(37 days)
Not Found
No
The document describes calibration materials for an immunoassay system and does not mention any AI or ML components in the device description or performance studies.
No
This device is for in vitro diagnostic use, specifically for the verification of calibration of assays, not for treating or diagnosing patients directly.
No
The device is a calibration material used to verify the calibration of diagnostic assays, not a diagnostic device itself. Its purpose is to ensure the accuracy of other diagnostic systems.
No
The device description clearly states that the device contains physical vials with human serum and other materials, indicating it is a physical product, not software only.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use/Indications for Use: The very first sentence for each product explicitly states "for in vitro diagnostic use". This is the primary indicator.
- Device Description: The description details materials used (human serum, human albumin based matrix) and their purpose in verifying calibration of diagnostic assays.
- Intended User/Care Setting: "For prescription use only" suggests use in a clinical or laboratory setting, consistent with IVDs.
- Summary of Performance Studies: The performance studies describe testing using patient serum samples and commercially available controls, which are typical for validating IVDs.
- Predicate Device(s): The listed predicate devices are also described as "Calibration Verification Material (CVM)", further indicating the device's role in the diagnostic process.
The device is a calibration verification material used to ensure the accuracy of other in vitro diagnostic assays (IMMULITE Cortisol, Folic Acid, and Vitamin B12 assays). This function is integral to the reliable performance of diagnostic testing.
N/A
Intended Use / Indications for Use
The IMMULITE® Cortisol Calibration Material is for in vitro diagnostic use in the verification of calibration of the IMMULITE Cortisol assay on the IMMULITE 2000 systems.
The IMMULITE® Folic Acid Calibration Material is for in vitro diagnostic use in the verification of calibration of the IMMULITE Folic Acid assay on the IMMULITE 2000 systems.
The IMMULITE® Vitanin B12 Calibration Material is for in vitro diagnostic use in the verification of calibration of the IMMULITE Vitamin B12 assay on the IMMULITE 2000 systems.
Product codes (comma separated list FDA assigned to the subject device)
JJX
Device Description
The IMMULITE® 2000 Cortisol Calibration Verification Material (CVM) contains one set of four vials each 3mL. CVM1 contains human serum with preservatives and CVM2, CVM3, and CVM4 contain various levels of cortisol in human serum with preservatives.
The IMMULITE® 2000 Folic Acid Calibration Verification Material (CVM) contains one set of four vials, 3mL (CVM1) each. when reconstituted. CVM1 contains a lyophilized human albumin based matrix with preservatives. CVM2. CVM3. and CVM4 contain various levels of lyophilized folic acid in human albumin based matrix with preservatives.
IMMULITE® 2000 Vitamin B12 Calibration Verification Material (CVM) contains one set of four vials, 3 mL each, when reconstituted. CVM1 contains a lyophilized human albumin-based matrix with preservatives. CVM2, CVM3, and CVM4 contain various levels of lyophilized Vitamin B12 in human albumin-based matrix with preservatives.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
For prescription use only
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Stability Study:
IMMULITE® 2000 Cortisol Calibration Verification Material (CVM):
- Real-time shelf life: 14 months when stored at -20°C prior to opening.
- Open component (in-use or open vial) claim: 8 hours at ambient or room temperature (15-25°C) after opening.
- Protocol: CVM study protocols run as part of calibrator stability testing. Stability calibrators/CVMs run in duplicate at time points: LCOCVM1, LCOCVM2, LCOCVM3, LCOCVM4 tested at 1, 182, 365, 425 days.
- Open component testing: Results determined from a 2-point adjustment. IMMULITE 2000 Cortisol (L2KCO) kit lot 362, lot 090 CVMs tested at 2-hourly intervals for up to 9 hours at ambient or room temperature (15-25°C).
- Acceptance Criteria:
- Part 1 (guideline): dose value of stability calibrators/CVM to fall between +/-20% of assigned dose for CVM level 2, +/-15% of assigned dose for level 3, and +/-19% of assigned dose for level 4.
- Part 2 (review limits): dose value of controls to be within 2 Standard Deviations (SD) of control target value when generated from stability calibrator curve. If Part 1 not met, additional data review conducted using Part 2 criteria.
IMMULITE® 2000 Folic Acid Calibration Verification Material (CVM):
- Real-time shelf life: 3.5 years when stored at -20°C prior to opening.
- Open component (in-use or open vial) claim: 8 hours at ambient or room temperature (15-25°C) after reconstitution.
- Protocol: CVM study protocols run as part of calibrator stability testing. Stability CVMs and reference CVMs run in duplicate at time points: LFOCVM1, LFOCVM2, LFOCVM3, LFOCVM4 tested at 1, 912, 1095, 1280 days.
- Open component testing: Results determined from a 2-point adjustment. IMMULITE 2000 Folic Acid (L2KFO) kit lot 446, lot 090B Folic Acid CVMs tested at 2-hourly intervals for up to 9 hours at ambient or room temperature (15-25°C).
- Acceptance Criteria:
- Part 1 (guideline): dose value of stability CVMs to fall between +/-15% of assigned dose for CVM level 2, within +/-10% of assigned dose for level 3, and within +/-12% of assigned dose for CVM level 4.
- Part 2 (review limits): dose value of controls to be within 2 Standard Deviations (SD) of control target value when generated from stability calibrator curve. If Part 1 not met, additional data review conducted using Part 2 criteria.
IMMULITE® 2000 Vitamin B12 Calibration Verification Material (CVM):
- Real-time shelf life: 69 months when stored at -20°C prior to opening.
- Open component (in-use or open vial) claim: 8 hours at room (ambient) temperature (15-25°C) after reconstitution.
- Protocol: CVM study protocols run as part of calibrator stability testing. Testing run in duplicate with time points: LVBCVM1, LVBCVM2, LVBCVM3, LVBCVM4 tested at 0, 67, 68, 69 months.
- Open component testing: Results determined from a 2-point adjustment. IMMULITE 2000 Vitamin B12 (L2KVB) kit lot 467, lot 108 Vitamin B12 CVMs tested at 2-hourly intervals for up to 9 hours at ambient or room temperature (15-25°C).
- Acceptance Criteria:
- Part 1 (guideline): dose value of stability calibrator/CVM to fall between +/-10% of assigned dose for CVM level 2 and level 3, and within +/-21% of assigned dose for CVM level 4.
- Part 2 (review limits): dose value of controls to be within 2SD of control target value when generated from stability calibrator curve. If Part 1 not met, additional data review conducted using Part 2 criteria.
Value Assignment Test:
IMMULITE® 2000 Cortisol Calibration Verification Material:
- Sample size/Data source: IMMULITE Cortisol CVMs are 4 level materials subset of 8 level Cortisol calibrators. Quality control performed by calculating recovery of patient samples, spiked patient samples, and controls. Six levels of commercially available controls and 25 patient serum samples, 3 spiked normal serum samples, and 5 normal serum samples used.
- Key Results: Value assigned using assigned reference calibrators. CVM dose values generated using curve generated by assigned reference calibrators. CVM values calculated based on recovered values for each run on each instrument independently and averaged across all systems. Expected assay range 1 to 50 g/dL. Guideline Range (95% confidence interval) for each CVM level established based on Target Mean and +/- 2 Standard Deviation (SD).
IMMULITE® 2000 Folic Acid Calibration Verification Material:
- Sample size/Data source: IMMULITE Folic Acid CVMs are 4 level materials subset of 8 level Folic Acid calibrators. Quality control performed by calculating recovery of patient samples and controls, controls must fall within their target ranges. Six levels of commercially available controls and 14 patient serum samples used.
- Key Results: Value assigned using assigned reference calibrators. CVM dose values generated using curve generated by assigned reference calibrators. CVM values calculated based on recovered values for each run on each instrument independently and averaged across all systems. Expected assay range 1 to 24 ng/mL. Guideline Range (95% confidence interval) for each CVM level established based on Target Mean and +/- 2 Standard Deviation (SD).
IMMULITE® 2000 Vitamin B12 Calibration Verification Material:
- Sample size/Data source: IMMULITE® 2000 Vitamin B12 CVMs are 4 level materials subset of 8 level Vitamin B12 calibrators. Quality control performed by calculating recovery of patient samples and controls. Eight levels of commercially available controls and 21 patient serum samples used.
- Key Results: Value assigned using internal standard and assigned reference calibrators. CVM dose values generated using curve generated by assigned reference calibrators. CVM values calculated based on recovered values for each instrument independently and averaged across all systems. Expected assay range 150 to 1000 pg/mL. Guideline Range (95% confidence interval) for each CVM level established based on Target Mean and +/- 2 Standard Deviation (SD).
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 862.1660 Quality control material (assayed and unassayed).
(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
0
1
510(k) Summary
Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence.
The assigned 510(k) Number: ______k1414444
| 1. Submitter
Mailing Address: | Siemens Healthcare Diagnostics Inc.
5210 Pacific Concourse Drive
Los Angeles, CA 90045 |
|------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:
Phone Number:
Fax Number:
E-mail Address:
Date Prepared: | Donna Velasquez
Regulatory Technical Specialist
(310)-645-8200 x7403
(310)-645-9999
donna.velasquez@siemens.com
May 30, 2014 |
| 2. Device Name
Proprietary Name:
Measurand:
Type of Test:
Regulation Section:
Classification:
Products Code:
Panel: | IMMULITE® 2000 Cortisol Calibration Verification Material
Quality Control materials for IMMULITE® 2000 Cortisol assay
Calibration Verification Material (CVM) for IMMULITE® 2000
Cortisol assay
21 CFR 862.1660, Quality Control Material
Class I Reserved
JJX - Single (Specified) Analyte Controls (Assayed and
Unassayed)
Clinical Chemistry (75) |
| 3. Predicate Device Name
Predicate 510(k) No: | IMMULITE® 2000 Total T3 Calibration Verification Material
(CVM)
K133124 |
| 4. Device Description: | The IMMULITE® 2000 Cortisol Calibration Verification Material
(CVM) contains one set of four vials each 3mL. CVM1 contains
human serum with preservatives and CVM2, CVM3, and CVM4
contain various levels of cortisol in human serum
with
preservatives. |
| 5. Intended Use:
Indication for Use: | See Indications for Use Statement below
The IMMULITE® Cortisol Calibration Verification Material
(CVM) is for in vitro diagnostic use in the verification of
calibration of the IMMULITE Cortisol assay on the IMMULITE
2000 systems |
| Special Conditions for
Use Statement(s):
Special Instrument
Requirements: | For prescription use only
IMMULITE® 2000 Systems |
1
6. Technological Characteristics and Substantial Equivalence Comparison with Predicate:
A comparison of the device features, intended use, and other · information demonstrates that the IMMULITE® 2000 Cortisol Calibration Verification Material (CVM) is substantially equivalent to the predicate device as summarized in Table 1.
SIMILARITIES | ||
---|---|---|
Candidate Device | ||
IMMULITE 2000 Cortisol CVM | Predicate Device | |
IMMULITE 2000 Total T3 CVM | ||
Intended | ||
Use | The IMMULITE® Cortisol Calibration | |
Verification Material (CVM) is for in | ||
vitro diagnostic use in the verification of | ||
calibration of the IMMULITE Cortisol | ||
assay on the IMMULITE 2000 systems. | The IMMULITE® Total T3 | |
Calibration Verification Material | ||
(CVM) is for in vitro diagnostic use in | ||
the verification of calibration of the | ||
IMMULITE Total T3 assay on the | ||
IMMULITE 2000 systems | ||
Storage | -20°C | Same |
Form | Liquid | Same |
Stability | Stable unopened until the | |
expiration date | Same | |
Levels | 4 | Same |
Matrix | Human Serum with preservatives | Same |
Use | Single Use Only | Same |
DIFFERENCES | ||
Candidate Device | ||
IMMULITE 2000 Cortisol CVM | Predicate Device | |
IMMULITE 2000 Total T4 CVM | ||
Analyte | Cortisol | Total T3 |
Table 1: Substantial Equivalence Comparison
7. Non-Clinical Performance Testing
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the candidate device.
Stability Summary:
The stability study was conducted to validate real-time shelf life and open component (in-use or open vial) claim for the IMMULITE 2000 Cortisol Calibration Material (CVM) to ensure that it maintains optimal product performance on IMMULITE 2000 platforms throughout the established shelf life of the CVM. The Calibration Verification Materials are stable up to 14 months when stored at -20°C prior to opening and stable for 8 hours at ambient or room temperature (15-25°C) after opening.
2
Stability Protocol Summary:
The CVM study protocols are run as part of the calibrator stability testing. The stability calibrators/CVMs are run in duplicate (as a minimum) at the time points shown in Table 2, and the dose value is determined from the reference calibrator curve.
| CVM Level | Time-Points
(Days) | | | |
|-----------|-----------------------|-----|-----|-----|
| LCOCVM1 | 1 | 182 | 365 | 425 |
| LCOCVM2 | 1 | 182 | 365 | 425 |
| LCOCVM3 | 1 | 182 | 365 | 425 |
| LCOCVM4 | 1 | 182 | 365 | 425 |
Table 2: Stability Protocol Summary
For Open Component testing, the results were determined from a 2-point adjustment. Using IMMULITE 2000 Cortisol (L2KCO) kit lot 362, lot 090 CVMs were tested at 2-hourly intervals for up to 9 hours at ambient or room temperature (15-25°C) conditions.
Stability Acceptance Criteria Summary:
The Acceptance Criteria for the Cortisol Calibration Verification Material (CVM) are in 2 parts. Part 1 consists of guideline acceptance criteria which require dose value of stability
calibrators/CVM to fall between ±20% of assigned dose for CVM level 2, ±15% of assigned dose for level 3 and ±19% of assigned dose for level 4. Part 2 review limits criteria which require dose value of the controls to be within 2 Standard Deviations (SD) of the control target value when generated from the stability calibrator curve. If the result is not within acceptable dose range of ±20% for level 2, ±15% of assigned dose for level 3 and within ±19% of assigned dose for level 4, then additional data review is conducted using part 2 criteria.
Traceability:
The IMMULITE Cortisol CVMs are traceable to an internal material which has been gravimetrically prepared. The CVMs are manufactured using qualified materials and measurement procedures.
Value Assignment:
The IMMULITE Cortisol CVMs are 4 level materials which are a subset of 8 level Cortisol calibrators. Calibrators are not commercialized but are used internally during manufacture and release testing of Cortisol reagents and two point adjustors. The IMMULITE calibrators and therefore CVMs are value assigned using assigned reference calibrators. The CVMs are manufactured using qualified materials and measurement procedures. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples, spiked patient samples and controls using the assigned CVM values. Six levels of commercially available controls and 25 patient serum samples, 3 spiked normal serum samples and 5 normal serum samples were used to validate calibrator/CVM value assignments.
3
Expected Values/Target Values/Reference Range:
Each CVM level was tested for a total of 27 replicates; 9 runs and 3 replicates per run, 3 different reagent kit lots and 8 IMMULITE 2000 systems. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The target values are provided in the IMMULITE® 2000 CVM Calibration Material lotspecific value card. The expected assay range is 1 to 50 g/dL. The target values in Table 3 can be considered as guidelines.
| Analyte target
levels | CVM Level | *Target
Mean
g/dL) | Standard
Deviation
(SD) | Guideline ±2SD
Range | |
|--------------------------|--------------|--------------------------|-------------------------------|-------------------------|------|
| | LCOCVM1 | 0.00 | - | 0. | 1.0 |
| | LCOCVM2 | 4.54 | 0.455 | 3. | 5.45 |
| | LCOCVM3 | 18.7 | 1.4 | 15 | 21.5 |
| | LCOCVM4 | 54.5 | 5.2 | 44 | 64.9 |
| Assay Range | 1 to 50 g/dL | | | | |
Table 3: Analyte Target Range Values
*Note: when CVMs are run by the customer, an actual value below and above the assay range will result when the customer programs the CVMs as calibration verifiers in the instrument software. If programmed as a patient or control, then the software will give values as the assay lower and upper range.
Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, total precision tabulated in the respective assay instructions for use may be considered as one factor when establishing local, acceptable ranges. The values provided above may be considered as guidelines. Value assignment is lot specific.
Standard/Guidance Documents Referenced:
CEN 13640 Stability Testing of In Vitro Diagnostic Reagents
Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators Guidance for Industry and FDA Staff - Assayed and Unassayed Quality Control Material
Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10
8. Conclusion:
The IMMULITE® 2000 Cortisol Calibration Verification Material is substantially equivalent to the predicate device intended for similar use. The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence, product safety and effectiveness. Based on the testing completed and the comparisons with predicate device, the IMMULITE® 2000 Cortisol Calibration Verification Material does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence.
4
510(k) Summary
Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence.
The assigned 510(k) Number: K141444
-
- Submitter Mailing Address:
Contact Person:
- Submitter Mailing Address:
Phone Number: Fax Number: E-mail Address: Date Prepared:
-
- Device Name Proprietary Name: Measurand: Type of Test:
Regulation Section: Classification: Products Code:
- Device Name Proprietary Name: Measurand: Type of Test:
Panel:
-
- Predicate Device Name
Predicate 510(k) No:
- Predicate Device Name
-
- Device Description:
-
- Intended Use: Indication for Use:
Special Conditions for Use Statement(s): Special Instrument Requirements:
- Intended Use: Indication for Use:
Siemens Healthcare Diagnostics Inc. 5210 Pacific Concourse Drive Los Angeles. CA 90045
Donna Velasquez Regulatory Technical Specialist (310) 645-8200 x7403 (310) 645-9999 donna.velasquez@siemens.com May 30, 2014
IMMULITE® 2000 Folic Acid Calibration Verification Material Quality Control materials for IMMULITE® 2000 Folic Acid assay Calibration Verification Material (CVM) for IMMULITE® 2000 Folic Acid assay 21 CFR 862.1660, Quality Control Material Class I Reserved JJX - Single (Specified) Analyte Controls (Assayed and Unassaved) Clinical Chemistry (75)
IMMULITE 2000 DHEA-SO4 Calibration Verification Material (CVM) K140258
The IMMULITE® 2000 Folic Acid Calibration Verification Material (CVM) contains one set of four vials, 3mL (CVM1) each. when reconstituted. CVM1 contains a lyophilized human albumin based matrix with preservatives. CVM2. CVM3. and CVM4 contain various levels of lyophilized folic acid in human albumin based matrix with preservatives.
See Indications for Use Statement below The IMMULITE® Folic Acid Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Folic Acid assay on the IMMULITE 2000 systems
For prescription use only
IMMULITE® 2000 Systems
5
-
- Technological Characteristics and Substantial Equivalence Comparison with Predicate:
A comparison of the device features, intended use, and other information demonstrates that the IMMULITE® 2000 Folic Acid Calibration Verification Material (CVM) is substantially equivalent to the predicate device as summarized in Table 1.
- Technological Characteristics and Substantial Equivalence Comparison with Predicate:
SIMILARITIES | ||||
---|---|---|---|---|
Candidate Device | ||||
IMMULITE 2000 Folic Acid CVM | Predicate Device | |||
IMMULITE 2000 DHEA-SO4 | ||||
CVM | ||||
Intended | ||||
Use | The IMMULITE® Folic Acid | |||
Calibration Verification Material (CVM) | ||||
is for in vitro diagnostic use in the | ||||
verification of calibration of the | ||||
IMMULITE Folic Acid assay on the | ||||
IMMULITE 2000 systems | The IMMULITE® DHEA-SO4 | |||
Calibration Verification Material | ||||
(CVM) is for in vitro diagnostic use | ||||
in the verification of calibration of | ||||
the IMMULITE DHEA-SO4 assay | ||||
on the IMMULITE 2000 systems | ||||
Form | Lyophilized | Same | ||
Levels | 4 | Same | ||
Stability | Stable unopened until the expiration date | Same | ||
Storage | -20°C | Same | ||
Use | Single Use Only | Same |
Table 1: Substantial Equivalence Comparison | ||||
---|---|---|---|---|
-- | --------------------------------------------- | -- | -- | -- |
DIFFERENCES | ||
---|---|---|
Candidate Device | ||
IMMULITE 2000 Folic Acid CVM | Predicate Device | |
IMMULITE 2000 DHEA-SO4 | ||
CVM | ||
Analyte | Folic Acid | DHEA-SO4 |
Matrix | Human Albumin with preservatives | Human Serum with Preservatives |
7. Non-Clinical Performance Testing
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the candidate device.
Stability Summary:
The stability study was conducted to validate real-time shelf life and open component (in-use or open vial) claim for the IMMULITE 2000 Folic Acid Calibration Verification Material (CVM) to ensure that it maintains optimal product performance on IMMULITE 2000 platforms throughout the established shelf life of the CVM. The Calibration Verification Materials are stable up to 3.5 years when stored at -20°C prior to opening, and for 8 hours at ambient or room temperature (15-25°C) after reconstitution.
Stability Protocol Summary:
1
The CVM study protocols are run as part of the calibrator stability testing. The stability CVMs and reference CVMs are run in duplicate (as a minimum) at the time points shown in Table 2 and the dose value determined from the reference calibrator curve.
6
| CVM Level | Time-Points
(Days) | | | |
|-----------|-----------------------|-----|------|------|
| LFOCVM1 | 1 | 912 | 1095 | 1280 |
| LFOCVM2- | 1 | 912 | 1095 | 1280 |
| LFOCVM3 | 1 | 912 | 1095 | 1280 |
| LFOCVM4 | 1 | 912 | 1095 | 1280 |
Table 2: Stability Protocol Summary
For Open Component testing, the results were determined from a 2-point adjustment. Using IMMULITE 2000 Folic Acid (L2KFO) kit lot 446, lot 090B Folic Acid CVMs were tested at 2hourly intervals for up to 9 hours at ambient or room temperature (15-25°C) conditions.
Stability Acceptance Criteria Summary:
The Acceptance Criteria for the IMMULITE Folic Acid Calibration Verification Material (CVM) are in 2 parts. Part 1 consists of guideline acceptance criteria which require dose value of stability CVMs to fall between ±15% of the assigned dose for CVM level 2, within ±10% of the assigned dose for level 3 and within ±12% of assigned dose for CVM level 4. Part 2 review limits criteria which require dose value of the controls to be within 2 Standard Deviations (SD) of the control target value when generated from the stability calibrator curve. If the result is not within acceptable dose range of ±15% of the assigned dose for CVM level 2, within ±10% of the assigned dose for level 3 and within ±12% of assigned dose for CVM level 4 then additional data review is conducted using part 2 criteria.
Traceability:
The IMMULITE Folic Acid CVMs are traceable to an internal material which has been gravimetrically prepared. The CVMs are manufactured using qualified materials and measurement procedures.
Value Assignment:
The IMMULITE Folic Acid CVMs are 4 level materials which are a subset of 8 level Folic Acid calibrators. Calibrators are not commercialized but are used internally during manufacture and release testing of Folic Acid reagents and two point adjustors. The IMMULITE calibrators and therefore CVMs are value assigned using assigned reference calibrators. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. The CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. The controls must fall within their target ranges. Six levels of commercially available controls and 14 patient serum samples were used to validate calibrator/CVM value assignments.
Expected Values/Target Values/Reference Range:
The CVMs are manufactured using qualified materials and measurement procedures. The Folic Acid CVMs were tested on 27 replicates in total comprised of 9 runs and 3 replicates per run on 6 IMMULITE 2000 systems and 3 different reagent kit lots. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. The CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The expected assay range is 1 to 24 ng/mL. The target values in Table 3 can be considered as guidelines.
7
Table 3: Analyte Target Range Levels
| Analyte target
levels | CVM Level | *Target
Mean
(ng/mL) | Standard
Deviation
(SD) | Guideline ±2SD Range
(ng/mL) |
|--------------------------|---------------|----------------------------|-------------------------------|---------------------------------|
| | LFOCVM1 | 0.00 | - | 0.00
1.0 |
| | LFOCVM2 | 3.11 | 0.28 | 2.55
3.67 |
| | LFOCVM3 | 12.4 | 0.75 | 10.9
13.9 |
| | LFOCVM4 | 26.3 | 1.6 | 23.1
29.5 |
| Assay Range | 1 to 24 ng/mL | | | |
*Note: when CVMs are run by the customer, an actual value below and above the assay range will result when the customer programs the CVMs as calibration verifiers in the instrument software. If programmed as a patient or control, then the software will give values as the assay lower and upper range.
Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, total precision tabulated in the respective assay instructions for use may be considered as one factor when establishing local, acceptable ranges. The values provided above may be considered as guidelines. Value assignment is lot specific.
Standard/Guidance Documents Referenced:
CEN 13640 Stability Testing of In Vitro Diagnostic Reagents Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators Guidance for Industry and FDA Staff – Assayed and Unassayed Quality Control Material
Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10
8. Conclusion:
The IMMULITE® 2000 Folic Acid Calibration Verification Material is substantially equivalent to the predicate device intended for similar use. The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence, product safety and effectiveness. Based on the testing completed and the comparisons with predicate device, the IMMULITE® 2000 Folic Acid Calibration Verification Material does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence.
8
510(k) Summary
Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence.
.
The assigned 510(k) Number: K141444
the assigned 510(k) Number: | K141444 |
---|---|
1. Submitter | |
Mailing Address: | Siemens Healthcare Diagnostics Inc. |
5210 Pacific Concourse Drive | |
Los Angeles, CA 90045 | |
Contact Person: | |
Phone Number: | |
Fax Number: | |
E-mail Address: | |
Date Prepared: | Donna Velasquez |
Regulatory Technical Specialist | |
(310)-645-8200 x7403 | |
(310)-645-9999 | |
donna.velasquez@siemens.com | |
May 30, 2014 | |
2. Device Name | |
Proprietary Name: | |
Measurand: | |
Type of Test: | |
Regulation Section: | |
Classification: | |
Products Code: | |
Panel: | IMMULITE® 2000 Vitamin B12 Calibration Verification Material |
Quality Control materials for IMMULITE 2000 Vitamin B12 | |
Calibration Verification Material (CVM) for IMMULITE® 2000 | |
Vitamin B12 assay | |
21 CFR 862.1660, Quality Control Material | |
Class I Reserved | |
JJX - Single (Specified) Analyte Controls (Assayed and Unassayed) | |
Clinical Chemistry (75) | |
3. Predicate Device Name | |
Predicate 510(k) No: | IMMULITE® 2000 DHEA-SO4 Calibration Verification Material |
K140258 | |
4. Device Description: | IMMULITE® 2000 Vitamin B12 Calibration Verification Material |
(CVM) contains one set of four vials, 3 mL each, when reconstituted. | |
CVM1 contains a lyophilized human albumin-based matrix with | |
preservatives. CVM2, CVM3, and CVM4 contain various levels of | |
lyophilized Vitamin B12 in human albumin-based matrix with | |
preservatives. | |
5. Intended Use: | |
Indication for Use: | See Indications for Use Statement below |
The IMMULITE® Vitamin B12 Calibration Verification Material | |
(CVM) is for in vitro diagnostic use in the verification of the | |
IMMULITE Vitamin B12 assay on the IMMULITE 2000 systems | |
Special Conditions for | |
Use Statement(s): | |
Special Instrument | |
Requirements: | For prescription use only |
IMMULITE® 2000 Systems |
9
-
- Technological Characteristics A comparison of the device features, intended use, and other information demonstrates that the IMMULITE® 2000 Vitamin B12 and Substantial Equivalence Calibration Verification Material (CVM) is substantially equivalent Comparison with Predicate: to the predicate device, as summarized in Table 1.
Table 1: Substantial Equivalence Comparison
SIMILARITIES | ||
---|---|---|
Candidate Device | ||
IMMULITE 2000 Vitamin B12 CVM | Predicate Device | |
IMMULITE 2000 DHEA-SO4 CVM | ||
Intended | ||
Use | The IMMULITE® Vitamin B12 | |
Calibration Verification Material | ||
(CVM) is for in vitro diagnostic use in | ||
the verification of calibration of the | ||
IMMULITE Vitamin B12 assay on the | ||
IMMULITE 2000 systems. | The IMMULITE® DHEA-SO4 | |
Calibration Verification Material | ||
(CVM) is for in vitro diagnostic use in | ||
the verification of calibration of the | ||
IMMULITE DHEA-SO4 assay on the | ||
IMMULITE 2000 systems | ||
Form | Lyophilized | Same |
Stability | Stable unopened until the | |
expiration date | Same | |
Storage | -20°C | Same |
Levels | 4 | Same |
Use | Single Use Only | Same |
DIFFERENCES | ||
---|---|---|
Candidate Device | ||
IMMULITE 2000 Vitamin B12 CVM | Predicate Device | |
IMMULITE 2000 DHEA-SO4 CVM | ||
Analyte | Vitamin B12 | DHEA-SO4 |
Matrix | Human Albumin with preservatives | Human Serum with preservatives |
7. Non-Clinical Performance Testing
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the candidate device.
Stability Summary:
The stability study was conducted to validate real-time shelf life and open component (in-use or open vial) claim for the IMMULITE 2000 Vitamin B12 CVM to ensure that it maintains optimal product performance on IMMULITE 2000 platforms throughout the established shelf life of the CVM. The Calibration Verification Materials are stable up to 69 months when stored at -20°C prior to opening and stable for 8 hours at room (ambient) temperature (15-25°C) after reconstitution.
Stability Protocol Summary:
The CVM study protocols are run as part of the calibrator stability testing. As summarized in Table 2, the testing was run in duplicate (at the minimum) and contains time-points at which the testing was performed.
10
| CVM Level | Time-Points
(Months) | | | |
|-----------|-------------------------|----|----|----|
| LVBCVM1 | 0 | 67 | 68 | 69 |
| LVBCVM2 | 0 | 67 | 68 | 69 |
| LVBCVM3 | 0 | 67 | 68 | 69 |
| LVBCVM4 | 0 | 67 | 68 | 69 |
For Open Component testing, the results were determined from a 2-point adjustment. Using IMMULITE 2000 Vitamin B12 (L2KVB) kit lot 467, lot 108 Vitamin B12 CVMs were tested at 2hourly intervals for up to 9 hours at ambient or room temperature (15-25°C) conditions.
Stability Acceptance Criteria Summary:
The Acceptance Criteria for the IMMULITE Vitamin B12 Calibration Verification Material (CVM) are in 2 parts. Part 1 consists of guideline acceptance criteria which require dose value of stability calibrator/CVM to fall between ±10% of assigned dose for CVM level 2 and level 3 and within ±21% of assigned dose for CVM level 4. Part 2 review limits criteria which require dose value of the controls to be within 2SD of the control target value when generated from the stability calibrator curve. If the result is not within acceptable dose range of ±10% for level 3 and ±21% for level 4, then additional data review is conducted using part 2 criteria.
Traceability:
The IMMULITE Vitamin B12 CVMs are traceable to an internal standard. The CVMs are manufactured using qualified materials and measurement procedures.
Value Assignment:
The IMMULITE® 2000 Vitamin B12 CVMs are 4 level materials which are a subset of 8 level Vitamin B12 calibrators. Calibrators are not commercialized but are used internally during manufacture and release testing of Vitamin B12 reagents and two point adjustors. The IMMULITE calibrators and therefore CVMs are value assigned reference calibrators. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. Eight levels of commercially available controls and 21 patient serum samples were used to validate CVM value assignments.
11
Expected Values/Target Values/Reference Range:
The Vitamin B12 CVMs were tested on 24 replicates in total comprised of 12 runs and 2 replicates per run, 3 IMMULITE 2000 and 5 IMMULITE/IMMULITE 1000 systems and 4 different reagent kit lots. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The expected assay range is 150 to 1000 pg/mL. The target values in Table 3 can be considered as guidelines.
| Analyte target
levels | Level | *Target Mean
(ng/mL) | Standard
Deviation
(SD) | Guideline ±2SD Range
(ng/mL) |
|--------------------------|-------------------|-------------------------|-------------------------------|---------------------------------|
| | LVBCVM1 | 0.00 | - | 0.00 - 150 |
| | LVBCVM2 | 234 | 23.5 | 187 - 281 |
| | LVBCVM3 | 579 | 34.5 | 510 - 648 |
| | LVBCVM4 | 1144 | 194.5 | 755 - 1533 |
| Assay Range | 150 to 1000 pg/mL | | | |
Table 3: Analyte Target Range Levels
*Note: when CVMs are run by the customer, an actual value below and above the assay range will result when the customer programs the CVMs as calibration verifiers in the instrument software. If programmed as a patient or control, then the software will give values as the assay lower and upper range.
Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, total precision tabulated in the respective assay instructions for use may be considered as one factor when establishing local, acceptable ranges. The values provided above may be considered as guidelines. Value assignment is lot specific.
Standard/Guidance Documents Referenced:
CEN 13640 Stability Testing of In Vitro Diagnostic Reagents
Guidance for Industry - Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators Guidance for Industry and FDA Staff - Assayed and Unassayed Quality Control Material
Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10
Conclusion: 9.
The IMMULITE® 2000 Vitamin B12 Calibration Material is substantially equivalent to the predicate device intended for similar use. The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence, product safety and effectiveness. Based on the testing completed and the comparisons with predicate device, the IMMULITE® 2000 Vitamin B12 Calibration Material does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence.
12
Image /page/12/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized caduceus, a symbol often associated with medicine and healthcare. The emblem is positioned to the right of the department's name, which is written in a circular arrangement around the emblem. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 9, 2014
SIEMENS HEALTHCARE DIAGNOSTICS, INC. C/O DONNA VELASQUEZ REGULATORY TECHNICAL SPECIALIST 5210 PACIFIC CONCURSE DRIVE LOS ANGELES CA 90045
Re: K141444
Trade/Device Name: IMMULITE® 2000 Cortisol Calibration Verification Material IMMULITE® 2000 Folic Acid Calibration Verification Material IMMULITE® 2000 Vitamin B12 Calibration Verification Material Regulation Number: 21 CFR 862.1660 Regulation Name: Quality Control Material (assayed and unassayed) Regulatory Class: I, Reserved Product Code: JJX Dated: June 3, 2014 Received: June 5, 2014
Dear Donna Velasquez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
13
Page 2 - Ms. Donna Velasquez
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resources/orYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Courtney H. Lias -S
Courtney H. Lias Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
14
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Dale: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K141444
Device Name
IMMULITE® 2000 Cortiso! Calibration Verification Material IMMULITE® 2000 Folic Acid Calibration Verification Material IMMULITE® 2000 Vitamin B12 Calibration Verification Material
Indications for Use (Describe)
The IMMULITE® Corlisol Calibration Material is for in vitro diagnostic use in the verification of calibration of the IMMULITE Cortisol assay on the IMMULITE 2000 systems.
The IMMULITE® Folic Acid Calibration Material is for in vitro diagnostic use in the verification of calibration of the IMMULITE Folic Acid assay on the IMMULITE 2000 systems.
The IMMULITE® Vitanin B12 Calibration Material is for in vitro diagnostic use in the verification of calibration of the IMMULITE Vitamin B12 assay on the IMMULITE 2000 systems.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpant C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
。
Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)
Yung W. Chan -S
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.4
The burden lime for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and mainlain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Orug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStall@lda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a curronly valid OMB number."