KangLi® pedicle screw spinal system is intended for posterior, non-cervical, pedicle fixation for the following indications: severe spondylolisthesis (grade 3 or 4) of the L5-Sl vertebrae; trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The device is to be used in skeletally mature patients, and for stabilization of the spine as an adjunct to fusion with bone graft. The levels of fixation are T8 – S1.
The spinal system consists of pedicle screws, nuts, rods and crosslink etc. It is made of titanium alloy (Ti6Al4V ELI), which meets ASTM F136, Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications, which are widely used for surgical implants with well-known biocompatibility. The proposed devices are provided non-sterile. It is required to be sterilized via autoclave method to reach a SAL of 106 by the hospital prior to surgery. The recommended sterilization method was validated per ISO 17665-1 Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices.
The provided text describes the 510(k) premarket notification for the KangLi® Pedicle Screw Spinal System. This is a medical device for spinal fixation, and the document focuses on demonstrating its substantial equivalence to a predicate device, rather than proving a new or novel performance against specific acceptance criteria for a diagnostic or AI-driven medical device.
Therefore, the information requested in your prompt (e.g., acceptance criteria for device performance like sensitivity/specificity, sample size for test sets, expert qualifications for ground truth, MRMC studies, standalone performance, training set details) is not applicable to this document because it pertains to a mechanical spinal implant, not a diagnostic or AI-assisted device assessed by such metrics.
The "Non-Clinical Testing" section (item 6) refers to bench tests to ensure the device meets design specifications and is substantially equivalent to the predicate. These tests are mechanical in nature, verifying physical properties and performance under stress, not diagnostic accuracy.
Here's a breakdown of what is available from the text relevant to mechanical equivalent performance:
1. A table of acceptance criteria and the reported device performance:
| Acceptance Criteria (Standard) | Reported Device Performance (Compliance) |
|---|---|
| ASTM F1717-13: Static compression bending test | Complies with standard |
| ASTM F1717-13: Dynamic compression bending test | Complies with standard |
| ASTM F1717-13: Static torsion test | Complies with standard |
| ISO 17665-1 (Sterilization validation) | Sterilization method validated per standard (SAL of 10-6) |
| ASTM F136 (Material biocompatibility) | Made of Ti6Al4V ELI meeting standard |
2. Sample sized used for the test set and the data provenance: Not applicable. These were bench tests, not clinical studies involving patient data or test sets for diagnostic performance.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for mechanical testing is established by engineering standards and measurements, not expert consensus on medical images or diagnoses.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. This refers to diagnostic study design, not mechanical testing.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not a software algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): For mechanical properties, the "ground truth" is defined by the objective physical measurements and material properties specified in the ASTM and ISO standards (e.g., strength parameters, biocompatibility, sterilization efficacy).
8. The sample size for the training set: Not applicable. No training set for an AI or diagnostic algorithm was used.
9. How the ground truth for the training set was established: Not applicable.
In summary, the provided document focuses on demonstrating substantial equivalence of a physical medical implant through adherence to established material and mechanical performance standards, not on diagnostic or AI performance metrics.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the eagle in a circular fashion.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 24, 2014
Suzhou Kangli Orthopaedics Instrument Company, Limited % Ms. Alice Gong Shanghai Yarui Consultant Company, Limited 503 Room, 8 Building, 600 Liu Zhou Road 200233 Shanghai China
Re: K142290
Trade/Device Name: KangLi® Pedicle Screw Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNH, MNI Dated: September 18, 2014 Received: September 29, 2014
Dear Ms. Gong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Ms. Alice Gong
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
K142290
Device Name
KangLi® Pedicle Screw Spinal System
KangLi® pedicle screw spinal system is intended for posterior, non-cervical, pedicle fixation for the following indications: severe spondylolisthesis (grade 3 or 4) of the L5-Sl vertebrae; trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The device is to be used in skeletally mature patients, and for stabilization of the spine as an adjunct to fusion with bone graft. The levels of fixation are T8 – S1.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Suzhou Kangli Orthopaedics Instrument Co., Ltd. The logo consists of two parts: a red graphic on the left and the company name in both Chinese and English on the right. The Chinese name is "苏州市康力骨科器械有限公司", and the English name is "SUZHOU KANGLI ORTHOPAEDICS INSTRUMENT CO., LTD."
Attachment 6 of Special 510(K) Submission:
510 (K) Summary
This 510(K) Summary of safety and effectiveness information is being submitted in accordance with requirement of 21 CFR807.92
-
- Date of Submission: Sept. 18, 2014
-
- Submitter / 510(K) Holder
Suzhou Kangli Orthopaedics Instrument Co., Ltd., Sha Zhou East Road, Zhangjiagang City, Jiangsu Province, China 215625
Contact Person: Miss Jin Hua Huang Tel: (86) 0512-58560562 Fax: (86) 0512-58358951 E-mail: hjh@kang-li.cn
3. Proposed Device Name
Trade name: KangLi® Pedicle Screw Spinal System Common name: Pedicle screw spinal system with crosslink
Classification Name: Pedicle screw spinal system Device Class: Class II Classification Panel: Orthopedic Panel Product Code: MNI, MNH Regulation Number: 21 CFR part 888.3070
4. Predicate Devices
510 (k) Number: K140053 Product Name: Pedicle screw spinal system Submitter: Suzhou Kangli Orthopaedics Instrument Co., Ltd.,
Device Description The spinal system consists of pedicle screws, nuts, rods and crosslink etc.
{4}------------------------------------------------
It is made of titanium alloy (Ti6Al4V ELI), which meets ASTM F136, Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications, which are widely used for surgical implants with well-known biocompatibility.
The proposed devices are provided non-sterile. It is required to be sterilized via autoclave method to reach a SAL of 106 by the hospital prior to surgery. The recommended sterilization method was validated per ISO 17665-1 Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices.
5. Indication for Use/Intended Use
KangLi pedicle screw spinal system is intended for posterior, non-cervical, pedicle fixation for the following indications: severe spondylolisthesis (grade 3 or 4) of the L5-Sl vertebrae; trauma (i.e. fracture or dislocation), spinal stenosis, curvatures (i.e. scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion. The device is to be used in skeletally mature patients, and for stabilization and immobilization of the spine as an adjunct to fusion with bone graft. The levels of fixation are T8 - S1.
6. Non-Clinical Testing
Bench tests were conducted to verify that proposed device meet all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that proposed device complies with the following standards:
ASTM F1717-13, Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model, including the following items:
-
- Static compression bending test:
-
- Dynamic compression bending test;
-
- Static torsion test.
7. Substantially Equivalent Conclusion
The KangLi® pedicle screw spinal system has same intended use than the predicate device and similar technological characteristics as the predicate device. The proposed device, the KangLi® pedicle screw spinal system with crosslink, is determined to be Substantially Equivalent (SE) to the predicate device, K140053 pedicle screw spinal system.