AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 3D Interstitial Ring Applicator Sets (60° and 90°) and the Ring Applicator Sets (45°, 60° and 90°) are intended for use for cancer treatment of the vagina, cervix, and uterus using HDR or PDR brachytherapy.

Device Description

The 3D Interstitial Ring Applicator Sets (60° and 90°), are applicators for intracavitary or intracavitary/interstitial brachytherapy. Brachytherapy is a form of radiotherapy using Gamma rays from a radioactive source placed at locations close to or within a tumor or other treatment area to a predefined treatment plan. The treatment plan defines the positions and times for the source to ensure the correct dose for the treatment area. The applicator acts to guide the radioactive source to the correct location or locations for treatment.

The Ring Applicator Sets (45°, 60° and 90°), are applicators for intracavitary brachytherapy. Brachytherapy is a form of radiotherapy using Gamma rays from a radioactive source placed at locations close to or within a tumor or other treatment area to a predefined treatment plan. The treatment plan defines the positions and times for the source to ensure the correct dose for the treatment area. The applicator acts to guide the radioactive source to the correct location or locations for treatment.

AI/ML Overview

This FDA 510(k) summary describes a new medical device, the Varian Medical Systems 3D Interstitial Ring Applicator Sets and Ring Applicator Sets, and claims substantial equivalence to a predicate device, the Mick Radio Nuclear Instruments, Inc. CT HDR Tandem/Ring Applicator With Rectal Ring Retractor.

Here's an analysis of the acceptance criteria and study information provided, structured as requested:

1. Table of Acceptance Criteria and Reported Device Performance

Feature/SpecificationPredicate Device Performance (K030110)New Device Reported PerformanceAcceptance Criteria Met (Yes/No)
Compatible AfterloaderGammaMed plus, GammaMed 12i(t), VariSourceGammaMedplus iX, GammaMedplusYes (compatible with listed afterloaders)
Intended UseHigh dose rate Brachytherapy treatment of the uterus and cervixCancer treatment of the vagina, cervix, and uterus using HDR or PDR brachytherapyYes (expanded but similar)
Indications for UseHigh Dose Rate irradiation of the uterus and cervixCancer treatment of the vagina, cervix, and uterus using HDR or PDR brachytherapyYes (expanded but similar)
DesignRing applicators 30°, 45°, 60°; 32mm diameter. Intrauterine tandems 30°, 45°, 60°; 20mm, 40mm, 60mm, 80mm lengths3D interstitial ring applicators Sets (60° and 90°), 30mm diameter. Intrauterine tandems 60° and 90°; 30mm, 40mm, 50mm, 60mm, 70mm, 80mm lengths. Holes for needle application. Ring applicators Sets (45°, 60°, 90°); 26mm, 30mm diameter. Intrauterine tandems 45°, 60°, 90°; 30mm, 40mm, 50mm, 60mm, 70mm, 80mm lengthsYes (design variations are for expanded utility)
MaterialsStainless Steel, PPSU, PEEK, TitaniumPEEK, TitaniumYes (different materials but assessed for biocompatibility)
PackingIndividualIndividualYes
SterilityProvided non sterileProvided non sterileYes
Sterilization MethodSteam sterilization (15 min @ 121°C, 5 min @ 134°C, 18 min @ 134°C)Steam sterilization (4 min @ 132°C, 3 min @ 134°C, 5 min @ 134°C)Yes (different parameters but assessed for effectiveness)
BiocompatibilityFull biocompatibilityFull biocompatibilityYes
Anatomical SitesUterus, cervixUterus, cervix, vaginaYes (expanded but similar)
Compatibility with environment/other devicesCT compatibleCT compatible, MR conditional for 1.5 and 3 TeslaYes (adds MR compatibility)
Where UsedBrachytherapy treatment roomBrachytherapy treatment roomYes
Device Functions CorrectlyImplicit for predicateDemonstrated through bench testingYes
Withstands Cycles of UseImplicit for predicateDemonstrated through bench testingYes
Radioactive Source Positional AccuracyImplicit for predicateDemonstrated through bench testing (accuracy required)Yes
Material unaffected by radiationImplicit for predicateDemonstrated through bench testingYes
Sterilization EffectivenessImplicit for predicateDemonstrated through bench testingYes
Usability (IEC 62366:2007)Implicit for predicateConformance to applicable requirements and specifications demonstrated through bench testingYes

2. Sample Size Used for the Test Set and Data Provenance

The submission form does not explicitly detail a "test set" in the context of clinical data for performance evaluation. Instead, it relies on bench testing for technical performance.

  • Sample Size for Bench Testing: Not specified in terms of number of devices or iterations, but implies sufficient testing to demonstrate functionality, durability, accuracy, and material integrity.
  • Data Provenance: The bench testing would inherently be prospective in nature, conducted by the manufacturer (Varian Medical Systems, Inc.) in Palo Alto, CA, USA, for the purpose of this submission. Since it's bench testing, country of origin is the manufacturing/testing location.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This submission relies on bench testing against engineering specifications and industry standards (like IEC 62366:2007) rather than clinical performance based on expert-determined ground truth. Therefore, no experts were used to establish a clinical ground truth for a test set.

4. Adjudication Method for the Test Set

Not applicable. Since no clinical "test set" requiring expert judgment for ground truth establishment was conducted, there was no adjudication method. Bench testing results would typically be assessed against predefined pass/fail criteria by engineering and quality personnel.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

Not applicable. This device is a physical medical instrument (an applicator for brachytherapy), not an AI/software-based diagnostic tool that would involve human readers or AI assistance. No MRMC study was conducted.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

Not applicable. This device is a physical medical instrument, not an algorithm or AI system. Its function is to physically guide a radioactive source.

7. The Type of Ground Truth Used

For the bench tests, the "ground truth" used was engineering specifications, design requirements, and applicable international standards (e.g., for afterloader compatibility, material properties, sterilization effectiveness, positional accuracy, usability to IEC 62366:2007 requirements). There was no clinical ground truth (like pathology, expert consensus on disease, or outcomes data) established for the performance evaluation in this submission.

8. The Sample Size for the Training Set

Not applicable. This is a physical medical device, not an AI/machine learning algorithm requiring a training set.

9. How the Ground Truth for the Training Set was Established

Not applicable for the same reason as above.

§ 892.5700 Remote controlled radionuclide applicator system.

(a)
Identification. A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.(b)
Classification. Class II.