(110 days)
The FloSpine Canaveral Pedicle System is intended for posterior, noncervical pedicle fixation as an adjunct to fusion in skeletally mature patients using autograft for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
When used for posterior non-cervical pediction in pediatric patients, the Canaveral Pedicle Screw System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The Canaveral Pedicle Screw System for pediatric use is intended to be used with autograft. Pediatric pedicle screw fixation is limited to the posterior approach.
The Canaveral Pedicle Screw System is indicated to provide the surgeon with a minimally invasive approach for posterior spine surgery.
The FloSpine Canaveral Pedicle Screw System comprises non-sterile, single use, titanium alloy components for creating a posterior spinal implant construct. The system attaches to the spine through a component system comprising of pedicle screw assemblies for open and minimally invasive procedures, rods, hooks and set screws. The system is designed to stabilize the spine during the fusion process. The screws are available as monoaxial, uniplanar and polyaxial screws in both cannulated and non-cannulated forms. The rods are available as straight and precurved rods and are available in 2 diameters and multiple lengths. Components are made of Ti6Al4V ELI, a titanium based alloy, which complies with ASTM F136. Some rods are made from cobalt chrome which complies with ASTM F1537.
The provided text is a 510(k) Summary for the Canaveral Pedicle Screw System. It outlines the device description, indications for use, and a comparison to predicate devices, along with performance standards for pre-clinical testing.
However, the document does not contain information on acceptance criteria, a study proving the device meets acceptance criteria, sample sizes, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, type of ground truth used, or details about training sets. The primary focus of this document is to establish substantial equivalence to predicate devices for regulatory clearance based on materials, function, sizes, and mechanical test results, as per ASTM F1717-10.
Therefore,Based on the provided document, the following is what can be extracted:
-
Table of acceptance criteria and the reported device performance:
The document refers to mechanical testing as per ASTM F1717-10, but does not provide specific acceptance criteria values or the device's reported performance against those criteria in a table format. It states that the device is substantially equivalent to predicate devices based on "mechanical test results." -
Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
This information is not available in the provided text. The document refers to "pre-clinical testing" for mechanical performance but does not specify sample sizes or data provenance. -
Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
This information is not applicable and not available. The document describes mechanical performance testing, not a study involving human assessment of images or clinical outcomes that would require expert-established ground truth. -
Adjudication method (e.g. 2+1, 3+1, none) for the test set:
This information is not applicable and not available. As above, the testing described is mechanical, not involving human adjudication. -
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This information is not applicable and not available. The device is a "Pedicle Screw Spinal System," a physical implant, not an AI-powered diagnostic or assistive tool for human readers. -
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This information is not applicable and not available, as the device is a physical implant, not an algorithm. -
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
The "ground truth" for the mechanical testing would be defined by the specified parameters and methodologies of the ASTM F1717-10 standard itself. It's about engineering specifications and material properties, not clinical ground truth. -
The sample size for the training set:
This information is not applicable and not available. The document describes a physical medical device, not an AI or machine learning model that would require a training set. -
How the ground truth for the training set was established:
This information is not applicable and not available, for the same reason as point 8.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, layered on top of each other. The profiles are connected to a wing-like shape, giving the impression of a bird in flight. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 27, 2014
FloSpine, LLC % Rich Jansen, Pharm.D. Silver Pine Consulting, LLC 11821 Bramble Cove Drive Fort Myers, Florida 33905
Re: K141850
Trade/Device Name: Canaveral Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MNH, MNI, KWP Dated: September 26, 2014 Received: September 29, 2014
Dear Dr. Jansen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm_ Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Ronald P. Jean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K141850
Device Name Canaveral Pedicle Screw System
Indications for Use (Describe)
The FloSpine Canaveral Pedicle System is intended for posterior, noncervical pedicle fixation as an adjunct to fusion in skeletally mature patients using autograft for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
When used for posterior non-cervical pediction in pediatric patients, the Canaveral Pedicle Screw System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The Canaveral Pedicle Screw System for pediatric use is intended to be used with autograft. Pediatric pedicle screw fixation is limited to the posterior approach.
The Canaveral Pedicle Screw System is indicated to provide the surgeon with a minimally invasive approach for posterior spine surgery.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary K141850
| Date Prepared: | October 21, 2014 |
|---|---|
| Contact: | James SpitlerFloSpine, LLC9288 Lake Serena Dr.Boca Raton, FL 33496561-705-3080 |
| Regulatory Contact: | Rich Jansen, Pharm. D.Silver Pine Consulting, LLCrichj@s-pineconsulting.com |
| Trade Names:Product Class:Classification: | Canaveral Pedicle Screw SystemClass III21 CFR §888.3070 Pedicle Screw Spinal System, and888.3050 Spinal Interlaminal Fixation Orthosis |
| Common Name:Product Codes:Panel Code: | Pedicle Screw SystemNKB, OSH, MNI, MNH, KWP87 |
Indications for Use:
The FloSpine Canaveral Pedicle System is intended for posterior, noncervical pedicle fixation as an adjunct to fusion in skeletally mature patients using autograft and/or allograft for the following indications: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.
When used for posterior non-cervical pedicle screw fixation in pediatric patients, the Canaveral Pedicle Screw System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The Canaveral Pedicle Screw System for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to the posterior approach.
The Canaveral Pedicle Screw System is indicated to provide the surgeon with a minimally invasive approach for posterior spine surgery.
Device Description:
The FloSpine Canaveral Pedicle Screw System comprises non-sterile, single use, titanium alloy components for creating a posterior spinal implant construct. The system attaches to the spine through a component system comprising of pedicle screw assemblies for open and minimally invasive procedures, rods, hooks and set screws. The system is designed to stabilize the spine during the fusion process. The screws are available as monoaxial, uniplanar and polyaxial screws in both cannulated and non-cannulated forms. The rods are available as straight and precurved rods and are available in 2 diameters and multiple lengths. Components are made of
{4}------------------------------------------------
Ti6Al4V ELI, a titanium based alloy, which complies with ASTM F136. Some rods are made from cobalt chrome which complies with ASTM F1537.
Predicate Device(s):
The FloSpine Pedicle Screw is substantially equivalent to the Synthes Matrix Pedicle Screw System (K120838)*, The Orthofix Matrix Spinal System (K120838) and the U&I Optimal System (K031585).
Performance Standards:
The pre-clinical testing performed includes static and dynamic compression bending, and static torsion per ASTM F1717-10.
Conclusion:
FloSpine concludes that the FloSpine Canaveral Pedicle Screw System is substantially equivalent to the predicate pedicle screw system in regards to indications for use, materials, function, sizes and mechanical test results and raises no new questions of safety or effectiveness.
§ 888.3070 Thoracolumbosacral pedicle screw system.
(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.