(238 days)
The FT4L method is an in vitro diagnostic test for the quantitative measurement of Free Thyroxine in human serum and plasma on the Dimension® EXL™ integrated chemistry system with LOCI® Module. Measurements of free thyroxine are used in the diagnosis and monitoring of thyroid disease.
The TSHL method is an in vitro diagnostic test for the quantitative measurement of Thyroid Stimulating Hormone (TSH, thyrotropin) in human serum and plasma on the Dimension® EXL™ integrated chemistry system with LOCI® Module. Measurements of TSH are used in the diagnosis and monitoring of thyroid disease.
The Dimension® LOCI Free Thyroxine Flex® reagent cartridge, FT4L and Dimension® LOCI Thyroid Stimulating Hormone Flex® reagent cartridge, TSHL are in vitro diagnostic tests for the quantitative measurement of Free Thyroxine and Thyroid Stimulating Hormone, respectively, in human serum and plasma on the Dimension® EXL™ integrated chemistry system with LOCI® Module. The submission is for the inclusion of pediatric reference intervals to the labeling (Package Inserts) of these assays. No changes were made to the reagents, device design, or manufacturing process.
The provided document is a 510(k) summary for the Siemens Healthcare Diagnostics Dimension® LOCI Free Thyroxine Flex® Reagent Cartridge (FT4L) and Dimension® LOCI Thyroid Stimulating Hormone Flex® Reagent Cartridge (TSHL). The focus of this submission is the addition of pediatric reference intervals to the existing devices, not a primary submission for a new device. Therefore, the information provided primarily pertains to the establishment of these pediatric reference intervals.
Here's an analysis based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the pediatric reference intervals are defined by the statistical methods used to establish them and the resulting intervals. The performance reported is the calculated pediatric reference intervals.
| Acceptance Criteria | Reported Device Performance (FT4L) | Reported Device Performance (TSHL) |
|---|---|---|
| Infant (01-23 months): Reference interval established using robust symmetrical method due to sample size < 120, validated against simulated non-parametric and robust symmetric intervals. | Reference Interval: 0.93 - 1.45 ng/dL | Reference Interval: 0.867 – 6.43 µIU/mL (Established using robust symmetrical method after log transformation due to skewed data) |
| Children (02-12 years): Reference interval established using non-parametric approach (2.5th and 97.5th percentiles). | Reference Interval: 0.82 - 1.40 ng/dL | Reference Interval: 0.704 - 4.01 µIU/mL |
| Adolescents (13-20 years): Reference interval established using non-parametric approach (2.5th and 97.5th percentiles). | Reference Interval: 0.78 - 1.34 ng/dL | Reference Interval: 0.516 - 4.13 µIU/mL |
| Confidence Intervals for Infant 95% Reference Intervals: 90% confidence intervals for lower and upper limits. | 90% CI of Lower Limit: 0.89 - 0.97 ng/dL 90% CI of Upper Limit: 1.41 - 1.50 ng/dL | 90% CI of Lower Limit: 0.710 - 1.04 µIU/mL 90% CI of Upper Limit: 5.40 - 7.54 µIU/mL |
2. Sample Size Used for the Test Set and the Data Provenance
The "test set" in this context refers to the patient samples used to establish the pediatric reference intervals.
- FT4L Assays: Data from a total of 411 patients
- Infants (01 – 23 months): 77 patients
- Children (02 – 12 years): 187 patients
- Adolescents (13 – 20 years): 147 patients
- TSHL Assays: Data from a total of 407 patients
- Infants (01 – 23 months): 75 patients
- Children (02 – 12 years): 185 patients
- Adolescents (13 – 20 years): 147 patients
Data Provenance: The document does not specify the country of origin. It also doesn't explicitly state whether the data was retrospective or prospective, but the phrasing "Data ... were analyzed to establish the reference intervals" suggests it was likely retrospective analysis of collected samples.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
The ground truth here is the definition of "healthy" individuals within the pediatric age groups to establish reference intervals. The document does not mention the use of experts to establish individual ground truths for each sample in the test set. Instead, it refers to a population used to establish the reference intervals. The underlying assumption is that these patients were deemed healthy for their age group. The statistical methods (CLSI C28-A3c, CLSI EP28-A3) mentioned are standard guidelines for establishing reference intervals in clinical laboratories, which are developed and endorsed by experts in the field of clinical pathology and laboratory medicine, but individual experts are not described as adjudicating each case.
4. Adjudication Method for the Test Set
No explicit adjudication method (e.g., 2+1, 3+1) is mentioned for the test set. The process involves collecting samples presumably from healthy individuals within the pediatric age groups and applying statistical methods guided by CLSI standards to determine the reference intervals.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This section is not applicable. The device (reagent cartridge) is an in vitro diagnostic test for quantitative measurement of hormones, not an imaging device or an AI-driven tool requiring human reader interpretation in the manner of an MRMC study. The study focuses on establishing reference intervals for a laboratory test.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This section is not applicable. The device is a laboratory reagent. The "performance" being evaluated here is the ability to accurately define reference intervals for existing assays, which is a statistical and clinical validation task, not an algorithm's standalone performance in the typical sense of AI. The assay itself works in a standalone (automated) manner on the Dimension® EXL™ system with LOCI® Module. The current submission's scope is the labeling modification to add pediatric reference intervals.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth for establishing pediatric reference intervals is based on healthy pediatric populations. The individuals included in the study for establishing reference intervals are assumed to be "healthy" (euthyroid, meaning normal thyroid function) for their age group. This is a common method for establishing reference ranges in laboratory medicine, following guidelines like CLSI C28-A3c. It's not individual expert consensus on specific cases, pathology reports, or long-term outcomes data, but rather a statistical representation of a healthy population.
8. The Sample Size for the Training Set
This submission is about establishing reference intervals, which traditionally does not involve a "training set" in the machine learning sense. The patient data described in point 2 is used directly to establish the reference intervals, rather than training an algorithm.
9. How the Ground Truth for the Training Set Was Established
As explained above, there isn't a "training set" in the typical AI/ML context. The patient samples used to establish the reference intervals are assumed to be from "healthy" individuals in the respective pediatric age groups, following standard clinical laboratory guidelines for reference interval studies.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human profile, composed of three overlapping shapes.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 26, 2014
SIEMENS HEALTHCARE DIAGNOSTICS INC. FRANCES DILLON SR. MANAGER, REGULATORY AFFAIRS PO BOX 6101, M/S 514 NEWARK DE 19714-6101
Re: K140842
Trade/Device Name: Dimension® LOCI Free Thyroxine Flex® Reagent Cartridge, FT4L: Dimension® LOCI Thyroid Stimulating Hormone Flex® Reagent Cartridge. TSHL Regulation Number: 21 CFR 862.1695 Regulation Name: Free thyroxine test system Regulatory Class: II Product Code: CEC. JLW Dated: November 21, 2014 Received: November 24, 2014
Dear Frances Dillon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{1}------------------------------------------------
electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Courtney
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K140842
Device Name
Dimension® LOCI Free Thyroxine Flex® reagent cartridge, FT4L
Indications for Use (Describe)
The FT4L method is an in vitro diagnostic test for the quantitative measurement of Free Thyroxine in human serum and plasma on the Dimension® EXL™ integrated chemistry system with LOCI® Module. Measurements of free thyroxine are used in the diagnosis and monitoring of thyroid disease.
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|---|
| ----------------------------------------------------------------------------------------------------------------- | ------------------------------------------------------------ |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Paperwork Reduction Act (PRA) Staff
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Type of Use (Select one or both, as applicable)
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K140842
Device Name
Dimension® LOCI Thyroid Stimulating Hormone Flex® reagent cartridge. TSHL
Indications for Use (Describe)
The TSHL method is an in vitro diagnostic test for the quantitative measurement of Thyroid Stimulating Hormone (TSH, thyrotropin) in human serum and plasma on the Dimension® EXL™ integrated chemistry system with LOCI® Module. Measurements of TSH are used in the diagnosis and monitoring of thyroid disease.
| ---------------------------------------------------------------------------------------------------------------------------------------------------- | ---------------------------------------------------------------------------------------------------------------------------------------------------- |
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|---|---|
| ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- | --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Type of Use (Select one or both, as applicable)
{4}------------------------------------------------
SIEMENS
Section 6. 510(k) Summary of Safety and Effectiveness
This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92 and the Safe Medical Device Act of 1990.
The assigned 510(k) Number is: K140842
1. Date Prepared
November 25, 2014
2. Applicant Information
| Contact: | Frances A. DillonSenior Manager, Regulatory Affairs |
|---|---|
| Address: | Siemens Healthcare Diagnostics Inc.P.O. Box 6101, M/S 514Newark, DE 19714 |
| Phone: | 302-631-6951 |
| Fax: | 302-631-6299 |
| Email: | frances.a.dillon@siemens.com |
Regulatory Information 3.
Table 1. Regulatory Information for Dimension® FT4L and TSHL Assays
| FT4L | TSHL | |
|---|---|---|
| Trade Name | Dimension® LOCI Free ThyroxineFlex® reagent cartridge, FT4L | Dimension® LOCI Thyroid StimulatingHormone Flex® reagent cartridge,TSHL |
| Model Numbers | RF610 | RF612 |
| Common Name | Radioimmunoassay,free thyroxine | Radioimmunoassay, thyroid stimulatinghormone |
| Classification Name | Free thyroxine test system | Thyroid stimulating hormone testsystem |
| FDA Classification | Class II | Class II |
| Review Panel | Clinical Chemistry | Clinical Chemistry |
| Product Code | CEC | JLW |
| Regulation Number | 862.1695 | 862.1690 |
{5}------------------------------------------------
Predicate Device Information 4.
The inclusion of pediatric reference intervals to the labeling (Package Inserts) of the Dimension® FT4L and TSHL assays does not require any other device modifications (i.e. no change to design or manufacturing process). No changes were made to the reagents. Therefore, as shown in the table below, the predicate and subject devices are the same.
| FT4L | TSHL | |
|---|---|---|
| Predicate Device | Dimension® LOCI Free ThyroxineFlex® reagent cartridge, FT4L | Dimension® LOCI Thyroid StimulatingHormone Flex® reagent cartridge,TSHL |
| Subject Device | Same (As above) | Same (As above) |
| Predicate 510(k) | K130276 | K081074 |
| Table 2. Summary of Predicate Devices for Dimension® FT4L and TSHL Assays | ||
|---|---|---|
| -- | -- | --------------------------------------------------------------------------- |
Substantial Equivalence Information 5.
The following table demonstrates substantial equivalence between the predicate Dimension® thyroid assays (with unmodified labeling) and Dimension® thyroid assays which have modified Instructions for Use (Package Inserts) including pediatric reference intervals.
| Item | Predicate Device(Unmodified Labeling) | Subject Device(With Pediatric Reference Intervals) |
|---|---|---|
| Analytes | Dimension® FT4L | Dimension® FT4L |
| Free Thyroxine (FT4) | Same | |
| Dimension® TSHL | Dimension® TSHL | |
| Thyroid Stimulating Hormone | Same | |
| Reagents | Dimension® FT4L | Dimension® FT4L |
| Dimension® LOCI Free ThyroxineFlex® reagent cartridge, FT4L | Same | |
| Dimension® TSHL | Dimension® TSHL | |
| Dimension® LOCI Thyroid StimulatingHormone Flex® reagent cartridge,TSHL | Same | |
| Instruments | Dimension® EXL™ with LOCI®Module and Dimension® EXL™ 200 | Same |
| Intended UseStatements | Dimension® FT4LThe FT4L method is an in vitrodiagnostic test for the quantitativemeasurement of Free Thyroxine inhuman serum and plasma on theDimension® EXL™ integratedchemistry system with LOCI® Module.Measurements of free thyroxine areused in the diagnosis and monitoring of | Dimension ® FT4LSame |
| Table 3. Summary of Substantial Equivalence for Dimension® FT4L and TSHL Assays | |||
|---|---|---|---|
| Item | Predicate Device(Unmodified Labeling) | Subject Device(With Pediatric Reference Intervals) | |
| thyroid disease. | |||
| Dimension® TSHL | Dimension® TSHL | ||
| The TSHL method is an in vitrodiagnostic test for the quantitativemeasurement of Thyroid StimulatingHormone (TSH, thyrotropin) in humanserum and plasma on the Dimension®EXL™ integrated chemistry systemwith LOCI® Module. Measurements ofTSH are used in the diagnosis andmonitoring of thyroid disease. | same | ||
| AnalyticalMeasuring Range(Assay Range) | Dimension® FT4L0.1 – 8.0 ng/dL [1.3 – 103 pmol/L] | Dimension® FT4LSame | |
| Dimension® TSHL | Dimension® TSHL | ||
| 0.007 – 100 μIU/mL [mIU/L] | Same | ||
| AdultReference Intervals | Dimension® FT4L0.76 – 1.46 ng/dL [9.8 – 18.8 pmol/L]The reference interval was transferredfrom that previously determined for theFT4 method on the Dimension Vista®System. It represents the central 95%of results determined non-parametrically from a population of 199healthy adults (140 males and 59females, 18 - 59 years of age). Theoriginal determination and transferenceto the FT4L method was done inaccordance with CLSI/NCCLS C28-A2. | Same | |
| Dimension® TSHL0.358 - 3.74 µIU/mL [mIU/L]The reference interval was transferredfrom that previously determined for theTSH method on the Dimension Vista®System. It represents the central 95%of results determined non-parametrically from a population of 297apparently healthy adults (187 malesand 110 females, 18 - 65 years ofage). The original determination andtransference to the TSHL method wasdone in accordance with CLSI/NCCLSC28-A2. | Same | ||
| Item | Predicate Device(Unmodified Labeling) | Subject Device(With Pediatric Reference Intervals) | |
| PediatricReference Intervals | Dimension® FT4LNone | Dimension® FT4LInfants (01 – 23 months)0.93 - 1.45 ng/dLChildren (02 – 12 years)0.82 - 1.40 ng/dLAdolescents (13 – 20 years)0.78 - 1.34 ng/dL | |
| Dimension® TSHL | Dimension® TSHLInfants (01 – 23 months)0.867 – 6.43 µIU/mLChildren (02 – 12 years)0.704 - 4.01 µIU/mLAdolescents (13 – 20 years)0.516 - 4.13 µIU/mL |
{6}------------------------------------------------
Table 3. Summary of Substantial Equivalence for Dimension® FT4L and TSHL Assays
{7}------------------------------------------------
Table 3. Summary of Substantial Equivalence for Dimension® FT4L and TSHL Assays
Standard/Guidance Document Reference 6.
Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline -- Third Edition (CLSI C28-A3c).
7. Pediatric Reference Intervals
Data from a total of 411 patients (77 infants, 187 children, 147 adolescents) were analyzed to establish the Dimension® FT4L assay reference intervals for the studied pediatric population. Data from a total of 407 patients (75 infants, 185 children, 147 adolescents) were analyzed to establish the Dimension® TSHL assay reference intervals for the studied pediatric population. Testing was performed on a Dimension® EXL™ with LM system. These pediatric reference intervals, as well as the previously-established euthyroid adult reference intervals and analytical measuring ranges, are presented below.
| Dimension® FT4L | Dimension® TSHL | |
|---|---|---|
| Infants (01 - 23M) | 0.93 - 1.45 ng/dL | 0.867 – 6.43 µIU/mL |
| Children (02 - 12Y) | 0.82 - 1.40 ng/dL | 0.704 - 4.01 µIU/mL |
| Adolescents (13 - 20Y) | 0.78 - 1.34 ng/dL | 0.516 - 4.13 µIU/mL |
| Euthyroid Adults* | 0.76 - 1.46 ng/dL | 0.358 - 3.74 µIU/mL |
| Assay Range* | 0.1 - 8.0 ng/dL | 0.007 - 100 µIU/mL |
| Table 4. Comparison of Pediatric and Adult 95% Reference Intervals | |||
|---|---|---|---|
- Information taken from existing Instructions for Use (Package Inserts)
A non-parametric approach was used to establish the reference intervals for children and adolescents where the 2.5" and 97.5" percentiles of the distribution of values were calculated. Because fewer than 120 patient samples were included for the infant age group, the reference intervals were calculated by using a robust measure of location and spread,
{8}------------------------------------------------
as developed by Horn and Pesce and recommended in CLSI EP28-A3c: Defining. Establishing, and Verifying Reference Intervals.
The robust symmetrical method was used for the Dimension FT4L analysis because the data was normally distributed. The reference interval calculated from the actual data using the robust symmetrical method was confirmed with two simulated reference intervals (nonparametric and robust symmetric). Each simulated reference interval was estimated as the means of the 95% upper and lower bounds from 1000 data sets. In each case, a normal distribution was fitted to the actual results, with either n=120 for the non-parametric analysis, or n=77 for the robust symmetric analysis.
For the Dimension TSHL assay, the sample distribution for infant subjects was highly positively skewed. As permitted by CLSI EP28-A3, the actual results were natural log transformed to produce a well-fitted normal distribution. The infant reference intervals were then calculated using the robust symmetric method on the log transformed data. The reference interval was confirmed with two simulated reference intervals (nonparametric and robust symmetric). Each simulated reference interval was estimated as the means of the 95% upper and lower bounds from 1000 data sets. In each case, a normal distribution was fitted to the actual results, with either n=120 for the non-parametric analysis, or n=75 for the robust symmetric analysis.
The 90% confidence intervals for the Infant 95% Reference Intervals are listed below.
| Infants (01 - 23M) | Dimension® FT4L | Dimension® TSHL |
|---|---|---|
| Calculation Method | Robust Symmetric | Robust Symmetric After LogTransformed |
| Reference Interval | 0.93 - 1.45 ng/dL | 0.867 - 6.43 µIU/mL |
| 90% Confidence Intervalof Lower Limit | 0.89 - 0.97 ng/dL | 0.710 - 1.04 µIU/mL |
| 90% Confidence Intervalof Upper Limit | 1.41 - 1.50 ng/dL | 5.40 - 7.54 µIU/mL |
Table 5. Infant Reference Interval - 90% Confidence Intervals for Upper and Lower Limits
Performance Characteristics 8.
The inclusion of pediatric reference intervals in the Instructions for Use (Package Inserts) does not necessitate the collection of additional analytical performance data as no changes were made to the reagents. Therefore, all analytical performance data previously reviewed for the Dimension® FT4L and TSHL assays continue to apply to these assays. All performance data is cross-referenced to the predicate 510(k) submissions for the FT4L and TSHL assays (K130276 and K081074).
Specifically, the following studies are not needed for the purpose of this submission:
- Precision/Reproducibility ■
- . Linearity
- l Calibrator/Assay Traceability
- . Calibrator/Assay Stability
- l Assay Cut-off
- " Method Comparison
- l Matrix Comparison
{9}------------------------------------------------
- י Sensitivity (Detection Limits, LoB, LoD, LoQ)
- l Analytical Specificity
Shelf-Life 9.
The inclusion of pediatric reference intervals in the Instructions for Use (Package Inserts) does not necessitate the collection of additional stability data as no changes were made to the reagents. Therefore, all stability methods, acceptance criteria and data previously reviewed for the Dimension® FT4L and TSHL assays continues to apply to these assays.
Specifically, the following stability studies are not needed for the purpose of this submission:
- י Shelf Life Stability
- l Onboard Stability
- l Open Vial Stability
10. Conclusions
The Dimension® FT4L and TSHL assays with pediatric reference intervals are substantially equivalent to the currently marketed Dimension® FT4L and TSHL assays.
No changes were made to the reagents. The inclusion of pediatric reference intervals in the Instructions for Use (Package Inserts) does not require a change in the device design or a change in the manufacturing process.
The addition of pediatric reference intervals for the Dimension® FT4L and TSHL assays is further supported by the following rationale:
-
- Testing of pediatric patients is within the established indications for use in the diagnosis and treatment of thyroid disease), as described in 21 CFR §862.1695 and 21 CFR §862.1690.
-
- The newly-established pediatric reference intervals are either within or are above the previously-established reference intervals for euthyroid (normal thyroid) adult populations and they are within the analytical measuring ranges of the Dimension® FT4L and TSHL assays. Therefore, the Dimension® FT4L and TSHL assays have appropriate analytical performance to test pediatric patients.
§ 862.1695 Free thyroxine test system.
(a)
Identification. A free thyroxine test system is a device intended to measure free (not protein bound) thyroxine (thyroid hormone) in serum or plasma. Levels of free thyroxine in plasma are thought to reflect the amount of thyroxine hormone available to the cells and may therefore determine the clinical metabolic status of thyroxine. Measurements obtained by this device are used in the diagnosis and treatment of thyroid diseases.(b)
Classification. Class II.