K Number
K151091

Validate with FDA (Live)

Date Cleared
2015-07-13

(81 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Vitrea Software Toshiba Package is an application package developed for use on Vitrea, a medical image processing software, which includes the following post-processing software applications.

CT/XA Cerebral Artery Morphological Analysis This software is intended to facilitate the extraction and seqmentation of user identified aneurysms on the cerebral arteries. The software can be used as an adjunct to diagnosis for the purposes of measurement of size and aspect ratio.

MR Wall Motion Tracking This application is intended to assist physicians with performing cardiac functional analysis based upon magnetic resonance images. It provides measurements of global and reqional myocardial function that is used for patients with suspected heart disease.

Device Description

Vitrea Software Toshiba Package, VSTP-001A, an application package developed for use on Vitrea, a medical image processing software, marketed by Vital Images. Inc. Vitrea Software Toshiba Package, VSTP-001A, includes two post processing applications, CT/XA Cerebral Artery Morphological Analysis and MR Wall Motion Tracking, which use brain and cardiac image data, obtained from CT/XA/MR systems, to assist physicians in performing specialized measurements and analysis.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Vitrea Software Toshiba Package, VSTP-001A, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The device includes two main applications: CT/XA Cerebral Artery Morphological Analysis and MR Wall Motion Tracking. The document describes the performance of a test but doesn't explicitly list "acceptance criteria" as a defined threshold value for each metric. Instead, it states that the studies demonstrated the device performed as intended and met required success ratios compared to manual methods or prior versions.

Application/FeatureAcceptance Criteria (Implicit)Reported Device Performance (as stated)
CT/XA Cerebral Artery Morphological AnalysisComparable to manual measurements and/or segmentations."CT/XA Cerebral Artery Morphological Analysis was comparable to manual measurements and/or segmentations."
Accurate extraction/display of aneurysm-shaped regions as well as measurement calculations."Bench studies were conducted using numerical phantoms to analyze the accuracy of extraction/display of aneurysm shaped regions as well as measurement calculations." and found it to be accurate.
MR Wall Motion TrackingMet the required success ratio for contour tracking process."The contour tracking process of the MR Wall Motion Tracking application met the required success ratio."
Accurate cardiac function and strain analysis."Bench studies were conducted using numerical phantoms... to analyze cardiac function and strain."

2. Sample Size Used for the Test Set and Data Provenance

  • CT/XA Cerebral Artery Morphological Analysis: The document mentions "clinical evaluations" and "bench studies using numerical phantoms." It does not specify the sample size for either the clinical evaluations or the numerical phantom studies.
  • MR Wall Motion Tracking: Similar to the CT/XA application, "clinical evaluations" and "bench studies using numerical phantoms" were conducted. The sample size is not specified for either.
  • Data Provenance: The document does not specify the country of origin or whether the data was retrospective or prospective. It only mentions "clinical evaluations."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not specify the number of experts used or their specific qualifications for establishing ground truth for the test set. It mentions "manual measurements and/or segmentations" for the CT/XA application, implying expert human input, but details are absent.

4. Adjudication Method for the Test Set

The document does not specify any adjudication method (e.g., 2+1, 3+1) for the test set.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, the document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study. It focuses on the device's performance compared to manual methods or prior versions, but not on how human readers improve with AI versus without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, standalone (algorithm only) performance seems to have been evaluated. The benchmarking using "numerical phantoms" and assessment of "accuracy of extraction/display" and "measurement calculations" for the CT/XA application, and "cardiac function and strain" for the MR application, suggests standalone testing. The statement about "contour tracking process... met the required success ratio" for MR Wall Motion Tracking also implies standalone algorithm evaluation.

7. The Type of Ground Truth Used

  • CT/XA Cerebral Artery Morphological Analysis: The ground truth appears to be based on expert human measurements and/or segmentations ("comparable to manual measurements and/or segmentations") and numerical phantoms for accuracy.
  • MR Wall Motion Tracking: The ground truth also seems to be based on numerical phantoms for accuracy analysis of cardiac function and strain, and an "intended" or "required success ratio" for contour tracking, which would likely be established against a reference standard or expert review.

8. The Sample Size for the Training Set

The document does not mention the sample size for the training set for either application. It focuses on verification/validation testing.

9. How the Ground Truth for the Training Set Was Established

The document does not provide information on how the ground truth for the training set was established, as it does not discuss the training process or dataset.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name in a circular arrangement around the perimeter. In the center is a stylized caduceus symbol, which is often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 13. 2015

Toshiba Medical Systems Corporation % Mr. Orlando Tadeo Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780

Re: K151091

Trade/Device Name: Vitrea Software Toshiba Package, VSTP-001 A Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 22, 2015 Received: April 23, 2015

Dear Mr. Tadeo:

This letter corrects our substantially equivalent letter of July 13, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael O'Hara

For

Robert Ochs, Ph.D Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K151091

Device Name

Vitrea Software Toshiba Package, VSTP-001A

Indications for Use (Describe)

Vitrea Software Toshiba Package is an application package developed for use on Vitrea, a medical image processing software, which includes the following post-processing software applications.

CT/XA Cerebral Artery Morphological Analysis This software is intended to facilitate the extraction and seqmentation of user identified aneurysms on the cerebral arteries. The software can be used as an adjunct to diagnosis for the purposes of measurement of size and aspect ratio.

MR Wall Motion Tracking

This application is intended to assist physicians with performing cardiac functional analysis based upon magnetic resonance images. It provides measurements of global and reqional myocardial function that is used for patients with suspected heart disease.

Type of Use (Select one or both, as applicable)

ك Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

TOSHIBA AMERICA MEDICAL SYSTEMS, INC. 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 730-5000

510(k) SUMMARY

    1. SUBMITTER'S NAME: Toshiba Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550

2. OFFICIAL CORRESPONDENT: Akinori Hatanaka Senior Manager, Regulatory Affairs and Vigilance

    1. ESTABLISHMENT REGISTRATION: 9614698

4. CONTACT PERSON:

Orlando Tadeo, Jr. Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive Tustin, CA 92780 (714) 669-7459

5. Date Prepared:

April 30, 2015 (Updated 07/07/2015)

6. TRADE NAME(S):

Vitrea Software Toshiba Package, VSTP-001A

    1. COMMON NAME:
      Radiological Image Processing Software

8. DEVICE CLASSIFICATION:

Class II (per 21 CFR 892.2050, Picture Archiving and Communications System)

9. PRODUCT CODE / DESCRIPTION: 90LLZ / Picture Archiving and Communications System

10. PERFORMANCE STANDARD:

None

{4}------------------------------------------------

11. PREDICATE DEVICES:

Device NameVitrea, Version 7.0Medical ImageProcessing SoftwareMRT-301/A5, VantageTitan 3T, v2.0(Reference Device)Aquilion ONE Vision,V7.0(Reference Device)
ApplicationNameCardiac AnalysisApplicationFor SuperconductingMRI System4D Cerebral ArteryMorphological Analysis
Marketed ByVital Images, Inc.Toshiba AmericaMedical SystemsToshiba AmericaMedical Systems
510(k) ClearanceNumberK150258K120487K142465
Clearance DateMarch 5, 2015May 23, 2012March 10, 2015
Product CodeLLZ, Picture Archivingand CommunicationsSystemLNH, MagneticResonance DiagnosticDeviceJAK, ComputedTomography X-RaySystem

12. REASON FOR SUBMISSION:

Adding existing software applications to an imaging workstation (Vitrea; Product Code LLZ). MR Wall Motion Tracking includes an improved Contour Tracking function which allows for heart wall strain analysis. CT/XA Cerebral Artery Morphological Analysis allows for the import of volume data from an interventional system in addition to the cleared CT volume analysis.

13. DEVICE DESCRIPTION:

Vitrea Software Toshiba Package, VSTP-001A, an application package developed for use on Vitrea, a medical image processing software, marketed by Vital Images. Inc. Vitrea Software Toshiba Package, VSTP-001A, includes two post processing applications, CT/XA Cerebral Artery Morphological Analysis and MR Wall Motion Tracking, which use brain and cardiac image data, obtained from CT/XA/MR systems, to assist physicians in performing specialized measurements and analysis.

14. INDICATIONS FOR USE:

Vitrea Software Toshiba Package is an application package developed for use on Vitrea®, a medical diagnostic system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. Vitrea has the following additional indications:

The CT/XA Cerebral Artery Morphological Analysis application is intended to facilitate the extraction and segmentation of user identified aneurysms on the cerebral arteries. The software can be used as an adjunct to diagnosis for the purposes of measurement of size and aspect ratio.

The MR Wall Motion Tracking application is intended to assist physicians with performing cardiac functional analysis based upon magnetic resonance images. It provides measurements of global and regional myocardial function that is used for patients with suspected heart disease.

{5}------------------------------------------------

15. SUBSTANTIAL EQUIVALENCE:

The software applications included in the Vitrea Software Toshiba Package, VSTP-001A, perform in a manner similar to and are intended for the same use as the predicate devices. These applications include modifications to the originally cleared devices, 4D Cerebral Artery Morphological Analysis (K142465) and Cardiac Analysis Application For Superconducting MRI System (K120487), in order to be available on the Vitrea, Medical Image Processing Software (K150258). The subject and predicate device applications are all post processing applications used to aid physicians with performing specialized measurement and analysis of CT, XA and MR image data.

Complete comparison tables are included in this submission. See below for a brief summary of technological characteristics of the software applications:

ItemCT/XA Cerebral ArteryMorphological Analysis4D Cerebral ArteryMorphological Analysis forAquilion ONE
510(k) ClearanceSubject DeviceK142465
Type of Input DataOne volume acquired by X-rayCT system or interventionalangiography systemMultiple volumes acquired by X-ray CT system
Anatomical RegionCerebral ArteryCerebral Artery
Segmentation(Extraction OfAneurysm-ShapedRegion)Semi-Automatic SegmentationWith One-Clicked Seed PointSemi-Automatic Segmentationwith One-Clicked Seed Point
Measurement(Distance, Angle,Length)AutomaticAutomatic
ltemMR Wall Motion TrackingCardiac Analysis ApplicationFor Superconducting MRISystem
510(k) ClearanceSubject DeviceK120487
Anatomical RegionCardiacCardiac
MyocardiumContour DetectionContour trackingContour detection applied to allphases individually
Global CardiacFunction AnalysisEjection Fraction, CardiacOutput, Volume Curves arecalculated from shapes ofmyocardial contours.Ejection Fraction, CardiacOutput, Volume Curves arecalculated from shapes ofmyocardial contours.
Regional WallMotion AnalysisWall motion, Wall thickness,Wall thickening are calculatedWall motion, Wall thickness,Wall thickening are calculated
Strain AnalysisStrain is calculatedNone

{6}------------------------------------------------

Result Display
Cine View,Polar Map Of MeasurementResultsTime-Curve Of RegionalMeasurementCine View,Polar Map Of MeasurementResultsTime-Curve Of RegionalMeasurement

16. SAFETY:

The device is designed and manufactured under the Quality System Requlations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC62304 and IEC62366.

17. TESTING

Risk analysis and verification/validation testing conducted through bench testing are included in this submission which demonstrates that the requirements for the applications have been met. Bench studies were conducted using numerical phantoms to analyze the accuracy of extraction/display of aneurysm shaped regions as well as measurement calculations and to analyze cardiac function and strain. Additionally, clinical evaluations were conducted to demonstrate that the CT/XA Cerebral Artery Morphological Analysis and MR Wall Motion Tracking applications performed as intended. The results confirmed that CT/XA Cerebral Artery Morphological Analysis was comparable to manual measurements and/or segmentations and that the contour tracking process of the MR Wall Motion Tracking application met the required success ratio.

Software Documentation for a Moderate Level of Concern, per the FDA quidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.

18. CONCLUSION

The software applications included in the Vitrea Software Toshiba Package, VSTP-001A, perform in a manner similar to and are intended for the same use as the predicate devices. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls and the performance data presented in this submission it is concluded that the subject device is substantially equivalent in safety and effectiveness to the predicate devices.

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).