K Number
K141250

Validate with FDA (Live)

Date Cleared
2014-06-11

(28 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The KARL STORZ Flexible Video-Uretero-Renoscope System is Indicated for endoscopic examination In the urinary tract and can be used to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures.

Device Description

The Flexible Video-Uretero-Renoscope System is used for visualization purposes during diagnostic and therapeutic procedures. The system components are the Flexible Video-Uretero-Renoscope and the Image 1 SPIES Camera Control Unit (CCU). The Flexible Video-Uretero-Renoscope uses an LED light integrated in the handle and fiber light guides to illuminate the cavity under examination. The video image is produced by a complementary metal-oxide-semiconductor (CMOS) imaging sensor is located at the tip of the insertion shaft. The imaging sensor transfers the video signal to the Image1 SPIES CCU via electronics in the handle. The Image1 HD CCU processes the sensor images and displays them on a standard HD display monitor. The modifications made to the Flexible Video-Uretero-Renoscope System (K131369) are the conversion to the Image 1 SPIES CCU (K135319), material changes for STERRAD compatibility, a reduction in the insertion portion from 700mm to 675mm, and revision of the reprocessing instructions.

AI/ML Overview

This document is a 510(k) summary for the KARL STORZ Flexible Video-Uretero-Renoscope System. It describes modifications to a previously cleared device (K131369) and asserts substantial equivalence to that predicate device.

Here's an analysis of the provided information regarding acceptance criteria and the study:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not present a formal table of acceptance criteria with corresponding performance results. Instead, it describes general compliance with standards and successful testing.

Acceptance Criteria CategoryDevice Performance (as reported)
Risk ManagementEvaluated according to ISO 14971 risk management process.
Functional and Performance TestingSystem was successfully tested for its functions and performance.
Optical CharacteristicsVerification of optical characteristics per ISO 8600 (image quality, illumination).
Electrical SafetyPerformed electrical safety testing per IEC 60601-1.
Electromagnetic Compatibility (EMC)Performed EMC testing per IEC 60601-1-2.
BiocompatibilityPerformed biocompatibility testing of patient contacting materials per ISO 10993.
Reprocessing (Cleaning/Sterilization)Validations conducted for the manual cleaning method and sterilization process.
Substantial EquivalenceModified system is substantially equivalent to its predicate devices (K131369), with similar indications for use, methods of operation, design, and materials (or substantially equivalent materials).

2. Sample Size Used for the Test Set and Data Provenance:

  • The document does not specify a sample size for any test set (e.g., number of devices tested, number of simulated uses, etc.).
  • The data provenance is non-clinical testing performed by the manufacturer, Karl Storz Endoscopy-America, Inc. There is no mention of country of origin of the data beyond implied internal testing by the applicant. The testing described is prospective in nature, as it's for a modified device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

  • This information is not applicable as the document explicitly states: "Clinical testing was not required to demonstrate substantial equivalence to the predicate device." Therefore, there was no clinical test set requiring expert-established ground truth.

4. Adjudication Method for the Test Set:

  • Not applicable for the reasons stated above; no clinical test set was performed.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. The document states "Clinical testing was not required."

6. If a Standalone (algorithm only without human-in-the-loop performance) was done:

  • This device is a physical medical instrument (Flexible Video-Uretero-Renoscope System) with an imaging sensor, not an AI algorithm. Therefore, the concept of "standalone (algorithm only without human-in-the-loop performance)" is not applicable. The testing described is for the device's physical and functional performance, optical characteristics, and safety.

7. The Type of Ground Truth Used:

  • For the non-clinical performance and safety testing, the "ground truth" was established by compliance with recognized international standards (e.g., ISO 14971, ISO 8600, IEC 60601-1, IEC 60601-1-2, ISO 10993) and the validated functional specifications of the device.

8. The Sample Size for the Training Set:

  • This device is a physical medical instrument, not an AI/ML algorithm. Therefore, the concept of a "training set" in the context of machine learning is not applicable.

9. How the Ground Truth for the Training Set Was Established:

  • As above, the concept of a "training set" is not applicable for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for Karl Storz Endoskope. The logo consists of the word "STORZ" in large, bold letters, with a circle in the middle of the "O". Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font. The logo is black and white.

KARL STORZ Premarket Notification Flexible Video-Uretero-Renoscope System 008_Summary of Safety and Effectiveness

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

JUN 1 1 2014

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.

Applicant:Karl Storz Endoscopy-America, Inc2151 E. Grand AvenueEI Segundo, CA 90245
Contact:Leigh SpottenRegulatory Affairs ManagerPhone: (424) 218-8738Fax: (424) 218-8519
Date of Preparation:May 08, 2014
Device Identification:Trade Name: Flexible UreteroscopeCommon Name: Flexible Video-Uretero-Renoscope SystemClassification Name: Ureteroscope and Accessories, Flexible/Rigid
Product Code:FGB
Regulation:21 CFR part 876.1500
Predicate Device(s):KARL STORZ Flexible Video-Uretero-Renoscope System,K131369
Device Description:The Flexible Video-Uretero-Renoscope System is used forvisualization purposes during diagnostic and therapeuticprocedures. The system components are the Flexible Video-Uretero-Renoscope and the Image 1 SPIES Camera ControlUnit (CCU). The Flexible Video-Uretero-Renoscope uses anLED light integrated in the handle and fiber light guides toilluminate the cavity under examination. The video image isproduced by a complementary metal-oxide-semiconductor(CMOS) imaging sensor is located at the tip of the insertionshaft. The imaging sensor transfers the video signal to theImage1 SPIES CCU via electronics in the handle. TheImage1 HD CCU processes the sensor images and displaysthem on a standard HD display monitor.The modifications made to the Flexible Video-Uretero-Renoscope System (K131369) are the conversion to the

{1}------------------------------------------------

Image 1 SPIES CCU (K135319), material changes for STERRAD compatibility, a reduction in the insertion portion from 700mm to 675mm, and revision of the reprocessing instructions.

Indications For Use: The KARL STORZ Flexible Video-Uretero-Renoscope System is indicated for endoscopic examination in the urinary tract and can be used to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures

Technological

Non-Clinical

Clinical

Conclusion:

Performance Data:

Performance Data:

Characteristics:

The Karl Storz modified Flexible Video-Uretero-Renoscope System has the same indications for use as the originally cleared Karl Storz Flexible Video-Uretero-Renoscope System (K131369). The methods of operation; design and materials used are either identical or substantially equivalent to existing legally marketed predicate devices

The modifications incorporated into the KARL STORZ Flexible Video-Uretero-Renoscope System have been evaluated according to ISO 14971 risk management process and the system was successfully tested for its functions and performance; including verification of optical characteristics per ISO 8600 (image quality, illumination). Safety testing was performed including electrical safety IEC 60601-1, electromagnetic compatibility per IEC 60601-1-2, and biocompatibility of the patient contacting materials per ISO 10993. Additional validations were conducted for the manual cleaning method, sterilization process.

Clinical testing was not required to demonstrate substantial equivalence to the predicate device.

The modified Flexible Video-Uretero-Renoscope System is substantially equivalent to its predicate devices. The nonclinical testing demonstrates that the device is as safe, as effective and performs as well as or better than the legally marketed devices.

. .

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and three bars underneath the eagle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 11, 2014

Karl Storz Endoscopy-America, Inc. Leigh Spotten Regulatory Affairs Manager 2151 E. Grand Avenue El Segundo, CA 90245-5017

Re: K141250

Trade/Device Name: Flexible Video-Uretero-Renoscope System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FGB Dated: May 13, 2014 Received: May 14, 2014

Dear Leigh Spotten,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{3}------------------------------------------------

Page 2 - Leigh Spotten

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

. ..

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Herbert P. Lerner -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): _K141250

Device Name: Flexible Video-Uretero-Renoscope System

Indications for Use:

The KARL STORZ Flexible Video-Uretero-Renoscope System is Indicated for endoscopic examination In the urinary tract and can be used to examine the interior of the kidney, and using additional accessories, to perform various diagnostic and therapeutic procedures.

:

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

..

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert P. Lerner-S
2014.06.11 17:46:51 -04'00'

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.