(189 days)
The Leva™ Spacer System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radios. These DDD patients may have up to Grade I spondylolisthesis at the involved level(s). The Leva™ Spacer System is to be used with autogenous bone graft and with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral body fusion device.
The Leva™ Spacer System is a lumbar intervertebral body fusion device fabricated from unalloyed commercially pure (Class II) titanium per ASTM F67. The device is available in both expandable and fixed configurations. The expandable device is provided in a collapsed form and is expanded to a predefined height in-situ using the Leva™ Inserter Instrument. Both the fixed and expandable implants are provided in different heights to accommodate the anatomical needs for a range of patients. The implants are designed to accommodate autogenous bone graft material. The implants have curved endplates to conform to the bony endplates of the patient and ridges on the endplates to resist expulsion.
I am sorry, but based on the provided text, I cannot extract any information about acceptance criteria or a study proving that a device meets those criteria. The document is a 510(k) premarket notification for a medical device called the Leva™ Spacer System. It describes the device, its intended use, and states that performance testing was conducted to demonstrate substantial equivalence to predicate devices, but it does not specify any acceptance criteria or detail the results of those tests in a way that would allow me to fill out the requested table or provide the other information.
Specifically:
- There is no table of acceptance criteria or reported device performance.
- There is no mention of sample sizes for test sets, data provenance, number or qualifications of experts, or adjudication methods for establishing ground truth.
- There is no information about multi-reader multi-case (MRMC) comparative effectiveness studies or standalone algorithm performance.
- The type of ground truth for any testing isn't specified (e.g., pathology, outcomes data).
- There is no information about the sample size for a training set or how ground truth for a training set was established, as this document pertains to a physical medical device, not an AI/ML algorithm.
The performance data section only lists the types of tests performed (e.g., Static and dynamic axial compression per ASTM F2077) but does not provide any results, acceptance criteria, or details regarding the studies themselves.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of a graphic representation of human profiles facing to the right, suggesting a sense of community and service.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 26, 2015
Spine Wave, Incorporated Ms. Roaida Johnson Senior Regulatory Affairs Manager 3 Enterprise Drive, Suite 210 Shelton, Connecticut 06484
Re: K141980
Trade/Device Name: Leva™ Spacer System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: January 2, 2015 Received: January 5, 2015
Dear Ms. Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Ms. Roaida Johnson
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Lori A. Wiggins -S
- for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known)
Device Name Leva(TM) Spacer System
Indications for Use (Describe)
The Leva™ Spacer System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radios. These DDD patients may have up to Grade I spondylolisthesis at the involved level(s). The Leva™ Spacer System is to be used with autogenous bone graft and with supplemental fixation. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral body fusion device.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
SE OFFOR FDA USE ONLY AND A
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the words "Spine Wave" in a bold, sans-serif font. The words are arranged horizontally, with "Spine" on the left and "Wave" on the right. A blue wave design is located under the word "Wave". The wave design adds a visual element to the logo.
510(k) Summary Leva™ Spacer System
1. Submitter Information
| Submitter: | Spine Wave, Inc. |
|---|---|
| Address: | Three Enterprise DriveSuite 210Shelton, CT 06484 |
| Telephone: | 203-712-1839 |
| Telefax: | 203-944-9493 |
| Contact: | Roaida R. Johnson |
| Date Prepared: | January 21, 2015 |
2. Device Information
| Trade Name: | LevaTM Spacer System |
|---|---|
| Common Name: | Intervertebral Body Fusion Device |
| Classification: | Class II (special controls) per 21 CFR 888.3080 |
| Classification Name: | Intervertebral Fusion Device with Bone Graft, Lumbar |
| Product Code: | MAX |
3. Purpose of Submission
The purpose of this submission is to gain clearance for a new intervertebral body fusion device.
4. Predicate Device Information
The Leva™ Spacer System described in this submission is substantially equivalent to the following predicates:
| Primary Predicate Device | Manufacturer | 510(k) No. |
|---|---|---|
| Distractable Wave Cage | AMT | K083626 |
| Additional Predicate Devices | Manufacturer | 510(k) No. |
| Patriot™ Spacer | Globus Medical | K072970 |
| Ray Threaded Fusion Cage | Stryker Spine | P950019 |
{4}------------------------------------------------
5. Device Description
The Leva™ Spacer System is a lumbar intervertebral body fusion device fabricated from unalloyed commercially pure (Class II) titanium per ASTM F67. The device is available in both expandable and fixed configurations. The expandable device is provided in a collapsed form and is expanded to a predefined height in-situ using the Leva™ Inserter Instrument. Both the fixed and expandable implants are provided in different heights to accommodate the anatomical needs for a range of patients. The implants are designed to accommodate autogenous bone graft material. The implants have curved endplates to conform to the bony endplates of the patient and ridges on the endplates to resist expulsion.
6. Intended Use
The Leva™ Spacer System is indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). The Leva™ Spacer System is to be used with autogenous bone graft and with supplemental fixation. Patients should have at least six (6) months of nonoperative treatment prior to treatment with an intervertebral body fusion device.
7. Comparison of Technological Characteristics
The substantial equivalence of the Leva™ Spacer System to the predicates is shown by similarity in intended use, indications for use, materials and performance.
8. Performance Data
The following testing was performed to demonstrate the substantial equivalence of the Leva™ Spacer System to its predicate:
- Static and dynamic axial compression (per ASTM F2077) ●
- Static and dynamic compression shear (per ASTM F2077) ●
- Subsidence (per ASTM F2267) .
- Cadaveric Validation .
9. Conclusion
Based on the indications for use, technological characteristics, performance testing and comparison to the predicates, the Leva™ Spacer System has been shown to be substantially equivalent to the predicate devices identified in this submission, and does not present any new issues of safety or effectiveness.
N/A