K Number
K142889
Date Cleared
2015-01-27

(116 days)

Product Code
Regulation Number
880.5970
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Celerity System is indicated for the positioning of Peripherally Inserted Catheters (PICC) in adult patients. It provides real-time catheter tip location by using the patient's cardiac electrical activity. The Celerity System is indicated for use as an alternative method to chest x-ray or fluoroscopy confirmation of PICC tip placement in adult patients.

Note: In general, devices that utilize ECG technique to observe P-Wave are limited, but not contraindicated, for patients where cardiac rhythms may change presentation of the P-Wave; including,

-Atrial fibrillation

-Atrial flutter

-Severe tachycardia

-Pacemaker-driven rhythm

  • Chronic obstructive pulmonary disease (COPD)

Such patients are easily identified prior to PCC insertion. Use of an additional confirmation method is necessary to confirm catheter tip location.

Device Description

The Celerity System includes the Celerity Monitor/Software, ECG Patient Cable, Remote Control Cable, Battery, Power Supply Cord and ECG Clip Cable (alligator clip). Procedural accessories including ECG Snap Leads, Surface Electrodes, Cable Cover and Prep Pads are provided as a convenience for the clinician.

AI/ML Overview

Here's an analysis of the provided text to extract information about the acceptance criteria and the study that proves the device meets them:

1. A table of acceptance criteria and the reported device performance

The provided text does not explicitly present a table of acceptance criteria with specific performance metrics. Instead, it refers to the device meeting "predetermined performance specifications" and suitability for its intended purpose. The core claim is that the device is suitable as an alternative to chest X-ray or fluoroscopy for PICC tip confirmation.

Acceptance Criteria (Implied)Reported Device Performance
PICC Tip Positioning Accuracy:"Suitability for its intended purpose" as an alternative to chest X-ray and fluoroscopy for PICC tip confirmation.
Safety and Effectiveness:"Demonstrates that the subject Celerity System meets predetermined performance specifications."
Usability/Human Factors:"Use related events noted in the studies have been adequately reviewed and addressed in order to ensure the safe, effective use of the device."
Compliance with Standards:Conformance to IEC 60601 (300 Edition), specifically IEC 60601-1 and IEC 60601-1-2.

2. Sample size used for the test set and the data provenance

The document mentions "Simulated Use / Human Factors Testing" but does not explicitly state the sample size (number of patients or simulated cases) used for this test set, nor does it specify the provenance (country of origin, retrospective or prospective) of the data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document.

4. Adjudication method for the test set

This information is not provided in the document.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The document describes the Celerity System as an "alternative method to chest X-ray or fluoroscopy confirmation." While this implies a comparison to traditional methods, it does not describe a multi-reader multi-case (MRMC) comparative effectiveness study focusing on human readers improving with AI assistance. The Celerity system itself determines the tip location, replacing the need for human interpretation of X-rays for that purpose. Therefore, an effect size of human readers improving with AI assistance is not applicable in the context described.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

The Celerity System provides "real-time catheter tip location by using the patient's cardiac electrical activity." This indicates a standalone performance where the device (algorithm) determines the tip location. The "Simulated Use / Human Factors Testing" would likely involve human interaction with the device using its standalone output, but the core function of tip determination is presented as automated.

7. The type of ground truth used

The document implies that the ground truth for evaluating tip positioning accuracy (even if not explicitly detailed) would be based on the established gold standard for PICC tip confirmation, which are chest X-ray or fluoroscopy. The device is presented as an alternative to these methods, suggesting they would serve as the comparator for accuracy.

8. The sample size for the training set

The document makes no mention of a training set or its sample size. The device appears to be based on physiological principles (P-wave changes) rather than a machine learning model that would require a distinct training set.

9. How the ground truth for the training set was established

As no training set is mentioned, this information is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal for the Department of Health & Human Services USA. The seal features the department's name in a circular arrangement around a stylized eagle. The eagle is depicted with its wings spread, symbolizing protection and service.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 5, 2015

AngioDynamics Inc. Mr. Michael P. Hanley Specialist, Global Regulatory Affairs 26 Forest Street Marlborough, MA 01752

Re: K142889

Trade/Device Name: CelerityTM PICCTip Confirmation System Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, implanted, long-term intravascular catheter Regulatory Class: II Product Code: LJS Dated: December 12, 2014 Received: December 15, 2014

Dear Mr. Hanley:

This letter corrects our substantially equivalent letter of January 27, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drue, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{1}------------------------------------------------

Page 2 - Mr. Michael P. Hanley

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part . 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina
Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K142889

Device Name Celerity PICC Tip Configuration System

Indications for Use (Describe)

The Celerity System is indicated for the positioning of Peripherally Inserted Catheters (PICC) in adult patients. It provides real-time catheter tip location by using the patient's cardiac electrical activity. The Celerity System is indicated for use as an alternative nethod to chest x-ray or fluoroscopy confirmation of PICC tip placement in adult patients.

Note: In general, devices that utilize ECG technique to observe P-Wave are limited, but not contraindicated, for patients where cardiac rhythms may change presentation of the P-Wave; including,

-Atrial fibrillation

-Atrial flutter

-Severe tachycardia

-Pacemaker-driven rhythm

  • Chronic obstructive pulmonary disease (COPD)

Such patients are easily identified prior to PCC insertion. Use of an additional confirmation method is necessary to confirm catheter tip location.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

10 - 11 - 11 FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "angiodynamics" in a sans-serif font. To the left of the word is a logo that appears to be a stylized, abstract shape. The logo is black and white and has a three-dimensional appearance.

510(k) Summary : K142889

:

Date Prepared: October 01, 2014

A.Submitter Information
Submitter:Angiodynamics Inc.26 Forest St
Marlborough, MA 01752
Tel:(508) 263-7983
Fax:(508) 263-9732
Contact:Michael Hanley, Specialist, Global
Regulatory Affairs
Email:MHanley@angiodynamics.com
Or:
Linda Varroso
AngioDynamics, Inc.
26 Forest Street
Marlborough, MA 01752
Tel:(508) 658-7967
Fax:(508) 263-9723
Email:lvarroso@angiodynamics.com
B.Trade NameTrade Name:Celerity™ PICC Tip Confirmation System
Common Name:PICC placement accessory
Classification Name:intravascular CatheterPercutaneous, implanted, long-term
Product Code:LJS, 21CFR 880.5970
Class:II
C.Predicate DevicesK140799 Celerity™ System
K093775 Sapiens ™ Tip Location System
Common Name: PICC placement accessory
Classification Name: Percutaneous, implanted, long-term intravascular
Catheter
Product Code: LJS, 21CFR 880.5970
Class: II
D.Device Description
The Celerity System includes the Celerity Monitor/Software, ECG PatientCable, Remote Control Cable, Battery, Power Supply Cord and ECG Clip
Cable (alligator clip). Procedural accessories including ECG Snap Leads,Surface Electrodes, Cable Cover and Prep Pads are provided as a
convenience for the clinician.
E.Intended Use
The Celerity System is intended to provide real time tip location
information of a central venous catheter by utilization of ECG to observe

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "angiodynamics" in black font. To the left of the word is a logo that appears to be a stylized representation of blood vessels or a similar biological structure. The logo is also in black and has a three-dimensional appearance.

P-wave changes as the tip approaches the right atrium of the heart via the superior vena cava.

F. Indications for Use

The Celerity System is indicated for the positioning of Peripherally Inserted Central Catheters (PICC) in adult patients. It provides real-time catheter tip location information by using the patient's cardiac electrical activity. The Celerity System is indicated for use as an alternative method to chest X-ray or fluoroscopy confirmation of PICC tip placement in adult patients.

Note: In general, devices that utilize ECG technique to observe P-wave are limited, but not contraindicated, for patients where cardiac rhythms may change presentation of the P-Wave: -Atrial fibrillation -Atrial flutter -Severe tachycardia -Pacemaker-driven rhythm -Chronic obstructive pulmonary disease (COPD)

Such patients are easily identified prior to PICC insertion. Use of an additional confirmation method is necessary to confirm catheter tip location.

Technological Characteristics G.

:

Technological Characteristics of the subject Celerity are equivalent with respect to the basic system design and function to that of the predicate device. The differences between the predicate and proposed devices do not raise new questions of safety or effectiveness.

H. Safety and Performance

The results of Design Verification and Validation activities were performed in accordance with Design Control requirements per 21 CFR 820.30 and demonstrate that the subject Celerity System meets predetermined performance specifications.

The performance evaluation plan included testing per the following recognized standards to assess conformance to IEC 60601 (300 Edition).

  • IEC 60601-1 Medical Electrical Equipment – Part 1: General Requirements for Safety Medical Electrical Equipment – Part 2: General IEC 60601-1-2 Requirements for Basic Safety and Essential Performance – Collateral Standard Electro Magnetic Compatibility - Requirements and Test

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the word "angiodynamics" in black font. To the left of the word is a logo that appears to be a stylized representation of interconnected shapes. The logo is also in black and is positioned to the left of the company name.

December 12, 2014 Re: K142889-Supplement

Human Factors Evaluation - Simulated Use Testing alternate to chest xray and fluoroscopy:

Simulated Use / Human Factors Testing has been conducted to evaluate the application of the Celerity electrocardiogram (ECG)-Based Tip Confirmation System (i.e., Celerity System) when used to confirm PICC tip position as an alternative method to chest X-ray or fluoroscopy confirmation. The use related events noted in the studies have been adequately reviewed and addressed in order to ensure the safe, effective use of the device as an alternate to x-ray techniques for confirmation of tip location of PICC. Based on the content of the proposed Celerity System's Risk Analysis / Use and Design FMEAs, and the content of the Instructions for Use, the Celerity System has demonstrated its suitability for its intended purpose.

l. Substantial Equivalence Conclusion

The proposed device is determined to be substantially equivalent to the predicate device based on:

  • The Intended Use and Indications for Use
  • . Operating principles/technology
  • Results of safety and performance testing ●
  • Responses to questions posed in FDA 510(k) "Substantial Equivalence" Decision-Making Flowchart

§ 880.5970 Percutaneous, implanted, long-term intravascular catheter.

(a)
Identification. A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”