K Number
K140841
Device Name
TABLO CARTRIDGE
Manufacturer
Date Cleared
2014-09-11

(162 days)

Product Code
Regulation Number
876.5820
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Tablo™ Cartridge is a single use, disposable arterial and venous bloodline set intended to provide extracorporeal access during hemodialysis. The Tablo™ Cartridge is compatible only with the Tablo™ Hemodialysis System.

Device Description

The Tablo™ Cartridge comprises a single-use, sterile Blood Tubing Set attached to an Organizer. The Organizer ensures proper placement of the Blood Tubing Set with the Tablo Hemodialysis System. The Blood Tubing Set consists of an arterial line, a venous line, an adaptor to connect the two lines, a saline line, a pressure transducer protector, a venous drip chamber, an arterial pressure pod, and a heparin syringe line. The Tablo Cartridge is developed exclusively for use with the Tablo Hemodialysis System. It is single-use, offered in one configuration, and provided sterile.

AI/ML Overview

This document is a 510(k) summary for the Tablo™ Cartridge, a blood tubing set used for hemodialysis. It does not present a study proving the device meets acceptance criteria in the way one might expect for a clinically-oriented AI/software device. Instead, it describes performance testing to demonstrate substantial equivalence to a predicate device for a hardware medical device.

Therefore, many of the requested fields (like sample size for test sets, number of experts, adjudication methods, MRMC studies, standalone performance for AI, and training set details) are not applicable to this type of submission.

Here's a breakdown based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria with corresponding performance metrics like sensitivity, specificity, or AUC, as would be common for AI/software devices. Instead, it lists types of performance testing and a general conclusion of compliance.

Acceptance Criteria CategoryReported Device Performance
Functional TestingMet all performance specifications. The minor differences in technological characteristics (organizer, inner diameter of blood pump segment, access point configuration, pressure measuring component) did not raise new or different questions of safety or effectiveness.
Biocompatibility TestingDemonstrated to be safe and did not raise new or different questions of safety or effectiveness.
Packaging and Shelf Life TestingDemonstrated to be safe and did not raise new or different questions of safety or effectiveness.
Sterilization ValidationDemonstrated to be safe and did not raise new or different questions of safety or effectiveness.
Pyrogenicity Testing (for blood fluid path)Demonstrated to be safe and did not raise new or different questions of safety or effectiveness.
Applicable Standards and FDA Guidance DocumentsComplies with applicable standards and FDA Guidance Documents.
Substantial EquivalenceDetermined to be substantially equivalent to the Fresenius Blood Tubing Set (K120823).

2. Sample size used for the test set and the data provenance

This information is Not Applicable to this document as it describes testing of a physical medical device (blood tubing set) rather than an AI/software device with a distinct "test set" of data. The testing mentioned would involve laboratory and engineering tests on physical devices.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is Not Applicable. Ground truth, in the context of expert review for medical images or data, is not relevant for the performance testing described for a blood tubing set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is Not Applicable. Adjudication methods are typically used for expert consensus on ambiguous cases in the context of diagnostic AI/software.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is Not Applicable. MRMC studies are specific to evaluating the impact of AI on human reader performance, which is not relevant for a physical device like a blood tubing set.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is Not Applicable. There is no algorithm discussed for standalone performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

This information is Not Applicable. The "ground truth" for the performance of a blood tubing set would be established by validated engineering specifications, material standards, and functional requirements rather than medical expert consensus or pathology.

8. The sample size for the training set

This information is Not Applicable. Training sets are relevant for machine learning algorithms, which are not described in this submission for a physical medical device.

9. How the ground truth for the training set was established

This information is Not Applicable. As there is no training set for an AI/ML algorithm, this question is not relevant.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name encircling a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to suggest community and care. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 10, 2014

Home Dialysis Plus Nancy Gallo Senior Vice President, Regulatory Affairs 257 Humboldt Court Sunnyvale, CA 94089

Re: K140841 Trade/Device Name: Tablo™ Cartridge Regulation Number: 21 CFR§ 876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: FJK Dated: August 8, 2014 Received: August 11, 2014

Dear Nancy Gallo,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K140841

Device Name Tablo™ Cartridge

Indications for Use (Describe)

The Tablo™ Cartridge is a single use, disposable arterial and venous bloodline set intended to provide extracorporeal access during hemodialysis. The Tablo™ Cartridge is compatible only with the Tablo™ Hemodialysis System.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

5. 510(k) Summary

This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR Part 807.92.

5.1 Submitter's Information

Submitter's Name:Moshe Pinto, CEO
Company:HDPlus
Address:257 Humboldt Ct, Sunnyvale, CA 94089
Contact Person:Nancy Gallo, Senior Vice-President, Regulatory Affairs
Phone:510.682.6505
Facsimile:408.329.9091
Email:ngallo@homedialysisplus.com
Date of Summary Preparation:March 31, 2014

5.2 Device Information

Trade Name:Tablo™ Cartridge
Common Name:Blood Tubing Set
Classification Name:Hemodialysis system and accessories
Classification Number:Class II per 21 CFR 876.5820
Product Code:FJK (Set, Tubing, Blood, With And Without Anti-Regurgitation Valve)
Classification Panel:Gastroenterology/Urology

Predicate Device Information 5.3

Fresenius Blood Tubing Set (K120823)

5.4 Device Description

The Tablo™ Cartridge comprises a single-use, sterile Blood Tubing Set attached to an Organizer. The Organizer ensures proper placement of the Blood Tubing Set with the Tablo Hemodialysis System. The Blood Tubing Set consists of an arterial line, a venous line, an adaptor to connect the two lines, a saline line, a pressure transducer protector, a venous drip chamber, an arterial pressure pod, and a heparin syringe line. The Tablo Cartridge is developed exclusively for use with the Tablo Hemodialysis System. It is single-use, offered in one configuration, and provided sterile.

{5}------------------------------------------------

5.5 Indications for Use

The Tablo Cartridge is a single use, disposable arterial and venous bloodline set intended to provide extracorporeal access during hemodialysis. The Tablo Cartridge is compatible only with the Tablo Hemodialysis System.

Technological Characteristics 5.6

The Tablo Cartridge (Subject Device) utilizes Fresenius Blood Tubing Set (K120823) as its Predicate.

5.6.1 Tablo Cartridge vs. Fresenius Blood Tubing Set (K120823):

The Subject and Predicate Devices have equivalent technological characteristics:

  • Intended Use – To provide extracorporeal access during hemodialysis.
  • Materials – Primary fluid path materials are Polyvinyl Chloride (PVC) and Polypropylene (PP).
  • Design & Construction - Polyvinyl Chloride (PVC) tubing of various lengths and diameters, with color-coded pinch clamps, color-coded injection ports, heparin line, saline line, and pressure monitoring components.
  • Sterility - Sterile, single use, non-pyrogenic.
  • Priming Volume Priming Volume of ≤ 300 ml ●

Minor differences exist in the technological characteristics of the Subject and Predicate Devices. None of the minor differences raise any new or different questions of safety or effectiveness. The differences include:

  • Subject Device is mounted on an organizer to facilitate interface with the machine's front panel. The Predicate does not have an organizer and the tubing is uncoiled and attached to the machine.
  • . The Inner Diameter (ID) of the Blood Pump Segment of Subject Device is smaller than that of the Predicate Device.
  • The default configuration of the access point used for both systems is a double needle configuration, though the Predicate is also equipped with the option of a single needle configuration.
  • The pressure measuring component of the Tablo Cartridge is different from that of the Predicate.

{6}------------------------------------------------

5.7 Performance Data

The following Performance Testing, developed in accordance with appropriate guidance documents and relevant standards, has been performed on the Subject Device to support the determination of substantial equivalence:

  • . Functional testing.
  • . Biocompatibility testing.
  • Packaging and shelf life testing.
  • Sterilization validation.
  • Pyrogenicity testing for the blood fluid path.

The results of all these tests demonstrated the Tablo Cartridge is safe and does not raise new or different questions of safety or effectiveness as compared to the Predicate.

5.8 Conclusion

The Performance Testing demonstrated the Tablo Cartridge meets all performance specifications and complies with applicable standards and FDA Guidance Documents. The Tablo Cartridge is substantially equivalent to the Predicate, and the minor differences in the technological characteristics of the Subject and the Predicate Devices do not raise any new or different questions of safety or effectiveness.

§ 876.5820 Hemodialysis system and accessories.

(a)
Identification. A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient.
(2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860).
(3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860).
(4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
(b)
Classification. (1) Class II (performance standards) for hemodialysis systems and all accessories directly associated with the extracorporeal blood system and the dialysate delivery system.(2) Class I for other accessories of the hemodialysis system remote from the extracorporeal blood system and the dialysate delivery system, such as the unpowered dialysis chair, hemodialysis start/stop tray, dialyzer holder set, and dialysis tie gun and ties. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.