(221 days)
The powered Electric Double Breast Pumps are intended to express and collect milk from the breast of a lactating woman. They are double pumps with a single pumping option and intended for single user.
The Electric Double Breast Pumps are designed and manufactured to express and collect milk from a lactating woman's breast. Both models, XB-8636 and XB-8703, are electrically powered, softwarecontrolled, digital single-user pumps.
When powered, the XB-8636 starts with stimulation mode for 120 seconds then automatically switches to expression mode for 30 minutes before automatic shutdown. This model has seven suction levels. When powered, the XB-8703 starts with stimulation mode for 180 seconds then automatically switches to expression mode for 30 minutes before automatic shutdown. This model has nine suction levels. Both models allow the user to adjust the suction levels when necessary.
This document is a 510(k) premarket notification for an Electric Double Breast Pump. It does not describe a study involving humans or AI for diagnostic purposes. Therefore, most of the requested information regarding acceptance criteria, study details, and AI performance will not be applicable.
However, I can extract the non-clinical acceptance criteria and the summary of non-clinical tests performed to meet those criteria.
1. A table of acceptance criteria and the reported device performance
Based on the document, the acceptance criteria are implicit in compliance with recognized standards and design specifications. The performance reported is that the device met these specifications and standards.
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| Safety and Essential Performance (Electrical) | Complies with IEC 60601-1: 2005 + CORR.1 (2006) + CORR.2 (2007) + AM1 (2012) |
| Home Healthcare Environment Compliance | Complies with IEC 60601-1-11: 2010 |
| Electromagnetic Compatibility (EMC) | Complies with IEC 60601-1-2: 2007 |
| Suction Curves | All test results complied with the design specifications of the proposed devices. |
| Backflow Prevention | No backflow occurred into the tubing or pump motor during testing, even if the bottle was over-filled. |
| Use-Life/Durability | No significant difference in suction strength or simulating milk suction flow rate between new-manufactured devices and devices after two-year use and 1000 cycles' disinfection. |
| Substantial Equivalence to Predicate Device | The differences in suction level, suction strength, cycle speed, and suction flow rate do not raise different questions of safety and effectiveness, and performance testing showed no adverse effects. |
2. Sample sized used for the test set and the data provenance
The document does not specify sample sizes for non-clinical tests (e.g., how many units were tested for use-life, backflow, or suction curves). The tests are entirely non-clinical/bench tests. Data provenance is implied to be from the manufacturer's internal testing in China (Wuxi Xinzhongrui Baby Supplies Co., Ltd).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This device is a breast pump, and the evaluation involves engineering and safety standards, not medical diagnostic "ground truth" established by experts in the context of a clinical study or AI performance evaluation.
4. Adjudication method for the test set
Not applicable, as this is not a study involving human or AI interpretation of data requiring adjudication.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No. This document pertains to the regulatory clearance of a physical medical device (breast pump) and does not involve AI or human readers for diagnostic interpretation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
No. This document does not describe the evaluation of an algorithm.
7. The type of ground truth used
For the non-clinical tests performed:
- Engineering specifications and recognized international standards: This serves as the "ground truth" or benchmark against which the device's performance was measured for safety and efficacy from an engineering perspective. For example, IEC standards define acceptable limits for electrical safety and EMC.
- Design Specifications: The manufacturer's own design specifications for parameters like suction curves, backflow prevention, and use-life were used.
8. The sample size for the training set
Not applicable. This is not an AI/machine learning study.
9. How the ground truth for the training set was established
Not applicable. This is not an AI/machine learning study.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized eagle or bird in flight, composed of three interconnected human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 21, 2015
Wuxi Xinzhongrui Baby Supplies Co., Ltd % Diana Hong General Manager Mid-link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 CN
Re: K151284
Trade/Device Name: Electric Double Breast Pumps Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: Class II Product Code: HGX Dated: November 17, 2015 Received: November 23, 2015
Dear Diana Hong,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K151284
Device Name Electric Double Breast Pumps
Indications for Use (Describe)
The powered Electric Double Breast Pumps are intended to express and collect milk from the breast of a lactating woman. They are double pumps with a single pumping option and intended for single user.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary K151284
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
-
- Date of Preparation: 12/07/2015
-
- Sponsor Identification
Wuxi Xinzhongrui Baby Supplies Co., Ltd No.117 Xinhua Road, Meicun, New District, Wuxi, Jiangsu, 214000, China
Contact Person: Shaozhong Chen Position: General Manager Tel: 86 510 81155228 Fax: 86 510 81155918 Email: jswxxzr@163.com
-
- Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Betty Xiao (Alternative Contact Person)
- Designated Submission Correspondent
Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
-
- Identification of Proposed Device
Trade Name: Electric Double Breast Pumps Common Name: Powered breast pump Models: XB-8636, XB-8703
- Identification of Proposed Device
Regulatory Information Classification Name: Pump, Breast, Powered; Classification: II; Product Code: HGX; Regulation Number: 21 CFR 884.5160; Review Panel: Obstetrics/Gynecology;
Intended Use Statement: The powered Electric Double Breast Pumps are intended to express and collect milk from the breast of a lactating woman. They are double pumps with a single pumping option and intended for single user.
Device Description
The Electric Double Breast Pumps are designed and manufactured to express and collect milk from a lactating woman's breast. Both models, XB-8636 and XB-8703, are electrically powered, softwarecontrolled, digital single-user pumps.
When powered, the XB-8636 starts with stimulation mode for 120 seconds then automatically switches to expression mode for 30 minutes before automatic shutdown. This model has seven suction levels. When powered, the XB-8703 starts with stimulation mode for 180 seconds then automatically switches to expression mode for 30 minutes before automatic shutdown. This model has nine suction levels. Both models allow the user to adjust the suction levels when necessary.
{4}------------------------------------------------
-
- Identification of Predicate Device
510(k) Number: K142479 Product Name: Megna Breast Pumps Model Name: M10
- Identification of Predicate Device
6. Substantially Equivalent (SE) Comparison
| Item | Proposed Devices | Predicate Device |
|---|---|---|
| Product Code | HGX | HGX |
| Regulation No. | 21 CFR 884.5160 | 21 CFR 884.5160 |
| Class | Class II | Class II |
| Indications for Use | The powered Electric Double BreastPumps are intended to express andcollect milk from the breast of alactating woman. They are doublepumps with a single pumping optionand intended for single user. | The powered Megna Breast Pumpsare intended to express and collectmilk from the breast of a lactatingwoman.The M5 model is a single pump. TheM7, M10, and M12 models are doublepumps with a single pumping option.All models are intended for singleusers. |
| Pump Type | Reciprocating Diaphragm | Reciprocating Diaphragm |
| Pumping Options | Single or Double | Single or Double |
| Cycling/SuctionControl Mechanism | Microprocessor | Microprocessor |
| Suction Levels | XB-8636: 7XB-8703: 9 | 9 |
| Suction Strength | Single & Double: 60 - 240 mmHg | Single: 60 - 297 mmHgDouble: 60 - 300 mmHg |
| Cycle Speed | XB-8636:Single: 36 - 103 cycles / minDouble: 25 - 76 cycles / minXB-8703:Single: 38 - 120 cycles / minDouble: 22 – 68 cycles / min | 38 – 139 cycles / min |
| Suction Flow Rate | XB-8636:Single & Double: 4 – 40 ml / minXB-8703:Single & Double: 4 – 60 ml / min | Single: 9 – 65 ml / minDouble: 10 – 67 ml / min |
| Power Supply | 6V DC Adaptor or 4 AA 1.5Vbatteries | 6V DC Adaptor |
Table 1 Comparison of Technology Characteristics
The differences in suction level, suction strength, cycle speed, and suction flow rate between the subject and predicate device do not raise different questions of safety and effectiveness.
{5}------------------------------------------------
-
- Summary of Non-Clinical Tests
Non-clinical tests were conducted to verify that the proposed device met all design specifications and was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- Summary of Non-Clinical Tests
-
. IEC 60601-1: 2005 + CORR.1 (2006) + CORR.2 (2007) + AM1 (2012), Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.
-
. IEC 60601-1-11: 2010, Medical Electrical Equipment - Part 1-11: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment.
-
. IEC 60601-1-2: 2007, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility -Requirements And Tests.
Additionally, the following non-clinical tests were conducted:
- Suction Curves of proposed devices were tested. All the tests results complied with the design . specifications of the proposed devices.
- . Back flow testing was conducted to ensure that even if the bottle is over-filled, no liquid will backflow into the tubing, and therefore no liquid can backflow into the pump motor. The test results showed that there was no backflow during the test.
- . Use-life testing was conducted on the proposed devices. The test result showed there was no significant difference in suction strength or simulating milk suction flow rate of new-manufactured devices and devices after two-year use and 1000 cycles' disinfection.
-
- Summary of Clinical Tests
Clinical testing was not required to demonstrate substantial equivalence of the Electric Double Breast Pumps to the predicate device.
-
- Substantially Equivalent (SE) Conclusion
The differences between the Electric Double Breast Pumps and its predicate device do not introduce a new intended use and do not raise different issues of safety and effectiveness. Performance testing demonstrated that no adverse effects have been introduced by these differences and that the device performs as intended.
- Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device.
§ 884.5160 Powered breast pump.
(a)
Identification. A powered breast pump in an electrically powered suction device used to express milk from the breast.(b)
Classification. Class II (performance standards).