K Number
K141201
Device Name
A&D MEDICAL UA-767F AND UA-767FAC DIGITAL BLOOD PRESSURE MONITOR
Manufacturer
Date Cleared
2014-10-08

(152 days)

Product Code
Regulation Number
870.1130
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The UA-767F and UA-767FAC are designed to be used by users who are older than twelve (12) years at home or clinics to monitor their blood pressure (systolic and diastolic) and pulse rate. The arm circumference range shall be between 8.7 inches (22.0 cm) and 17.7 inches (45.0 cm).
Device Description
UA-767F and UA-767FAC have the same design except UA-767FAC having an AC plug for an external AC adapter connection. Both devices use an inflatable cuff which is wrapped around the patient's upper arm. After the user pushes the "START" button, the cuff is inflated automatically by an internal pump of UA-767F and UA-767FAC. The systolic and diastolic blood pressures are determined by oscillometric method. The deflation rate is controlled by the internal exhaust valve. There is a quick exhaust mechanism so that the cuff pressure can be completely released urgently. There is a maximum pressure safety setting at 300mmHg. UA-767F and UA-767FAC will not inflate the cuff higher than 300mmHg. UA-767F and UA-767FAC will turn on an irregular heartbeat indicator if an irregular heartbeat was detected during the measurement process. At the end of the measurement, the systolic and diastolic pressures with pulse rate are shown on the LCD and stored in the device memory. The cuff is also deflated automatically to 0 mmHg at the same time.
More Information

Not Found

No
The description details a standard oscillometric blood pressure monitor with basic features like irregular heartbeat detection and memory storage. There is no mention of AI or ML algorithms for data analysis, prediction, or enhanced functionality.

No
The device is described as a blood pressure monitor, which is used for monitoring and measuring, not for treating or providing therapy.

Yes

The device is designed to monitor blood pressure (systolic and diastolic) and pulse rate, providing information about a user's physiological state, which is a key characteristic of a diagnostic device.

No

The device description clearly outlines hardware components such as an inflatable cuff, internal pump, internal exhaust valve, LCD, and device memory, indicating it is not a software-only device.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is to monitor blood pressure and pulse rate. This is a physiological measurement taken directly from the patient's body.
  • Device Description: The device uses an inflatable cuff and oscillometric method to measure blood pressure. This is a physical measurement technique, not a test performed on a biological sample in vitro (outside the body).
  • Lack of IVD Characteristics: There is no mention of analyzing biological samples (like blood, urine, etc.) or performing tests on these samples to diagnose a condition.

IVD devices are specifically designed to examine specimens derived from the human body to provide information for the diagnosis, prevention, monitoring, treatment, or alleviation of disease. This device does not fit that description.

N/A

Intended Use / Indications for Use

The UA-767F and UA-767FAC are designed to be used by users who are older than twelve (12) years at home or clinics to monitor their blood pressure (systolic and diastolic) and pulse rate. The arm circumference range shall be between 8.7 inches (22.0 cm) and 17.7 inches (45.0 cm).

Product codes

DXN

Device Description

UA-767F and UA-767FAC have the same design except UA-767FAC having an AC plug for an external AC adapter connection. Both devices use an inflatable cuff which is wrapped around the patient's upper arm. After the user pushes the "START" button, the cuff is inflated automatically by an internal pump of UA-767F and UA-767FAC. The systolic and diastolic blood pressures are determined by oscillometric method. The deflation rate is controlled by the internal exhaust valve. There is a quick exhaust mechanism so that the cuff pressure can be completely released urgently. There is a maximum pressure safety setting at 300mmHg. UA-767F and UA-767FAC will not inflate the cuff higher than 300mmHg. UA-767F and UA-767FAC will turn on an irregular heartbeat indicator if an irregular heartbeat was detected during the measurement process. At the end of the measurement, the systolic and diastolic pressures with pulse rate are shown on the LCD and stored in the device memory. The cuff is also deflated automatically to 0 mmHg at the same time.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

patient's upper arm

Indicated Patient Age Range

older than twelve (12) years

Intended User / Care Setting

users ... at home or clinics

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

We conducted design verification and design validation activities based on the comparison of the UA-767F and UA-767FAC with the predicate devices. Based on the changes, we conducted the appropriated test methodology and pass/fail criteria. After the tests were conducted, the test records were collected in the design history file (DHF).
UA-767F and UA-767FAC met all applicable requirements of the standards. None of the test demonstrated that the UA-767F and UA-767FAC bring new issues of safety and effectiveness.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K043217, K111686

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc)

Not Found

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).

0

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an image of a stylized caduceus, which is a symbol of medicine.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 8, 2014

A&D Engineering, Inc. c/o Mr. Jerry Wang Director of Engineering 1756 Automation Parkway San Jose, CA 95131

Re: K141201

Trade/Device Name: A&D Medical UA-767F and UA-767FAC Digital Blood Pressure Monitors Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system. Regulatory Class: Class II Product Code: DXN Dated: September 5, 2014 Received: September 8, 2014

Dear Mr. Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

1

Page 2 - Mr. Jerry Wang

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Melissa A. Torres -S

For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

2

Indications for Use

510(k) Number (if known) K141201

Device Name

A&D Medical UA-767F & UA-767FAC Digital Blood Pressure Monitors

Indications for Use (Describe)

The UA-767F and UA-767FAC are designed to be used by users who are older than twelve (12) years at home or clinics to monitor their blood pressure (systolic and diastolic) and pulse rate. The arm circumference range shall be between 8.7 inches (22.0 cm) and 17.7 inches (45.0 cm).

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY CONSULT OF Connection

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

3

K141201 510(k) Summarv

This summary of 510(k) safety and effective information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.

1. Date Prepared

September 30, 2014

2. Submitter's Information

A&D Engineering, Inc. Mr. Jerry Wang 1756 Automation Parkway, San Jose, CA 95131 Tel: 408-518-5113 Fax: 408-635-2313 Email: jwang@andonline.com

3. Device Information

| Proprietary Name: | A&D Medical UA-767F and UA-767FAC Digital Blood
Pressure Monitors |
|----------------------|-------------------------------------------------------------------------------------|
| Common/Usual Name: | Blood Pressure Monitor |
| Classification name: | Non-invasive blood pressure measurement System
21 CFR 870-1130, Class II, 74DXN. |

4. Predicate Devices

A&D Model UA-767PBT Digital Blood Pressure Monitor with 510(k) number K043217 A&D Model UA-1000 Family Digital Blood Pressure Monitors with 510(k) number K111686 Predicate devices are designed and manufactured by the same company and facilities as the modified devices UA-767F and UA-767FAC.

ട. Device Description - Technological and Operational Characteristics

UA-767F and UA-767FAC have the same design except UA-767FAC having an AC plug for an external AC adapter connection. Both devices use an inflatable cuff which is wrapped around the patient's upper arm. After the user pushes the "START" button, the cuff is inflated automatically by an internal pump of UA-767F and UA-767FAC. The systolic and diastolic blood pressures are determined by oscillometric method. The deflation rate is controlled by the internal exhaust valve. There is a quick exhaust mechanism so that the cuff pressure can be completely released urgently. There is a maximum pressure safety setting at 300mmHg. UA-767F and UA-767FAC will not inflate the cuff higher than 300mmHg. UA-767F and UA-767FAC will turn on an irregular heartbeat indicator if an irregular heartbeat was detected during the measurement process. At the end of the measurement, the systolic and diastolic pressures with pulse rate are shown on the LCD and stored in the device memory. The cuff is also deflated automatically to 0 mmHg at the same time.

4

Indications of Use 6.

The UA-767F and UA-767FAC are designed to be used by users who are older than twelve (12) years at home or clinics to monitor their blood pressure (systolic and diastolic) and pulse rate. The arm circumference range shall be between 8.7 inches (22.0 cm) to 17.7 inches (45.0 cm).

This indication for use of A&D Medial UA-767F and UA-767FAC as described in the labeling and FDA IFU form 3881 are the same as their predicated devices, UA-1000 family and UA-767PBT.

7. Summary of Substantial Equivalence

Modifications made from the predicate devices:

  • . Change the plastic molds so UA-767F & UA-767FAC have a different appearance
  • Modify memory size and add %IHB feature in memory average display. Remove ● AM / PM function.
  • Enable up to four users to share the same device.
  • . Replace the static capacity type pressure sensor to semiconductor pressure sensor.
  • . Remove the following functions from UA-1000 family - Voice, Tricheck, and AM/PM measurement average.
  • Remove the Bluetooth wireless communication from UA-767PBT.
  • Add a new wide range cuff, model UA-420, to cover arm size 22 to 42 cm. The predicate UA-290 (23 to 37 cm) and UA-291 (31 to 45 cm) cuffs remain working.

| Parameter | Predicate Devices
(UA-1000 family & UA-767PBT) | Modified Devices
(UA-767F & UA-767FAC) |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Power source | 4 AA size batteries and AC
adaptor as an option | No Change – the same |
| Battery Life | 4 months with daily measurement | No change – the same |
| Measurement
Method | Oscillometric Method | No change – the same |
| Measurement
Range | BP : 20 to 280 mmHg
Pulse : 30 to 200 beats/min | BP: 0 -299 mmHg
Pulse: 30 to 200 beats/min |
| Accuracy | BP : +/- 3mmHg or +/- 2% of
measured value, whichever is
greater
Pulse : +/- 5 % (pulse) | BP : +/- 3mmHg
Pulse : +/- 5 % |
| Pressurization
Source | Automatic internal pump | No change – the same |
| Cuff Deflation
Method | UA-767PBT – Standard exhaust
valve
UA-1000 family – Constant speed
electrical controlled exhaust valve
(ECEV method) | No change – the same as
UA-767PBT |
| Display Type | Liquid crystal display | No change – the same |
| Cuff Attachment
Method | | |
| By plastic hose connected to
monitor | No change – the same | |
| IHB (Irregular
Heartbeats
Detection) | More than +/-25% to the mean
interval of all pulse intervals | No change – the same |
| Operating
Environment | 50°F (10°C) to 104°F (40°C)
30 %RH to 85% RH | 50°F (10°C) to 104°F (40°C)
15 %RH to 85% RH |
| Storage
Environment | 14°F (-20°C) to 140°F (60°C)
10 %RH to 95% RH | No change – the same |
| Pressure Indicator | UA-1000 family (Yes) under USA
JUC VII guideline | No change – the same |
| Number of User | One | Up to four |
| Data Memory Size
with Time & Date | 90 memories for UA-1000 family
40 memories for UA-767PBT | Change to 60 memories for
each user with %IHB feature |
| Pressure Sensor
design | Static electricity capacity type | Semiconductor type |
| Dimensions in mm | UA-767PBT: 163.7 x 111 x 66.7
UA-1000 Family: 140 x 60 x 105 | 96 (W) x 68 (H) x 130 (D) |
| Weight | UA-1010–265g
UA-1020–285g
UA-1030T–300g without batteries | 240g without batteries |
| Cuff Design | UA-767PBT – D-ring cuffs
UA-1000 family – U-shape cuffs | No change – the same as
UA-767PBT |
| Arm Size | UA-289 small cuff : 16 to 24 cm
UA-290 medium cuff : 23 to 37
cm
UA-291 large cuff : 31 to 45 cm | No change – UA-290 and
UA-291.
New cuff model – UA-420 :
22 to 42cm. |
| Clock (Time/Date) | Yes for UA-1020 and UA-1030T | No change – the same |
| Talking | Yes – UA-1030T, No - others | Removed |
| AM/PM | Yes – UA-1020 & UA-1030T | Removed |
| Wireless Radio
Connectivity | UA-1000 family (No), UA-
767PBT (Yes) | Removed |
| Personal PC
Analysis Software | UA-1000 family (No), UA-
767PBT (Yes) | Removed |

Product Specification Comparison

5

Key Features Comparison

ParameterPredicate DevicesModified Devices
(UA-1000 family & UA-(UA-767F & UA-
767PBT)767FAC)
Field serviceNot allowedNo Change - the same
Automatic zero at "START"YesNo Change - the same
Manual zero adjustmentNot allowedNo Change - the same
CalibrationNot allowed in the fieldNo Change - the same
SterilizationNot neededNo Change - the same

6

Discussion of standards used in the design verification and design validation 8.

We conducted design verification and design validation activities based on the comparison of the UA-767F and UA-767FAC with the predicate devices. Based on the changes, we conducted the appropriated test methodology and pass/fail criteria. After the tests were conducted, the test records were collected in the design history file (DHF).

A&D Medical follows FDA regulation and international standards in our medical device development and manufacturing processes. The following standards were used to demonstrate compliance to FDA recognized consensus standards for the UA-767F and UA-767FAC devices.

  • . ANSI/AAMI/IEC 80601-2-30:2009 Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated noninvasive sphygmomanometers.
  • . AAMI/ANSI 60601-1:2006 Medical electrical equipment. General requirements for basic safety and essential performance
  • AAMI/ANSI/IEC 60601-1-2:2007 Medical electrical equipment. General ● requirements for basic safety and essential performance. Collateral standard. Electromagnetic compatibility. Requirements and tests
  • . AAMI/ANSI/ISO 14971:2012 Medical devices. Application of risk management to medical devices

There is no new material involved from the predicate devices to UA-767F and UA-767FAC. So, there is no need to conduct any biocompatibility tests.

UA-767F and UA-767FAC met all applicable requirements of the standards. None of the test demonstrated that the UA-767F and UA-767FAC bring new issues of safety and effectiveness.

Substantial Equivalence Conclusion:

UA-767F and UA-767FAC Digital blood pressure monitors have the following similarities to the predicate devices, UA-1000 family and UA-767PBT digital blood pressure monitors, which previously received the 510(k) clearance.

| Comparison Area | Predicate Devices
UA-767PBT & UA-1000 | Subject Devices
(UA-767F & UA-767FAC) |
|-------------------|----------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Indication of Use | Measure blood pressure (systolic and diastolic) and pulse rate. | Measure blood pressure (systolic and diastolic) and pulse rate.
(identical to both) |
| Intended Use | Measure systolic, diastolic and pulse rate for adults at home and clinics. | Measure systolic, diastolic and pulse rate for adults at home and clinics.
(identical to both) |

7

| Measurement
Method | Oscillometric Method | Oscillometric Method
(identical to both) |
|-----------------------|--------------------------------------------|-------------------------------------------------------------------|
| Inflation Method | Automatic internal pump | Automatic internal pump
(identical to both) |
| Deflation Method | Standard exhaust valve | Standard exhaust valve
(identical to UA-767PBT) |
| Materials | PVC, nylon, metal. | PVC, nylon, metal
(identical to both) |
| Energy source | AA batteries with AC adapter as an option. | AA batteries with AC adapter as an option.
(identical to both) |

As a conclusion, the intended use of the modified device, UA-767F and UA-767FAC as described in its labeling, has not changed as a result of the modifications. The fundamental scientific technology of the modified device, UA-767F and UA-767FAC, has not changed, either. The results of the previous activities demonstrate that the A&D Medical UA-767F and UA-767FAC Digital Blood Pressure Monitors are as safe, as effective, and performs as well as or better than the predicate devices.