K Number
K141465
Device Name
CARDIOGARD EMBOLI PROTECTION CANNULA
Date Cleared
2015-01-09

(220 days)

Product Code
Regulation Number
870.4210
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The CardioGard Emboli Protection Cannula combines the function of a standard arterial cannula with an added suction mechanism to capture debris that may result from cardiac surgery. The CardioGard Emboli Protection Cannula is intended for perfusion of the ascending aorta during short term (≤ 6 hours) cardiopulmonary bypass (CPB) procedures. The CardioGard suction lumen is intended for the removal of particulate emboli during surgical procedures that require CPB.
Device Description
The CardioGard Emboli Protection Cannula, shown below, is a disposable 24 French double lumen arterial cannula. The cannula functions to deliver oxygenated blood to the heart during procedures requiring cardiopulmonary bypass (CPB), while at the same time suctioning blood and embolic matter away from the surgical field. The arterial cannula is inserted centrally in the ascending aorta. The CardioGard Emboli Protection Cannula features a tip configuration which diffuses oxygenated blood from the heart-lung machine to the ascending aorta through the Cannula Outlet, while also aspirating blood and embolic matter through the Suction Lumen Inlet. The flow rates through the two cannula lumens are carefully controlled so that emboli are suctioned back to the CPB machine for filtration while still enabling sufficient blood flow into the aorta.
More Information

Not Found

No
The description focuses on the mechanical function of the cannula and suction mechanism, with no mention of AI or ML.

Yes
The device is intended to treat a pathological condition (particulate emboli during cardiac surgery) and provides a therapeutic benefit (removal of emboli to reduce new brain lesions).

No

The device is an arterial cannula designed to capture debris and deliver oxygenated blood during cardiac surgery. Its primary function is therapeutic/interventional (removing emboli and providing perfusion), not to diagnose a condition.

No

The device is a physical cannula with a suction mechanism, clearly described as a disposable hardware component used during surgery.

Based on the provided text, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is for "perfusion of the ascending aorta during short term (≤ 6 hours) cardiopulmonary bypass (CPB) procedures" and "removal of particulate emboli during surgical procedures that require CPB." This describes a device used in vivo (within the body) during a surgical procedure.
  • Device Description: The description details a "double lumen arterial cannula" that "delivers oxygenated blood" and "suctioning blood and embolic matter away from the surgical field." This is a surgical tool used directly on the patient.
  • Lack of IVD Characteristics: There is no mention of the device being used to examine specimens in vitro (outside the body) to provide information for diagnosis, monitoring, or screening. The device's function is entirely focused on managing blood flow and removing debris during surgery.

Therefore, the CardioGard Emboli Protection Cannula is a surgical device, not an In Vitro Diagnostic device.

N/A

Intended Use / Indications for Use

The CardioGard Emboli Protection Cannula combines the function of a standard arterial cannula with an added suction mechanism to capture debris that may result from cardiac surgery. The CardioGard Emboli Protection Cannula is intended for perfusion of the ascending aorta during short term (≤ 6 hours) cardiopulmonary bypass (CPB) procedures. The CardioGard suction lumen is intended for the removal of particulate emboli during surgical procedures that require CPB.

Product codes (comma separated list FDA assigned to the subject device)

DWF

Device Description

The CardioGard Emboli Protection Cannula, shown below, is a disposable 24 French double lumen arterial cannula. The cannula functions to deliver oxygenated blood to the heart during procedures requiring cardiopulmonary bypass (CPB), while at the same time suctioning blood and embolic matter away from the surgical field. The arterial cannula is inserted centrally in the ascending aorta.

The CardioGard Emboli Protection Cannula features a tip configuration which diffuses oxygenated blood from the heart-lung machine to the ascending aorta through the Cannula Outlet, while also aspirating blood and embolic matter through the Suction Lumen Inlet. The flow rates through the two cannula lumens are carefully controlled so that emboli are suctioned back to the CPB machine for filtration while still enabling sufficient blood flow into the aorta.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

ascending aorta

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Nonclinical Data:
Sterilization Validation: EtO sterilization was validated to an SAL of 10° using the overkill method, half-cycle technique in accordance with EN ISO 11135-1:2008 Sterilization of health care products - Ethylene Oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices. The LAL test was used to verify that the EtO sterilized CardioGard is pyrogen-free.

Shelf Life Testing: Functionality and package integrity testing was conducted on finished, EtO sterilized devices after accelerated aging equivalent to 2 years. All tests passed, supporting a 2 year labelled shelf life. Functionality tests included:

  • Visual Inspection and Dimensional Verification
  • Back Pressure
  • Pressure Drop
  • Air Leakage
  • Liquid Leakage
  • Force at Break

Package tests included:

  • Peel Strength (Tensile) Test
  • Burst Test
  • Dye Penetration Test

Biocompatibility Testing - Biocompatibility testing was conducted in accordance with ISO 10993-1:2009, "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing," for externally communicating devices in contact with circulating blood for limited (

§ 870.4210 Cardiopulmonary bypass vascular catheter, cannula, or tubing.

(a)
Identification. A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.(b)
Classification. Class II (performance standards).

0

Public Health Service

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus, which is a symbol of medicine, with three lines forming the wings and a snake winding around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 9, 2015

Cardiogard Medical Ltd % Sheila Hemeon-Heyer President Heyer Regulatory Solutions 125 Cherry Lane Armherst, Massachusetts 01002

Re: K141465

Trade/Device Name: Cardiogard Emboli Protection Cannula Regulation Number: 21 CFR 870.4210 Regulation Name: Cardiopulmonary Bypass Vascular Catheter, Cannula, Or Tubing Regulatory Class: Class II Product Code: DWF Dated: December 10, 2014 Received: December 11, 2014

Dear Sheila Hemeon-Heyer,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

1

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

M.G. Hillebrand

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

2

510(k) Number: K141465

CardioGard Emboli Protection Cannula Device Name:

Indications for Use:

The CardioGard Emboli Protection Cannula combines the function of a standard arterial cannula with an added suction mechanism to capture debris that may result from cardiac surgery. The CardioGard Emboli Protection Cannula is intended for perfusion of the ascending aorta during short term (≤ 6 hours) cardiopulmonary bypass (CPB) procedures. The CardioGard suction lumen is intended for the removal of particulate emboli during surgical procedures that require CPB.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

3

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

| A. Submitter: | Heyer Regulatory Solutions LLC
P.O. Box 2151
Amherst, MA 01004-2151
Contact: Sheila Hemeon-Heyer
Sheila@heyer-regulatory.com |

---------------------------------------------------------------------------------------------------------------------------------------------------------
  • B. Manufacturer/ CardioGard Medical Ltd. 510(k) Applicant: 6 Yoni Netanyahu Street 6037604 Or-Yehuda lsrael Contact: Walid Haddad, PhD Chief Operating Officer +972 (3) 546 7163 walid@cardiogard.com
  • C. Date Prepared: January 9, 2015

Device Name and Classification Information: D.

Trade Name:CardioGard Emboli Protection Cannula
Classification Name:Catheter, Cannula and Tubing, Vascular,
Cardiopulmonary Bypass
Product Code, CFR:DWF, 21 CFR 870.4210
Review Panel:Cardiovascular Devices
Class:II
Predicate Device(s):Embol-X Access Device / Aortic Cannula,
cleared under K102420 and K020693, and
Embol-X Intra-Aortic Filter cleared under
K062429, K031946, and K022071

F. Device Description:

The CardioGard Emboli Protection Cannula, shown below, is a disposable 24 French double lumen arterial cannula. The cannula functions to deliver oxygenated blood to the heart during procedures requiring cardiopulmonary

4

bypass (CPB), while at the same time suctioning blood and embolic matter away from the surgical field. The arterial cannula is inserted centrally in the ascending aorta.

The CardioGard Emboli Protection Cannula features a tip configuration which diffuses oxygenated blood from the heart-lung machine to the ascending aorta through the Cannula Outlet, while also aspirating blood and embolic matter through the Suction Lumen Inlet. The flow rates through the two cannula lumens are carefully controlled so that emboli are suctioned back to the CPB machine for filtration while still enabling sufficient blood flow into the aorta.

Image /page/4/Picture/3 description: The image shows two diagrams related to the CardioGard device. The left diagram, labeled "CardioGard Features," illustrates the components of the CardioGard device, including the suction tube, main tube, suction lumen inlet, and cannula outlet. The right diagram, titled "CardioGard Inserted in the Heart-Lung Circuit," depicts how the CardioGard device is integrated into the heart-lung circuit during a medical procedure.

G. Indication for Use:

The CardioGard Emboli Protection Cannula combines the function of a standard arterial cannula with an added suction mechanism to capture debris that may result from cardiac surgery.

The CardioGard Emboli Protection Cannula is intended for perfusion of the ascending aorta during short term (≤ 6 hours) cardiopulmonary bypass (CPB) procedures. The CardioGard suction lumen is intended for the removal of particulate emboli during surgical procedures that require CPB.

5

H. Technical Comparison with Predicate Devices and Discussion of Similarities and Differences

Information for the predicate device was obtained from the 510(k) Summary and other publicly available sources.

| | CardioGard Emboli Protection Cannula | Embol-X Access Device / Aortic
Cannula
(K102420, K020693) | Embol-X Intra-Aortic Filter
(K062429, K031946, K022071) |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended
use | The CardioGard Emboli Protection
Cannula combines the function of a
standard arterial cannula with an added
suction mechanism to capture debris that
may result from cardiac surgery. The
CardioGard Emboli Protection Cannula is
intended for perfusion of the ascending
aorta during short term (≤ 6 hours)
cardiopulmonary bypass (CPB)
procedures. The CardioGard suction
lumen is intended for the removal of
particulate emboli during surgical
procedures that require CPB. | The EMBOL-X Access Device/Aortic
Cannula is indicated for the perfusion of
the ascending aorta during short-term (≤ 6
hours) cardiopulmonary bypass (CPB)
surgery where procedures may require the
hemostatic introduction and removal of
compatible intravascular devices into the
vascular system. | The Embol-X Intra-aortic filter is
indicated for use with the Embol-X
Access Device/Aortic Cannula in first
time, non-emergent cardiac surgery
procedures requiring aortic
crossclamp, to capture and remove
particulate emboli from the ascending
aorta and heart in patients aged 18
years and older. |
| Use
duration | For the length of the CPB surgery (≤ 6
hours) | For the length of the CPB surgery (≤ 6
hours) | The filter may remain in situ for up to
60 minutes. |
| Design | Two lumen, curved tip with two ports.
One port is for aortic perfusion of
oxygenated blood to the patient. The
second port is for suction of embolic
particles from the aortic arch. | Two lumen, curved tip with two ports. One
port is for aortic perfusion of oxygenated
blood to the patient. The second port is for
insertion of the EMBOL-X filter device,
intended to remove embolic particles from
the aortic arch. | A cartridge that locks into the filter
lumen of the EMBOL-X Cannula. The
filter then opens to fill the diameter of
the ascending aorta. The filter is
intended to be inserted temporarily
during CPB. Filters are manufactured
in 5 sizes to conform to patient
anatomy. |
| | CardioGard Emboli Protection Cannula | Embol-X Access Device / Aortic
Cannula
(K102420, K020693) | Embol-X Intra-Aortic Filter
(K062429, K031946, K022071) |
| Dimensions | Tip Size: 24Fr
Length: 30cm
Main tube diameter: 3/8"
Suction tube diameter: 1/4" | Tip Size: 24Fr; 20 Fr effective flow
Length: 28 cm
Main tube diameter: 3/8"
Embolic filter port diameter: unknown | Five sizes available for aortas from 2.2 cm to 4.0 cm. Filter opens to fill the diameter of the ascending aorta. |
| Schematic
of device | Image: CardioGard Emboli Protection Cannula | Image: Embol-X Access Device / Aortic Cannula | Image: Embol-X Intra-Aortic Filter |
| | CardioGard Emboli Protection Cannula | Embol-X Access Device / Aortic
Cannula
(K102420, K020693) | Embol-X Intra-Aortic Filter
(K062429, K031946, K022071) |
| Pressure
Drop | Image: CardioGard Emboli Protection Cannula Pressure Drop Graph | Image: Embol-X Access Device / Aortic Cannula Pressure Drop Graph | N/A |
| Biological
Safety | Tip material: PVC
Tubing: PVC + Nirosta.
Connector: ABS
Biocompatibility has been confirmed in
accordance with testing under ISO10993. | Polymeric tubes
Biocompatible | Conventional medical grade materials
and processes
Biocompatible |
| Sterilization | EtO sterilized, single-use, disposable,
non-pyrogenic | EtO sterilized, single-use, disposable, non-
pyrogenic | Gamma radiation, single-use,
disposable, non-pyrogenic |

6

7

8

Discussion of Similarities and Differences

Intended Use: The intended use of the CardioGard Cannula is essentially the same as that of the combined Embol-X Aortic Cannula and Intra-Aortic filter. The Embol-X Intra-Aortic Filter can only be used with the Embol-X Aortic Cannula, and the two devices are sold together as the Embol-X Glide System. Both the CardioGard and Embol-X devices are intended for arterial perfusion and embolic particle capture during cardiopulmonary bypass (CPB) surgery.

Use Duration: The CardioGard Cannula is similar to the Embol-X Cannula in that both can be used for the entire duration of the CPB surgery (≤ 6 hours). The duration of use for the Embol-X filter is limited to 60 minutes, and it is intended primarily for use during periods of aortic manipulation (i.e., clamps removal).

Design and Dimensions: The designs of the CardioGard and Embol-X Cannulas are very similar. Both have curved tips with two ports: one for arterial perfusion and one for removal of embolic particles. Both devices are 24Fr with 3/8" diameter for the main tube, and similar tube lengths. Cannula insertion for both devices is per standard surgical practice for arterial perfusion cannulas. The main difference between the two device designs is the method of embolic particle removal: The CardioGard uses suction through the second portal located behind the perfusion portal, while the Embol-X uses a mesh filter inserted into the aorta through the second portal located behind the perfusion portal. In vitro side-by-side comparison testing of the two devices demonstrated the substantial equivalence of these two methods of embolic particle removal.

Pressure Drop:

The pressure drop (difference between the inlet and outlet pressures) testing was conducted for both the CardioGard and Embol-X cannulas under the same protocol. This in vitro side-by-side comparison pressure drop testing demonstrated that the CardioGard cannulas have a smaller, and therefore better, pressure drop as compared to the predicate device at all measured flow rates (3, 4, 5, and 6 l/min).

Biological Safety: The CardioGard Cannula is substantially equivalent to the Embol-X devices in that all are made of materials commonly used in medical devices and are biocompatible for short term contact with circulating blood.

9

Sterilization: The CardioGard Cannula is substantially equivalent to the Embol-X devices in that all are provided sterile, are non-pvrogenic, and are intended for single use only.

. Nonclinical Data:

The following nonclinical testing was provided in this 510(k):

Sterilization Validation: EtO sterilization was validated to an SAL of 10° using the overkill method, half-cycle technique in accordance with EN ISO 11135-1:2008 Sterilization of health care products - Ethylene Oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices. The LAL test was used to verify that the EtO sterilized CardioGard is pyrogen-free.

Shelf Life Testing: Functionality and package integrity testing was conducted on finished, EtO sterilized devices after accelerated aging equivalent to 2 years. All tests passed, supporting a 2 year labelled shelf life. Functionality tests included:

  • Visual Inspection and Dimensional Verification ●
  • Back Pressure ●
  • Pressure Drop .
  • Air Leakage ●
  • Liquid Leakage ●
  • Force at Break ●

Package tests included:

  • . Peel Strength (Tensile) Test
  • Burst Test ●
  • Dye Penetration Test .

Biocompatibility Testing -Biocompatibility testing was conducted in accordance with ISO 10993-1:2009, "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing," for externally communicating devices in contact with circulating blood for limited (