(34 days)
The Immunalysis Buprenorphine Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 5ng/mL. The assay is intended for use in laboratories for the qualitative and semi-quantitative analysis of Buprenorphine in human urine with automated clinical chemistry analyzers. This assay is calibrated against Buprenorphine. This in-vitro diagnostic device is for prescription use only.
The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GC-MS or permitting laboratories to establish quality control procedures.
The Immunalysis Buprenorphine Urine Enzyme Immunoassay Kit provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography / Mass Spectroscopy (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
Immunalysis Buprenorphine Urine Controls:
The Immunalysis Buprenorphine Urine Controls are used as control materials in the Immunalysis Buprenorphine Urine Enzyme Immunoassay.
Immunalysis Buprenorphine Urine Calibrators:
The Immunalysis Buprenorphine Urine Calibrators are used as calibrators in the Immunalysis Buprenorphine Urine Enzyme Immunoassay for the qualitative and semiquantitative determination of Buprenorphine in urine on automated clinical chemistry analyzers.
The assay consists of antibody/ substrate reagent and enzyme conjugate reagent. The antibody/ substrate reagent includes rabbit antibodies to Buprenorphine, glucose-6phosphate (G6P) and nicotinamide adenine dinucleotide (NAD) in Tris buffer with Sodium Azide as a preservative. The enzyme conjugate reagent includes buprenorphine derivative labeled with glucose-6-phosphate dehydrogenase (G6PDH) in Tris buffer with Sodium Azide as a preservative. Calibrators and controls are sold separately. Reagents are liquid, ready to use
The buprenorphine calibrator and controls consists of a single calibrator at 5ng/mL, a control set containing a LOW control at 3.75ng/mL and a HIGH control at 6.25ng/mL and a calibrator set containing a negative calibrator, a Level 1 calibrator at 5ng/mL, a Level 2 calibrator at 10ng/mL, a Level 3 calibrator at 20ng/mL and a Level 4 calibrator at 40ng/mL.
Here's a breakdown of the acceptance criteria and study details for the Immunalysis Buprenorphine Urine Enzyme Immunoassay, based on the provided 510(k) summary:
This device is an in-vitro diagnostic (IVD) test, not an AI/ML-based device that interprets images or signals using algorithms learned from data. Therefore, many standard AI/ML evaluation criteria, such as MRMC studies, expert adjudication, or separate training/test sets with human-in-the-loop performance, are not applicable here. The "acceptance criteria" for this device are defined by its analytical performance metrics, which show it functions as intended for its intended use.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly listed with numerical targets in the same clear format as for an AI/ML device (e.g., "sensitivity > 90%"). Instead, the acceptability is demonstrated by the device meeting certain performance characteristics (precision, specificity, linearity, non-interference) that are deemed adequate for its intended use as a qualitative and semi-quantitative drug test. The FDA's substantial equivalence determination implies these criteria were met.
The performance studies and their results serve as the "reported device performance" and, by implication, demonstrate meeting the acceptance criteria.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Precision/Cutoff Characterization: Consistent and accurate determination of positive/negative results around the 5 ng/mL cutoff, and reproducibility of measurements. | Qualitative: 100% agreement for concentrations ≤ 3.75 ng/mL (Negative) and ≥ 6.25 ng/mL (Positive). At the 5 ng/mL cutoff, 48/80 (60%) were Negative and 32/80 (40%) were Positive. This demonstrates the cutoff serves as a boundary. Semi-Quantitative: 100% agreement for concentrations ≤ 3.75 ng/mL (Negative) and ≥ 6.25 ng/mL (Positive). At the 5 ng/mL cutoff, 33/80 (41.25%) were Negative and 47/80 (58.75%) were Positive. Precision was evaluated over 20 days, 2 runs/day in duplicate (N=80). |
| Specificity and Cross-Reactivity: Minimal interference from structurally similar compounds and no interference from other common drugs or compounds. | Structurally Related Compounds: Buprenorphine (100% cross-reactivity), NorBuprenorphine (90.91%), Buprenorphine Glucuronide (0.17%), NorBuprenorphine Glucuronide (0.13%). All other listed opioids and common drugs (e.g., 6-Acetyl morphine, Codeine, Heroin, Morphine, Methadone, Oxycodone, etc.) showed "N.D." (Not Detected/<0.05%) for both qualitative and semi-quantitative modes. |
| Interference: No significant interference from common exogenous or endogenous compounds, or variations in pH and specific gravity. | Structurally Non-Similar Compounds: No interference observed from 90+ listed compounds (e.g., Acetaminophen, Alprazolam, Caffeine, Ibuprofen, etc.) at high concentrations (100,000 ng/mL or 500,000 ng/mL), with correct negative results at -25% cutoff and positive results at +25% cutoff. Endogenous Compounds: Most endogenous compounds (Acetone, Ascorbic Acid, Bilirubin, Creatinine, Ethanol, Glucose, Hemoglobin, Human Serum Albumin, Oxalic Acid, Sodium Azide, Sodium Chloride, Sodium Fluoride, Urea) showed no interference. Exceptions: Boric Acid (1% w/v) and Riboflavin (0.0075 g/dL) caused false negative results at +25% cutoff. This became a known limitation in the labeling. pH Effect: No interference between pH 3.0 and 11.0. Specific Gravity Effect: No interference between specific gravity 1.000 and 1.030. |
| Linearity/Recovery: Accurate measurement across the expected analytical range for semi-quantitative analysis. | Recovery ranged from 77% (at 55 ng/mL) to 105% (at 35 ng/mL), demonstrating reasonable linearity for semi-quantitative purposes. Expected concentrations ranged from 5 ng/mL to 55 ng/mL. |
| Method Comparison: Agreement with a accepted confirmatory method (LC/MS). | Qualitative: 100% agreement between the Test Device and LC/MS Confirmation for both positive (40 samples) and negative (40 samples) results. Semi-Quantitative: 100% agreement between the Test Device and LC/MS Confirmation for both positive (40 samples) and negative (40 samples) results. |
| Stability: Reagents, calibrators, and controls maintain performance over specified storage and use periods. | Closed Accelerated Stability: Supports 1-year expiration for reagents and 12 months for calibrators/controls. All tested levels remained within specifications for up to 40 days of accelerated testing. Open/On-Board Stability (Reagents): Supports 28-day open vial expiration. All replicates for tested levels remained within specifications for up to 28 days. Calibrator and Control Stability (Open Vial): Supports 6-month open vial expiration. All calibrator and control levels were within specifications for up to 13 days of accelerated testing. Real-time studies are ongoing. |
| Calibrator and Control Traceability & Value Assignment: Accuracy and consistency of calibrator and control values. | Calibrators and controls are traced to a commercially available standard solution from Cerilliant Chemicals. Values are assigned based on Mass Spectrometry results, with adjustments and retesting if outside acceptable ranges. |
2. Sample Sizes Used for the Test Set and Data Provenance
- Precision/Cutoff Characterization: 80 determinations per concentration level (20 days, 2 runs per day in duplicate) for 9 concentration levels resulting in a total of 720 measurements.
- Specificity and Cross-Reactivity: Not explicitly stated as a "sample size" of patient samples, but structurally similar compounds and other drugs were spiked into drug-free urine. Around 30+ compounds were tested at specific concentrations.
- Interference: Not explicitly stated as a "sample size" of patient samples, but various non-similar compounds, endogenous compounds, and different pH/specific gravity levels were tested by spiking into drug-free urine containing target analyte at ±25% of cutoff. Around 90+ non-similar compounds and 17 endogenous compounds were tested. pH was tested at 9 levels, and specific gravity at 8 levels.
- Linearity/Recovery: A drug-free urine pool spiked with a high concentration, then serially diluted to create 11 different concentration levels.
- Method Comparison: 80 "unaltered, anonymous and discarded clinical urine samples" (40 positive, 40 negative based on LC/MS).
- Stability Studies: Conducted with a sufficient number of replicates over timepoints to support the claimed stability. Specific "sample sizes" (e.g., number of batches or vials) are not detailed but the methodology (e.g., "replicates") implies sufficient testing.
Data Provenance:
- Country of Origin: Not explicitly stated for patient samples. The manufacturer, Immunalysis Corporation, is based in Pomona, CA, USA.
- Retrospective or Prospective:
- Precision, Specificity, Interference, Linearity: Lab-based studies, likely prospective, using controlled spiked samples.
- Method Comparison: "Unaltered, anonymous and discarded clinical urine samples obtained from clinical testing laboratories." This suggests retrospective use of existing clinical samples.
- Stability: Prospective, controlled lab studies.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts
Not applicable in the typical sense for this IVD device.
For IVD devices like this immunoassay, the "ground truth" for the test set is established by:
- Preparation of controlled samples: For precision, specificity, interference, and linearity studies, samples are prepared in a laboratory with known concentrations of the analyte or interfering substances. This is the "ground truth" for these studies.
- Reference Method Analysis: For method comparison, the "ground truth" is established by a more definitive analytical method, in this case, Gas Chromatography/Mass Spectrometry (GC-MS) or Liquid Chromatography / Mass Spectrometry (LC/MS). These are established analytical chemistry techniques, not subject to expert interpretation in the same way as, for example, a radiological image. Therefore, human experts are not used to establish this kind of ground truth.
4. Adjudication Method for the Test Set
Not applicable. As the ground truth is established by known sample concentrations or definitive analytical methods (GC-MS/LC/MS), there is no need for human expert adjudication.
5. If a Multi-reader Multi-case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an automated in-vitro diagnostic immunoassay for chemical analysis of urine, not an AI/ML-based diagnostic aid for human readers (e.g., radiologists, pathologists). It does not involve "human readers" interpreting results that would then be assisted by AI. The device provides a direct analytical result.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Essentially, yes, but not for an "algorithm" in the AI sense. The primary performance data (Precision, Specificity, Interference, Linearity, Method Comparison) demonstrates the device's analytical performance on its own, producing qualitative and semi-quantitative results for Buprenorphine in urine. It operates as a "standalone" analytical instrument providing results.
7. The Type of Ground Truth Used
- Controlled Spike-ins: For Precision, Specificity/Cross-Reactivity, Interference, and Linearity, the ground truth was established by manufacturing urine samples with known, precisely measured concentrations of Buprenorphine or other compounds.
- Reference Standard (LC/MS Confirmation): For the Method Comparison study, the ground truth was established by Liquid Chromatography / Mass Spectrometry (LC/MS), which is considered the preferred confirmatory method for drug testing in urine.
8. The Sample Size for the Training Set
Not applicable in the AI/ML sense. This is a traditional enzyme immunoassay, not an AI/ML device that requires a training set to learn relationships from data. Its "training" involves chemical reagent formulation and instrument calibration.
9. How the Ground Truth for the Training Set Was Established
Not applicable in the AI/ML sense. For a chemical assay, the equivalent of "establishing ground truth for training" would involve:
- Chemical Formulation: Developing the immunoassay reagents (antibodies, enzymes, substrates) based on known chemical principles to specifically react with Buprenorphine.
- Calibration: Using precisely measured standard solutions (calibrators) with known concentrations of Buprenorphine to establish the assay's response curve. The document states: "Calibrators and controls are manufactured and are tested by mass spectrometry. If any of the analytes are out of the acceptable range, then the calibrator or control is adjusted and retested. Values are assigned to the calibrator and controls once the Mass spectrometry results are within the acceptable ranges." This process essentially establishes the "ground truth" for the device's internal calibration.
{0}------------------------------------------------
510(k) SUMMARY
JUL 0 1 2014
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(c).
- A. Contact Information
| 1. | Manufacturer: | Immunalysis Corporation |
|---|---|---|
| 2. | Contact Name: | Joseph Ginete |
| 3. | Contact Title: | Regulatory Affairs Specialist |
| 4. | Address: | 829 Towne Center Drive Pomona, CA 91767 |
| 5. | Phone: | (909) 482-0840 |
| 6. | Fax: | (909) 482-0850 |
| 7. | Email: | jginete@immunalysis.com |
| 8. | Summary prepared on: | June 30, 2014 |
| B. Device Information | ||
| 1. Trade Name: | Immunalysis Buprenorphine Urine Enzyme Immunoassay | |
| Immunalysis Buprenorphine Urine Controls | ||
| Immunalysis Buprenorphine Urine Calibrators | ||
| 2. Common Name: | Immunalysis Buprenorphine Urine Enzyme Immunoassay | |
| Immunalysis Buprenorphine Urine Controls | ||
| Immunalysis Buprenorphine Urine Calibrators | ||
| 3. Device Classification: | II | |
| 4. | Regulation Number: | CFR 862.3650 Opiate Test System |
| CFR 862.3200 Clinical Toxicology Calibrator | ||
| CFR 862.3280 Clinical Toxicology Control Materials | ||
| 5. Panel: | Toxicology(91) | |
| 6. | Product Code: | DJG |
| DLJ | ||
| LAS | ||
| C. Legally Marketed Device to Which We are Claiming Equivalence (807.92(A)(3)) | ||
| 1. | Predicate Device: | CEDIA Buprenorphine Assay |
| 1. | Predicate Device: | CEDIA Buprenorphine AssayCEDIA Buprenorphine ControlsCEDIA Buprenorphine Calibrators |
|---|---|---|
| 2. | Predicate Company: | Microgenics |
| 3. | Predicate K Number: | K040316 |
1
{1}------------------------------------------------
D. Device Description
The assay consists of antibody/ substrate reagent and enzyme conjugate reagent. The antibody/ substrate reagent includes rabbit antibodies to Buprenorphine, glucose-6phosphate (G6P) and nicotinamide adenine dinucleotide (NAD) in Tris buffer with Sodium Azide as a preservative. The enzyme conjugate reagent includes buprenorphine derivative labeled with glucose-6-phosphate dehydrogenase (G6PDH) in Tris buffer with Sodium Azide as a preservative. Calibrators and controls are sold separately. Reagents are liquid, ready to use
The buprenorphine calibrator and controls consists of a single calibrator at 5ng/mL, a control set containing a LOW control at 3.75ng/mL and a HIGH control at 6.25ng/mL and a calibrator set containing a negative calibrator, a Level 1 calibrator at 5ng/mL, a Level 2 calibrator at 10ng/mL, a Level 3 calibrator at 20ng/mL and a Level 4 calibrator at 40ng/mL.
E. Intended Use
Immunalysis Buprenorphine Urine Enzyme Immunoassay:
The Immunalysis Buprenorphine Urine Enzyme Immunoassay is a homogeneous enzyme immunoassay with a cutoff of 5ng/mL. The assay is intended for use in laboratories for the qualitative and semi-quantitative analysis of Buprenorphine in human urine with automated clinical chemistry analyzers. This assay is calibrated against Buprenorphine. This in-vitro diagnostic device is for prescription use only.
The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as GC-MS or permitting laboratories to establish quality control procedures.
The Immunalysis Buprenorphine Urine Enzyme Immunoassay Kit provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liquid Chromatography / Mass Spectroscopy (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.
Immunalysis Buprenorphine Urine Controls:
The Immunalysis Buprenorphine Urine Controls are used as control materials in the Immunalysis Buprenorphine Urine Enzyme Immunoassay.
Immunalysis Buprenorphine Urine Calibrators:
The Immunalysis Buprenorphine Urine Calibrators are used as calibrators in the Immunalysis Buprenorphine Urine Enzyme Immunoassay for the qualitative and semiquantitative determination of Buprenorphine in urine on automated clinical chemistry analyzers.
{2}------------------------------------------------
Comparison of the new device with the predicate device F
| Item | Predicate Device (K040316) | Test Device |
|---|---|---|
| Intended Use | For the qualitative and semi-quantitative determination of thepresence of buprenorphine in humanurine at a cutoff of 5 ng/ml | For the qualitative and semi-quantitative determination of thepresence of buprenorphine in humanurine at a cutoff of 5 ng/ml |
| Type of Product | Analytical Reagents | Analytical Reagents |
| Measured Analytes | Buprenorphine | Buprenorphine |
| Test Matrix | Urine | Urine |
| Cutoff Levels | 5ng/mL of Buprenorphine | 5ng/mL of Buprenorphine |
| Test System | Homogenous Enzyme Immunoassay | Homogenous Enzyme Immunoassay |
| Materials | Buffer 1, Buffer 2, Lyophilized Reagentla and Lyophilized Reagent | Antibody/ Substrate Reagents andEnzyme Labeled Conjugate |
| Mass SpectroscopyConfirmation | Required for preliminary positiveanalytical results | Required for preliminary positiveanalytical results |
| Antibody | Mouse monoclonal anti-buprenorphinederivative | Rabbit Monoclonal Antibody toBuprenorphine |
| Storage | 2 - 8°C until expiration date | 2 - 8°C until expiration date |
| Calibrator Form | Liquid | Liquid |
| Calibrator Levels | One (1) Level (5ng/mL) | One (1) Level (5ng/mL) |
| Control Set Levels | Two (2) Levels (3ng/mL and 7ng/mL) | Two (2) Levels (3.75ng/mL and6.25ng/mL) |
| Calibrator SetLevels | Five (5) Levels (0, 5, 20, 50 and 75ng/mL) | Five (5) Levels (0, 5, 10, 20 and 40ng/mL) |
G. The following laboratory performance studies were performed to determine substantial equivalence of the Immunalysis Buprenorphine Urine Enzyme Immunoassay to the predicate
-
- Precision/ Cutoff Characterization Study was performed for 20 days, 2 runs per day in duplicate (N=80) on concentration of ±25%, ±50%, ±75% and ±100% of the cutoff. The study verified that the cutoff serves as a boundary between a negative and positive interpretation of a qualitative result. In addition, it also verified the product performance relative to the ability of the device to produce the same value during repeated measurements. The instrument used for this test was a Beckman Coulter AU 400e.
- a. The following is a summary table of the Qualitative Analysis for the 5ng/mL cutoff test data results.
| Concentration (ng/mL) | % of cutoff | # of determinations | Result |
|---|---|---|---|
| 0 | -100% | 80 | 80 Negative |
| 1.25 | -75% | 80 | 80 Negative |
| 2.5 | -50% | 80 | 80 Negative |
| 3.75 | -25% | 80 | 80 Negative |
| 5 | Cutoff | 80 | 48 Negative/32 Positive |
| 6.25 | +25% | 80 | 80 Positive |
| 7.5 | +50% | 80 | 80 Positive |
| 8.75 | +75% | 80 | 80 Positive |
| 10 | +100% | 80 | 80 Positive |
{3}------------------------------------------------
- b. The follówing is a summary table of the Semi-Quantitative Analysis for the 5ng/mL cutoff test data results.
| Semi-Quantitative Analysis (for 5ng/mL cutoff) | |||
|---|---|---|---|
| Concentration (ng/mL) | % of cutoff | # of determinations | Result |
| 0 | -100% | 80 | 80 Negative |
| 1.25 | -75% | 80 | 80 Negative |
| 2.5 | -50% | 80 | 80 Negative |
| 3.75 | -25% | 80 | 80 Negative |
| 5 | Cutoff | 80 | 33 Negative/ 47 Positive |
| 6.25 | +25% | 80 | 80 Positive |
| 7.5 | +50% | 80 | 80 Positive |
| 8.75 | +75% | 80 | 80 Positive |
| 10 | +100% | 80 | 80 Positive |
- Specificity and Cross-Reactivity - Structurally similar compounds were spiked into drug free urine at levels that will yield a result that is equivalent to the cutoff. The study verified assay performance relative to the ability of the device to exclusively determine certain drugs. The instrument used for this test was a Beckman Coulter AU 400e.
| Structurally Related Compounds - Qualitative | |||
|---|---|---|---|
| Compound | Concentration Tested (ng/mL) | Result | Cross-Reactivity (%) |
| Buprenorphine | 5 | N/A | 100.00 |
| NorBuprenorphine | 5.5 | POS | 90.91 |
| Buprenorphine Glucuronide | 3,000 | POS | 0.17 |
| NorBuprenorphine Glucuronide | 4,000 | POS | 0.13 |
| 6-Acetyl morphine | 100,000 | NEG | N.D. |
| Codeine | 100,000 | NEG | N.D. |
| Dihydrocodeine | 100,000 | NEG | N.D. |
| EDDP | 100000 | NEG | N.D. |
| EMDP | 100,000 | NEG | N.D. |
| Ethyl Morphine | 100,000 | NEG | N.D. |
| Heroin | 100,000 | NEG | N.D. |
| Hydrocodone | 100,000 | NEG | N.D. |
| Hydromorphone | 100,000 | NEG | N.D. |
| LAAM | 100,000 | NEG | N.D. |
| Levorphanol | 100,000 | NEG | N.D. |
| Methadone | 100,000 | NEG | N.D. |
| Meperidine | 100,000 | NEG | N.D. |
| Morphine 3 Glucuronide | 100,000 | NEG | N.D. |
| Morphine 6 Glucuronide | 100,000 | NEG | N.D. |
| Morphine | 100,000 | NEG | N.D. |
| Nalorphine | 100,000 | NEG | N.D. |
| Naloxone | 100,000 | NEG | N.D. |
| Naltrexone | 100,000 | NEG | N.D. |
| Norpropoxyphene | 100,000 | NEG | N.D. |
| Oxycodone | 100,000 | NEG | N.D. |
| Oxymorphone | 100,000 | NEG | N.D. |
| Diacetyl Morphine | 100,000 | NEG | N.D. |
- a. The qualitative result summary table is outlined below:
N.D. = Not Detected (<0.05%)
{4}------------------------------------------------
b. The semi-quantitative result summary table is outlined below:
| Structurally Related Compounds - Semi-Quantitative | ||
|---|---|---|
| Compound | Concentration Tested (ng/mL) | Cross-Reactivity (%) |
| Buprenorphine | 5 | 100.00 |
| NorBuprenorphine | 5.5 | 90.91 |
| Buprenorphine Glucuronide | 3,000 | 0.17 |
| NorBuprenorphine Glucuronide | 4,000 | 0.13 |
| 6-Acetyl morphine | 100,000 | N.D. |
| Codeine | 100,000 | N.D. |
| Dihydrocodeine | 100,000 | N.D. |
| EDDP | 100000 | N.D. |
| EMDP | 100,000 | N.D. |
| Ethyl Morphine | 100,000 | N.D. |
| Heroin | 100,000 | N.D. |
| Hydrocodone | 100,000 | N.D. |
| Hydromorphone | 100,000 | N.D. |
| LAAM | 100,000 | N.D. |
| Levorphanol | 100,000 | N.D. |
| Methadone | 100,000 | N.D. |
| Meperidine | 100,000 | N.D. |
| Morphine 3 Glucuronide | 100,000 | N.D. |
| Morphine 6 Glucuronide | 100,000 | N.D. |
| Morphine | 100,000 | N.D. |
| Nalorphine | 100,000 | N.D. |
| Naloxone | 100,000 | N.D. |
| Naltrexone | 100,000 | N.D. |
| Norpropoxyphene | 100,000 | N.D. |
| Oxycodone | 100,000 | N.D. |
| Oxymorphone | 100,000 | N.D. |
| Diacetyl Morphine | 100,000 | N.D. |
- Interference - Structurally non-similar compounds, endogenous compounds, the effect of pH and the effect of specific gravity was evaluated by spiking the potential interferent into drug free urine containing the target analyte at ±25% of the cutoff. Boric Acid and Riboflavin caused a false negative response at the concentrations tested. All other potential interferents analyzed verified that assay performance is unaffected by externally ingested compounds or an internally existing physiological condition. The instrument used for this test was a Beckman Coulter AU 400e.
- a. The following is a summary table of the structurally non-similar compounds for the 5ng/mL cutoff
| Structurally Non-Similar Compounds (for 5ng/mL cutoff) | |||||
|---|---|---|---|---|---|
| Compound | Concentration Tested (ng/mL) | -25% Cutoff (3.75ng/mL) | +25% Cutoff (6.25ng/mL) | ||
| Result | Interference? | Result | Interference? | ||
| 4-Bromo-2,5,Dimethoxyphenethylamine | 100,000 | Negative | No | Positive | No |
| 6-Acetylcodeine | 100,000 | Negative | No | Positive | No |
| 7-Aminoclonazepam | 100,000 | Negative | No | Positive | No |
| 7-Aminoflurazepam | 100,000 | Negative | No | Positive | No |
| 7-Aminonitrazepam | 100,000 | Negative | No | Positive | No |
| Acetaminophen | 500,000 | Negative | No | Positive | No |
| Acetylsalicylic Acid | 500,000 | Negative | No | Positive | No |
| Alprazolam | 100,000 | Negative | No | Positive | No |
| Amitriptyline | 100,000 | Negative | No | Positive | No |
| Amobarbital | 100,000 | Negative | No | Positive | No |
| Structurally Non-Similar Compounds (for 5ng/mL cutoff) | |||||
| Compound | Concentration Tested (ng/mL) | -25% Cutoff (3.75ng/mL) | +25% Cutoff (6.25ng/mL) | ||
| Result | Interference? | Result | Interference? | ||
| S-(+) Amphetamine | 100,000 | Negative | No | Positive | No |
| Benzoylecgonine | 500,000 | Negative | No | Positive | No |
| Benzylpiperazine | 100,000 | Negative | No | Positive | No |
| Bromazepam | 100,000 | Negative | No | Positive | No |
| Bupropion | 100,000 | Negative | No | Positive | No |
| Butabarbital | 100,000 | Negative | No | Positive | No |
| Caffeine | 500,000 | Negative | No | Positive | No |
| Cannabidiol | 100,000 | Negative | No | Positive | No |
| Cannabinol | 100,000 | Negative | No | Positive | No |
| Carbamazeprine | 100,000 | Negative | No | Positive | No |
| Carisoprodol | 100,000 | Negative | No | Positive | No |
| Chlordiazepoxide | 100,000 | Negative | No | Positive | No |
| Chlorpromazine | 100,000 | Negative | No | Positive | No |
| Clobazam | 100,000 | Negative | No | Positive | No |
| Clomipramine | 100,000 | Negative | No | Positive | No |
| Clonazepam | 100,000 | Negative | No | Positive | No |
| Cocaine | 100,000 | Negative | No | Positive | No |
| Codeine | 100,000 | Negative | No | Positive | No |
| Cotinine | 100,000 | Negative | No | Positive | No |
| Cyclobenzaprine | 100,000 | Negative | No | Positive | No |
| Delta-9-THC | 100,000 | Negative | No | Positive | No |
| Demoxepam | 100,000 | Negative | No | Positive | No |
| Desakylflurazepam | 100,000 | Negative | No | Positive | No |
| Desipramine | 100,000 | Negative | No | Positive | No |
| Diazepam | 100,000 | Negative | No | Positive | No |
| Dihydrocodeine | 100,000 | Negative | No | Positive | No |
| Diphenhydramine | 500,000 | Negative | No | Positive | No |
| Doxepin | 100,000 | Negative | No | Positive | No |
| Ecgonine | 100,000 | Negative | No | Positive | No |
| Ecgonine methyl ester | 100,000 | Negative | No | Positive | No |
| EDDP | 100,000 | Negative | No | Positive | No |
| 1R,2S(-)-Ephedrine | 100,000 | Negative | No | Positive | No |
| 1S,2R(+)-Ephedrine | 100,000 | Negative | No | Positive | No |
| EtG | 100,000 | Negative | No | Positive | No |
| Fenfluramine | 100,000 | Negative | No | Positive | No |
| Fentanyl | 100,000 | Negative | No | Positive | No |
| Flunitrazepam | 100,000 | Negative | No | Positive | No |
| Fluoxetine | 100,000 | Negative | No | Positive | No |
| Flurazepam | 100,000 | Negative | No | Positive | No |
| Hexobarbital | 100,000 | Negative | No | Positive | No |
| Ibuprofen | 100,000 | Negative | No | Positive | No |
| Imipramine | 100,000 | Negative | No | Positive | No |
| Ketamine | 100,000 | Negative | No | Positive | No |
| Lamotrignine | 100,000 | Negative | No | Positive | No |
| Lidocaine | 100,000 | Negative | No | Positive | No |
| Lorazepam | 100,000 | Negative | No | Positive | No |
| Compound | ConcentrationTested (ng/mL) | -25% Cutoff (3.75ng/mL)Result | Interference? | +25% Cutoff (6.25ng/mL)Result | Interference? |
| Lorazepam Glucuronide | 50,000 | Negative | No | Positive | No |
| Lormetazepam | 100,000 | Negative | No | Positive | No |
| LSD | 100,000 | Negative | No | Positive | No |
| Maprotiline | 100,000 | Negative | No | Positive | No |
| (+)-MDA | 100,000 | Negative | No | Positive | No |
| MDEA | 100,000 | Negative | No | Positive | No |
| MDMA | 100,000 | Negative | No | Positive | No |
| Meperidine | 100,000 | Negative | No | Positive | No |
| Meprobamate | 100,000 | Negative | No | Positive | No |
| Methadone | 500,000 | Negative | No | Positive | No |
| S(+)-Methamphetamine | 500,000 | Negative | No | Positive | No |
| Methaquolone | 100,000 | Negative | No | Positive | No |
| Methylphenidate | 100,000 | Negative | No | Positive | No |
| Midazolam | 100,000 | Negative | No | Positive | No |
| Naproxen | 100,000 | Negative | No | Positive | No |
| N-desmethyltapentadol | 100,000 | Negative | No | Positive | No |
| Nitrazepam | 100,000 | Negative | No | Positive | No |
| Nordiazepam | 100,000 | Negative | No | Positive | No |
| Norcodeine | 100,000 | Negative | No | Positive | No |
| Normorphine | 100,000 | Negative | No | Positive | No |
| Norpseudoephedrine | 100,000 | Negative | No | Positive | No |
| Nortriptyline | 100,000 | Negative | No | Positive | No |
| Oxazepam | 100,000 | Negative | No | Positive | No |
| Oxazepam Glucuronide | 50,000 | Negative | No | Positive | No |
| PCP | 100,000 | Negative | No | Positive | No |
| Pentazocine | 100,000 | Negative | No | Positive | No |
| Pentobarbital | 100,000 | Negative | No | Positive | No |
| Phenobarbital | 100,000 | Negative | No | Positive | No |
| Phentermine | 100,000 | Negative | No | Positive | No |
| Phenylephrine | 100,000 | Negative | No | Positive | No |
| Phenytoine | 100,000 | Negative | No | Positive | No |
| PMA | 100,000 | Negative | No | Positive | No |
| PPA | 100,000 | Negative | No | Positive | No |
| Propoxyphene | 100,000 | Negative | No | Positive | No |
| Propranolol | 100,000 | Negative | No | Positive | No |
| Protriptyline | 100,000 | Negative | No | Positive | No |
| R,R(-)-Pseudoephedrine | 100,000 | Negative | No | Positive | No |
| S,S(+)-Pseudoephedrine | 100,000 | Negative | No | Positive | No |
| Ranitidine | 100,000 | Negative | No | Positive | No |
| Ritalinic Acid | 100,000 | Negative | No | Positive | No |
| Salicylic Acid | 100,000 | Negative | No | Positive | No |
| Secobarbital | 100,000 | Negative | No | Positive | No |
| Sertraline | 100,000 | Negative | No | Positive | No |
| Sufentanil Citrate | 100,000 | Negative | No | Positive | No |
| Temazepam | 100,000 | Negative | No | Positive | No |
| Structurally Non-Similar Compounds (for 5ng/mL cutoff) | |||||
| Compound | ConcentrationTested (ng/mL) | -25% Cutoff (3.75ng/mL) | +25% Cutoff (6.25ng/mL) | ||
| Result | Interference? | Result | Interference? | ||
| 11-nor-9 carboxy THC | 100,000 | Negative | No | Positive | No |
| Theophylline | 100,000 | Negative | No | Positive | No |
| Thioridazine | 100,000 | Negative | No | Positive | No |
| Tramadol | 100,000 | Negative | No | Positive | No |
| Trazodone | 100,000 | Negative | No | Positive | No |
| Triazolam | 100,000 | Negative | No | Positive | No |
| Trifluoromethylphenyl-piperazine | 100,000 | Negative | No | Positive | No |
| Trimipramine | 5,000 | Negative | No | Positive | No |
| Venlafaxine | 100,000 | Negative | No | Positive | No |
| Zolpidem Tartrate | 100,000 | Negative | No | Positive | No |
Immunalysis Corporation
{5}------------------------------------------------
Immunalysis Corporation
:
{6}------------------------------------------------
Immunalysis Corporation
{7}------------------------------------------------
b. The following is a summary table of the endogenous compounds results for the 5ng/mL cutoff
| Endogenous Compounds (for 5ng/mL cutoff) | |||||
|---|---|---|---|---|---|
| Compound | ConcentrationTested (ng/mL) | -25% Cutoff (3.75ng/mL) | +25% Cutoff (6.25ng/mL) | ||
| Result | Interference? | Result | Interference? | ||
| Acetone | 1.0 g/dL | Negative | No | Positive | No |
| Ascorbic Acid | 1.5 g/dL | Negative | No | Positive | No |
| Bilirubin | 0.002 g/dL | Negative | No | Positive | No |
| Boric Acid | 1% w/v | Negative | No | Negative | Yes |
| Creatinine | 0.5 g/dL | Negative | No | Positive | No |
| Ethanol | 1.0 g/dL | Negative | No | Positive | No |
| Galactose | 0.01 g/dL | Negative | No | Positive | No |
| γ-Globulin | 0.5 g/dL | Negative | No | Positive | No |
| Glucose | 2.0 g/dL | Negative | No | Positive | No |
| Hemoglobin | 0.300 g/dL | Negative | No | Positive | No |
| Human Serum Albumin | 0.5 g/dL | Negative | No | Positive | No |
| Oxalic Acid | 0.1 g/dL | Negative | No | Positive | No |
| Riboflavin | 0.0075 g/dL | Negative | No | Negative | Yes |
| Sodium Azide | 1% w/v | Negative | No | Positive | No |
| Sodium Chloride | 6.0 g/dL | Negative | No | Positive | No |
| Sodium Flouride | 1% w/v | Negative | No | Positive | No |
| Urea | 6.0 g/dL | Negative | No | Positive | No |
c. Boric Acid and Riboflavin interfere with the assay and the limitations have been added to the labeling regarding these two compounds.
d. The following is a summary table of the effect of pH results for the 5ng/mL cutoff
| Effect of pH (for 5ng/mL cutoff) | |||||
|---|---|---|---|---|---|
| Test Parameter | Value | -25% Cutoff (3.75ng/mL) | +25% Cutoff (6.25ng/mL) | ||
| Result | Interference? | Result | Interference? | ||
| pH | 3.0 | Negative | No | Positive | No |
| pH | 4.0 | Negative | No | Positive | No |
| pH | 5.0 | Negative | No | Positive | No |
| pH | 6.0 | Negative | No | Positive | No |
| pH | 7.0 | Negative | No | Positive | No |
| pH | 8.0 | Negative | No | Positive | No |
| pH | 9.0 | Negative | No | Positive | No |
| pH | 10.0 | Negative | No | Positive | No |
| pH | 11.0 | Negative | No | Positive | No |
{8}------------------------------------------------
| Effect of Specific Gravity (for 5ng/mL cutoff) - Qualitative | |||||
|---|---|---|---|---|---|
| Test Parameter | Value | -25% Cutoff (3.75ng/mL) | +25% Cutoff (6.25ng/mL) | ||
| Result | Interference? | Result | Interference? | ||
| Specific Gravity | 1.000 | Negative | No | Positive | No |
| Specific Gravity | 1.002 | Negative | No | Positive | No |
| Specific Gravity | 1.005 | Negative | No | Positive | No |
| Specific Gravity | 1.010 | Negative | No | Positive | No |
| Specific Gravity | 1.015 | Negative | No | Positive | No |
| Specific Gravity | 1.020 | Negative | No | Positive | No |
| Specific Gravity | 1.025 | Negative | No | Positive | No |
| Specific Gravity | 1.030 | Negative | No | Positive | No |
- e. The following is a summary table of the effect of specific gravity result for the Engiml
-
- Linearity/ Recovery A drug free urine pool was spiked with a high concentration of the target analyte as a high value specimen. Additional pools were made by serially diluting the high value specimen. The study verified assay linearity in the semi-quantitative mode. The instrument used for this test was a Beckman Coulter AU 400e.
| Linearity/ Recovery | ||
|---|---|---|
| Expected Concentration (ng/mL) | Mean Concentration (ng/mL) | Recovery (%) |
| 5 | 5.0 | 100 |
| 10 | 9.9 | 99 |
| 15 | 13.7 | 91 |
| 20 | 20.2 | 101 |
| 25 | 24.2 | 97 |
| 30 | 29.5 | 98 |
| 35 | 36.7 | 105 |
| 40 | 39.3 | 98 |
| 45 | 40.8 | 91 |
| 50 | 40.9 | 82 |
| 55 | 42.4 | 77 |
| a. Summary results are listed in the following table | |
|---|---|
| ------------------------------------------------------ | -- |
- Method Comparison - Unaltered, anonymous and discarded clinical urine samples obtained from clinical testing laboratories were analyzed with the test device. The study verified that the product performance can be verified by Mass Spectrometry. The instrument used for this test was a Beckman Coulter AU 400e and an Agilent 6430 Liquid Chromatography Tandem Mass Spectrometry.
- a. The following is a comparison table of qualitative assay performance for the 5ng/mL cutoff
| LC/MS Confirmation | |||
|---|---|---|---|
| (+) | (-) | ||
| Test Device | (+) | 40 | 0 |
| (-) | 0 | 40 |
- b. The following is a summary table of qualitative assay performance for the 5ng/mL cutoff
| Type | <2.5ng/mL | 2.5~4.9 ng/mL | 5~7.5 ng/mL | >7.5 ng/mL | Agreement (%) |
|---|---|---|---|---|---|
| Qualitative/ Positive | 0 | 0 | 4 | 36 | 100% |
| Qualitative/ Negative | 36 | 4 | 0 | 0 | 100% |
{9}------------------------------------------------
- c. The following is a comparison table of semi-quantitative assay performance for the 5ng/mL cutoff
| LC/MS Confirmation | |||
|---|---|---|---|
| (+) | (-) | ||
| Test Device | (+) | 40 | 0 |
| (-) | 0 | 40 |
- d. The following is a summary table of semi-quantitative assay performance for the 5nq/mL cutoff
| Assay Performance verified by LC/MS - 5ng/mL Cutoff | |||||
|---|---|---|---|---|---|
| Type | Buprenorphine Concentration | Agreement (%) | |||
| <2.5ng/mL | 2.5 ~ 4.9 ng/mL | 5 ~ 7.5 ng/mL | >7.5 ng/mL | ||
| Semi-Quantitative/ Positive | 0 | 0 | 4 | 36 | 100% |
| Semi-Quantitative / Negative | 36 | 4 | 0 | 0 | 100% |
- Stability -
a. A closed accelerated stability study was performed on reagents, calibrators and controls at 25°C to establish the initial expiration dating. The stability study supported an initial expiration date of 1 year for reagents. This stability study supported an initial expiration date of 12 months for calibrators and controls. The instrument used for this test was a Beckman Coulter AU 400e.
-
- The following is a summary of the qualitative stability data. The 0 and 3.75ng/mL levels were negative in comparison to the 5ng/mL cutoff for Day 0. 2. 8. 16. 24. 32 and 40. The 6.25ng/mL level was positive in comparison to the 5ng/mL cutoff for Day 0, 2, 8, 16, 24, 32 and 40. This accelerated stability study was performed to establish initial expiration dating. Real time stability studies are ongoing.
-
- The following is a summary of the semi-quantitative stabilility data for the 5ng/mL cutoff. The 3.75ng/mL level was negative in comparison to the 5ng/mL cutoff for Day 0, 2, 8, 16, 24, 32 and 40. The 6.25ng/mL level was positive in comparison to the 5ng/mL cutoff for Day 0, 2, 8, 16, 24, 32 and 40. This accelerated stability study was performed to establish initial expiration dating. Real time stability studies are ongoing.
- b. An open/ on-board stability study was performed on reagents to establish expiration dating when reagents are opened and stored on board the instrument at 2°C to 8°C. The stability study supported an initial open vial expiration date of 28 days. The instrument used for this test was a Beckman Coulter AU 400e.
-
- The following is a summary of the qualitative open/ on-board stability data for the 5ng/mL cutoff. All replicates for the 3.75ng/mL level were negative in comparison to the 5ng/mL cutoff for Day 0, 7, 14, 21 and 28. All replicates of the 6.25ng/mL level were positive in comparison to the 5ng/mL cutoff for Day 0, 7, 14, 21 and 28.
-
- The following is a summary of the semi-quantitative open/ onboard stability data for the 5ng/mL cutoff. The mean of the replicates for the 3.75ng/mL level were negative in comparison to the 5ng/mL cutoff for Day 0, 7, 14, 21 and 28. The mean of the replicates of the 6.25ng/mL level were positive in comparison to the 5ng/mL cutoff for Day 0, 7, 14, 21 and 28.
-
-
- Calibrator and Control Traceability - all components of the calibrator and controls have been traced to a commercially available standard solution from Cerilliant Chemicals.
{10}------------------------------------------------
-
- Calibrator and Control Stability An open accelerated stability study was performed at 37°C to establish the initial open vial expiration dating. The stability study supported an initial open vial expiration date of 6 months. The instrument used for this test was a Beckman Coulter AU 400e. All calibrator levels (5, 10, 20 and 40ng/mL) and control levels (3.75 and 6.25ng/mL) were within specifications for Day 0, 3, 7, 10 and 13. This accelerated stability study was performed to establish initial expiration dating. Real time stability studies are ongoing.
- မှ.
စာ Calibrator and Control Value Assignment - calibrators and controls are manufactured and are tested by mass spectrometry. If any of the analytes are out of the acceptable range, then the calibrator or control is adjusted and retested. Values are assigned to the calibrator and controls once the Mass spectrometry results are within the acceptable ranges.
H. Conclusion
The information provided in this pre-market notification demonstrates that the Immunalysis Buprenorphine Urine Enzyme Immunoassay is substantially equivalent to the legally marketed predicate device for its general intended use.
{11}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avente Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002
July 1, 2014
IMMUNALYSIS CORPORATION JOSEPH GINETE REGULATORY AFFAIRS SPECIALIST 829 TOWNE CENTER DR. POMONA CA 91767
Re: K141406
Trade/Device Name: Immunalysis Buprenorphine Urine Enzyme Immunoassay Immunalysis Buprenorphine Urine Controls Immunalysis Buprenorphine Urine Calibrators
Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: II Product Code: DJG. DLJ. LAS Dated: May 23, 2014 Received: May 28, 2014
Dear Mr. Joseph Ginete:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,
If your device is classified (see above) into cither class II (Snecial Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Parts 801 and 809): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{12}------------------------------------------------
Page 2-Mr. Ginete
electronic product radiation control provisions (Scctions 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809). please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.lda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Courtney H. Lias -S
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{13}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known) K141406
Device Name
Immunalysis Burrenorphine Urine Enzyme Immunalysis Buprenorphine Urine Controls and Calibrators
Indications for Use (Describe)
Immunalysis Buprenorphine Urine Enzyme Immunoassay:
The Inmunalysis Buprenorphine Urine Enzyne Inmunoassay is a homogeneous corzyme immunoassay with a cutoff of Sng/mL. The assay is intended for use in laboratories for the quantitative and semi-quantitative analysis of Buprenorphine with automated clinical chemistry analyzers. This assay is calibrated against Burrenorphine. This in-vitro diagnostic device is for preseription use only.
The semi-quanitative mode is for purposes of enabling laboratories to determine an appropriate dilution of the speciment for confirmation by a confirmatory method such as GC-MS or permitting laboratories to establish quality control procedures. The Immunalysis Buprenorphine Urine Enryne Immunoassay Kit provides only a preliminary analytical test result. A more specific alternate chemical must be used in order to obtain a confirmed analytical result. Gas Chromatography/ Mass Spectrometry (GC-MS) or Liguid Chromatography / Mass Spectrospopy (LC/MS) is the preferred confirmatory method. Clinical consideration and professional judgment should be any drug of abuse test result, particularly when preliminary positive results are used.
Immunalysis Buprenorphine Urine Controls:
The Immunalysis Buprenomhine Urine Control naterials in the Inimunalysis Buprenorphine Urine Enzyne Immunoassay.
Immunalysis Buprenorphine Urine Calibrators:
The Innunalysis Buprenomhine Urine Calibrators are used as calibrators in the Immunalysis Bunrenomhine Urine Enzyme Immunoassy for the qualitative and semination of Burgnomhine in urine on automated clinical chemistry analyzers.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
enise Johnson-lyles -
§ 862.3650 Opiate test system.
(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).