K Number
K142771

Validate with FDA (Live)

Date Cleared
2014-12-22

(88 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

PVC films form a barrier to body fluids and bloodborne Pathogens. The PVC rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a medical procedure. Poly (vinyl chloride) glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151 requirements.

AI/ML Overview

This looks like a 510(k) premarket notification for patient examination gloves, which are Class I medical devices. For these types of devices, the regulatory approval heavily relies on demonstrating "substantial equivalence" to a predicate device already on the market, rather than a clinical effectiveness study of the kind you might see for an AI diagnostic device.

Therefore, many of the questions you've asked (about AI assistance, ground truth, expert opinions for labelling, effect size, etc.) are not applicable to this document as it describes a physical, non-AI medical device (gloves).

However, I can extract the relevant information regarding acceptance criteria and the study (testing) used to prove the device met those criteria, based on the provided text.


1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicsAcceptance Criteria (Standard)Reported Device Performance
DimensionASTM standard D 5250-06(Reapproved 2011)Meets
- Length≥ 230mm min (from ASTM D5250-06)230mm min for all sizes
- Width (Small)80-90 mm (from ASTM D5250-06)80-85 mm
- Width (Medium)90-100 mm (from ASTM D5250-06)94-100 mm
- Width (Large)100-110 mm (from ASTM D5250-06)102-108 mm
- Width (X Large)110-120 mm (from ASTM D5250-06)113-116 mm
Physical PropertiesASTM standard D 5250-06(Reapproved 2011)Meets
- Elongation (Before/After aging)≥ 300% (from ASTM D5250-06)390-420%
- Tensile Strength (Before/After aging)≥ 14MPa (from ASTM D5250-06)15-20 MPa
Freedom from Pinholes21 CFR 800.20, ASTM D 5250-06(Reapproved 2011), ASTM D 5151-06(Reapproved 2011)Meets, Holes at Inspection Level I AQL 2.5 (per ASTM D5151-06)
Powder ResidualASTM D 5250-06 (Reapproved 2011) & D6124-06 (Reaffirmation 2011)Meets, <2mg/glove (Results generated values below 2mg)
BiocompatibilityPrimary Skin Irritation (ISO 10993-10: 2010-08-01)Passes (Not a Primary Skin Irritation)
Dermal sensitization (ISO 10993-10: 2010-08-01)Passes (Not a Dermal sensitization)

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for the test set for each of these physical and biocompatibility evaluations. It refers to meeting ASTM standards, which would imply that the testing was conducted according to the sampling plans specified within those standards.

The data provenance is from non-clinical tests conducted by the manufacturer, Xuanhua Metallurgical Industry Co., Ltd., likely in China, where the company is located. The tests were performed to demonstrate that the device meets established ASTM and ISO standards, and FDA regulations (21 CFR 800.20). These are retrospective tests in the sense that they are conducted on manufactured product batches to ensure compliance.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

Not applicable. For physical device testing against standards, "ground truth" is established by the specifications in the standards themselves, measured objectively using calibrated equipment and defined procedures, not by expert consensus in a medical diagnostic sense. Biocompatibility tests are also performed in a laboratory setting with established protocols.

4. Adjudication Method for the Test Set

Not applicable. The tests are objective measurements against defined physical and chemical criteria. There is no human adjudication process described or implied for these types of tests.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted diagnostic device.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is not an AI device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The "ground truth" for this device's performance is adherence to recognized national and international standards and regulations for medical gloves. Specifically:

  • Physical Properties: ASTM D5250-06 (Reapproved 2011)
  • Freedom from Pinholes: 21 CFR 800.20, ASTM D5151-06 (Reapproved 2011)
  • Powder Residual: ASTM D6124-06 (Reaffirmation 2011)
  • Biocompatibility: ISO 10993-10:2010-08-01 (for Primary Skin Irritation and Dermal Sensitization)

8. The Sample Size for the Training Set

Not applicable. This is not an AI device, so there is no concept of a "training set" in the machine learning sense. The manufacturer's production process would be "trained" or optimized to consistently produce gloves meeting these standards.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As above, this is not an AI device. The establishment of "ground truth" (i.e., the standards themselves) for medical devices is through a consensus process by standards bodies (like ASTM and ISO) and regulatory bodies (like FDA), based on scientific evidence, clinical experience, and risk assessment for patient and user safety and efficacy.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white seal for the Department of Health and Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three faces in profile, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 22, 2014

Xuanhua Metallurgical Industry Co., Ltd C/O Mr. Chu Xiaoan Rm. F302 Bldg., 41. Jing Cheng Ya Ju. Courtyard 6 of Southern Dou Ge Zhuang Beijing, 100121 CN Chaoyang District, Beijing 100121 CHINA

Re: K142771

Trade/Device Name: Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LYZ Dated: December 1, 2014 Received: December 11, 2014

Dear Mr. Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Xiaoan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Susan Runno, DDS, MA

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K142771

Device Name

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored)

Indications for Use (Describe)

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section C 510(k) Summary

510(K) Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: K142771

Premarket Notification [510(k)] Summary

1.0 Submitter:

Submitter's name :Xuanhua Metallurgical Industry Co., Ltd.
Submitter's address :Bridge North, Weisan Rd, East Ring Rd, High-Tec District, Zhangjiakou City, Hebei Province.075000,P.R.China
Phone number :86-313-6216205
Fax number :86-313-5809316
Name of contact person:Mr.Dong Yanlong
Date of preparation :2014-09-14

2.0 Name of the Device

Device Name:Powder Free Vinyl Patient ExaminationGloves, Clear (non-colored)
Proprietary/Trade name:XUANYE Powder Free Vinyl PatientExamination Gloves, Clear (non-colored)
Common Name:Exam gloves
Classification Name:Patient examination glove
Device Classification:I
Regulation Number:21 CFR 880.6250
Panel:General Hospital (80)
Product Code:LYZ

3.0 Predicate device

Device Name:Powder-Free Vinyl Patient Examination Glove(Non-colored)
Company name:Zhang Jia Gang Fengyuan Plastic Product Co.Ltd.
510(K) Number:K091663.

4.0 Device Description:

4.1 How the device functions:

PVC films form a barrier to body fluids and bloodborne Pathogens

4.2 Scientific concepts that form the basis for the device

The PVC rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a

{4}------------------------------------------------

medical procedure.

Physical and performance characteristics such as design, materials and 4.3 physical properties:

Poly (vinyl chloride) glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D5250 and ASTM D5151 requirements.

5.0 Device Intended Use (Indication for use):

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

6.0 Summary of the Technological Characteristics of the Device:

The Powder Free Vinyl Patient Examination Gloves, Clear (non-colored), non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

CharacteristicsStandardDevice performance
DimensionASTM standard D5250-06(Reapproved 2011).Meets
Physical PropertiesASTM standard D5250-06(Reapproved 2011).Meets
Freedom frompinholes21 CFR 800.20Meets
Powder ResidualASTM standard D 5250-06(Reapproved 2011).andD6124-06(Reaffirmation 2011)Meets<2mg/glove
BiocompatibilityPrimary Skin Irritation in rabbitsISO 10993-10: 2010-08-01PassesNot a Primary SkinIrritation
Dermal sensitization in theguinea pigISO 10993-10: 2010-08-01PassesNot a Dermalsensitization

7.0 Substantial Equivalent Based on Assessment of Nan-Clinical Performance Data:

Powder Free Vinyl Patient Examination Gloves, Clear (non-colored), meet requirements per ASTM D5250-06 (Reapproved 2011), per ASTM D6124-06 Reaffirmation 2011), per 21 CFR 800.20 and ISO 10993-10:2010-08-01.

The performance test data of the non clinical tests that support a determination of substantial equivalent is the same as mentioned immediately above.

8.0 Substantial Equivalent Based on Assessment of Clinical Performance Data:

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

Substantial Equivalence Comparison: 9.0

{5}------------------------------------------------

Features&DescriptionPredicate DeviceSubject DeviceResultofComparison
CompanyZhang Jia Gang FengyuanPlastic Product Co.Ltd.XuanhuaMetallurgicalIndustry Co., Ltd.--
510(K) NumberK091663K142771
Product namePowder Free Vinyl PatientExamination Gloves, Clear(Non-colored)Powder Free Vinyl PatientExamination Gloves, Clear(non-colored)same
Product CodeLYZLYZsame
SizeSmall/ Medium/Large/X largeSmall/ Medium/Large/X largesame
Intend for usePowder free Vinyl PatientExamination Gloves,Clear(Non-colored)is adisposable device intendedfor medical purposes that isworn on the examiner's handor finger to preventcontamination betweenpatient and examiner.Powder free Vinyl PatientExamination Gloves, Clear(Non-colored) is adisposable device intendedfor medical purposes that isworn on the examiner's handor finger to preventcontamination betweenpatient and examiner.same
Device Descriptionand SpecificationsMeets ASTM D5250-06(Reapproved 2011)Meets ASTM D5250-06(Reapproved 2011)same
Dimensions--LengthMeets ASTMD5250 -06(Reapproved 2011)≥230mm min230mm min for all sizessame
Dimensions-- WidthMeets ASTM D5250-06(Reapproved 2011)Small80-90 mmMedium90-100mmLarge100-110mmXlarge110-120 mmSmall80-85 mmMedium94-100mmLarge102-108mmX large113-116 mmsame
Dimensions--ThicknessMeets ASTM D5250-06(Reapproved 2011)Finger 0.05mm min.Palm 0.08mm min.Thickness (mm) min.Finger 0.09-0.11Palm 0.09-0.12same
Physical PropertiesMeets ASTM D 5250-06(Reapproved 2011)Before aging/after agingElongation ≥300%Tensile Strength≥ 14MPaBefore aging/after agingElongation :390-420%Tensile Strength:15-20 MPasame
FreedomfromPinholesMeets21 CFR 800.20ASTM D5250-06(Reapproved 2011)ASTM D 5151-06(Reapproved 2011)Meets ASTMD5151-06(Reapproved 2011)Holes atInspection Level I AQL2.5same
Residual PowderMeets ASTMD 6124-06(Reapproved 2011)below 2mg of residualpowderMeets ASTMD 6124-06(Reapproved 2011)Results generated valuesbelow 2mg of residualpowdersame
Materials used tofabricate thedevicesPVCPVCsame
Dusting orDonning Powder:PUPUsame
Dusting orDonning Powder:namePUSurface Coating Agentsame
Compareperformance datasupportingsubstantialequivalenceMeetsASTM D5151-06(Reapproved 2011)ASTM D5250-06(Reapproved 2011)ASTM D6124-06(Reaffirmation 2011)MeetsASTM D5151-06(Reapproved 2011)ASTM D5250-06(Reapproved 2011)ASTM D6124-06(Reaffirmation 2011)same
Single Patient UseSingle Patient UseSingle Patient Usesame
BiocompatibilityUnder the conditions of thisstudy, the test article was anon-irritant or non-sensitizer.SKIN IRRITATIONDERMAL andSENSITIZATION STUDIESMeets ISO10993-10:2002/Amd.1:2006Under the conditions of thisstudy, the test article was anon-irritant or non-sensitizer.SKIN IRRITATIONDERMAL andSENSITIZATIONSTUDIES Meets ISO10993-10 Third Edition2010-08-01same
Labeling for thelegally marketeddevice to whichsubstantialequivalence isclaimed.-Powder Free-Patient Examination Glove-Single Use Only- Manufactured For:- Lot-Powder Free-Patient Examination Glove-Single Use Only- Manufactured For:- Lotsame

{6}------------------------------------------------

Substantial Equivalence Comparison: 10.0

It can be concluded that the Powder Free Vinyl Patient Examination Gloves, Clear (non-colored) meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL meet labeling claims.

It can be concluded that the Powder Free Vinyl Patient Examination Gloves, Clear is as safe, as effective, and performs as well as the predicate device, (non-colored) Powder-Free Vinyl Patient Examination Glove (Non-colored) Zhang Jia Gang Fengyuan Plastic Product Co., Ltd. K091663.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.