K Number
K141715
Device Name
GLOVE PORT
Manufacturer
Date Cleared
2015-03-25

(273 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Glove Port is intended to provide access for multiple instruments and/or endoscope to the abdominal cavity through a single incision during minimally invasive laparoscopic surgery.

Device Description

The Glove Port is intended to provide access for multiple instruments and/or endoscope to the abdominal cavity through a single incision during minimally invasive laparoscopic surgery. This device is single use and sterilized.

The GLOVE PORT is laparoscopic instrument port which retracts a small abdominal incision to allow multiple laparoscopic instruments to pass through to the abdomen.

AI/ML Overview

The document describes a 510(k) premarket notification for the "Glove Port" device, which is an endoscopy surgery instrument. The submission aims to demonstrate substantial equivalence to legally marketed predicate devices.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the bench testing performed to demonstrate functionality and mechanical safety. The reported device performance is that these tests showed the device is "as safe, as effective and performs in a substantially equivalent manner to the predicate device."

Acceptance Criteria (Implied from Bench Tests)Reported Device Performance
Comparative leak rate (without instruments)Substantially equivalent to predicate devices
Comparative leak rate (with instruments)Substantially equivalent to predicate devices
Comparative leak rate (after vigorous manipulation of instruments)Substantially equivalent to predicate devices
Insufflation flow rateSubstantially equivalent to predicate devices
Insertion-withdrawal forces of instrumentsSubstantially equivalent to predicate devices
Determination of minimum size of skin incisionSubstantially equivalent to predicate devices
Evaluation of device fixationSubstantially equivalent to predicate devices

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample size used for the test set or the data provenance (e.g., country of origin of the data, retrospective or prospective). It simply states that "Bench testing is performed."

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not mention the use of experts to establish ground truth for the bench tests. The tests appear to be physical or mechanical assessments rather than clinical evaluations requiring expert interpretation.

4. Adjudication Method for the Test Set

Not applicable. There is no mention of adjudication for the bench tests.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done. The study described is bench testing, not a clinical study involving human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. The "Glove Port" is a physical medical device, not an algorithm or AI.

7. The Type of Ground Truth Used

For the bench tests, the "ground truth" would be the established functional and safety specifications for such devices, determined by engineering, regulatory, and mechanical principles. The comparison is made against the performance of predicate devices.

8. The Sample Size for the Training Set

Not applicable. The "Glove Port" is a physical medical device, not an AI or algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure in profile, with three faces overlapping to create a sense of depth and community. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 25, 2015

NELIS % Mr. Peter Chung Plus Global 300 Atwood Street Pittsburgh, Pennsylvania 15213

Re: K141715

Trade/Device Name: Glove Port Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OTJ Dated: February 10, 2015 Received: February 23, 2015

Dear Mr. Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K141715

Device Name Glove Port

Indications for Use (Describe)

The Glove Port is intended to provide access for multiple instruments and/or endoscope to the abdominal cavity through a single incision during minimally invasive laparoscopic surgery.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

[as required by 807.92(c)]

1. Applicant

    1. Company : NELIS
    1. Address :1005, 201-Dong Bucheon Techno Park Ssangyong 3-cha Samjeong-dong Ojeong-gu, Bucheon-si Gyeongggi-do, 421-742 Korea
    1. Tel : +82(32)624-1697,
    1. Fax : +82(32)624-1699
    1. Homepage : www.nelis.co.kr
    1. Contact person : Peter Chung, 412-687-3976
    1. Contact person address : 300, Atwood Street, Pittsburgh, PA, 15213, USA
    1. Submission date : Apr. 23, 2014

2. Device Information

    1. Trade Name : GLOVE PORT
    1. Common Name : Endoscopy Surgery Instrument
    1. Classification Name : Laparoscope, general & plastic surgery
    1. Product Code : OTJ
    1. Regulation Number : 876.1500
    1. Class of device : Class II
    1. Panel : General & Plastic Surgery

3. The legally marketed predicate devices :

4. Device description :

The Glove Port is intended to provide access for multiple instruments and/or endoscope to the abdominal cavity through a single incision during minimally invasive laparoscopic surgery. This device is single use and sterilized.

5. Intended Use :

The Glove Port is intended to provide access for multiple instruments and/or endoscope to the abdominal cavity through a single incision during minimally invasive laparoscopic surgery.

6. Technological Characteristics :

The GLOVE PORT is laparoscopic instrument port which retracts a small abdominal incision to allow multiple laparoscopic instruments to pass through to the abdomen .

7. Performance data:

Bench testing is performed to demonstrate the functionality and mechanical safety as following items

  • a. comparative leak rate test to evaluate the leak rate without instruments, with instruments, and after vigorous manipulation of instruments vs. predicate device(s)
  • b. insufflation flow rate
  • c. insertion-withdrawal forces of instruments
  • d. determination of minimum size of skin incision
  • e. evaluation of GLOVE PORT device fixation

{4}------------------------------------------------

8. Predicate device comparison table

ManufacturerNELISCovidienAdvanced SurgicalConcepts
510(k) No.K141715K073719K093372
Indicationfor useThe Glove Port is intended to provide access for multiple instruments and/or endoscope to the abdominal cavity through a single incision during minimally invasive laparoscopic surgery.The ASC TriPort Laparoscopic Access Device is intended for use as a multiple instrument and/or camera port during minimally invasive abdominal laparoscopic surgery.The SILSTM Port is indicated for multiple instrument or camera access to the abdominal cavity through a single incision for performing minimally invasive laparoscopic procedures.
ProductnameEndoscopy Surgery InstrumentSurgical TrocarLaparoscopicAccessory
TradenameGlove PortOriginalGlove PortHGlove Port AGlove PortATSILSTM PortASC TriPortLaparoscopicAccess Device
Model60ea14ea33ea9ea3 ea4ea
ProductconfigurationImage: Glove Port OriginalImage: Glove Port HImage: Glove Port AImage: Glove Port ATImage: SILSTM PortImage: ASC TriPort Laparoscopic Access Device
MaterialPolyurethanePolyurethanePolyurethanePolyurethaneSiliconeSilicone
Port No.3~4 ea3 ea3~4 ea
Absolutesize10mm~25mm25mm10mm~25mm
Installationandabdominalwall innerfixationWound retractor ring being fixed inside the abdominal wall inside.Non-fixedDistal ring being fixed in the abdominal wall inside.
SterilizationmethodEO Gas sterilizationGamma sterilization(Cobalt-60)Gamma sterilization(Cobalt-60)
CertificationCE0120, ISO 9001:2008, ISO 13485:2003, Registration of KFDARegistered 510(k)by FDARegistered 510(k)by FDA

9. Conclusion:

The Device is investigated for function and effectiveness to compare the operation of function between Glove Port and predicate devices. Comparison results demonstrate that the specifications and performance of the device are same as functional and effective as the legally marketed predicate device.

Therefore, it is concluded that Glove Port is substantially equivalent to the legally marketed predicate device.

The performance tests demonstrated that GLOVE PORT is as safe, as effective and performs in a substantially equivalent manner to the predicate device.

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.