K Number
K143729
Manufacturer
Date Cleared
2015-08-18

(232 days)

Product Code
Regulation Number
880.6850
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SterilContainer S is a reusable sterilization container system (consisting of perforated lids with filter retention plates, and single use polypropylene filters) intended to enclose another medical device that is to be sterilized by a healthcare provider. This container system has been validated with stainless steel and Teflon lumens, hinged, and knurled instruments. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container system is compatible for use with the V-PRO 60 Low Temperature Sterilization System's Lumen and Flexible Cycles. The SterilContainer S includes accessories such as silicone mats, baskets, trays, racks, eyepiece holders and sleeve holders. The Steril Container S was demonstrated to maintain the sterility of its contents for 180 days following successful sterilization.

Device Description

The SterilContainer S is a container system that will allow for sterilization and storage of medical devices. This container system is for use in low-temperature sterilization technology such as with the Steris V-PRO 60 Sterilization System. The SterilContainer Srigid containers are made from non-anodized Aluminum and utilize disposable (single use) polypropylene filters. The SterilContainer S includes accessories such as mats, baskets, trays, holders, organizers, filters, indicator cards and tamper-evident locks.

AI/ML Overview

The provided text describes the acceptance criteria and the study results for the SterilContainer™ S sterilization container system.

Here's the breakdown of the information requested:

1. Table of acceptance criteria and the reported device performance:

Acceptance Criteria / Performance PropertyReported Device Performance (SterilContainer™ S)
Sterilization EfficacyDemonstrated a 6 log reduction to no growth in a half-cycle validation, supporting a sterility assurance level (SAL) of 10-6 in a full cycle validation.
Whole Package Microbial Aerosol ChallengeAfter exposure to a defined amount of aerosol microorganisms, contents maintained sterility.
Event Related Sterility MaintenanceDemonstrated the ability to provide an effective barrier for maintaining sterility of the contents after processing, followed by a 180-day event-related storage under conditions simulating hospital sterile package handling and storage.
Material CompatibilityAfter 100 cycles of processing, no visible or functional changes were observed.
Simulated UseA worst-case dual channel flexible endoscope was reproducibly sterilized under worst-case simulated use testing conditions in the Flexible Cycle.
Sterility Maintenance Duration180 days (explicitly stated in the "Indications for Use" and "510(k) Summary").

2. Sample size used for the test set and the data provenance:

The document does not specify the exact sample size for each performance test (e.g., number of containers tested for sterilization efficacy, or event-related sterility maintenance). It uses terms like "a defined amount of aerosol microorganisms" or "a worst case dual channel flexible endoscope."

The data provenance is nonclinical testing performed by a "qualified testing laboratory." It does not provide information on the country of origin of the data or whether it was retrospective or prospective, as these terms are typically relevant for clinical studies involving human or animal subjects, not for device performance validation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided as this is a nonclinical performance study of a sterilization container, not a diagnostic device requiring expert interpretation of results to establish ground truth. The "ground truth" here is determined by objective measures (e.g., microbial growth, physical changes, sterility tests).

4. Adjudication method for the test set:

This information is not applicable and therefore not provided. Adjudication methods (like 2+1, 3+1) are used for resolving discrepancies in expert interpretations, which is not relevant for this type of nonclinical device performance testing.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance:

This information is not applicable and therefore not provided. The device described is a sterilization container, not an AI-powered diagnostic tool. Therefore, MRMC studies and AI assistance metrics are irrelevant.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This information is not applicable as the device is a physical sterilization container, not a software algorithm.

7. The type of ground truth used:

The ground truth used in this study is based on objective laboratory measurements and tests designed to assess:

  • Microbial reduction and absence of growth: For sterilization efficacy and microbial aerosol challenge.
  • Maintenance of sterility: For event-related sterility maintenance.
  • Absence of visible or functional changes: For material compatibility.
  • Successful sterilization under specified conditions: For simulated use.

This is not expert consensus, pathology, or outcomes data in the traditional sense, but rather defined physical and biological endpoints.

8. The sample size for the training set:

This information is not applicable as the device is a physical sterilization container and does not involve machine learning or a training set.

9. How the ground truth for the training set was established:

This information is not applicable for the same reason as above (no training set for a physical device).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of a human figure in profile, with three faces overlapping to suggest a sense of community or interconnectedness. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 18, 2015

Aesculap, Inc. Denise R. Adams, RAC Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, PA 18034

Re: K143729

Trade/Device Name: SterilContainer™ S Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization wrap containers, trays, cassettes & other accessory Regulatory Class: II Product Code: KCT Dated: July 22, 2015 Received: July 23, 2015

Dear Ms. Adams:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Ms. Adams

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K143729

Device Name SterilContainer S

Indications for Use (Describe)

The SterilContainer S is a reusable sterilization container system (consisting of perforated lids with filter retention plates, and single use polypropylene filters) intended to enclose another medical device that is to be sterilized by a healthcare provider. This container system has been validated with stainless steel and Teflon lumens, hinged, and knurled instruments. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container system is compatible for use with the V-PRO 60 Low Temperature Sterilization System's Lumen and Flexible Cycles. The SterilContainer S includes accessories such as silicone mats, baskets, trays, racks, eyepiece holders and sleeve holders.

The Steril Container S was demonstrated to maintain the sterility of its contents for 180 days following successful sterilization.

LidBottomDescriptionTotalloadedcontainerweight (lbs)Intended Container Load
JM440JM440Full Size 90mm (4 ½")*21.46• Lumened and non-lumened devices withdiffusion-restricted spaces such as the hingedportion of forceps and scissors
JM489JM441Full Size 120mm (5 ½")*21.46
JM442Full Size 135mm (6")*21.46• Medical devices, including single, dual andtriple channeled rigid and semi-rigidendoscopes, with the following configurations:
JM789JM740¾ Size 90mm (4 ¼")*13.85• single or dual lumen devices with stainlesslumen(s) that is (are) ≥ 0.77 mm internaldiameter (ID) and ≤ 410 mm length
JM741¾ Size 120mm (5 ½")*13.85• triple lumen devices with stainless steellumens that are
JM742¾ Size 135mm (6")*13.85≥ 1.2 mm ID and ≤ 275 mm length
JM389JM340½ Size 90mm (4 ¼")*13.85
JM341½ Size 120mm (5 ½")*13.85≥ 1.8 mm ID and ≤ 310 mm length
JM342½ Size 135mm (6")*13.85≥ 2.8 mm ID and ≤ 317 mm length
JM020JM021Extra Long Mini 73mm(3")**7.65
JM174JM188Mini 67mm(2 ⅝")**7.65

V-PRO 60 Sterilizer Lumen Cycle

  • The validation chamber load consisted of one containing a basket iid, mat, accessories, 12 lumens, and metal and non-metal medical devices

** The validation chamber load consisted of two containing a basket and basket lid, mat, accessories, and metal and non-metal medical devices for a total chamber load weight of 15.30 lbs. Each container held six (6) lumens for a total of 12 total lumens per load.

{3}------------------------------------------------

V-PRO 60 Sterilizer Non Lumen Cycle

LidBottomDescriptionTotalloadedcontainerweight (lbs)Intended Container Load
JM489JM440Full Size 90mm (4 ¼”)*20.0
JM489JM441Full Size 120mm (5 ½”)*20.0
JM489JM442Full Size 135mm (6”)*20.0
JM789JM740¾ Size 90mm (4 ¼”)*13.85Non-lumened devices includingdevices with stainless steel ortitanium diffusion-restricted spacessuch as the hinged portion of forcepsand scissors.
JM789JM741¾ Size 120mm (5 ½”)*13.85
JM789JM742¾ Size 135mm (6”)*13.85
JM389JM340½ Size 90mm (4 ¼”)*13.85
JM389JM341½ Size 120mm (5 ½”)*13.85
JM389JM342½ Size 135mm (6”)*13.85
JM020JM021Extra Long Mini 73mm (3")**7.64
JM174JM188Mini 67mm (2 ⅝")**7.64
  • The validation chamber load consisted of one container with a basket lid, mat, accessories, and metal and nonmetal medical devices

** The validation chamber load consisted of two containers with a basket lid, mat, accessories, and metal and nonmetal medical devices for a total chamber load weight of 15.30 lbs.

LidBottomDescription*Totalloadedcontainerweight(lbs)Intended Container Load
JM489JM440Full Size 90mm (4 ¼")21.46One flexible surgical endoscope or
JM441Full Size 120mm (5 ½")21.46bronchoscope with a light cord (if not integral
JM442Full Size 135mm (6")21.46to the endoscope) and mat without any
JM789JM740¾ Size 90mm (4 ¼")13.85additional load. The flexible endoscopes may
JM741¾ Size 120mm (5 ½")13.85contain:
JM742¾ Size 135mm (6")13.85· single or dual lumen devices with lumens
JM389JM340½ Size 90mm (4 ¼")13.85that are ≥ 1 mm ID and ≤ 990 mm length
JM341½ Size 120mm (5 ½")13.85
JM342½ Size 135mm (6")13.85

V-PRO 60 Sterilizer Flexible Cycle

  • The validation chamber load consisted of one container with a basket and lid, mat, accessories, three (3) 1 x 1000mm lumens, one flexible endoscope, and one light cable.

{4}------------------------------------------------

AccessoriesV-PRO 60
Stainless Steel baskets,basket lids, and dividersYes
Instrument OrganizationSystem (Silicone andStainless Steel racks,brackets, holders, andclamps)Yes
Silicone matsYes
Stainless Steel racks, trays,holders, clamps, bracketsand platformsYes

V-PRO 60 Sterilizer Compatible SterilContainer S Accessories

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

AESCULAP

510(k) SUMMARY (as required by 21 CFR 807.92) K143729

SPONSOR:Aesculap, Inc.3773 Corporate ParkwayCenter Valley, PA 18034Establishment Registration Number: 2916714
CONTACT:Denise R. Adams, RAC610-984-9076 (phone)610-791-6882 (fax)denise.adams@aesculap.com (email)
DATE:July 7, 2015
TRADE NAME:SterilContainer™ S
COMMON NAME:Sterilization container
CLASSIFICATION NAME:Sterilization wrap containers, trays, cassettes & other accessories(21 CFR 880.6850, Product Code KCT)
DEVICE CLASS:Class II per 21 CFR 880.6850
PREDICATE:K093649, Aesculap SterilContainer S cleared for use in V-PRO 1and V-PRO 1 Plus

DEVICE DESCRIPTION

The SterilContainer S is a container system that will allow for sterilization and storage of medical devices. This container system is for use in low-temperature sterilization technology such as with the Steris V-PRO 60 Sterilization System. The SterilContainer Srigid containers are made from non-anodized Aluminum and utilize disposable (single use) polypropylene filters. The SterilContainer S includes accessories such as mats, baskets, trays, holders, organizers, filters, indicator cards and tamper-evident locks.

INDICATIONS FOR USE

The SterilContainer S is a reusable sterilization container system (consisting of perforated bottoms and perforated lids with filter retention plates, and single use polypropylene filters) intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. This container system has been validated with stainless steel and Teflon lumens, hinged, and knurled instruments. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container system is compatible for use with the V-PRO 60 Low Temperature Sterilization System's Lumen, Non Lumen and Flexible Cycles. The SterilContainer Sincludes accessories such as silicone mats, baskets, trays, racks, eyepiece holders and sleeve holders.

Aesculap, Inc. 3773 Corporate Parkway Center Valley, PA 18034 Phone: 800-258-1946 www.aesculapusa.com

{6}------------------------------------------------

AESCULAP

The SterilContainer S was demonstrated to maintain the sterility of its contents for 180 days following successful sterilization.

LidBottomDescriptionTotalloadedcontainerweight (lbs)Intended Container Load
JM440Full Size 90mm (4 ¼")*21.46Lumened and non-lumened devices with
JM489JM441Full Size 120mm (5 ½")*21.46diffusion-restricted spaces such as the hingedportion of forceps and scissors
JM442Full Size 135mm (6")*21.46• Medical devices, including single, dual and
JM789JM740¾ Size 90mm (4 ¼")*13.85triple channeled rigid and semi-rigidendoscopes, with the following configurations:
JM741¾ Size 120mm (5 ½")*13.85• single or dual lumen devices with stainlesslumen(s) that is (are) ≥ 0.77 mm internaldiameter (ID) and ≤ 410 mm length
JM742¾ Size 135mm (6")*13.85
JM389JM340½ Size 90mm (4 ¼")*13.85• triple lumen devices with stainless steellumens that are
JM341½ Size 120mm (5 ½")*13.85≥ 1.2 mm ID and ≤ 275 mm length
JM342½ Size 135mm (6")*13.85≥ 1.8 mm ID and ≤ 310 mm length≥ 2.8 mm ID and ≤ 317 mm length
JM020JM021Extra Long Mini 73mm(3")**7.65
JM174JM188Mini 67mm(2 ⅝")**7.65

V-PRO 60 Sterilizer Lumen Cycle

  • The validation chamber load consisted of one containing a basket and basket lid, mat, accessories, 12 lumens, and metal and non-metal medical devices

**The validation chamber load consisted of two containers containing a basket lid, mat, accessories, and metal and non-metal medical devices for a total chamber load weight of 15.30 lbs. Each container held six (6) lumens for a total of 12 total lumens per load

LidBottomDescriptionTotalloadedcontainerweight (lbs)Intended Container Load
JM440Full Size 90mm (4 ¼”)*20.0
JM489JM441Full Size 120mm (5 ½”)*20.0
JM442Full Size 135mm (6”)*20.0
JM740¾ Size 90mm (4 ¼”)*13.85Non-lumened devices includingdevices with stainless steel ortitanium diffusion-restricted spacessuch as the hinged portion of forcepsand scissors.
JM789JM741¾ Size 120mm (5 ½”)*13.85
JM742¾ Size 135mm (6”)*13.85
JM340½ Size 90mm (4 ¼”)*13.85
JM389JM341½ Size 120mm (5 ½”)*13.85
JM342½ Size 135mm (6”)*13.85
JM020JM021Extra Long Mini 73mm (3”)**7.64
JM174JM188Mini 67mm (2 ⅝” )**7.64

V-PRO 60 Sterilizer Non Lumen Cycle

  • The validation chamber load consisted of one container with a basket lid, mat, accessories, and metal and non-metal medical devices

** The validation chamber load consisted of two containers with a basket lid, mat, accessories, and metal and non-metal medical devices for a total chamber load weight of 15.30 lbs.

Aesculap, Inc. 3773 Corporate Parkway Center Valley, PA 18034 Phone: 800-258-1946 www.aesculapusa.com

{7}------------------------------------------------

AESCULAP.

LidBottomDescription*Totalloadedcontainerweight(lbs)Intended Container Load
JM489JM440Full Size 90mm (4 ¼")21.46One flexible surgical endoscope orbronchoscope with a light cord (if not integralto the endoscope) and mat without anyadditional load. The flexible endoscopes maycontain:
JM441Full Size 120mm (5 ½")21.46
JM442Full Size 135mm (6")21.46
JM789JM740¾ Size 90mm(4 ¼")13.85additional load. The flexible endoscopes maycontain:• single or dual lumen devices with lumensthat are ≥ 1 mm ID and ≤ 990 mm length
JM741¾ Size 120mm(5 ½")13.85
JM742¾ Size 135mm (6")13.85
JM389JM340½ Size 90mm(4 ¼")13.85
JM341½ Size 120mm(5 ½")13.85
JM342½ Size 135mm (6")13.85

V-PRO 60 Sterilizer Flexible Cycle

  • The validation chamber load consisted of one container with a basket and lid, mat, accessories, three (3) 1 x 1000mm lumens, one flexible endoscope, and one light cable.

V-PRO 60 Sterilizer Compatible SterilContainer S Accessories

AccessoriesV-PRO 60
Stainless Steel baskets, basket lids,and dividersYes
Instrument Organization System(Silicone and Stainless Steel racks,brackets, holders, and clamps)Yes
Silicone matsYes
Stainless Steel racks, trays, holders,clamps, brackets, and platformsYes

1

.

.

{8}------------------------------------------------

AESCULAP.

TECHNOLIGICAL CHARACTERISTICS (compared to predicates)

The SterilContainer S is for use in low-temperature sterilization technology such as the Steris V-PRO 60. The SterilContainer S is the same container system as was cleared in K093649. The materials and design have not changed.

SystemSterilContainer S(K143729)SterilContainer S(K093649)
Intended UseThe SterilContainer S is a reusablesterilization container system(consisting of perforated bottomsand perforated lids with filterretention plates, and single usepolypropylene filters) intended to beused to enclose another medicaldevice that is to be sterilized by ahealthcare provider. This containersystem has been validated withstainless steel and Teflon lumens,hinged, and knurled instruments. Itis intended to allow sterilization ofthe enclosed device and alsomaintain sterility of the encloseddevice until used. This containersystem is compatible for use with theV-PRO 60 Low TemperatureSterilization System's Lumen, NonLumen and Flexible Cycles. TheSterilContainer S includesaccessories such as silicone mats,baskets, trays, racks, eyepieceholders and sleeve holders.The Aesculap Sterilcontainer S is areusable sterilization containersystem intended to be used toenclose another medical device thatis to be sterilized by a healthcareprovider. It is intended to allowsterilization of the enclosed deviceand also maintain sterility of theenclosed device until used. Thiscontainer system is compatible foruse with the V-PRO 1and V-PRO Plus Systems.
SterilizerSteris V-PRO 60Steris V-PRO 1 and V-PRO 1 Plus
MaterialNon-anodized aluminumNon-anodized aluminum
Container typePerforatedPerforated
Filter typePolypropylenePolypropylene

Aesculap, Inc. 3773 Corporate Parkway Center Valley, PA 18034 Phone: 800-258-1946 www.aesculapusa.com

{9}------------------------------------------------

AESCULAP.

PERFORMANCE DATA

The SterilContainer S has been fully validated for the Steris V-PRO 60 Sterilizer. These validations were conducted by a qualified testing laboratory. The performance testing demonstrates substantial equivalence to the predicate devices.

Performance PropertiesResults
Sterilization EfficacyTesting demonstrated a 6 log reduction to no growth in a halfcycle validation. This testing supports a sterility assurance level(SAL) of 10-6 in a full cycle validation.
Whole Package MicrobialAerosol ChallengeAfter exposure to a defined amount of aerosolmicroorganisms contents maintained sterility
Event Related SterilityMaintenanceTesting demonstrated the ability to provide an effectivebarrier for maintaining sterility of the contents afterprocessing followed by a 180 day event related storageunder conditions which simulate hospital sterile packagehandling and storage conditions.
Material CompatibilityAfter 100 cycles of processing no visible or functionalchanges were observed
Simulated UseA worst case dual channel flexible endoscope wasreproducibly sterilized under worst case simulated usetesting conditions in the Flexible Cycle.

CONCLUSION

Based on the nonclinical tests performed the subject device is as safe, as effective and performs at least as safely and effectively as the legally marketed predicate device (K093649), Class II (21 CFR 880.6850), product code KCT.

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).