K Number
K141820
Date Cleared
2014-10-21

(106 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Indicated for use in selective cannulation of the biliary ducts including, but not limited to the common bile, cystic, right and left hepatic ducts. The guidewires are designed to be used during endoscopic biliary procedures for catheter introduction and exchanges.

The Jagtail™ High Performance Guidewire Extension attached to the extendable Jagwire™ 260 cm Guidewire is designed for use during endoscopic biliary procedures for non-rapid exchange catheter introduction and exchanges. The attachment of the Jagtail High Performance Guidewire Extension to the extendable Jagwire 260 cm guidewire creates an extended guidewire that can be used to exchange out a non-rapid exchange biliary catheter without removing the original guidewire from the associated duct. When the exchange is completed, the Jagtail High Performance Guidewire Extension can be detached and the original, extendable Jagwire 260 cm guidewire can be used in a conventional manner.

Device Description

Jagwire™ High Performance Guidewire, and Jagwire™ High Performance Extendable Guidewire:
The Boston Scientific Jagwire™ High Performance Guidewire is constructed of a metal alloy core, which is encapsulated in a striped PTFE jacket with a 5 cm hydrophilic radiopaque distal tip. The striped jacket provides endoscopically visible movement markings.

Jagtail™ High Performance Guidewire Extension:
The Boston Scientific Jagtail™ High Performance Guidewire Extension is a 200 cm long guidewire extension constructed of a stainless steel core with a connector at the distal end. The shaft is coated with a PTFE sleeve. The Jagtail High Performance Guidewire Extension is exclusively compatible with the Jagwire High Performance Extendable Guidewire, 260 cm. The distal end of the Jagtail High Performance Guidewire Extension is inserted into the proximal end of the extendable Jagwire 260 cm guidewire with the use of the alignment tool. The alignment tool is then removed from the Jagtail High Performance Guidewire Extension.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device called "Jagwire™ High Performance Guidewire" and its extensions. It is not an AI/ML device, therefore, a lot of the metrics that you requested are not relevant here. However, I will extract and provide the relevant information.

The document discusses the regulatory approval for a guidewire, which is a physical medical device used in endoscopic biliary procedures. The "acceptance criteria" and "study" described in the document relate to the physical and biological performance of this guidewire, rather than an AI/ML algorithm's diagnostic or predictive capabilities.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategorySpecific Test/Performance CharacteristicReported Device Performance
BiocompatibilityCytotoxicityAcceptable results
SensitizationAcceptable results
Intracutaneous ReactivityAcceptable results
LatexAcceptable results
USP Physiochemical testingAcceptable results
Physical PerformanceTip Adhesion Shear StrengthMeets Product Specifications
End Product Tip TensileMeets Product Specifications
Tip Column StrengthMeets Product Specifications
Dimensional Test for Tip Outer DiameterMeets Product Specifications
Tip Cantilever StrengthMeets Product Specifications
Atraumatic TipMeets Product Specifications
Appearance and CleanlinessMeets Product Specifications

2. Sample size used for the test set and the data provenance:

The document does not specify a "sample size" in the context of an algorithm's test set or data provenance (e.g., country of origin, retrospective/prospective) because this applies to a physical device, not a data-driven model.

  • Sample Size for Testing: Not explicitly stated as "sample size" in the context of a dataset. Instead, it refers to multiple units of the physical device or its components tested.
  • Data Provenance: Not applicable as it's not a data-driven study, but rather a performance evaluation of a manufactured device.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not applicable. "Ground truth" in the context of expert consensus is relevant for AI/ML diagnostic studies, not for the physical performance testing of a medical guidewire. The performance was assessed against established engineering and biocompatibility standards.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

This information is not applicable. Adjudication methods are used in diagnostic studies to resolve discrepancies in expert opinions for ground truth establishment. For physical device testing, the results are typically quantitative measurements against predefined specifications.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This is not an AI/ML device, so no MRMC comparative effectiveness study involving human readers or AI assistance was performed.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

This is not an AI/ML device, so no standalone algorithm performance was evaluated.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

The "ground truth" for this device's performance is established by engineering specifications, material science standards (e.g., biocompatibility standards like ISO 10993-1), and mechanical performance metrics. The device is deemed acceptable if it meets these predefined physical and biological requirements.

8. The sample size for the training set:

This information is not applicable. The device is a physical product, not an AI/ML model, so there is no "training set."

9. How the ground truth for the training set was established:

This information is not applicable as there is no training set for a physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing right, with flowing lines connecting them. The profiles are meant to represent the department's focus on health and human well-being.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 14, 2014

Boston Scientific Corp. Allison Baillie Regulatory Affairs Specialist II 100 Boston Scientific Way Marlborough, MA 01752

K141820 Re:

Trade/Device Name: Jagwire™ High Performance Guidewire Jagwire™ High Performance Extendable Guidewire Jagwire™ High Performance Guidewire Extension Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: OCY Dated: July 3. 2014 Received: July 7, 2014

Dear Allison Baillie,

This letter corrects our substantially equivalent letter of October 21, 2014.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Joyce M. Whang -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141820

Device Name

Jagwire™ High Performance Guidewire, Jagwire™ High Performance Extendable Guidewire

Indications for Use (Describe)

Indicated for use in selective cannulation of the biliary ducts including, but not limited to the common bile, cystic, right and left hepatic ducts. The guidewires are designed to be used during endoscopic biliary procedures for catheter introduction and exchanges.

Type of Use (Select one or both, as applicable):

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K141820

Device Name

Jagtail™ High Performance Guidewire Extension

Indications for Use (Describe)

The Jagtail™ High Performance Guidewire Extension attached to the extendable Jagwire™ 260 cm Guidewire is designed for use during endoscopic biliary procedures for non-rapid exchange catheter introduction and exchanges. The attachment of the Jagtail High Performance Guidewire Extendable Jagwire 260 cm guidewire creates an extended guidewire that can be used to exchange out a non-rapid exchange biliary catheter without removing the original guidewire from the associated duct. When the exchange is completed, the Jagail High Performance Guidewire Extension can be detached and the original, extendable Jagwire 260 cm guidewire can be used in a conventional manner.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K141820 Pg. 1 of 3

SECTION 5 510(k) SUMMARY

1. Submitter

Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752 Telephone: 508-683-4942 Fax: 508-683-5939

Contact: Allison Baillie Regulatory Affairs Specialist II Date Prepared: October 17, 2014

2. Device

Trade Names:Jagwire™ High Performance Guidewire,Jagwire™ High Performance Extendable Guidewire,Jagtail™ High Performance Guidewire Extension
Common Name:Endoscopic Guidewire
Classification Name:Endoscope and accessories
Regulation Number:876.1500
Product Code:OCY
Classification:Class II

3. Predicate Devices

Trade Name:Jagwire™ High Performance Guidewire,Jagwire™ High Performance Extendable Guidewire,Jagtail™ High Performance Guidewire Extension
Manufacturer andClearance Number:Boston Scientific Corporation, K960186
Common Name:Endoscopic Guidewire
Classification Name:Endoscope and accessories
Regulation Number:876.1500
Product Code:EZB, OCY
Classification:Class II (previously 510(k) exempt under Procode EZB)
Trade Name:Sensor Guidewire
Manufacturer andClearance Number:Boston Scientific Corporation, 510(k) Exempt
Common Name:Endoscopic Guidewire
Classification Name:Endoscope and accessories
Regulation Number:876.1500
Product Code:EZB
Classification:Class II (previously 510(k) exempt under Procode EZB)

{5}------------------------------------------------

4. Device Description

Jagwire™ High Performance Guidewire, and Jagwire™ High Performance Extendable Guidewire:

The Boston Scientific Jagwire™ High Performance Guidewire is constructed of a metal alloy core, which is encapsulated in a striped PTFE jacket with a 5 cm hydrophilic radiopaque distal tip. The striped jacket provides endoscopically visible movement markings.

Jagtail™ High Performance Guidewire Extension:

The Boston Scientific Jagtail™ High Performance Guidewire Extension is a 200 cm long guidewire extension constructed of a stainless steel core with a connector at the distal end. The shaft is coated with a PTFE sleeve. The Jagtail High Performance Guidewire Extension is exclusively compatible with the Jagwire High Performance Extendable Guidewire, 260 cm. The distal end of the Jagtail High Performance Guidewire Extension is inserted into the proximal end of the extendable Jagwire 260 cm guidewire with the use of the alignment tool. The alignment tool is then removed from the Jagtail High Performance Guidewire Extension.

5. Indication for Use:

Jagwire™ High Performance Guidewire, and Jagwire™ High Performance Extendable Guidewire:

Indicated for use in selective cannulation of the biliary ducts including, but not limited to the common bile, cystic, right and left hepatic ducts. The guidewires are designed to be used during endoscopic biliary procedures for catheter introduction and exchanges.

Jagtail™ High Performance Guidewire Extension:

The Jagtail™ High Performance Guidewire Extension attached to the extendable Jagwire™ 260 cm Guidewire is designed for use during endoscopic biliary procedures for non-rapid exchange catheter introduction and exchanges. The attachment of the Jagtail High Performance Guidewire Extension to the extendable Jagwire 260 cm guidewire creates an extended guidewire that can be used to exchange out a non-rapid exchange biliary catheter without removing the original guidewire from the associated duct. When the exchange is completed, the Jagtail High Performance Guidewire Extension can be detached and the original, extendable Jagwire 260 cm guidewire can be used in a conventional manner.

{6}------------------------------------------------

6. Technological Characteristics:

The materials used to attach the Tungsten/Pebax tip of the proposed Jagwire™ Guidewires have changed slightly. The current adhesive primer material used to attach the current Jagwire guidewire tip has been discontinued by the supplier.

Two options for replacement a material for the discontinued adhesive primer material are proposed. One option is a direct replacement of the discontinued adhesive primer material. The second option is a material that is coextruded with the guidewire tip, which has also simplified the tip attachment process.

With the exception of the changes in the tip attachment materials, the proposed Jagwire™ guidewires are identical in design, materials and intended use to the currently commercialized Jagwire™M guidewires. There were no changes made to the Jagtail™ High Performance Guidewire Extension.

7. Performance Data:

Non-clinical testing was successfully performed on the proposed Jagwire™ Guidewires as described below.

Biocompatibility Testing Summary:

The proposed Jagwire Guidewires were assessed in accordance with ISO 10993-1, and are classified as surface, limited contact duration (<24hours), mucous membrane contact devices. All required testing was completed, which included testing for: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Latex and USP Physiochemical testing was completed with acceptable results.

Performance Testing Summary:

Non-clinical comparative performance bench testing was successfully completed to demonstrate that the new adhesive primer, and new resin (coextruded with the tip of the proposed Jagwire) meets the Product Specifications of the currently marketed Jagwire Guidewires. This testing included: Tip Adhesion Shear Strength, End Product Tip Tensile, Tip Column Strength, Dimensional Test for Tip Outer Diameter, Tip Cantilever Strength, Atraumatic Tip, and Appearance and Cleanliness.

8. Conclusion:

Boston Scientific has demonstrated that the proposed Jagwire™ Guidewires are substantially equivalent to the currently marketed Jagwire™ Guidewires.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.