(232 days)
The Navigator™ AE is an infiltration pump used with the Cellfina handpiece to cause a flow of dilute Lidocaine from a bag for subcutaneous delivery of anesthetic in a manner controlled manually by a healthcare professional. The Navigator is not intended for use as an IV infusion pump.
The Navigator AE is an optional fluid delivery mechanism for use with the Cellfina Aesthetic handpiece. The device is comprised of the following major components: (1) a fluid delivery module (pump drive unit); (2) a disposable cassette that is designed for single use; and, (3) a wired USB foot pedal. The Navigator AE components interface with a commercially available single-use handpiece and tubing set that is discarded following use on a specific patient.
The provided text describes the acceptance criteria and performance testing of the Navigator Aesthetic Delivery System (Navigator AE), an infiltration pump.
Here's the breakdown of the information requested:
1. A table of acceptance criteria and the reported device performance
The document doesn't present a direct table of acceptance criteria with corresponding performance metrics in a concise format. Instead, it describes various performance tests and the overall conclusion that the device "functions as designed and can be operated by the user as intended," and that "The performance data demonstrates that the Navigator Aesthetic Delivery System is substantially equivalent to the named predicates."
However, we can infer the acceptance criteria from the "Performance Testing Summary" and list the reported performance implicitly:
Acceptance Criteria (Inferred from Performance Testing Summary) | Reported Device Performance (Implicit) |
---|---|
Functionality | |
Characterization of system functionality over expected conditions of use (fluidic properties, environmental characteristics) | Testing confirmed system functionality. |
Flow rate characterization, accuracy of volumes dispensed, flow profiles, flow rate accuracy | Testing confirmed flow rate characterization, accuracy of volumes dispensed, flow profiles, and flow rate accuracy. |
Occlusion testing | Occlusion testing was performed. |
Integrity of fluid pathway components | Integrity of fluid pathway components was confirmed. |
Reliability assessment | Reliability assessment was performed. |
Shipping studies per ISTA 2a | Shipping studies per ISTA 2a were performed. |
Software | |
Development in accordance with EN 62304:2006 | Software was developed in accordance with EN 62304:2006 and evaluated. |
Documentation for Major Level of Concern software provided | Documentation for Major Level of Concern software was provided. |
Software verification and validation (including static analysis) | Software verification and validation testing, including software static analysis, was performed. |
Zero unresolved software anomalies | Zero unresolved software anomalies were reported. |
Electrical Safety and Electromagnetic Compatibility | |
Verification through testing in accordance with UL60601-1, IEC 60601-2-24, IEC 60601-1-2 | Electrical safety and electromagnetic compatibility were verified through testing in accordance with these standards. |
Biological and Drug Compatibility | |
Biocompatibility per ISO 10993-1 (Cytotoxicity, Sensitization, Hemolysis, Irritation, Acute Systemic Toxicity) | Biocompatibility testing was performed per ISO 10993-1, covering Cytotoxicity, Sensitization, Hemolysis, Irritation, and Acute Systemic Toxicity. |
Extractables/leachables testing per ISO 10993-18 | Extractables/leachables testing was performed per ISO 10993-18. |
Drug compatibility and particulate analysis per USP | Drug compatibility and particulate analysis were performed per USP . |
Sterility and Shelf Life | |
Sterilization validation methodology in accordance with ISO 11135 | Sterilization validation methodology in accordance with ISO 11135 was described. |
Ethylene oxide residuals for ethylene oxide and ethylene chlorohydrin specification limits per ISO 10993-7 | Ethylene oxide residuals specification limits were established per ISO 10993-7. |
Sterility Assurance Level of 10^-6 | Sterility Assurance Level is 10^-6. |
Non-pyrogenic claims verified through Kinetic Chromogenic LAL and Material-Mediated Rabbit Pyrogen testing per ISO 10993-11 | Non-pyrogenic claims were verified through specified testing. |
Shelf Life of 6 months established with accelerated aging data per ASTM F1980 | A Shelf Life of 6 months was established with accelerated aging data. |
Package integrity and functional attributes remain within specifications | Testing verified package integrity and functional attributes remained within established specifications. |
Human Factors | |
Simulated use/human factors studies | A simulated use study of human factors was conducted with intended users in the intended use environment. |
Evaluation of device performance, possible use error, user perception of difficulties, and critical tasks/use scenarios | The study assessed these aspects and demonstrated that the Navigator AE functions as designed and can be operated by the user as intended. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not explicitly state the sample sizes for the "test set" for most of the performance criteria. The testing discussed seems to involve engineering, laboratory, and simulated use studies rather than clinical data with a traditional "test set" of patient cases.
- Functionality: Number of units or test conditions are not specified.
- Software: "Zero unresolved software anomalies were reported," but the exact scope or sample size of testing is not detailed.
- Electrical Safety and EMC: Not specified.
- Biological and Drug Compatibility: Not specified (e.g., number of samples for each test).
- Sterility and Shelf Life: Not specified for sterilization validation, but a "Shelf Life of 6 months" was established.
- Human Factors: A "simulated use study" was conducted with "intended users," but the number of users or sessions is not specified.
Data Provenance: The studies appear to be laboratory-based and simulated use studies, rather than retrospective or prospective human clinical data from specific countries.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This information is not provided in the document. The studies described are primarily engineering, software, and laboratory tests, and a simulated human factors study. There is no mention of experts establishing a "ground truth" for a test set in the context of clinical evaluation or image interpretation. A "Clinical Evaluation was determined not to be required for the Navigator AE."
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This information is not applicable/not provided. The performance tests are technical validations (e.g., flow rate accuracy, electrical safety, software V&V) and a simulated human factors study, not clinical studies requiring adjudication of outcomes or diagnoses.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI, vs without AI assistance
There is no mention of an MRMC study or AI assistance in the document. The Navigator AE is an infusion pump, a hardware device, and not an AI-powered diagnostic or assistive tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This concept is not applicable as the device is an infusion pump. However, much of the performance testing (functionality, software, electrical safety, biological compatibility, sterility) can be considered "standalone" in the sense that it evaluates the device's inherent technical performance without a human operator performing clinical tasks. The human factors study involved human interaction, but its purpose was to evaluate usability and potential use errors.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The "ground truth" for the Navigator AE's performance is established by engineering specifications, recognized national and international standards (e.g., ISO, IEC, UL, ASTM, USP), and documented design requirements. For example:
- Fluid flow accuracy tested against defined flow rate and volume specifications.
- Software performance verified against design requirements and EN 62304.
- Electrical safety tested against UL60601-1, IEC 60601-2-24, and IEC 60601-1-2.
- Biocompatibility against ISO 10993-1.
- Sterility against ISO 11135 and a specified Sterility Assurance Level.
For the human factors study, the "ground truth" would be the expected safe and effective operation of the device by intended users according to its design and instructions, with deviations representing potential use errors.
8. The sample size for the training set
This information is not provided and is largely not applicable in the context of device performance testing for an infusion pump. The device does not appear to involve machine learning or AI models that require a "training set." The development process follows engineering and software development methodologies (e.g., V&V per EN 62304), not AI model training.
9. How the ground truth for the training set was established
As there is no mention of a training set for machine learning/AI, this question is not applicable. The "ground truth" for the device's design and operation is established by engineering principles, regulatory requirements, and international standards, as described in point 7.
§ 880.5725 Infusion pump.
(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).