(303 days)
The BellaFuse™ is an implant intended to fill bony voids or gaps of the skeletal system, i.e. posterolateral spine. These osseous defects are surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. Bellafuse™ resorbs and is replaced with bone during the healing process.
Device Identification and Materials of Use:
BellaFuse™ is a resorbable bone void filler, combining Human Demineralized Bone Matrix (DBM) with gelatin. The primary component of this BVF product is demineralized particle bone, which is derived from human donor cortical bone. The additional porcine gelatin is a biocompatible component which maintains the shape and enhances flexibility.
Device Characteristics:
This product is provided in several flexible sheet sizes ranging from 10x10x2.5mm to 50x100x5.0mm. It is supplied sterile for single patient use.
Body Contact:
BellaFuse™ is a permanent resorbable device, implanted in bone tissue.
I am sorry, but the provided text does not contain the detailed information necessary to complete the table and answer the questions regarding acceptance criteria, device performance, and study specifics as outlined in your request.
The document is a 510(k) premarket notification summary for a medical device called BellaFuse™. It discusses the device's intended use, description, mechanism of action, and a summary of testing for serological, biocompatibility, osteoconduction, and osteoinductive potential. However, it does not provide:
- Specific numerical acceptance criteria for performance. It states the device "was tested successfully to fully assess the performance to grow bone" and "was shown to have osteoinductive potential," but doesn't quantify these successes with specific metrics or thresholds.
- Detailed reported device performance outcomes. While it mentions new bone growth, it doesn't provide statistical results like sensitivity, specificity, accuracy, or quantitative measures of bone growth in the context of specific acceptance criteria.
- Sample sizes for test sets, training sets, or data provenance. Only a sample size of "nude rat muscle pouch model" (n=?) is mentioned for osteoinductive potential, and no other sample sizes are given.
- Information about experts, ground truth establishment, adjudication methods, or MRMC studies.
- Specifics on how the ground truth for training or test sets was established.
Therefore, I cannot populate the table or answer most of the questions based on the provided text.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, stacked on top of each other. The profiles are connected by a flowing line that forms the hair and neck of each figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Hans Biomed Corporation % Ms. Patsy J. Trisler, JD, RAC Regulatory Consultant Trisler Consulting 5600 Wisconsin Avenue, #509 Chevy Chase, Maryland 20815
Re: K151271 Trade/Device Name: BellaFuse™ Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV, MBP Dated: February 1, 2016 Received: February 1. 2016
Dear Ms. Trisler:
This letter corrects our substantially equivalent letter of March 11, 2016.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
April 22, 2016
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K151271
Device Name
BellaFuse™
Indications for Use (Describe)
BellaFuse™ is an implant intended to fill bony voids or gaps of the skeletal system i.e., posterolateral spine. These osseous defects are surgically created of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. BellaFuse™ resorbs and is replaced with bone during the healing process.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
{3}------------------------------------------------
Section 5.0 510(k) SUMMARY
| Submitter Name: | Hans Biomed Corp. |
|---|---|
| Submitter Address: | 64, Yuseong-Daero 1628Beon-Gil, Yuseong-Gu,Korea, Republic of, 305-811 |
| Contact Person: | Ms. Lucy Choi |
| Phone Number: | 0082 2 466 2266 |
| Fax Number: | 0082 2 463 1554 |
| Date Prepared: | May 13, 2015 |
| Device Trade Name: | BellaFuse™ |
| Device Common andClassification Name: | Resorbable Calcium Salt Bone Void Filler |
| Classification Number: | 21 CFR 888.3045 |
| Product Code: | MQV, MBP |
| Device Class: | II |
| Predicate Devices: | Primary: K043421, RTI Allograft Strip IC, RegenerationTechnologies, Inc.K062205, DBX® Strip, Musculoskeletal Transplant FoundationK113728, SurFuse™ Gel & Putty, ExFuse™ Gel & Putty, HansBiomed Corp. |
| Indications for UseStatement: | The BellaFuse™ is an implant intended to fill bony voids or gapsof the skeletal system, i.e. posterolateral spine. These osseousdefects are surgically created or the result of traumatic injury to thebone and are not intrinsic to the stability of the bony structure.Bellafuse™ resorbs and is replaced with bone during the healingprocess. |
| DEVICE DESCRIPION: | Device Identification and Materials of Use:BellaFuse™ is a resorbable bone void filler, combining HumanDemineralized Bone Matrix (DBM) with gelatin. The primarycomponent of this BVF product is demineralized particle bone,which is derived from human donor cortical bone. The additionalporcine gelatin is a biocompatible component which maintains theshape and enhances flexibility.Device Characteristics:This product is provided in several flexible sheet sizes rangingfrom 10x10x2.5mm to 50x100x5.0mm. It is supplied sterile forsingle patient use.Body Contact:BellaFuse™ is a permanent resorbable device, implanted in bonetissue. |
{4}------------------------------------------------
Mechanism of Action:
This device resorbs over time and remodels providing an osteoconductive scaffold for regeneration of new bone. In addition, because it is composed primarily of DBM, it has osteoinductive potential.
Environment of Use:
BellaFuse™ is for use only in institutional health care or hospital environments.
SUMMARY OF TESTING:
Serological Testing & Biocompatibility:
The donor bone is obtained from AATB-certified tissue banks in the United States and screened for:
- Antibodies to the Human Immunodeficiency Virus, Type 1 . and Type 2 (anti-HIV-1 and anti-HIV-2);
- . Nucleic Acid Test (NAT) for HIV-1;
- . Hepatitis B Surface Antigen (HBsAg);
- . Total antibodies to Hepatitis B core antigen (anti-HBctotal):
- . Antibodies to the Hepatitis C Virus (anti-HCV);
- . Nucleic Acid Test (NAT) for HCV;
- Syphilis .
The manufacturing and sterilization processes were assessed for viral inactivation potency by a validation assessment for viral inactivation test Transmissible gastroenteritis virus (TGEV), Pseudorabies virus (PRV), Porcine retavirus (PRoV), Porcine Parvovirus (PPV), Hepatitis A virus (HAV). The validation assessment observed complete inactivation of inoculated viral titers.
Biocompatibility testing, according to ISO 10993, has been performed and the device has been shown to be safe, non-toxic and biocompatible.
Performance: Osteoconduction and Performance as a Bone Void Filler:
BellaFuse™ was tested successfully to fully assess the performance to grow bone in the in vivo rabbit spinal model.
Osteoinductive Potential:
The devicehas been tested in vivo in the athymic (nude) rat muscle pouch model and was shown to have osteoinductive potential, in that new bone grew within the muscle tissue. The osteoinductive potential also was evaluated with a surrogate, in vitro BMP-2 ELISA, assay. Results from that assay were correlated with results from the same lots in which bone successfully formed in the athymic rat. Each lot of the BellaFuse™ device will be evaluated for osteoinductive potential using the in vitro assay.
{5}------------------------------------------------
Osteoinductive assay results observed in surrogate assessments should not be interpreted to predict clinical performance in human subjects.
COMPARISON TO BellaFuse™ has the same intended use as all predicate devices. PREDICATE DEVICES: It contains the same base osteoconductive material, DBM, as all
predicates. It is provided in the same flexible sheet/strip form and contains the same porcine gelatin, as the primary RTI Allograft Strip IC and supporting predicate DBX® Strip predicates.
It is manufactured in the same facility and is sterilized the same as the third predicate, SurFuse™ Gel & Putty, ExFuse™ Gel & Putty.
SUBSTANTIAL The comparisons summarized above and the study data presented in the 510(k) lead to the conclusion that the BellaFuse™ bone void EQUIVALENCE filler device is substantially equivalent to the primary and CONCLUSION: supporting predicate devices.
§ 888.3045 Resorbable calcium salt bone void filler device.
(a)
Identification. A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.