K Number
K143570
Device Name
WIRION
Manufacturer
Date Cleared
2015-06-04

(169 days)

Product Code
Regulation Number
870.1250
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The WIRION™ is indicated for use as an embolic protection system to contain and remove embolic material (thrombus/ debris) while performing angioplasty and stenting procedures in carotid arteries.

The diameter of the vessel at the site of filter basket placement should be between 3.5 mm to 6.0 mm. WIRION™ may be used with commercially available 0.014" guide wires.

Device Description

WIRION™ is a embolic protection System comprised of an independent Filter Unit that can be delivered, locked and deployed on commercially marketed guide wires, according to physician preference, anywhere on the wire. The WIRION™ is a rapid exchange system for single use by a single operator.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the WIRION™ Embolic Protection System, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document explicitly states that "All tests met their predetermined acceptance criteria," but it does not provide a specific table of these criteria. Instead, it details the observed performance of the primary and secondary endpoints in the clinical study.

Endpoint CategoryAcceptance Criteria (Implied/Performance Goal)Reported Device Performance
Primary EndpointMACCE rate < 0.0015 (P value vs. historic control)MACCE rate: 3.3% (P value = 0.0008)
Secondary Endpoints (Performance)High success rate (>95%) for all functionsDevice success: 99.2%
Clinical success: 97.5%
Angiographic success: 99.2%
Procedural success: 98.3%
Secondary Endpoints (Complications)Low incidence of complicationsAccess site complications: 1.7%
Neurological events: 4.1%

Study Details:

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 120 patients
  • Data Provenance: Clinical studies. The text does not explicitly state the country of origin but implies a multicenter study, and it mentions "historic control" which suggests a comparison to existing data from potentially various sources. The study was prospective, open label, and single arm.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not provide details on the number or specific qualifications of experts used to establish ground truth for the clinical study. MACCE (Major Adverse Cardiac and Cerebrovascular Events) rates and other clinical endpoints are typically determined by clinical outcomes and adjudicated by study investigators or independent clinical events committees, but the specific structure is not described here.

4. Adjudication Method for the Test Set

The document does not explicitly state the adjudication method (e.g., 2+1, 3+1). Clinical trials typically use independent adjudication committees for primary endpoints like MACCE, but this detail is not provided.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not conducted or mentioned. This device is an embolic protection system (a physical medical device), not an AI-powered diagnostic tool.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

No, a standalone algorithm performance study was not done. This is a physical medical device. The performance refers to its function within the human body during a medical procedure, not an algorithm's output.

7. The Type of Ground Truth Used

The ground truth for the clinical study was based on outcomes data (MACCE rates, device success, clinical success, angiographic success, procedural success, complications) observed in human patients during the study.

8. The Sample Size for the Training Set

The document does not mention a "training set" in the context of an algorithm or AI. This is a physical medical device, not a software algorithm that would require a training set. The clinical study described involved 120 patients.

9. How the Ground Truth for the Training Set Was Established

As there is no "training set" for an algorithm mentioned, this question is not applicable. For the clinical study, the ground truth was established by clinical observation and assessment of patient outcomes.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 4, 2015

Gardia Medical Ltd. Ohad Haas VP Quality Assurance & Regulatory Affairs 2 Ha-Eshel Street, PO Box 3081 Caesarea Industrial Park 38900, Israel

Re: K143570

Trade/Device Name: WIRION Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NTE Dated: May 7, 2015 Received: May 11, 2015

Dear Ohad Haas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K143570

Device Name WIRIONTM

Indications for Use (Describe)

The WIRION™ is indicated for use as an embolic protection system to contain and remove embolic material (thrombus/ debris) while performing angioplasty and stenting procedures in carotid arteries.

The diameter of the vessel at the site of filter basket placement should be between 3.5 mm to 6.0 mm. WIRION™ may be used with commercially available 0.014" guide wires.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over The Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Gardia Medical. The logo features the word "GARDIA" in a bold, sans-serif font, with the word "MEDICAL" in a smaller font size underneath. Above the text is an abstract graphic element composed of two overlapping shapes. The top shape is a light blue, semi-circular pattern of dots, while the bottom shape is a red, triangular shape also composed of dots.

TRADITIONAL 510(K) SUMMARY WIRION™ Embolic Protection System

Date of summary:June 1, 2015
510(k) Number:K143570
Applicant's Name:Gardia Medical Ltd.2 Ha-Eshel St. P.O. Box 3081Caesarea Industrial Park 38900, IsraelTel.: +972-4-6277166, Ext. 107Fax: +972-4-6277266
Contact Person:Ohad HaasVP QA & RAPhone: (972) 4-6277166, Ext. 303Fax: (972) -4-6277266ohadh@gradiamedical.com
Trade Name:WIRION™
Common name:Embolic Protection System
Classification:Name: Embolic Protection SystemDescription: Cardiovascular Percutaneous catheterProduct Code: NTERegulation Number: 870.1250Class: IIClassification Panel: Cardiovascular Percutaneous catheter
Predicate Devices:Substantial equivalence to the following predicates is claimed:1. FilterWire EZ™ Embolic Protection System; BostonScientific, cleared under 510(k) number K063313.2. AngioGuard RX Emboli Capture Guide wire SystemCordis Corporation, a Johnson & Johnson Companycleared under 510(k) number K101651.

Device Description:

WIRION™ is a embolic protection System comprised of an independent Filter Unit that can be delivered, locked and deployed on commercially marketed guide wires, according to physician preference, anywhere on the wire. The WIRION™ is a rapid exchange system for single use by a single operator.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Gardia Medical. The logo features the company name in blue, with "GARDIA" in a larger, bolder font above "MEDICAL" in a smaller font. Above the text is a graphic element composed of two overlapping shapes made of small circles. The shape on the left is blue, and the shape on the right is red.

Indication for Use Statement:

The WIRION™ is indicated for use as an embolic protection system to contain and remove embolic material (thrombus/debris) while performing angioplasty and stenting procedures in carotid arteries.

The diameter of the vessel at the site of filter basket placement should be between 3.5 mm to 6.0 mm. WIRION™ may be used with commercially available 0.014" guide wires.

Technological characteristics and Substantial Equivalence:

A comparison between the WIRION™ Embolic Protection System and its predicates was performed in order to support substantial equivalence determination.

The substantial equivalency was concluded based on the device's intended use, system components, functional characteristics, principle of operation, technological characteristics, sterilization, materials and biocompatibility.

The conclusion of the comparison analysis is that the WIRION™ Embolic Protection System is Substantially Equivalent to the predicate devices.

Performance Testing:

Clinical and non-clinical tests were conducted. The appropriate tests performed in order to determine substantial equivalency were completed.

These include testing in accordance with FDA Guidance for coronary and carotid embolic protection devices - premarket notification submissions (February 15, 2008) and additional tests as determined during WIRION™s V&V activities. All tests met their predetermined acceptance criteria and the device was found substantially equivalent to the predicate devices for its intended use.

Non-clinical Data:

The following tests have been performed:

  • a) Embolic Capture Efficiency and Retrieval Ability
  • b) Stent Compatibility
  • c) Simulated Use
  • d) Resistance to filter rupture during removal of a fully loaded filter
  • e) Flow Characteristics
  • f) Radial Outward Force
  • g) Tip Flexibility
  • h) Tensile Strength
  • Torque Strength and Response i)
  • j) Dimensional Verification
  • k) Catheter Coating Integrity
  • I) Corrosion Resistance
  • m) Fatigue
  • n) Radio-opacity
  • o) EtO Sterilization Validation for at least SAL 10-6.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Gardia Medical. The logo features a stylized graphic element above the company name. The graphic element is composed of two overlapping shapes: a blue, semi-transparent shape on the left and a red shape on the right, both with a dotted pattern. Below the graphic, the word "GARDIA" is written in a bold, blue sans-serif font, and below that, the word "MEDICAL" is written in a smaller, lighter blue font.

  • p) Shelf Life/Package Integrity Tests
  • q) Biocompatibility according to ISO 10993-1:2009 and FDA Blue Book Memorandum, G95-1.

Animal Data:

Two animal studies were conducted using swine model.

Safety parameters such as vessel injury, thrombus formation, neurological effects, device integrity and hemolysis were evaluated.

Performance parameters including usability, navigation, delivery, placement and retrieval of the system and the ability to capture embolic particles were evaluated as well. The device performed comparably to the predicate device. The animal studies results support the substantial equivalence of the WIRION EPS.

Clinical Data

Clinical data from the WIRION™ was collected from clinical studies and commercial use.

Open label single arm feasibility study aimed at demonstrating the safety and performances of the device used during treatment of carotid artery lesions undergoing stenting was conducted.

The WISE study: Multicenter non-randomized prospective, open label, single arm study with comparison to historic control was conducted on 120 patients.

The primary endpoint was MACCE rate and the secondary endpoint evaluated the device performances.

The data demonstrated that the observed MACCE rate was 3.3% while the MACCE rate of the historic control was 6.3%. The obtained P value was 0.0008, less than the 0.0015 which is the performance goal of the study thus the WIRION system met the primary end point.

Secondary endpoint (performances) included device success, clinical success, access site complications, angiographic and procedural success rate. All functions had high success rate of more than 95%. The endpoints related to incidents and complications, were low resulting in 1.7% and 4.1% as summarized below: Device success: 99.2%; Clinical success: 97.5%; Angiographic success: 99.2%; Procedural success: 98.3%; Access site complications: 1.7%; Neurological events: 4.1%. The device performed well and met all specifications and intended use requirements therefore secondary endpoint was met as well.

Conclusions:

The evaluation of the WIRION™ Embolic Protection System performance tests and biocompatibility studies demonstrated that the device performs as intended and that it is as safe and as effective as the predicate devices.

In light of the above, we believe it is substantially equivalent to the FilterWire EZ™ Embolic Protection System (K063313) and to the AngioGuard RX Emboli Capture Guidewire System (K101651) and that it is appropriate for its intended use.

§ 870.1250 Percutaneous catheter.

(a)
Identification. A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.(b)
Classification. Class II (performance standards).