-ray markers, or Titanium alloy (Ti6Al4V ELI, ASTM F136). The Amendia Interbody Fusion System implants are available in a range of sizes and shapes, and are designed to accommodate variations in surgical approach and patient anatomy. Each cage has a hollow center to allow placement of autograft. Ridges on the superior and inferior surfaces of the device help to grip the endplates and prevent expulsion.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for AMENDIA. The logo is in blue and consists of the word "AMENDIA" in a sans-serif font. The "A" is in a square. Below the word "AMENDIA" is the tagline "creating balanced solutions" in a smaller font.
Indications and Intended use:
The Amendia Lumbar Interbody Fusion Devices are indicated for intervertebral body spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one level or two contiquous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level. Amendia Lumbar Interbody Fusion Devices are to be used with autogenous bone graft and supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.
Summary of Technological Characteristics:
The subject devices are substantially equivalent to the predicate devices as well as other similar devices cleared by FDA for commercial distribution in the United States. The Subject Device was shown to have the same technological characteristics as its predicate devices through comparison of characteristics including design, intended use, material composition, and function, Both the subject and predicate lumbar devices are interbody devices designed to contain graff material and facilitate fusion between two vertebral bodies in the lumbar region of the spine.
Summary of Performance Testing:
Non-clinical mechanical testing for the Subject Device was performed on the worst case subject device. Testing included static and dynamic axial compression and compression shear (ASTM F2077), subsidence (ASTM F2267), and expulsion testing. Performance testing demonstrated the Subject Device is substantially equivalent to the predicate device.
Conclusion:
Based on the comparison to predicate devices, the Subject Device has been shown to be substantially equivalent to legally marketed predicate devices.
N/A