K Number
K150609
Date Cleared
2015-11-16

(251 days)

Product Code
Regulation Number
878.4460
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A latex surgeon's glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

Device Description

The proposed device, Latex Surgeon's Gloves (Powdered and Powder Free) is a sterilized and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination. The proposed device is made of natural rubber latex, per standard ASTM D3577-0961, the rubber surgical gloves classification is: "Type I - gloves compounded primarily from natural rubber latex". The proposed device includes Powdered and Powder Free Latex Surgeon's Gloves, and variations of different size. All variations share the same color, creamy white. The proposed device is provided radiation sterilized to achieve the Sterility Assurance Level (SAL) of 10-6 and place in a sterility maintenance package to ensure a shelf life of 3 years.

AI/ML Overview

This document describes the performance testing for "Annuy Latex Surgeon's Gloves (Powdered and Powder Free)" to demonstrate substantial equivalence to a predicate device (K130301). The study is a non-clinical bench test study.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The device performance is compared against various ASTM and ISO standards. The tables below summarize the acceptance criteria (standards) and the reported device performance for both powdered and powder-free gloves.

For Powdered Latex Surgeon's Gloves:

CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
Dimension (ASTM 3577-09e1)
Palm widthSize 6: 76±6 mm, 6 1/2: 83±6 mm, 7: 89±6 mm, 7 1/2: 95±6 mm, 8: 102±6 mm, 8 1/2: 108±6 mm, 9: 114±6 mmSize 6: 7680 mm, 6 1/2: 8288 mm, 7: 8894 mm, 7 1/2: 99101 mm, 8: 100105 mm, 8 1/2: 107111 mm, 9: 112~115 mm (All reported values fall within the standard's range)
Overall Length265 mm minSize 6: 267272 mm, 6 1/2: 271275 mm, 7: 272278 mm, 7 1/2: 272281 mm, 8: 278282 mm, 8 1/2: 280284 mm, 9: 286~289 mm (All reported values exceed the minimum)
Finger Thickness0.10 mm min0.10~0.13 mm range across sizes (All reported values meet or exceed the minimum)
Palm Thickness0.10 mm min0.10~0.13 mm range across sizes (All reported values meet or exceed the minimum)
Cuff thickness0.10 mm min0.10~0.13 mm range across sizes (All reported values meet or exceed the minimum)
Physical Properties (ASTM 3577-09e1)
- Tensile Strength (Before Aging)24 MPa, minReported values for Powdered Gloves are not explicitly listed in the "Test Data" column, but the "Substantially Equivalent Comparison Conclusion" table confirms "Meets ASTM D3577-09e1" and lists the same acceptance criteria as the predicate. The overall conclusion states tests "demonstrated that the proposed device complies with the following standards".
- Ultimate Elongation (Before Aging)750% minSame as above.
- Stress at 500% Elongation5.5 MPa, maxSame as above.
- Tensile Strength (After Aging)18 MPa, min (at 70±2 °C, 166±2 hrs)Same as above.
- Ultimate Elongation (After Aging)560% min (at 70±2 °C, 166±2 hrs)Same as above.
Freedom from Holes (ASTM D5151-11)Standard Acceptance Criteria: Holes at Inspection Level I, AQL 1.5Passed Standard Acceptance Criteria: Holes at Inspection Level I, AQL 1.5
Powder Amount (ASTM D6124-11)≤ 15 mg/dm²1.43 mg/dm², Meets ASTM D6124-06 (Reapproved 2011), Results generated values below 15 mg/dm² of Powder amount.
Protein Content (ASTM D5712-10)≤ 200 µg/dm² Aqueous extractable proteinAqueous extractable protein: 132 mcg/dm², Meets ASTM D5712-10, Results generated values below 200 µg/dm² of aqueous extractable protein content.
Biocompatibility (ISO 10993-10:2010-08-01)
- Primary Skin IrritationNot an irritantUnder the condition of the test, not an irritant.
- Dermal SensitizationNot a sensitizerUnder the condition of the test, not a sensitizer.
Sterilization Validation (ISO 11137-1:2006)Sterility Assurance Level (SAL) of 10-6Sterility Assurance Level (SAL) of 10-6

For Powder-free Latex Surgeon's Gloves:

CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
Dimension (ASTM 3577-09e1)
Palm widthSize 6: 76±6 mm, 6 1/2: 83±6 mm, 7: 89±6 mm, 7 1/2: 95±6 mm, 8: 102±6 mm, 8 1/2: 108±6 mm, 9: 114±6 mmSize 6: 7681 mm, 6 1/2: 8187 mm, 7: 8793 mm, 7 1/2: 99101 mm, 8: 100104 mm, 8 1/2: 106110 mm, 9: 111~115 mm (All reported values fall within the standard's range)
Overall Length265 mm minSize 6: 268275 mm, 6 1/2: 270276 mm, 7: 270278 mm, 7 1/2: 272281 mm, 8: 278284 mm, 8 1/2: 280285 mm, 9: 284~289 mm (All reported values exceed the minimum)
Finger Thickness0.10 mm min0.10~0.13 mm range across sizes (All reported values meet or exceed the minimum)
Palm Thickness0.10 mm min0.10~0.13 mm range across sizes (All reported values meet or exceed the minimum)
Cuff thickness0.10 mm min0.10~0.13 mm range across sizes (All reported values meet or exceed the minimum)
Physical Properties (ASTM 3577-09e1)
- Tensile Strength (Before Aging)24 MPa, min24.1-40.8 MPa (Reported values exceed the minimum)
- Ultimate Elongation (Before Aging)750% min961%-1239% (Reported values exceed the minimum)
- Stress at 500% Elongation5.5 MPa, max4.3~4.7 MPa (Reported values are below the maximum)
- Tensile Strength (After Aging)18 MPa, min (at 70±2 °C, 166±2 hrs)20.8-27.4 MPa (Reported values exceed the minimum)
- Ultimate Elongation (After Aging)560% min (at 70±2 °C, 166±2 hrs)761%-1103% (Reported values exceed the minimum)
Freedom from Holes (ASTM D5151-11)Standard Acceptance Criteria: Holes at Inspection Level I, AQL 1.5Passed Standard Acceptance Criteria: Holes at Inspection Level I, AQL 1.5
Powder residue (ASTM D6124-11)≤ 2 mg/glove0.6 mg/glove, Meets ASTM D6124-06 (Reapproved 2011), Results generated values below 2 mg of residual powder.
Protein Content (ASTM D5712-10)≤ 200 µg/dm² Aqueous extractable proteinAqueous extractable protein: 61.6 mcg/dm², Meets ASTM D5712-10, Results generated values below 200 µg/dm² of Aqueous extractable protein content.
Biocompatibility (ISO 10993-10:2010-08-01)
- Primary Skin IrritationNot an irritantUnder the condition of the test, not an irritant.
- Dermal SensitizationNot a sensitizerUnder the condition of the test, not a sensitizer.
Sterilization Validation (ISO 11137-1:2006)Sterility Assurance Level (SAL) of 10-6Sterility Assurance Level (SAL) of 10-6

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not explicitly state the exact sample size for each test. For example, for "Palm width," it provides a range of measured values (e.g., 76~80 mm for size 6) rather than listing individual measurements or a specific count of gloves tested. However, the tests are stated to comply with referenced ASTM and ISO standards, which typically specify minimum sample sizes for such evaluations.
  • Data Provenance: The testing was conducted by Anhui Anyu Latex Products Co., Ltd in China. It is a retrospective study since the data is collected and presented after the manufacturing and testing of the gloves.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This document describes non-clinical, bench testing of physical and chemical properties of a medical device (surgical gloves). As such, "ground truth" is established by adherence to recognized international standards (ASTM and ISO). There are no human experts used to establish a clinical ground truth for this type of test.

4. Adjudication Method for the Test Set

Not applicable. The product performance is evaluated against objective, standardized criteria defined by ASTM and ISO specifications, not subjective human assessment requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. This is a non-clinical bench test report for physical and chemical properties of gloves, not a clinical study involving human readers or cases.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

This refers to a medical device, specifically surgical gloves, and not an algorithm or AI. Therefore, this question is not applicable. The performance described here is the inherent physical and chemical properties of the manufactured glove itself, which are measured using various testing methodologies.

7. The Type of Ground Truth Used

The "ground truth" for the test results is established by objective, quantitative measurements against the pre-defined specifications and acceptance criteria outlined in the referenced ASTM and ISO standards (e.g., ASTM D3577-09e1 for dimensions and physical properties, ASTM D5151-11 for freedom from holes, ASTM D6124-11 for powder amount, ASTM D5712-10 for protein content, ISO 10993-10 for biocompatibility, and ISO 11137-1:2006 for sterilization).

8. The Sample Size for the Training Set

Not applicable. This is a report on the testing of a manufactured medical device, not an AI/ML algorithm development. There is no concept of a "training set" in this context.

9. How the Ground Truth for the Training Set was Established

Not applicable. As explained above, this is not an AI/ML algorithm study.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure in profile, with three faces overlapping to symbolize the department's focus on health and well-being. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 16, 2015

Anhui Anyu Latex Products Co., Ltd. c/o Doris Dong Shanghai CV Technology Co., Ltd. Room 1706. No. 128 Songle Rd., Songjiang Area Shanghai 201600 CHINA

Re: K150609

Trade/Device Name: Annuy Latex Surgeon's Gloves (Powdered and Powder Free) Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's glove Regulatory Class: Class I Product Code: KGO Dated: October 6, 2015 Received: October 14, 2015

Dear Ms. Dong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

Device Name

Annuy Latex Surgeon's Gloves (Powdered and Powder Free)

Indications for Use (Describe)

A latex surgeon's glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows a logo with a blue circle on the left side and a green design on the right side. The green design appears to be a stylized representation of leaves or feathers. The background is a light blue color, providing a soft contrast to the logo elements.

Section 5 510(k) Summarv

[As required by 21 CFR 807.92]

1. Submission Information: 510(k) Number: K150609 Sept 30th, 2015 Date: Type of 510(k) Submission: Traditional Basis for 510(k) Submission: New device Submitter/Manufacturer: Anhui Anyu Latex Products Co., Ltd East Qinji Road & North Xinghua Road, Bengbu, Anhui, 233010 China Contactor: Doris Dong (Consultant) Shanghai CV Technology Co., Ltd. Room 1706, No. 128 Songle Rd., Songjiang Area, Shanghai, 201600 China E-mail: doris d@126.com Tel: 86 21-31261348

2. Device Description: Proprietary Name: Annuy Latex Surgeon's Gloves (Powdered and Powder Free) Common Name: Surgical Gloves Classification Name: surgeon's gloves Product Code: Surgeon's Glove - 79KGO Device Class: T Regulation Number: 21 CFR 878,4460 Review Panel: General & Plastic Surgery Indications for use: A latex surgeon's glove is a device made of natural rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. Device Description: The proposed device, Latex Surgeon's Gloves (Powdered and Powder Free) is a sterilized and disposable medical glove intended to be worn by operating room personnel to protect a surgical wound from contamination. The proposed device is made of natural rubber latex, per standard ASTM D3577-0961, the rubber surgical gloves classification is: "Type I - gloves compounded primarily from natural rubber latex" . The proposed device includes Powdered and Powder Free Latex Surgeon's Gloves, and variations of different size. All variations share the same color, creamy white. The proposed device is provided radiation sterilized to achieve the Sterility Assurance Level (SAL) of 106 and place in a sterility maintenance package to ensure a shelf life of 3 years.

3. Predicate Device Identification

510(k) Number: K130301 Clearing date: August 22, 2013 Product Name: Latex Surgeon's Gloves (Powdered and Powder Free)

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows a logo with a circular design. The left half of the circle is blue, while the right half is green. Inside the circle, there are curved lines that resemble a plant or a wave. The lines are green and appear to be growing or flowing from the bottom of the circle towards the top.

Manufacturer: ELIMEDICAL DEVICES (FUJIAN) INC.

4. Performance Test Data Summary

• The Powdered Latex Surgeon's Gloves

CharacteristicsDimensionStandardTest Data
ASTM 3577-09e1
Palm widthSize(mm)(mm)
676±676~80
61/283±682~88
789±688~94
71/295±699~101
8102±6100~105
81/2108±6107~111
9114±6112~115
Overall LengthSize(mm)(mm)
6265 min267~272
61/2265 min271~275
7265 min272~278
71/2265 min272~281
8265 min278~282
81/2265 min280~284
9265 min286~289
Finger ThicknessSize(mm)(mm)
60.10 min0.10~0.12
61/20.10 min0.10~0.13
70.10 min0.11~0.13
71/20.10 min0.10~0.13
80.10 min0.10~0.13
81/20.10 min0.10~0.12
90.10 min0.10~0.13
Palm ThicknessSize(mm)(mm)
60.10 min0.10~0.12
61/20.10 min0.10~0.13
70.10 min0.11~0.13
71/20.10 min0.10~0.13
80.10 min0.10~0.13
81/20.10 min0.11~0.13
90.10 min0.11~0.13
Cuff thicknessSize(mm)(mm)
60.10 min0.10~0.12
61/20.10 min0.10~0.12
70.10 min0.11~0.13
71/20.10 min0.10~0.13
80.10 min0.10~0.12
81/20.10 min0.10~0.13
90.10 min0.10~0.13
Physical PropertiesASTM 3577-09e1
Before AgingAfter aging at 70±2 °C166±2 hrs
- Tensile Strength24MPa,min18MPa,min
- Ultimate Elongation750% min560% min
- Stress at 500%Elongation5.5MPa,max--
Freedom from HolesASTM D5151-11
Passed Standard Acceptance CriteriaHoles at Inspection Level I, AQL 1.5
Powder AmountASTM D6124-11, ≤ 15mg/dm²
1.43 mg/dm², Meets ASTMD6124-06(Reapproved 2011), Results generatedvalues below 15mg/dm² of Powder amount
Protein ContentASTM D5712-10, ≤200µg/dm² Aqueousextractable protein
Aqueous extractable protein: 132 mcg /dm²,Meets ASTM D5712-10, Results generatedvalues below 200µg/dm² of aqueous extractableprotein content
BiocompatibilityPrimary Skin Irritation in rabbits ISO10993-10:2010-08-01
Under the condition of the test, not an irritant.
Dermal sensitization in the guinea pig ISO10993-10:2010-08-01
Under the condition of the test, not a sensitizer.
SterilizationValidationISO 11137-1:2006Sterility Assurance Level (SAL) of 10-6

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows a logo with a stylized design. The logo features a blue semi-circle at the top, with green curved lines extending upwards from the bottom, resembling blades of grass or stylized waves. The overall design is simple and modern, with a focus on natural elements.

• The Powder-free Latex Surgeon's Gloves

CharacteristicsStandardTest Data
DimensionASTM 3577-09e1
Palm widthSize(mm)(mm)
6$76\pm6$76~81
61/2$83\pm6$81~87
7$89\pm6$87~93
71/2$95\pm6$99~101
8$102\pm6$100~104
81/2$108\pm6$106~110
9$114\pm6$111~115
Overall LengthSize(mm)(mm)
6265 min268~275
61/2265 min270~276
7265 min270~278
71/2265 min272~281
8265 min278~284
81/2265 min280~285
9265 min284~289
Finger ThicknessSize(mm)(mm)
60.10 min0.10~0.12
61/20.10 min0.10~0.12
70.10 min0.11~0.12
71/20.10 min0.10~0.13
80.10 min0.10~0.12
81/20.10 min0.10~0.12
90.10 min0.10~0.13
Palm ThicknessSize(mm)(mm)
60.10 min0.10~0.13
61/20.10 min0.10~0.13
70.10 min0.10~0.12
71/20.10 min0.10~0.13
80.10 min0.11~0.12
81/20.10 min0.11~0.12
90.10 min0.10~0.12
Cuff thicknessSize(mm)(mm)
60.10 min0.10~0.12
61/20.10 min0.10~0.12
70.10 min0.10~0.12
71/20.10 min0.10~0.13
80.10 min0.10~0.12
81/20.10 min0.10~0.12
90.10 min0.10~0.12
Physical PropertiesASTM 3577-09e1
Before AgingAfter aging at 70±2 °C166±2 hrsBefore AgingAfter aging at 70±2 °C166±2 hrs
- Tensile Strength24MPa,min18MPa,min24.1-40.8Mpa20.8-27.4Mpa
- Ultimate Elongation750% min560% min961%-1239%761%-1103%
- Stress at 500%Elongation5.5MPa,max--4.3~4.7Mpa--
Freedom from HolesASTM D5151-11Passed Standard Acceptance CriteriaHoles at Inspection Level I, AQL 1.5
Powder residueASTM D6124-11, ≤2mg/glove0.6 mg/glove, Meets ASTMD6124-06(Reapproved 2011),Results generatedvalues below 2mg of residual powder
Protein ContentASTM D5712-10, ≤200µg/dm² Aqueousextractable proteinAqueous extractable protein:61.6 mcg/dm²,Meets ASTM D5712-10, Results generatedvalues below 200µg/dm² of Aqueous extractableprotein content

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows a logo with a circular shape. The logo is divided into two main sections: a blue upper section and a green lower section. The green section features a stylized representation of plant leaves or blades of grass, creating a natural or environmental feel. The overall design is simple and modern, suggesting a focus on nature, growth, or sustainability.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows a logo with a stylized design. The logo features a combination of blue and green elements. On the left side, there is a blue semi-circle. To the right of the semi-circle, there are several curved green lines that resemble blades of grass or stylized leaves. The overall design is simple and modern.

BiocompatibilityPrimary Skin Irritation in rabbits ISO10993-10:2010-08-01Under the condition of the test, not an irritant.
Dermal sensitization in the guinea pig ISO10993-10:2010-08-01Under the condition of the test, not a sensitizer.
SterilizationValidationISO 11137-1:2006Sterility Assurance Level (SAL) of 10-6

5. Non-Clinical Test Conclusion

Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • ASTM D3577-09e1, Standard Specification for Rubber Surgical Gloves

  • ASTM D5151-06 (Reapproved 2011), Standard Test Method for Detection of Holes in Medical Gloves

    • ASTM D6124-06, (Reaffirmation 2011) Standard Test Method for Residual Powder on Medical Gloves
  • ASTM D6499-12, Standard Test Method for the Immunological Measurement of Antigenic Protein in Natural Rubber and Its Products

  • ASTM D5712-10. Standard Test Method for the Analysis of Aqueous Extractable Protein in Natural Rubber and Its Products Using the Modified Lowry Method

  • ISO 11137-1:06, Sterilization of Health Care Products - Radiation - Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process For Medical Devices (Sterility) * ISO 11137-2:13, Sterilization of Health Care Products - Radiation - Part 2: Establishing the Sterilization Dose. (Sterility)

  • ASTM F1929-12, Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration (Sterility)

  • ASTM F88/F88M-09, Standard Test Method for Seal Strength of Flexible Barrier Materials (Sterility)

  • ASTM F1608:09, Standard Test Method For Microbial Ranking of Porous Packaging Materials (Exposure Chamber Method) (Sterility)

  • ASTM D7161-05(Reapproved 2010),Standard Practice for Determination of Real Time Expiration Dating of Mature Medical Gloves Stored Under Typical Warehouse Conditions

  • ISO 10993-10:10, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization (Biocompatibility)

6. Substantial Equivalent Based on Assessment of Clinical Performance Data:

Clinical data was not including in this submission.

7. Substantially Equivalent Comparison Conclusion

· For Powdered Latex Surgeon's Gloves

ParametersNew DevicePredicate Device
1510(k) NumberK150609K130301
2Marketingclearance date--AUG 22,2013
3Device NameAnnuy Powdered Latex Surgeon's GlovesPowdered Latex Surgeon's Gloves
4ManufacturerAnhui Anyu Latex Products Co., LtdELIMEDICAL DEVICES (FUJIAN) INC.
5Product CodeKGOKGO
6Regulation No.21 CFR 878.446021 CFR 878.4460
7ClassII
8Device descriptionAnnuy Powdered Latex Surgeon's Gloves areType I - gloves compounded primarily fromnatural rubber latex, and variations of differentsize. All variations share the same color,creamy white.Powdered Latex Surgeon's Gloves are TypeI - gloves compounded primarily fromnatural rubber latex, and variations ofdifferent size. All variations share the samecolor, creamy white.
9Intended useA latex surgeon's glove is a device made ofnatural rubber intended to be worn byoperating room personnel to protect a surgicalwound from contamination.A latex surgeon's glove is a device made ofnatural rubber intended to be worn byoperating room personnel to protect asurgical wound from contamination.
10MaterialNatural Rubber LatexNatural Rubber Latex
11DesignSingle useSterile(Gamma radiation, SAL:10-6)PowderedCuffedSingle useSterile(Gamma radiation, SAL:10-6)PowderedCuffed
12SizeProvided in sizes 6,61/2, 7, 71/2, 8, 81/2,9Provided in sizes 6,61/2, 7, 71/2, 8, 81/2,9
13SpecificationsPowdered Latex Examination Gloves MeetASTM D3577-09e1Powdered Latex Examination Gloves MeetASTM D3577-09e1
14Dimensions- Overall LengthMeets ASTM D3577-09e1265mm minMeets ASTM D3577-09e1265mm min
15Dimensions- Palm WidthMeets ASTM D3577-09e16 76±6mm61/2 83±6mm7 89±6mm71/2 95±6mm8 102±6mm81/2 108±6mm9 114±6mmMeets ASTM D3577-09e16 76±6mm61/2 83±6mm7 89±6mm71/2 95±6mm8 102±6mm81/2 108±6mm9 114±6mm
16Dimensions- Finger ThicknessMeets ASTM D3577-09e10.10mm minMeets ASTM D3577-09e10.10mm min
Dimensions- Palm ThicknessMeets ASTM D3577-09e10.10mm minMeets ASTM D3577-09e10.10mm min
Dimensions- Cuff ThicknessMeets ASTM D3577-09e10.10mm minMeets ASTM D3577-09e10.10mm min
17Physical PropertiesMeets ASTM D3577-09e1Before Aging:- Tensile Strength: 24Mpa min- Ultimate Elongation: 750% min- Stress at 500% Elongation: 5.5Mpa maxMeets ASTM D3578-10After aging at 70±2 oC 166±2 hrs:- Tensile Strength: 18Mpa min- Ultimate Elongation: 560% minMeets ASTM D3577-09e1Before Aging:- Tensile Strength: 24Mpa min- Ultimate Elongation: 750% min- Stress at 500% Elongation: 5.5Mpa maxMeets ASTM D3578-10After aging at 70±2 oC 166±2 hrs:- Tensile Strength: 18Mpa min- Ultimate Elongation: 560% min
18Freedom fromMeets ASTM D3577-09e1 and ASTMMeets ASTM D3577-09e1 and ASTM
EAST QINJI ROAD & NORTH XINGHUA ROAD, BENGBU, ANHUI, CHINA
HolesD5151-11D5151-11
Holes at Inspection Level I, AQL 1.5Holes at Inspection Level I, AQL 1.5
19Powder AmountMeets ASTM D6124-11Below 15mg/dm² of residual powderMeets ASTM D6124-11Below 15mg/dm² of residual powder
20Protein ContentMeets ASTM D5712-10,Below 200µg/dm² of aqueous extractableprotein contentMeets ASTM D5712-10,Below 200µg/dm² of aqueous extractableprotein content
21Absorbable donningor dusting powderCornstarch,Powdered with absorbable dusting powder,U.S.P.Cornstarch,Powdered with absorbable dusting powder,U.S.P.
22BiocompatibilityMeets ISO 10993-10:2010;Skin Irritation: Under the condition of thetest, not an irritant.Skin Sensitization: Under the condition of thetest, not a sensitizer.Meets ISO 10993-10:2010;Skin Irritation: Under the condition of thetest, not an irritant.Skin Sensitization: Under the condition ofthe test, not a sensitizer.
23ColorCreamy white, no colorantCreamy white, no colorant
24Labeling FeaturesInclude the required labeling:• Surgical Gloves• Sterile• Single Use Only• Powdered• Natural rubber latex• Device color: Creamy white• Allergy warning• Contents, Size, REF, Lot Number, MFGdate, expiration data• Storage conditions• Manufacturer Name and AddressInclude the required labeling:• Surgical Gloves• Sterile• Single Use Only• Powdered• Natural rubber latex• Device color: Creamy white• Allergy warning• Contents, Size, Lot Number, MFG date• Storage conditions• Manufacturer and distributor Name andAddress

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows a logo with a circular design. The left half of the circle is blue, while the right half is green. Within the green portion, there are several curved lines that resemble blades of grass or leaves. The background is a light blue gradient.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows a logo with a stylized design. The logo features a circular shape with two distinct sections. The upper section is colored in a light blue hue, while the lower section is in a darker green shade. Within the logo, there are curved lines that resemble a plant or a wave, adding a dynamic and organic feel to the design.

• For Powder-free Latex Surgeon's Gloves

ParametersNew DevicePredicate Device
1510(k) NumberK150609K130301
2Marketing clearance date--AUG 22,2013
3Device NameAnnuy Powder-free Latex Surgeon's GlovesPowder-free Latex Surgeon's Gloves
4ManufacturerAnhui Anyu Latex Products Co., LtdELIMEDICAL DEVICES (FUJIAN) INC.
5Product CodeKGOKGO
6Regulation No.21 CFR 878.446021 CFR 878.4460
7ClassII
8Device descriptionAnnuy Powder-free Latex Surgeon's Gloves are Type I - gloves compounded primarily from natural rubber latex, and variations of different size. All variations share the same color, creamy white.Powder-free Latex Surgeon's Gloves are Type I - gloves compounded primarily from natural rubber latex, and variations of different size. All variations share the same color, creamy white.
9Intended useA latex surgeon's glove is a device made ofA latex surgeon's glove is a device made of
natural rubber intended to be worn byoperating room personnel to protect a surgicalwound from contamination.natural rubber intended to be worn byoperating room personnel to protect a surgicalwound from contamination.
10MaterialNatural Rubber LatexNatural Rubber Latex
11DesignSingle useSterile(Gamma radiation, SAL:10-6)Powder-freeCuffedSingle useSterile(Gamma radiation, SAL:10-6)Powder-freeCuffed
12SizeProvided in sizes 6,61/2, 7, 71/2, 8, 81/2,9Provided in sizes 6,61/2, 7, 71/2, 8, 81/2,9
13SpecificationsPowdered Latex Examination Gloves MeetASTM D3577-09e1Powdered Latex Examination Gloves MeetASTM D3577-09e1
14Dimensions- Overall LengthMeets ASTM D3577-09e1265mm minMeets ASTM D3577-09e1265mm min
15Dimensions- Palm WidthMeets ASTM D3577-09e16             76±6mm61/2            83±6mm7             89±6mm71/2            95±6mm8             102±6mm81/2           108±6mm9             114±6mm
16Dimensions- Finger ThicknessMeets ASTM D3577-09e10.10mm minMeets ASTM D3577-09e10.10mm min
Dimensions- Palm ThicknessMeets ASTM D3577-09e10.10mm minMeets ASTM D3577-09e10.10mm min
Dimensions- Cuff ThicknessMeets ASTM D3577-09e10.10mm minMeets ASTM D3577-09e10.10mm min
17Physical PropertiesMeets ASTM D3577-09e1Before Aging:- Tensile Strength: 24Mpa min- Ultimate Elongation: 750% min- Stress at 500% Elongation: 5.5Mpa maxMeets ASTM D3578-10After aging at 70±2 °C 166±2 hrs:- Tensile Strength: 18Mpa min- Ultimate Elongation: 560% minMeets ASTM D3577-09e1Before Aging:- Tensile Strength: 24Mpa min- Ultimate Elongation: 750% min- Stress at 500% Elongation: 5.5Mpa maxMeets ASTM D3578-10After aging at 70±2 °C 166±2 hrs:- Tensile Strength: 18Mpa min- Ultimate Elongation: 560% min
18Freedom fromHolesMeets ASTM D3577-09e1 and ASTMD5151-11Holes at Inspection Level I, AQL 1.5Meets ASTM D3577-09e1 and ASTMD5151-11Holes at Inspection Level I, AQL 1.5
19Powder residueMeets ASTM D6124-11Below 2mg/glove of residual powderMeets ASTM D6124-11Below 2mg/glove of residual powder
20Protein ContentMeets ASTM D5712-10,Below 200µg/dm2 of aqueous extractableMeets ASTM D5712-10,Below 200µg/dm2 of aqueous extractable
protein contentprotein content
21Polymer CoatingIt is coated with a polyurethane based coating on the inner side to facilitate donning.Polymer name: polyurethane & polyacrylic acidIt is coated with a polyurethane based coating on the inner side to facilitate donning.Polymer name: polyurethane & polyacrylic acid
22BiocompatibilityMeets ISO 10993-10:2010;Skin Irritation: Under the condition of the test, not an irritant.Skin Sensitization: Under the condition of the test, not a sensitizer.Meets ISO 10993-10:2010;Skin Irritation: Under the condition of the test, not an irritant.Skin Sensitization: Under the condition of the test, not a sensitizer.
23ColorCreamy white, no colorantCreamy white, no colorant
24Labeling FeaturesInclude the required labeling:• Surgical Gloves• Sterile• Single Use Only• Powder-free• Natural rubber latex• Device color: Creamy white• Allergy warning• Contents, Size, Lot Number, Lot Number, MFG date, expiration data• Storage conditions• Manufacturer Name and AddressInclude the required labeling:• Surgical Gloves• Sterile• Single Use Only• Powder-free• Natural rubber latex• Device color: Creamy white• Allergy warning• Contents, Size, Lot Number, MFG date• Storage conditions• Manufacturer and distributor Name and Address

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows a logo with a blue circle on the left side. Inside the circle, there are green, stylized shapes that resemble leaves or blades of grass. The green shapes are arranged in a way that suggests growth or movement, with the tallest shape in the center and smaller shapes on either side.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image shows a logo with a circular shape. The left side of the circle is blue, while the right side features green lines that resemble blades of grass or leaves. The green lines are arranged in a fan-like pattern, creating a sense of growth or nature.

Discussion: Based on the above comparison, the Annuy Latex Surgeon's Gloves (Powdered and Powder Free) and the predicate device have same technology characteristics such as material, design, intended use, specification and performance features.

Both the subject device and predicate device meet the technology characteristics of ASTM D3577 and ISO10993-10 standards. From the tests, results generated values that,

Annuy Powdered Latex Surgeon's Gloves contain no more than 15mg/dm² powder and no more than 200µg/dm² extractable protein.

Annuy Powder-free Latex Surgeon's Gloves contain no more than 2mg/glove of residual powder and no more than 200ug/dm2 extractable protein.

Conclusion: The Annuy Latex Surgeon's Gloves (Powdered and Powder Free) manufactured by Anhui Anyu Latex Products Co., Ltd are substantially equivalent to the predicate device (K130301).

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).