K Number
K143073

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-03-25

(149 days)

Product Code
Regulation Number
870.4350
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Affinity NT Hollow Fiber Oxygenator with Plasma Resistant Fiber is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

Model CB511:
The Affinity NT Hollow Fiber Oxygenator with Plasma Resistant Fiber with Carmeda BioActive Surface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

Model 511T:
The Affinity NT Hollow Fiber Oxygenator with Plasma Resistant Fiber with Trillium Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

Device Description

The Medtronic Affinity NT Hollow Fiber Oxygenator is a single use gas exchange device with plasma resistant fiber. It is designed to have the blood flow outside the fiber and includes an integral heat exchanger. The oxygenator is available uncoated or bonded on its blood contacting surfaces with either Carmeda BioActive Surface or Trillium Biosurface. The device is single-use, nontoxic, nonpyrogenic, and supplied sterile for clinical use.

AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the Affinity NT Oxygenator based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance Criteria (Implied)Reported Device Performance
BiocompatibilityBiocompatible according to ISO 10993-1, externally communicating, circulating blood contact, limited durationPass
Positive Pressure Integrity TestingMaintains appropriate seal on the device, no leaksPass
Gas Transfer (Oxygen and Carbon Dioxide)Transfers gas to specificationPass
Pressure Drop (Blood and Gas)Meets pressure drop requirementsPass
Plasma BreakthroughMeets Plasma Breakthrough requirementsPass
Blood Trauma (Hemolysis, Platelets, WBC)Meets Blood Trauma requirementsPass
Pressure IntegrityMeets Pressure Integrity requirementsPass
Heat Exchanger PerformanceMeets heat exchanger efficiency requirementsPass
Coating TestingMeets coating requirementsPass

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes for each specific bench test performed. It refers to "bench testing" and "real-time aging testing" but does not quantify the number of devices or data points used for these tests.

  • Data Provenance: The studies were conducted internally by Medtronic ("Bench testing was used to verify the performance characteristics of this device. In addition, real-time aging testing for devices incorporating the previously implemented changes was completed..."). The country of origin is not specified, but the submitter (Medtronic) is based in Minneapolis, Minnesota, USA. The testing is retrospective in the sense that it's performed on manufactured devices to demonstrate compliance, not prospective clinical trials.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

This information is not applicable as the document describes bench testing of a medical device (an oxygenator) rather than a diagnostic AI algorithm that requires expert-established ground truth from medical images or patient data. The "ground truth" for these tests would be the established engineering specifications and performance standards for oxygenators.

4. Adjudication Method for the Test Set

This information is not applicable for the same reason as point 3. Bench testing results are typically objectively measured against predefined specifications, not subject to human adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for evaluating the impact of AI on human reader performance, typically in diagnostic imaging. The Affinity NT Oxygenator is a physical medical device, not an AI software.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

No, a standalone performance evaluation of an algorithm was not done. This device is not an AI algorithm. The performance evaluation presented is for the physical device itself.

7. The Type of Ground Truth Used

The ground truth used for these performance tests is engineering specifications and established performance requirements for cardiopulmonary bypass oxygenators. For example, "Demonstrates the devices transfer gas to specification" implies a predefined target range or minimum for gas transfer efficiency. Similarly, "Demonstrates the devices meet Plasma Breakthrough requirements" refers to conforming to a pre-established threshold for plasma leakage.

8. The Sample Size for the Training Set

This information is not applicable. The device is a physical medical device, not an AI model that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as point 8.

{0}------------------------------------------------

Image /page/0/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or fabric.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 25, 2015

Medtronic, Inc. Jessica Sixberry Principal Regulatory Affairs Specialist 7611 Northland Drive Minneapolis, Minnesota 55428

Re: K143073

Trade/Device Name: Affinity NT Oxygenator, Affinity NT Oxygenator with Trillium Biosurface, Affinity NT Oxygenator with Carmeda Biosurface Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ Dated: February 13, 2015 Received: February 18, 2015

Dear Ms. Sixberry,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

MA-Kille

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K143073

Device Name Affinity NT Hollow Fiber Oxygenator

Indications for Use (Describe)

Model 511:

The Affinity NT Hollow Fiber Oxygenator with Plasma Resistant Fiber is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

Model CB511:

The Affinity NT Hollow Fiber Oxygenator with Plasma Resistant Fiber with Carmeda BioActive Surface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

Model 511T:

The Affinity NT Hollow Fiber Oxygenator with Plasma Resistant Fiber with Trillium Biosurface is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Medtronic logo. On the left side of the logo is a gray circle with a human figure inside. The human figure is in three different positions, as if they are rotating. To the right of the circle is the word "Medtronic" in blue, bold letters.

510(k) Summary

Date Prepared:March 18, 2015
Submitter:Medtronic, Inc.
Medtronic Perfusion Systems
7611 Northland Drive
Minneapolis, MN 55428
Establishment Registration Number: 2184009
Contact Person:Jessica Sixberry
Principal Regulatory Affairs Specialist
Phone: (763) 514-9849
Fax:
Email: jessica.m.sixberry@medtronic.com

Device Name and Classification:

Trade Name:Affinity NT Oxygenator Uncoated or with Carmeda® BioActive
Surface or with Trillium® Biosurface
Common Name:Oxygenator
Classification Name:Cardiopulmonary bypass Oxygenator
Classification Panel:Cardiovascular
Regulation Number:21 CFR 870.4350
Product Code:DTZ
Classification:Class II

Predicate Devices

Medtronic Affinity Oxygenator (K932252) Medtronic Affinity Oxygenator with Trillium Biosurface (K973760) Medtronic Affinity Oxygenator with Carmeda BioActive Surface (K000430)

Device Description

The Medtronic Affinity NT Hollow Fiber Oxygenator is a single use gas exchange device with plasma resistant fiber. It is designed to have the blood flow outside the fiber and includes an integral heat exchanger. The oxygenator is available uncoated or bonded on its blood contacting surfaces with either Carmeda BioActive Surface or Trillium Biosurface. The device is single-use, nontoxic, nonpyrogenic, and supplied sterile for clinical use.

The Affinity NT Hollow Fiber Oxygenator with Plasma Resistant Fiber is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

{4}------------------------------------------------

Affinity NT Oxygenator is designed to be an integral part of the cardiopulmonary heart lung bypass circuit for use during cardiac surgery. Blood that comes from the patient is delivered through a blood pump to the oxygenator and other auxiliary devices, and back to the patient.

The purpose of this 510(k) Notification was to notify the FDA of an alternate material formulation for the luer caps used on the access port and the sample port as well as report previous changes on the Affinity NT Oxygenator models identified in the table below.

ModelsDescription
511Affinity NT Oxygenator
CB511Affinity NT Oxygenator with Carmeda BioActive Surface
511TAffinity NT Oxygenator with Trillium Biosurface

Indications for Use

The Affinity NT Hollow Fiber Oxygenator with Plasma Resistant Fiber (with Carmeda BioActive surface or Trillium Biosurface) is intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood and to cool or warm the blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

Comparison to Predicate Devices

The Affinity NT Oxygenators have the same intended use, design and materials, and principles of operation and technology when compared to the predicate Affinity Oxygenators (including previously implemented changes).

  • Intended Use: The intended use is the same as the predicate devices.
  • Design: The design is the same as the predicate devices.
  • Materials: The materials of the Affinity NT Oxygenator are the same or equivalent to . the materials used in the predicate devices.
  • Principles of Operation and Technology: The principles of operation are the same as ● the predicate devices.
  • Performance: The performance of the device is the same as the predicate device. ●

Summary of Performance Data

Bench testing was used to verify the performance characteristics of this device. In addition, real-time aging testing for devices incorporating the previously implemented changes was completed and supports substantial equivalence. Clinical testing was not required to establish substantial equivalence.

{5}------------------------------------------------

The following performance tests were conducted:

TestingDescriptionResult
Subject of 510(k)BiocompatibilityEnsures alternate material for the luer cap isbiocompatible according to ISO 10993-1,externally communicating, circulating bloodcontact, limited durationPass
Positive PressureIntegrity TestingEnsures luer caps made with the alternatematerial maintains appropriate seal on thedevice ensure the device does not leakPass
Previously Implemented Changes (real time aging)Gas Transfer(oxygen and carbondioxide)Demonstrates the devices transfer gas tospecificationPass
Pressure Drop(Blood and Gas)Demonstrates the devices meet pressure droprequirementsPass
PlasmaBreakthroughDemonstrates the devices meet PlasmaBreakthrough requirementsPass
Blood Trauma(hemolysis,Platelets, WhiteBlood Cells)Demonstrates the devices meet BloodTrauma requirementsPass
Pressure IntegrityDemonstrates the devices meet PressureIntegrity requirementsPass
Heat ExchangerPerformanceDemonstrates the devices meet the heatexchanger efficiency requirementsPass
Coating TestingCoating testing demonstrates that the coateddevices meet the coating requirementsPass

Conclusion

Medtronic has demonstrated that the Affinity NT Oxygenators are substantially equivalent to the predicate devices based upon design, test results, and indications for use.

§ 870.4350 Cardiopulmonary bypass oxygenator.

(a)
Identification. A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”