K Number
K141146

Validate with FDA (Live)

Device Name
INVISIPORT
Date Cleared
2015-05-01

(361 days)

Product Code
Regulation Number
880.5965
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Invisiport is indicated for any patient requiring repeated access of the vascular system for delivery of medications, nutritional supplementation, fluids, blood products, or the sampling of blood, or for power injection of contrast media when used with a power-injectable Huber needle or infusion set.

Device Description

The predicate device and subject device are injection ports. An open ended radiopaque catheter is preattached to the predicate and subject devices.

This 510(k) pre-market notification supports the addition of a catheter size variation to the predicate device. The modification to the predicate device to create the subject device is a change in the attached catheter. The predicate device attached catheter is substantially a 6 Fr catheter. The subject device attached catheter is a 4.4 Fr catheter. This change affects the attached catheter.

AI/ML Overview

The provided text is a 510(k) summary for a medical device called Invisiport™, specifically addressing a modification to an existing predicate device. The primary modification is a change in the attached catheter from 6 Fr to 4.4 Fr.

Here's an analysis based on your request, using the information available in the document:

1. A table of acceptance criteria and the reported device performance

The document provides a Table 1: Subject Device Performance Testing Summary which lists tests, a summary of requirements (acceptance criteria surrogates), and the result (Pass).

TestSummary of Requirement (Acceptance Criteria)Reported Device Performance (Result)
1. VisualVerify markings on OD with numbers at applicable increments.Pass
2. DimensionalVerify critical dimensions.Pass
3. Clearance ParametersVerify clearance volume.Pass
4. Port to Catheter Connection TestLateral and Axial (Dry): No leaks after static load.Lateral and Axial (Wet): No leaks after port exposed to wet medium and a static load.Cycle Lateral and Axial: No leaks after cyclic load.Creep: No leaks after minimal load.Load to Failure, Axial and Lateral.Pass
5. Port Leak TestWithstand static pressure without leaking for both intermittent and continuous pressure testing.Pass
6. Power Injection Flow RateVerify flow rate under power injection conditions with a set viscosity and flow.Pass

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the sample sizes used for each of the performance tests. It also does not explicitly state the country of origin of the data or whether the tests were retrospective or prospective. Given that these are bench tests for a 510(k) submission, they are inherently prospective, conducted specifically for this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This section is not applicable (N/A). The document describes bench performance tests, not studies involving expert evaluation of clinical data or images to establish ground truth. The "ground truth" for these tests is the physical performance against defined engineering requirements, verified by laboratory measurements.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is not applicable (N/A). Adjudication methods like 2+1 or 3+1 typically refer to the process of resolving discrepancies among multiple human readers in clinical studies or image interpretations. The tests described are bench tests with objective pass/fail criteria.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable (N/A). This document describes a 510(k) submission for a physical medical device (an infusion port and catheter), not an AI-powered diagnostic tool. Therefore, a multi-reader multi-case (MRMC) comparative effectiveness study involving human readers and AI assistance was not performed.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This section is not applicable (N/A). As mentioned above, this is a physical medical device, not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The ground truth for these performance tests is based on engineering specifications and established test methods. For example, for "Port Leak Test," the ground truth is "Withstand static pressure without leaking." This is verified through direct physical measurement against a predetermined threshold or standard.

8. The sample size for the training set

This section is not applicable (N/A). There is no "training set" mentioned in the context of this device's performance evaluation. Training sets are relevant for machine learning or AI models.

9. How the ground truth for the training set was established

This section is not applicable (N/A). As there is no training set for an AI model, the method for establishing its ground truth is irrelevant here.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its wing and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 1, 2015

Stealth Therapeutics, Inc. c/o Mr. Gary Syring Quality & Regulatory Associates, LLC 800 Levanger Lane Stoughton, WI 53589

Re: K141146

Trade/Device Name: Invisiport™ Regulation Number: 21 CFR 880.5965 Regulation Name: Subcutaneous, implanted, intravascular infusion port and catheter Regulatory Class: II Product Code: LJT Dated: March 31, 2015 Received: April 2, 2015

Dear Mr. Syring:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Gary Syring

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Tina
Kiang-S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K141146

Device Name Invisiport

Indications for Use (Describe)

The Invisiport is indicated for any patient requiring repeated access of the vascular system for delivery of medications, nutritional supplementation, fluids, blood products, or the sampling of blood, or for power injection of contrast media when used with a power-injectable Huber needle or infusion set.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

] Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Stealth Therapeutics Inc. The logo has the word "STEALTH" in large, bold, sans-serif font. Below the word "STEALTH" is the text "THERAPEUTICS INC." in a smaller, lighter font. There is a curved line above the word "STEALTH" and a burst of light to the right of the word.

K141146

510(k) Summary

This summary is provided to support the 510(k) notification for the subject Invisiport manufactured for Stealth Therapeutics, Inc.

Contact Name:Samuel Adams
Company Name:Stealth Therapeutics, Inc.
Address:101 Nob Hill Road, Suite 100Madison, WI 53713
Phone:(608) 217-2685
Date Summary Prepared:April 29, 2015
Trade Name:InvisiportTM
Common Name:Implantable Infusion Port
Classification Name:Subcutaneous, Implanted, Intravascular Infusion Port and Catheter21 CFR 880.5965, Product Code LJT
Predicate Device:K110407, InvisiportTM, manufactured for Stealth Therapeutics, Inc.

Product Description

The predicate device and subject device are injection ports. An open ended radiopaque catheter is preattached to the predicate and subject devices.

This 510(k) pre-market notification supports the addition of a catheter size variation to the predicate device. The modification to the predicate device to create the subject device is a change in the attached catheter. The predicate device attached catheter is substantially a 6 Fr catheter. The subject device attached catheter is a 4.4 Fr catheter. This change affects the attached catheter.

Intended Use of the Device

The intended use of the subject device is the same as the predicate device:

The Invisiport™ is indicated for any patient requiring repeated access of the vascular system for delivery of medications, nutritional supplementation, fluids, blood products, or the sampling of blood, or for power injection of contrast media when used with a power-injectable Huber needle or infusion set.

Summary of Technological Characteristics

The following table provides a side-by-side comparison of the subject device to the predicate device.

{4}------------------------------------------------

Table 1: Substantial Equivalence Technical Characteristics
FeatureSubject DevicePredicate DeviceComment
Indications forUseThe Invisiport™is indicated for any patient requiringrepeated access of the vascular systemfor delivery of medications, nutritionalsupplementation, fluids, blood, bloodproducts, or the sampling of blood, or forpower injection of contrast when usedwith a power-injectable Huber needle orinfusion set.The Invisiport™is indicated for any patient requiringrepeated access of the vascular systemfor delivery of medications, nutritionalsupplementation, fluids, blood, bloodproducts, or the sampling of blood, or forpower injection of contrast when usedwith a power-injectable Huber needle orinfusion set.Same
Power InjectionYesYesSame
Power InjectionInput PressureLimit300 psi300 psiSame
Power InjectionFlow Rate of11.8 cP fluid4.5 ml/sec5 ml/secBoth devices are powerinjectable at pressures up to 300psi.The subject device with a4.4 Fr catheter supports amaximum of 4.5 ml/sec powerinjection flow rate of an 11.8 cPfluid with a 300 psi powerinjection limit. The predicatedevice with a 6 Fr cathetersupports a 5 ml/sec powerinjection flow rate of an 11.8 cPfluid with a 300 psi powerinjection limit.The difference in flow rate isdue to the power injector'sability to support a flow ratewith an 11.8 cP fluid at 300 psiapplied pressure. The smaller4.4 Fr catheter supports a lowerflow rate at this 300 psi pressurelimit.
Port AccessHospital/clinic licensed health careproviderHospital/clinic licensed health careproviderSame
Location ofimplantPeripheral or thoracicPeripheral or thoracicSame
DesignSeptum/port with integrated catheterSeptum/port with integrated catheterSame
AttachedCatheter Length53.3 cm53.3 cmSame
Catheter Diameter1.47 mm (4.4 French)2.0 mm (6 French)A smaller diameter cathetervariation is available.

Performance tests to demonstrate substantial equivalency

The risks presented by a smaller catheter diameter have been evaluated. Applicable tests recommended in the FDA Guidance: Guidance on 510(k) Submissions for Implanted Infusion Ports and the tests performed to support substantial equivalence of the predicate device was evaluated. The following table identifies the tests performed on the subject device and results.

{5}------------------------------------------------

Table 1: Subject Device Performance Testing Summary
TestSummary of RequirementResult
1. VisualVerify markings on OD with numbers at applicable increments.Pass
2. DimensionalVerify critical dimensions.
3. ClearanceParametersVerify clearance volume.
4. Port to CatheterConnection TestLateral and Axial (Dry): No leaks after static load.Lateral and Axial (Wet): No leaks after port exposed to wet medium and a staticload.Cycle Lateral and Axial: No leaks after cyclic load.Creep: No leaks after minimal load.Load to Failure, Axial and Lateral.Pass
5. Port Leak TestWithstand static pressure without leaking for both intermittent and continuouspressure testing.Pass
6. Power InjectionFlow RateVerify flow rate under power injection conditions with a set viscosity and flow.Pass

All tests passed with acceptable results on the subject device.

No clinical data are required to confirm substantial equivalence of the subject device. Bench evaluations of potentially affected performance are adequate to support substantial equivalence.

Conclusions

The subject device is a modification of the predicate device and meets all established acceptance criteria for performance testing. The intended use and technology of the subject device is the same as the predicate device.

§ 880.5965 Subcutaneous, implanted, intravascular infusion port and catheter.

(a)
Identification. A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.