K Number
K141327

Validate with FDA (Live)

Device Name
MINIMA S SYSTEM
Date Cleared
2014-08-06

(77 days)

Product Code
Regulation Number
888.3353
Age Range
All
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Minima S System is indicated for use in partial or total hip arthroplasty and is intended for press-fit (uncemented) use. When used in total hip arthroplasty, the Minima S Stems are intended for use with compatible femoral heads and acetabular components. When used in partial hip arthroplasty, the Minima S stems are intended for use with Lock Bipolar Heads. Hip arthroplasty is intended for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

  • non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis and hip dysplasia;
  • rheumatoid arthritis;
  • treatment of femoral head and neck fractures; .
  • revisions in cases of good remaining femoral bone stock.
Device Description

The Minima S System is intended for partial or total hip arthroplasty in cementless use. It is a monolithic collarless stem available in 9 sizes (#4-#12) in standard and lateralized versions.
When used in total hip arthroplasty Minima S stems are coupled to:

  • Biolox Delta femoral heads (object of this submission) or Limacorporate Femoral Heads (K112158) articulating with Limacorporate Cemented Cups (K112158) or
  • Biolox Delta heads (object of this submission) or Limacorporate Femoral Heads (K112158) articulating with Delta TT Acetabular System (K112898).
    When used in partial hip arthroplasty the Minima S femoral stem is coupled to Lock Bipolar Heads (Limacorporate K112158).
    The Minima S stem is made of Ti6Al4V and it has a plasma sprayed titanium coating in the proximal area (ASTM F1472 - ISO 5832-3). The stem is characterized by a 12/14 conical taper to be coupled to Limacorporate Femoral Heads, Biolox Delta femoral heads or Lock Bipolar Heads. In addition necks are lowered to reduce accidental contact between stem and acetabular cups and they are mirror-polished to reduce abrasion of the UHMWPE cups in case of abnormal contact. The stem has a rectangular section and is characterized by a "V" shaped A-P profile to improve adaptability to the most common bone morphologies and to facilitate insertion of the stem in the canal.
    The Minima S System stems is available in standard and lateralized versions with different CCD angles (131° and 134°), offsets and stem lengths.
    Biolox Delta Heads devices are used by surgeons to replace the head of the femur during total or partial hip surgery. They are characterized by a spherical shape and are coupled with the acetabular cup (K112158, K112898) inserted in the acetabulum, in total hip replacement, or with Lock Bipolar Heads (K112158) in partial hip replacement.
    Biolox Delta heads are coupled with the Minima S stems by means of a 12/14 Morse taper. Biolox Delta heads are made of Biolox Delta ceramic and are available with 28, 32 and 36 mm of diameters and in sizes (offsets) S, M, L and XL (XL size available only for head size 36).
AI/ML Overview

This document is a 510(k) premarket notification for the Minima S System, a hip replacement device. It primarily focuses on demonstrating substantial equivalence to previously marketed devices rather than presenting the results of a primary clinical study for the Minima S System itself. As such, information regarding acceptance criteria, device performance, and study design elements typically found in a clinical trial report is not present in this document for the Minima S System.

However, the document does describe non-clinical testing performed to demonstrate substantial equivalence. Based on the provided text, here's an analysis of the relevant sections, acknowledging that many requested fields cannot be filled due to the nature of this submission:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Non-Clinical)Reported Device Performance (Non-Clinical)
Minima S System (Stems):
Fatigue resistance according to ISO 7206-6Tests results demonstrated the device's ability to perform under expected clinical conditions.
Fatigue resistance according to ISO 7206-4Tests results demonstrated the device's ability to perform under expected clinical conditions. (Included evaluation of pull-out force for ceramic heads).
Titanium Plasma Spray coating conformity to FDA guideline and referenced standardsCoating was characterized to verify conformity.
Biolox Delta Heads:
Burst strength according to FDA Guidance Document for the Preparation of Premarket Notifications for Ceramic Ball Hip SystemsTests results demonstrated the device's ability to perform under expected clinical conditions.
Fatigue according to FDA Guidance Document for the Preparation of Premarket Notifications for Ceramic Ball Hip SystemsTests results demonstrated the device's ability to perform under expected clinical conditions.
Post-fatigue evaluation according to FDA Guidance Document for the Preparation of Premarket Notifications for Ceramic Ball Hip SystemsTests results demonstrated the device's ability to perform under expected clinical conditions.

Explanation: The document states that the non-clinical tests demonstrated the devices' ability to perform under expected clinical conditions, implying that the acceptance criteria for these tests were met. However, the specific quantitative pass/fail criteria (e.g., maximum deflection, number of cycles survived, minimum pull-out force values, specific burst pressure) are not detailed in this summary.


Information Not Available or Not Applicable Based on the Provided Document:

The following information is specifically requested for "the study that proves the device meets the acceptance criteria." Since this document details a 510(k) submission based on substantial equivalence and non-clinical testing, and explicitly states "Clinical testing was not necessary," most of the requested fields related to clinical studies, human reader performance, and ground truth establishment from patient data are not applicable and are therefore left blank or marked as "N/A."

  1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size: Not specified for the non-clinical tests. Non-clinical mechanical tests typically use a smaller number of physical samples (e.g., n=3 or n=6 per test condition) rather than a "test set" of patient data.
    • Data Provenance: N/A (Non-clinical mechanical testing).
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: N/A (No clinical test set requiring expert ground truth in this document).

  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: N/A (No clinical test set requiring adjudication in this document).

  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document refers to a hip replacement system, not an AI diagnostic device.

  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. This document refers to a hip replacement system, not an AI diagnostic device.

  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): N/A. For the non-clinical tests, the "ground truth" would be the engineering specifications and performance limits defined by the referenced ISO standards and FDA guidance documents.

  7. The sample size for the training set: Not applicable. This document refers to non-clinical testing for a physical medical device, not a machine learning model.

  8. How the ground truth for the training set was established: Not applicable. This document refers to non-clinical testing for a physical medical device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 6, 2014

Limacorporate S.p.A. % Stephen Peoples. VMD Peoples and Associates Consulting LLC 5010 Lodge Pole Lane Fort Wayne, Indiana 46814

Re: K141327 Trade/Device Name: Minima S System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO, JDI, KWY, KWZ, LPH, MBL Dated: June 11, 2014 Received: June 13, 2014

Dear Dr. Peoples:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Dr. Stephen Peoples

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Lori A. Wiggins

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known): K141327 Device Name: Minima S System Indications for Use:

Minima S System Indications for Use

The Minima S System is indicated for use in partial or total hip arthroplasty and is intended for press-fit (uncemented) use. When used in total hip arthroplasty, the Minima S Stems are intended for use with compatible femoral heads and acetabular components. When used in partial hip arthroplasty, the Minima S stems are intended for use with Lock Bipolar Heads. Hip arthroplasty is intended for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

  • · non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis and hip dysplasia;
  • rheumatoid arthritis;
  • treatment of femoral head and neck fractures; .
  • · revisions in cases of good remaining femoral bone stock.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

Elizabeth D%rank -S

Division of Orthopedic Devices

{3}------------------------------------------------

Summary of Safety and Effectiveness

Date: July 16, 2014

Manufacturer: Limacorporate S.p.A. Via Nazionale, 52 33038 - Villanova di San Daniele Udine - Italy

U.S. Contact Person: Dr. Stephen J. Peoples Principal Consultant Phone: 260-645-0327 FAX: +39 0432945512

ProductCommonNameProductCodeRegulation and Classification Name
Minima SSystemTotal orHemiHipProsthesisLZOHip joint metal/ceramic/polymer semi-constrained cemented or nonporousuncemented prosthesis per 21 CFR888.3353
JDIHip joint metal/polymer semi-constrainedcemented prosthesis per 21 CFR888.3350
KWYHip joint femoral (hemi-hip)metal/polymer cemented or uncementedprosthesis per 21 CFR 888.3390
KWZHip joint metal/polymer constrainedcemented or uncemented prosthesis per21 CFR 888.3310
LPHHip joint metal/polymer/metal semi-constrained porous-coated uncementedprosthesis per 21 CFR 888.3358
MBLHip joint metal/polymer/metal semi-constrained porous-coated uncementedprosthesis per 21 CFR 888.3358

Description:

The Minima S System is intended for partial or total hip arthroplasty in cementless use. It is a monolithic collarless stem available in 9 sizes (#4-#12) in standard and lateralized versions.

When used in total hip arthroplasty Minima S stems are coupled to:

  • Biolox Delta femoral heads (object of this submission) or Limacorporate Femoral ● Heads (K112158) articulating with Limacorporate Cemented Cups (K112158) or
  • Biolox Delta heads (object of this submission) or Limacorporate Femoral Heads (K112158) articulating with Delta TT Acetabular System (K112898).

{4}------------------------------------------------

When used in partial hip arthroplasty the Minima S femoral stem is coupled to Lock Bipolar Heads (Limacorporate K112158).

The Minima S stem is made of Ti6Al4V and it has a plasma sprayed titanium coating in the proximal area (ASTM F1472 - ISO 5832-3). The stem is characterized by a 12/14 conical taper to be coupled to Limacorporate Femoral Heads, Biolox Delta femoral heads or Lock Bipolar Heads. In addition necks are lowered to reduce accidental contact between stem and acetabular cups and they are mirror-polished to reduce abrasion of the UHMWPE cups in case of abnormal contact. The stem has a rectangular section and is characterized by a "V" shaped A-P profile to improve adaptability to the most common bone morphologies and to facilitate insertion of the stem in the canal.

The Minima S System stems is available in standard and lateralized versions with different CCD angles (131° and 134°), offsets and stem lengths.

Biolox Delta Heads devices are used by surgeons to replace the head of the femur during total or partial hip surgery. They are characterized by a spherical shape and are coupled with the acetabular cup (K112158, K112898) inserted in the acetabulum, in total hip replacement, or with Lock Bipolar Heads (K112158) in partial hip replacement.

Biolox Delta heads are coupled with the Minima S stems by means of a 12/14 Morse taper. Biolox Delta heads are made of Biolox Delta ceramic and are available with 28, 32 and 36 mm of diameters and in sizes (offsets) S, M, L and XL (XL size available only for head size 36).

Intended Use:

The Minima S System stems are indicated for use in partial or total hip arthroplasty and are intended for press-fit (uncemented) use. When used in total hip arthroplasty, the Minima S Stems are intended for use with compatible femoral heads and acetabular components. When used in partial hip arthroplasty, the Minima S stems are intended for use with Lock Bipolar Heads.

Hip arthroplasty is intended for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

  • · non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis and hip dysplasia;
  • · rheumatoid arthritis;
  • treatment of femoral head and neck fractures;
  • · revisions in cases of good remaining femoral bone stock.

Predicate Devices:

For the Minima S Stem:

  • H-MAX S stems (Limacorporate, K112091): ●
  • Encore Linear Hip (DJO, K972791 and K991325); ●
  • Modulus Femoral Hip System (Limacorporate, K112158); ●
  • Taperloc 12/14 Taper Femoral Component (BIOMET, K043537). ●

{5}------------------------------------------------

For the Biolox Delta Heads:

  • Femoral Heads (Limacorporate, K112158):
  • Articul/eze (DJO, K883460 and K980513); ●
  • Biolox Delta Ceramic Femoral Heads (Zimmer, K130899); ●
  • Biolox Delta Ceramic Femoral Heads (Smith & Nephew, K100412, K083762). ●

Comparable Features to Predicate Device(s):

The Minima S stems are similar to the predicate devices in terms of intended use, indications, design and materials. The Minima S stems can be used in total hip arthroplasty as all predicates and they can be used also in partial hip arthroplasty as the Modulus Femoral Hip System (K112158). All stems are intended for cementless use.

The Minima S stems has a design which is similar to predicates with the exception that Modulus stems are made of 2-pieces coupled through conical taper. Like predicates, the subject device is available in standard and lateralized versions, with different CCD angles, offsets and sizes.

The components of the Minima S stems are manufactured from the same materials as the predicate devices.

The same considerations can be done for the Biolox Delta Heads which are similar for intended use, design and materials to predicates

Non-Clinical Testing:

The Minima S System was tested for fatigue resistance. Mechanical tests were performed on worst case components or constructs. The following tests were performed to demonstrate substancial equivalency of the Minima S System to predicate devices:

  • Fatigue Test according to ISO 7206-6; ●
  • Fatigue Tests according to ISO 7206-4 (at the end of the test the pull-out force for ceramic heads has been evaluated).

The tests results demonstrated the device's ability to perform under expected clinical conditions.

In addition the Titanium Plasma Spray coating has been characterized to verify the conformity to FDA guideline and referenced standards.

Biolox Delta heads have been tested for burst strength, fatigue and post-fatigue evaluation in accord to FDA Guidance Document for the Preparation of Premarket Notifications for Ceramic Ball Hip Systems. The tests results demonstrated the device's ability to perform under expected clinical conditions.

{6}------------------------------------------------

Clinical Testing:

Clinical testing was not necessary to demonstrate substantial equivalence of the Minima
S System to the predicate devices.

§ 888.3353 Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis.

(a)
Identification. A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.(b)
Classification. Class II.