(242 days)
The CLINITEK Novus® Automated Urine Chemistry Analyzer is a fully automated urinalysis instrument. The CLINITEK Novus analyzer is intended to read Siemens Healthcare Diagnostics CLINITEK Novus Cassettes, as well as determine urine specific gravity and urine clarity. The CLINITEK Novus 10 Urinalysis Cassette is intended for the seniquantitative measurement of the following parameters in urine: bilirubin, blood (occult), glucose, ketone (acetoacetic acid), leukocytes, nitrite (qualitative), pH, protein, color, and urobilinogen.
These measurements are used to assist diagnosis in the following areas:
- · Carbohydrate metabolism (such as diabetes mellitus)
- · Kidney function
- · Liver function
- · Metabolic disorders
- · Urinary tract infection
For in vitro diagnostic use.
The CLINITEK Novus® Calibration Kit is intended to be used with the CLINITEK Novus Urinalysis Cassette to calibrate the CLINITEK Novus Automated Urine Chemistry analyzer. This product is for professional in vitro diagnostic use.
The CLINITEK Novus® system is a fully automated urine chemistry analyzer that is designed for use with the CLINITEK Novus® 10 Urinalysis Cassette. The analyzer automates the process of urine strip testing by dispensing urine samples onto a test pad, and then by reading the color and intensity of light reflected from the reacted test pads, converts the results into clinically meaningful units.
Test results from the test pads are qualitative and semi-quantitative; specific gravity is measured by a refractometer assembly.
Here's an analysis of the acceptance criteria and study detailed in the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not explicitly state pre-defined acceptance criteria (e.g., "The device must achieve X% agreement"). However, it does present the results of comparison studies against a predicate device and states that "The results for overall percent agreement and within one level agreement met acceptance criteria." This implies that acceptable thresholds were achieved. Based on the presented data, we can infer the achieved performance. For specific analytes, the values vary, but they consistently show high exact match and very high agreement within +/- 1 block.
| Analyte | Performance Metric | CLINITEK Novus Result (All Sites Combined) | Implied/Achieved Acceptance Criteria (based on "met acceptance criteria") |
|---|---|---|---|
| Bilirubin | % exact match (range) | 64.4% - 99.8% | High % (exact match) |
| % within 1 block (range) | 98.2% - 100.0% | Very high % (within 1 block) | |
| Blood | % exact match (range) | 81.5% - 97.8% | High % (exact match) |
| % within 1 block (range) | 100.0% (except one at 99.7%) | Very high % (within 1 block) | |
| Glucose | % exact match (range) | 88.5% - 99.6% | High % (exact match) |
| % within 1 block (range) | 100.0% | Very high % (within 1 block) | |
| Ketone | % exact match (range) | 88.2% - 100.0% | High % (exact match) |
| % within 1 block (range) | 99.2% - 100.0% | Very high % (within 1 block) | |
| Leukocyte | % exact match (range) | 81.1% - 96.6% | High % (exact match) |
| % within 1 block (range) | 99.9% - 100.0% | Very high % (within 1 block) | |
| Nitrite | % exact match (range) | 99.3% - 100.0% | High % (exact match) |
| % within 1 block (range) | 100.0% | Very high % (within 1 block) | |
| pH | % exact match (range) | 45.4% - 99.8% | High % (exact match) |
| % within 1 block (range) | 97.5% - 100.0% | Very high % (within 1 block) | |
| Protein | % exact match (range) | 89.9% - 98.7% | High % (exact match) |
| % within 1 block (range) | 99.9% - 100.0% | Very high % (within 1 block) | |
| Urobilinogen | % exact match (range) (sites 1-3) | 85.6% - 99.9% | High % (exact match) |
| % within 1 block (range) (sites 1-3) | 100.0% | Very high % (within 1 block) | |
| Urobilinogen | % exact match (range) (site 4) | 93.8% - 100.0% | High % (exact match) |
| % within 1 block (range) (site 4) | 100.0% | Very high % (within 1 block) | |
| Specific Gravity | R2 | 0.983 | High R2 (correlation) |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size:
- For the method comparison study (comparing CLINITEK Novus to the predicate CLINITEK Atlas), up to 2773 specimens were tested across 3 sites for most analytes.
- For Urobilinogen, an additional fourth site was involved due to a high number of contrived samples in the initial three sites. This fourth site tested 261 samples (187 negative, and then 48, 10, 12, 4 for various positive levels).
- Data Provenance: The document does not explicitly state the country of origin. The study was conducted "across 3 sites" (and later a fourth), implying a multi-center study within the scope of the device manufacturer's operations/clinical trials. It is a prospective comparison study as it compared the new device (CLINITEK Novus) against an existing predicate device (CLINITEK Atlas) using actual patient specimens.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The context of this device being an automated urine chemistry analyzer comparing its readings to a predicate automated urine chemistry analyzer suggests that the "ground truth" for the test set was established by the results from the predicate device (CLINITEK Atlas), not human experts. Therefore, the concept of a number of experts and their qualifications for establishing ground truth is not directly applicable in this context. The predicate device's established performance serves as the reference.
4. Adjudication Method for the Test Set
Since the comparison is primarily between two automated devices, and the "ground truth" is derived from the predicate device's readings, a typical adjudication method involving human readers (like 2+1 or 3+1 consensus) is not applicable. The data presented is a direct comparison of the readings obtained from the CLINITEK Novus analyzer against the readings obtained from the CLINITEK Atlas analyzer.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance?
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study involving human readers and AI assistance was not performed. This device is an automated in vitro diagnostic (IVD) analyzer, not an AI-powered image analysis tool meant to assist human readers. Its primary function is to automatically measure chemical parameters in urine.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done?
Yes, the performance reported is a standalone performance of the CLINITEK Novus Automated Urine Chemistry Analyzer. It assesses the device's ability to accurately measure urine parameters independently, by comparing its results directly to those of a predicate automated device. There is no human-in-the-loop component in the reported performance evaluation.
7. The Type of Ground Truth Used
The primary "ground truth" used for the method comparison study was the results obtained from the predicate device, the CLINITEK Atlas Automated Urine Chemistry Analyzer. For the detection limit study, "ground truth" was established by spiking negative pooled human urine with known quantities of analytes.
8. The Sample Size for the Training Set
The document describes studies comparing the new device to a predicate device and analytical performance characteristics. It does not explicitly mention a "training set" or a machine learning/AI model training process for this IVD device. The description focuses on demonstrating substantial equivalence to a predicate device through analytical and method comparison studies. If the device uses algorithms derived from prior data, that information is not detailed in this summary.
9. How the Ground Truth for the Training Set Was Established
Since a "training set" is not explicitly mentioned in the context of machine learning/AI for this device, the method for establishing its ground truth is not applicable/not provided in this document. The device's operation appears to be based on established reflectance photometry and refractive index principles for chemical analysis, rather than a learned model from a training set.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 17, 2014
SIEMENS HEALTHCARE DIAGNOSTICS, INC. SUSAN TIBEDO SR. MANAGER REGULATORY AFFAIRS 2 EDGEWATER DRIVE NORWOOD MA 02062
Re: K140717
Trade/Device Name: CLINITEK Novus Automated Urine Chemistry Analyzer CLINITEK Novus 10 Urinalysis Cassette, CLINITEK Novus Calibration Kit Regulation Number: 21 CFR 864.6550 Regulation Name: Occult Blood Test Regulatory Class: II Product Code: JIO, KQO, JJB, JIL, JIN, LJX, JMT, CEN, JIR, JRE, CDM, JIX Dated: September 29, 2014 Received: October 1, 2014
Dear Ms. Susan Tibedo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the
{1}------------------------------------------------
electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Courtney
Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K140717
Device Name CLINITEK Novus® Automated Urine Chemistry Analyzer CLINITEK Novus® 10 Urinalysis Cassette CLINITEK Novus® Calibration Kit
Indications for Use (Describe)
The CLINITEK Novus® Automated Urine Chemistry Analyzer is a fully automated urinalysis instrument. The CLINITEK Novus analyzer is intended to read Siemens Healthcare Diagnostics CLINITEK Novus Cassettes, as well as determine urine specific gravity and urine clarity. The CLINITEK Novus 10 Urinalysis Cassette is intended for the seniquantitative measurement of the following parameters in urine: bilirubin, blood (occult), glucose, ketone (acetoacetic acid), leukocytes, nitrite (qualitative), pH, protein, color, and urobilinogen.
These measurements are used to assist diagnosis in the following areas:
- · Carbohydrate metabolism (such as diabetes mellitus)
- · Kidney function
- · Liver function
- · Metabolic disorders
- · Urinary tract infection
For in vitro diagnostic use.
The CLINITEK Novus® Calibration Kit is intended to be used with the CLINITEK Novus Urinalysis Cassette to calibrate the CLINITEK Novus Automated Urine Chemistry analyzer. This product is for professional in vitro diagnostic use.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary (K140717)
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
| 1.0 | Submitter Information | |
|---|---|---|
| Owner | Siemens Healthcare Diagnostics, Inc.2 Edgewater DriveNorwood, MA 02062 | |
| Contact | Primary: Susan TibedoSenior Manager Regulatory Affairssusan.tibedo@siemens.com781-269-3888 TEL | |
| Secondary: Noor MalkiVP, Clinical, Regulatory and Quality - POC781-269-3401 TEL | ||
| Date Summary Prepared | November 12, 2014 | |
| 2.0 | Device Information | |
| Proprietary Names | CLINITEK Novus® Automated Urine ChemistryAnalyzerCLINITEK Novus® 10 Urinalysis CassetteCLINITEK Novus® Calibration Kit | |
| Common Name | Automated Urine Chemistry Analyzer | |
| Main Classification Name | Automated urinalysis system21 CFR 862.2900, Class IProduct Code KQO |
Subsequent Classifications
| Classification | ||||
|---|---|---|---|---|
| Device | RegulationSection | DeviceClass | ProductCode | Test |
| Automatedurinalysissystem | 21 CFR 862.2900 | 1 | KQO | Automated Urine ChemistryAnalyzer |
| Bilirubin | 21 CFR 862.1115 | 1 | JJB | Urinary bilirubin and its conjugates(non-quantitative) test system |
| Blood (occult) | 21 CFR 864.6550 | 2 | JIO | Occult blood test |
| Glucose | 21 CFR 862.1340 | 2 | JIL | Urinary glucose (non-quantitative)test system |
| Ketone(acetoaceticacid) | 21 CFR 862.1435 | 1 | JIN | Ketones (non-quantitative) testsystem |
| Leukocytes | 21 CFR 864.7675 | 1 | LJX | Leukocyte peroxidase test |
Page 1 of 19
{4}------------------------------------------------
| Classification | ||||
|---|---|---|---|---|
| Device | RegulationSection | DeviceClass | ProductCode | Test |
| Nitrite | 21 CFR 862.1510 | 1 | JMT | Nitrite (non-quantitative) testsystem |
| pH | 21 CFR 862.1550 | 1 | CEN | Urinary pH (nonquantitative) testsystem |
| Protein | 21 CFR 862.1645 | 1 | JIR | Urinary protein or albumin (non-quantitative) test system |
| Specificgravity | 21 CFR 862.2800 | 1 | JRE | Refractometer for clinical use |
| Urobilinogen | 21 CFR 862.1785 | 1 | CDM | Urinary urobilinogen (non-quantitative) test system |
| Calibrator | 21 CFR 862.1150 | 2 | JIX | Calibrator, multi-analyte mixture |
3.0 Predicate Device
| Predicate | |
|---|---|
| Device Name | CLINITEK Atlas |
| Common Name | Automated Urine Chemistry Analyzer |
| 510(k) Number | K932674 |
| Manufacturer | Siemens Healthcare Diagnostics |
4.0 Device Description
The CLINITEK Novus® system is a fully automated urine chemistry analyzer that is designed for use with the CLINITEK Novus® 10 Urinalysis Cassette. The analyzer automates the process of urine strip testing by dispensing urine samples onto a test pad, and then by reading the color and intensity of light reflected from the reacted test pads, converts the results into clinically meaningful units.
Test results from the test pads are qualitative and semi-quantitative; specific gravity is measured by a refractometer assembly.
| Analyte | Abbreviation | ConventionalSemi-Quantitative | SISemi-Quantitative | Qualitative |
|---|---|---|---|---|
| Bilirubin | BIL | Negative | Negative | Negative |
| Small | Small | 1+ | ||
| Moderate | Moderate | 2+ | ||
| Large | Large | 3+ | ||
| Analyte | Abbreviation | ConventionalSemi-Quantitative | SISemi-Quantitative | Qualitative |
| Blood | BLO | Negative | Negative | Negative |
| Trace | Trace | Trace | ||
| NHT | NHT | NHT | ||
| Small | Small | 1+ | ||
| Moderate | Moderate | 2+ | ||
| Large | Large | 3+ | ||
| Glucose | GLU | Negative | Negative | Negative |
| 100 mg/dL | 5.5 mmol/L | Trace | ||
| 250 mg/dL | 14 mmol/L | 1+ | ||
| 500 mg/dL | 28 mmol/L | 2+ | ||
| 1000 mg/dL | >=55 mmol/L | 3+ | ||
| Ketone | KET | Negative | Negative | Negative |
| Trace | Trace | Trace | ||
| 15 mg/dL | 1.5 mmol/L | 1+ | ||
| 40 mg/dL | 3.9 mmol/L | 2+ | ||
| 80 mg/dL | 7.8 mmol/L | 3+ | ||
| >=160 mg/dL | >=15.6 mmol/L | 4+ | ||
| Leukocyte | LEU | Negative | Ca 15 cells/µL | Negative |
| Trace | Ca 15 cells/µL | Trace | ||
| Small | Ca 70 cells/µL | 1+ | ||
| Moderate | Ca 125 cells/µL | 2+ | ||
| Large | Ca 500 cells/µL | 3+ | ||
| Nitrite | NIT | Negative | Negative | Negative |
| Positive | Positive | Positive | ||
| pH | pH | 5.0 | 5.0 | 5.0 |
| 5.5 | 5.5 | 5.5 | ||
| 6.0 | 6.0 | 6.0 | ||
| 6.5 | 6.5 | 6.5 | ||
| 7.0 | 7.0 | 7.0 | ||
| 7.5 | 7.5 | 7.5 | ||
| 8.0 | 8.0 | 8.0 | ||
| 8.5 | 8.5 | 8.5 | ||
| >=9.0 | >=9.0 | >=9.0 | ||
| Analyte | Abbreviation | ConventionalSemi-Quantitative | SISemi-Quantitative | Qualitative |
| Protein | PRO | Negative | Negative | Negative |
| Trace | Trace | Trace | ||
| 30 mg/dL | 0.3 g/L | 1+ | ||
| 100 mg/dL | 1.0 g/L | 2+ | ||
| 300 mg/dL | 3.0 g/L | 3+ | ||
| >=1000 mg/dL | >=10.0 g/L | 4+ | ||
| SpecificGravity(0.005 units) | SG | <=1.005 | <=1.005 | <=1.005 |
| 1.010 | 1.010 | 1.010 | ||
| 1.015 | 1.015 | 1.015 | ||
| 1.020 | 1.020 | 1.020 | ||
| 1.025 | 1.025 | 1.025 | ||
| >=1.030 | >=1.030 | >=1.030 | ||
| SpecificGravity(0.001 units) | SG | 1.000 to <=1.099 in increments of 0.001 | ||
| Urobilinogen | UROUBG (SI) | 0.2 E.U./dL(mg/dL) | 3.2 µmol/L | 0.2 E.U./dL |
| 1 E.U./dL | 16 µmol/L | 1 E.U./dL | ||
| 2 E.U./dL | 33 µmol/L | 2 E.U./dL | ||
| 4 E.U./dL | 66 µmol/L | 4 E.U./dL | ||
| >=8 E.U./dL | >=131µmol/L | >=8 E.U./dL | ||
| Color | COL | Yellow | Yellow | Yellow |
| Dark Yellow | Dark Yellow | Dark Yellow | ||
| Orange | Orange | Orange | ||
| Red | Red | Red | ||
| Green | Green | Green | ||
| Other | Other | Other | ||
| Clarity | CLA | Clear | Clear | Clear |
| Cloudy | Cloudy | Cloudy | ||
| Turbid | Turbid | Turbid | ||
| Sl. Cloudy | Sl. Cloudy | Sl. Cloudy |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
5.0 Intended Use Statement
The CLINITEK Novus® Automated Urine Chemistry Analyzer is a fully automated urinalysis instrument for clinical laboratory use. The CLINITEK Novus analyzer is intended to read Siemens Healthcare Diagnostics CLINITEK Novus Cassettes, as well as determine urine specific gravity and urine clarity.
The CLINITEK Novus 10 Urinalysis Cassette is intended for the semi-quantitative measurement of the following parameters in urine: bilirubin, blood (occult), glucose, ketone (acetoacetic acid), leukocytes, nitrite (qualitative), pH, protein, color, and urobilinogen.
These measurements are used to assist diagnosis in the following areas:
- · Carbohydrate metabolism (such as diabetes mellitus)
- Kidney function
- Liver function
- Metabolic disorders
- Urinary tract infection
For in vitro diagnostic use.
The CLINITEK Novus® Calibration Kit is intended to be used with the CLINITEK Novus Urinalysis Cassette to calibrate the CLINITEK Novus Automated Urine Chemistry Analyzer. This product is for professional in vitro diagnostic use.
| 6.0 | Summary Comparison of Technological Characteristics | |||
|---|---|---|---|---|
| ----- | ----------------------------------------------------- | -- | -- | -- |
| Feature | Candidate DeviceCLINITEK Novus analyzer | Predicate DeviceCLINITEK Atlas analyzer |
|---|---|---|
| Intended Use | Automated Urine Chemistry Analyzerintended for professional in vitrodiagnostic use in clinical laboratories.The analyzer is intended for themeasurement of the followingcomponents in urine: bilirubin, blood(occult), glucose, ketone (acetoaceticacid), leukocytes, nitrite, pH, protein,urobilinogen and specific gravity. | Same |
| UrinalysisStripsAnalyticalMethod | Light is reflected at specificwavelengths from the test pad "readarea". Reported results depend uponthe degree of color change andintensity in the pad which is directlyrelated to the concentration of theanalyte in the urine. | Same |
| Feature | Candidate DeviceCLINITEK Novus analyzer | Predicate DeviceCLINITEK Atlas analyzer |
| SpecificGravity | Fiber optic refractive index method. | Same |
| InstrumentOpticalSystem | Two pairs of LEDs illuminate the testpads and a camera in the analyzerelectronically analyzes the color andintensity of the light reflected from thereacted test pads. | Reflectance photometerconsisting of halogen lamp,fiber optics and lens anddetector.Light travels onto a lenswhere it is focused onto thedetector, and the intensity isconverted into electricalimpulses to generateclinically meaningful units. |
| ReagentFormat | Cassette contains reagent test padsorganized in strips on a test card format. | Roll containing reagent testpads in strip format. |
| ReagentTest PadChemistry | Each reagent test pad consists ofimpregnated paper with chemicalsspecific to measuring the analyte. | Same |
| ReagentStorage andStability | Storage temperature: 15-30°CShelf life: 12 monthsOpen shelf life: 14 days | Same |
| CalibrationMethod | The system performs a cleaning cycleand dry pad calibration. Four, liquid,ready-to-use calibrators are used tocalibrate the SG sensor and certain testpads. | Same without dry padcalibration |
| QualityCheck | Conducts a quality check to assess if thetest pad has been exposed to highhumidity. | None |
| Data Storage | 7,500 Patient results200 Calibration results400 Quality control results | 1000 Patient results200 QC / calibrator results |
| Connectivity | LIS, HIS and remote-diagnosis ready.Optional module enables hospitalnetwork (such as HIS/LIS) connectivity.This also allows copying a selectedsetup across multiple instruments via thenetwork. | Same |
| Barcode | Internal barcode readerOptional handheld barcode reader -automatic entry of control lot andexpiration | Same |
| UserInterface | Integrated color touch screen | Keypad and 12-line display |
{8}------------------------------------------------
6.0 Summary Comparison of Technological Characteristics
{9}------------------------------------------------
Calibrators
Similarities and Differences
| Feature | Candidate DeviceCLINITEK Novus Calibration Kit | Predicate DeviceCLINITEK Atlas CalibrationKit (K932674) |
|---|---|---|
| Intended Use | In vitro diagnostic product for thecalibration of CLINITEK AutomatedUrine Chemistry Analyzer | Same |
| Preparation | Ready to use | Same |
| Matrix | Liquid | Same |
| Number of calibrators | Four: Calibrator #1, #2, #3 and #4 | Same |
| Cleaning cycle | Includes a cleaning cycle beforecalibration | Does not include a cleaning cycle before calibration |
| Stability | Shelf-life: 18 months stored at 2-8°COpen bottle: 18 months stored at 2-8°C | Same |
| Traceability | Calibrators have been tested usingmaterials or calibrated devices whichare traceable to the National Instituteof Standards and Technology inconformance with standard EN17511:2001. | Same |
| Value Assignment | Calibrators are prepared to achievetargeted analyte values. Thecalibrators have fixed values and arenot lot specific. Calibrators mustmeet internal specifications beforerelease. | Same |
7.0 Test Principle
The CLINITEK Novus analyzer utilizes an optical system that has a digital imager. The concentration of each analyte is measured by the color change that develops when a sample is deposited on a test pad. The imager records the color and intensity of the light reflected off the reacted tested pads, which are illuminated by two pairs of LED lights. The optical system converts this data into clinically meaningful results.
Specific gravity (SG) is determined by the fiber optic refractive index method. The SG sensor measures the amount of light passing through the fiber optic at one end. Because the refractive index is proportional to specific gravity, the light measured correlates to the specific gravity of the sample yielding a quantitative result.
{10}------------------------------------------------
8.0 Performance Characteristics
Analytical Performance
a. Precision / Reproducibility
Precision was evaluated on 3 CLINITEK Novus analyzers over 10 days of testing. Two runs per day were completed with 2 replicates per run across 1 lot of CLINITEK Novus reagent urinalysis cassettes. Commercially available controls were used providing a high positive control, a low clinically significant positive control and a negative control for evaluating the percent agreement. The results are shown in the following tables:
| Analyte | Control Limits | Exact Block Agreement (%) | Agreement within ±1 Block (%) |
|---|---|---|---|
| Bilirubin | Negative | 120/120 (100%) | 120/120 (100%) |
| Blood | Negative | 119/120 (99.2%) | 120/120 (100%) |
| Glucose | Negative | 120/120 (100%) | 120/120 (100%) |
| Ketone | Negative | 120/120 (100%) | 120/120 (100%) |
| Leukocyte | Negative | 120/120 (100%) | 120/120 (100%) |
| Nitrite | Negative | 120/120 (100%) | 120/120 (100%) |
| pH | < 6.5 | 120/120 (100%) | 120/120 (100%) |
| Protein | Negative | 120/120 (100%) | 120/120 (100%) |
| SG | < 1.005 | 120/120 (100%) | 120/120 (100%) |
| Urobilinogen | < 1.0 mg/dL | 114/120 (95%) | 120/120 (100%) |
Table 1. Reproducibility (Within Device) for Low / Negative Control
Table 2. Reproducibility (Within Device) for Low Positive Control
| Analyte | Control Limits | Exact BlockAgreement | Agreement within ±1 Block |
|---|---|---|---|
| Bilirubin | Moderate | 120/120 (100%) | 120/120 (100%) |
| Blood | Trace | 120/120 (100%) | 120/120 (100%) |
| Glucose | 100 mg/dL | 119/120 (99.2%) | 120/120 (100%) |
| Ketone | 15 mg/dL | 120/120 (100%) | 120/120 (100%) |
| Leukocyte | Small | 109/120 (90.8%) | 120/120 (100%) |
| рн | 7.0-8.0 | 120/120 (100%) | 120/120 (100%) |
| Protein | 30 mg/dL | 120/120 (100%) | 120/120 (100%) |
| ട്ട | 1.006 — 1.019 | 120/120 (100%) | 120/120 (100%) |
| Urobilinogen | 2 mg/dL | 120/120 (100%) | 120/120 (100%) |
Table 3. Reproducibility (Within Device) High Positive Control
| Analyte | Control Limits | Exact BlockAgreement | Agreement within ±1 Block |
|---|---|---|---|
| Bilirubin | Large | 120/120 (100%) | 120/120 (100%) |
| Blood | Large | 120/120 (100%) | 120/120 (100%) |
| Glucose | 1000 mg/dL | 117/120 (97.5%) | 120/120 (100%) |
| Ketone | 80 mg/dL | 108/120 (90%) | 120/120 (100%) |
| Leukocyte | Large | 120/120 (100%) | 120/120 (100%) |
| Nitrite | Positive | 120/120 (100%) | 120/120 (100%) |
Siemens Healthcare Diagnostics CLINITEK Novus® Automated Urine Chemistry Analyzer / Traditional 510(k) Page 8 of 19
{11}------------------------------------------------
| Analyte | Control Limits | Exact BlockAgreement | Agreement within ±1 Block |
|---|---|---|---|
| pH | >9.0 | 120/120 (100%) | 120/120 (100%) |
| Protein | 300 mg/dL | 120/120 (100%) | 120/120 (100%) |
| SG | >/= 1.020 | 120/120 (100%) | 120/120 (100%) |
| Urobilinogen | 4 mg/dL | 115/120 (95.8%) | 120/120 (100%) |
| Table 3. Reproducibility (Within Device) High Positive Control |
|---|
b. Detection Limit
The sensitivity of the CLINITEK Novus urinalysis cassette reagents was evaluated by spiking negative pooled human urine with known quantity of analytes. The analytes were spiked at multiple concentrations around the transition value from negative to positive. Multiple samples were tested at each concentration on the CLINITEK Novus analyzer. Sensitivity was defined as the concentration in which ≥ 55% of the results were positive. The lower limits of analyte detection for each reagent test are shown in the following table:
| Analyte | Limit of Detection(Sensitivity) |
|---|---|
| Bilirubin | 0.5 mg/dL |
| Blood | 0.013 mg/dL |
| Glucose | 36 mg/dL |
| Ketone | 3.6 mg/dL |
| Leukocyte | 6.0 cells/µL |
| Nitrite | 0.06 mg/dL |
| Protein | 10.8 mg/dL |
| Urobilinogen | 0.24 mg/dL |
c. Analytical Specificity
Urine samples were spiked with a high concentration of potentially interfering substances. Substances that resulted in a change in the reported clinical block output were serially diluted to determine the concentration at which the substance interfered. The results are in the following table:
{12}------------------------------------------------
| InterferingSubstance | Initial ConcTested | BIL | BLO | GLU | KET | LEU | NIT | PRO | URO |
|---|---|---|---|---|---|---|---|---|---|
| Acetaminophen | 40 mg/dL | None | None | None | None | None | None | None | None |
| Acetylcysteine | 270 mg/dL | FalseNegative at135 mg/dL | FalseNegative at67.5 mg/dL | FalseNegative at67.5 mg/dL | FalsePositive at67.5 mg/dL | None | None | None | FalseNegative at67.5 mg/dL |
| Ascorbic Acid | 100 mg/dL | FalseNegative at25 mg/dL | FalseNegative at25 mg/dL | FalseNegative at25 mg/dL | None | None | FalseNegative at100 mg/dL | None | None |
| AmmoniumChloride | 200 mg/dL | None | None | None | None | None | None | None | None |
| Boric Acid | 1000 mg/dL | None | None | None | FalseNegative at500 mg/dL | None | None | None | None |
| CalciumChloride | 100 mg/dL | None | None | None | None | FalseNegative at100 mg/dL | None | None | None |
| Captopril | 50 mg/dL | None | FalseNegative at12.5 mg/dL | None | FalsePositive at12.5 mg/dL | None | None | None | FalseNegative at25 mg/dL |
| Cefoxitin | 66 mg/dL | None | None | None | None | None | None | None | None |
| Cephalosporin | 80 mg/dL | None | None | None | None | None | None | None | None |
| Chlorhexidine | 150 mg/dL | None | FalseNegative at112.5 mg/dL | None | None | FalseNegative at37.5 mg/dL | None | FalsePositive at37.5 mg/dL | FalseNegative at75 mg/dL |
| Chloroquine | 40 mg/dL | None | FalseNegative at40 mg/dL | None | None | FalseNegative at20 mg/dL | None | FalsePositive at20 mg/dL | None |
| InterferingSubstance | Initial ConcTested | BIL | BLO | GLU | KET | LEU | NIT | PRO | URO |
| Citric Acid | 80 mg/dL | FalseNegative at40 mg/dL | None | None | None | None | None | None | None |
| Curcuma | 300 mg/dL | FalseNegative at75 mg/dL | FalseNegative at75 mg/dL | None | FalsePositive at75 mg/dL | FalsePositive at75 mg/dL | FalsePositive at75 mg/dL | None | FalseNegative/Positive at75 mg/dL |
| Formalin | 370 mg/dL | None | FalseNegative at185 mg/dL | None | FalseNegative at92.5 mg/dL | FalsePositive at92.5 mg/dL | FalseNegative at185 mg/dL | None | FalseNegative at92.5 mg/dL |
| Galactose | 20 mg/dL | None | None | None | None | None | None | None | None |
| GentamicinSulfate | 6 mg/dL | None | None | None | None | None | None | None | None |
| Glucose | 5 g/dL | None | None | NA | None | FalseNegative at1250 mg/dL | None | None | FalseNegative at2500 mg/dL |
| Glycine | 400 mg/dL | None | None | None | None | None | None | None | None |
| Glyburide | 15 ug/mL | None | None | None | None | None | None | None | None |
| Hemoglobin | 5 mg/dL | None | None | None | None | None | None | None | None |
| Hypochlorite | 2.60% | FalseNegative at1.3% | FalsePositive at0.65% | FalsePositive at1.3% | FalsePositive at1.3% | FalseNegative at0.65% | None | None | FalseNegative at0.65% |
| Hydrochloro-thiazide | 6 mg/mL | None | None | None | None | None | None | None | FalseNegative at6 mg/dL |
| Indican | 50 mg/dL | None | None | None | None | None | None | None | None |
| Imipenem | 100 mg/dL | None | None | None | FalsePositive at50 mg/dL | None | None | None | None |
| InterferingSubstance | Initial ConcTested | BIL | BLO | GLU | KET | LEU | NIT | PRO | URO |
| Keflex | 150 mg/dL | None | None | None | None | None | None | None | None |
| Keflin | 50 mg/dL | None | None | None | None | None | None | None | None |
| Lactose | 20 mg/dL | None | None | None | None | None | None | None | FalseNegative at20 mg/dL |
| Levadopa | 75 mg/dL | None | FalseNegative at18.8 mg/dL | None | FalsePositive at56 mg/dL | None | None | None | None |
| Lodine | 100 mg/dL | None | None | None | None | None | None | None | None |
| Meropenem | 100 mg/dL | None | None | None | None | None | None | None | FalseNegative at100 mg/dL |
| Mesna | 80 mg/dL | FalseNegative at20 mg/dL | FalseNegative at20 mg/dL | FalseNegative at20 mg/dL | FalsePositive at20 mg/dL | None | None | None | FalseNegative at20 mg/dL |
| Metformin | 4 mg/mL | None | None | None | None | None | None | None | None |
| Nitrofurantoin | 120 mg/dL | None | None | None | None | None | None | None | FalseNegative at120 mg/dL |
| Oxalic Acid | 95 mg/dL | None | None | None | None | None | FalseNegative at71 mg/dL | None | None |
| P-aminosalicyclicacid | 250 mg/dL | None | None | None | None | None | None | None | FalseNegative at62.5 mg/dL |
| Penicillin | 300 mg/dL | None | None | None | None | None | None | None | None |
| Pyridium | 200 mg/dL | None | None | None | None | None | FalsePositive at50 mg/dL | FalsePositive at150 mg/dL | FalsePositive at100 mg/dL |
| InterferingSubstance | Initial ConcTested | BIL | BLO | GLU | KET | LEU | NIT | PRO | URO |
| PotassiumChloride | 1200 mg/dL | None | None | None | None | None | None | None | None |
| Quinidine | 30 mg/dL | None | None | None | None | None | None | None | None |
| Riboflavin | 120 mg/dL | None | None | None | None | None | FalsePositive at30 mg/dL | None | None |
| Sodium Nitrite | 3 mg/dL | None | None | None | None | None | NA | None | None |
| Sulfamethoxazole | 70 mg/dL | None | None | None | None | None | None | None | FalseNegative at17.5 mg/dL |
| Tagamet | 10 mg/dL | None | None | None | None | None | None | None | FalseNegative at5 mg/dL |
| Tetracycline | 12 mg/dL | None | None | None | None | None | None | None | None |
| pH > 9 | NA | None | None | None | None | None | None | None | None |
| MicrobialPeroxidase | NA | None | FalsePositive | None | None | None | None | None | None |
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
{16}------------------------------------------------
Comparison Studies
d. Method Comparison with Predicate Device
The method comparison study compared the clinical results of the CLINITEK Novus analyzer to the predicate device. The study was conducted across 3 sites with up to 2773 specimens tested. The results for overall percent agreement and within one level agreement met acceptance criteria. In the following tables, results are shown for all sites combined.
| BIL | Predicate Result | ||||
|---|---|---|---|---|---|
| Negative | Small | Moderate | Large | ||
| CLINITEKNovusResult | Negative | 2234 | 104 | 1 | 1 |
| Small | 5 | 190 | 15 | 1 | |
| Moderate | 1 | 112 | 7 | ||
| Large | 102 | ||||
| Total | 2239 | 295 | 128 | 111 | |
| % exact match | 99.8% | 64.4% | 87.5% | 91.9% | |
| % within 1 block | 100.0% | 100.0% | 99.2% | 98.2% |
| BLO | Predicate Result | |||||
|---|---|---|---|---|---|---|
| Negative | Trace | Small | Moderate | Large | ||
| CLINITEKNovusResult | Negative | 1489 | 4 | 1 | ||
| Trace/NHT | 34 | 97 | 12 | |||
| Small | 18 | 127 | 10 | |||
| Moderate | 5 | 125 | 14 | |||
| Large | 1 | 204 | ||||
| Total | 1523 | 119 | 145 | 136 | 218 | |
| % exact match | 97.8% | 81.5% | 87.6% | 91.9% | 93.6% | |
| % within 1 block | 100.0% | 100.0% | 99.7% | 100.0% | 100.0% |
{17}------------------------------------------------
| GLU | Predicate Result | |||||
|---|---|---|---|---|---|---|
| Negative | 100mg/dL | 250mg/dL | 500mg/dL | 1000mg/dL | ||
| Negative | 1563 | |||||
| 100mg/dL | 7 | 147 | ||||
| CLINITEKNovusResult | 250mg/dL | 19 | 95 | |||
| 500mg/dL | 8 | 88 | 4 | |||
| 1000mg/dL | 4 | 183 | ||||
| Total | 1570 | 166 | 103 | 92 | 187 | |
| % exact match | 99.6% | 88.5% | 92.2% | 95.7% | 97.9% | |
| % within 1 block | 100.0% | 100.0% | 100.0% | 100.0% | 100.0% |
| KET | Predicate Result | ||||||
|---|---|---|---|---|---|---|---|
| Negative | Trace | 15 mg/dL | 40 mg/dL | 80 mg/dL | 160 mg/dL | ||
| Negative | 1670 | 14 | 1* | ||||
| CLINITEKNovusResult | Trace | 2 | 158 | 9 | |||
| 15 mg/dL | 2 | 82 | 6 | ||||
| 40 mg/dL | 2 | 117 | 1 | ||||
| 80 mg/dL | 66 | ||||||
| 160 mg/dL | 5 | 28 | |||||
| Total | 1672 | 174 | 93 | 124 | 72 | 28 | |
| % exact match | 99.9% | 90.8% | 88.2% | 94.4% | 91.7% | 100.0% | |
| % within 1 block | 100.0% | 100.0% | 100.0% | 99.2% | 100.0% | 100.0% |
- This is a highly colored sample. Predicate labeling cautions testing of highly colored samples as they may yield a falsely elevated result.
{18}------------------------------------------------
| LEU | CLINITEK Atlas Result | |||||
|---|---|---|---|---|---|---|
| Negative | Trace | Small | Moderate | Large | ||
| CLINITEKNovusResult | Negative | 1600 | 5 | 1 | ||
| Trace | 79 | 80 | 13 | |||
| Small | 1 | 10 | 129 | 3 | ||
| Moderate | 16 | 85 | 5 | |||
| Large | 6 | 157 | ||||
| Total | 1680 | 95 | 159 | 94 | 162 | |
| % exact match | 95.0% | 85.7% | 81.1% | 90.4% | 96.6% | |
| % within 1 block | 99.9% | 100.0% | 99.7% | 100.0% | 100.0% |
| NIT | Predicate Result | |
|---|---|---|
| Negative | Positive | |
| CLINITEKNovusResult | 1707 | |
| 12 | 399 | |
| Total | 1719 | 399 |
| % exact match | 99.3% | 100.0% |
| % within 1 block | 100.0% | 100.0% |
| pH* | Predicate Result | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| 5 | 5.5 | 6 | 6.5 | 7 | 7.5 | 8 | 8.5 | ||
| CLINITEKNovusResult | 5 | 85 | 121 | ||||||
| 5.5 | 1 | 203 | 213 | ||||||
| 6 | 1 | 426 | 19 | ||||||
| 6.5 | 26 | 399 | 10 | ||||||
| 7 | 33 | 376 | 2 | ||||||
| 7.5 | 106 | 178 | |||||||
| 8 | 1 | 64 | 44 | 1 | |||||
| 8.5 | 1 | 51 | 49 | ||||||
| >=9.0 | 2 | 361 | |||||||
| Total | 86 | 325 | 666 | 451 | 492 | 245 | 97 | 411 | |
| % exact match | 98.8% | 62.5% | 64.0% | 88.5% | 76.4% | 72.7% | 45.4% | 99.8% | |
| % within 1 block | 100.0% | 100.0% | 99.9% | 100.0% | 100.0% | 99.6% | 97.5% | 100.0% |
*Results on CLINITEK Novus of ≥8.5 were considered a match to a pH of 8.5 on the predicate.
{19}------------------------------------------------
| PRO | Predicate Result | ||||||
|---|---|---|---|---|---|---|---|
| Negative | Trace | 30mg/dL | 100mg/dL | 300mg/dL | 1000mg/dL | ||
| Negative | 1511 | 2 | |||||
| CLINITEKNovusResult | Trace | 24 | 207 | 4 | |||
| 30mg/dL | 20 | 184 | 3 | ||||
| 100mg/dL | 1 | 7 | 145 | 1 | |||
| 300mg/dL | 7 | 77 | 8 | ||||
| 1000mg/dL | 71 | ||||||
| Total | 1536 | 229 | 195 | 155 | 78 | 79 | |
| % exact match | 98.4% | 90.4% | 94.4% | 93.5% | 98.7% | 89.9% | |
| % within 1 block | 99.9% | 100.0% | 100.0% | 100.0% | 100.0% | 100.0% |
| URO | Predicate Result | |||||
|---|---|---|---|---|---|---|
| 0.2 mg/dL | 1 mg/dL | 2 mg/dL | 4 mg/dL | 8mg/dL | ||
| CLINITEKNovusResult | 0.2mg/dL | 1982 | 53 | |||
| 1 mg/dL | 2 | 327 | 15 | |||
| 2 mg/dL | 2 | 209 | 10 | |||
| 4 mg/dL | 5 | 116 | 3 | |||
| 8 mg/dL | 4 | 45 | ||||
| Total | 1984 | 382 | 229 | 130 | 48 | |
| % exact match | 99.9% | 85.6% | 91.3% | 89.2% | 93.8% | |
| % within 1 block | 100.0% | 100.0% | 100.0% | 100.0% | 100.0% |
{20}------------------------------------------------
As the number of contrived samples for URO exceeded 15%, testing was also performed at a fourth site. For this fourth site, less than 5% of the samples were contrived. Results are in the following table:
| URO | Predicate Result | |||||
|---|---|---|---|---|---|---|
| 0.2mg/dL | 1 mg/dL | 2 mg/dL | 4 mg/dL | 8 mg/dL | ||
| CLINITEKNovusResult | 0.2 mg/dL | 187 | 2 | |||
| 1 mg/dL | 46 | |||||
| 2 mg/dL | 10 | |||||
| 4 mg/dL | 12 | |||||
| 8 mg/dL | 4 | |||||
| Total | 187 | 48 | 10 | 12 | 4 | |
| % exact match | 100.0% | 95.8% | 100.0% | 100.0% | 100.0% | |
| % within 1 level | 100.0% | 100.0% | 100.0% | 100.0% | 100.0% |
Overall Specific Gravity Slope and Intercept
| Site | N | R2 | Slope | Intercept |
|---|---|---|---|---|
| Overall | 1993 | 0.983 | 1.020 | -0.019 |
- c. Matrix Comparison
Not applicable. The device is for urine samples only.
Clinical Studies
a. Clinical Sensitivity Not applicable
b. Clinical Specificity Not applicable
Clinical Cut-off
Not applicable
Expected Values / Reference Range
Complete details of the expected analyte values are provided in the labeling.
9.0 Instrument Name:
CLINITEK Novus Automated Urine Chemistry Analyzer
{21}------------------------------------------------
10.0 System Descriptions
Modes of Operation
Single and continuous testing.
Specimen Identification
An internal barcode reader scans the barcode labels on the sample tubes in the racks or sample identification may be entered manually using the touch screen keyboard.
Specimen Sampling and Handling
The CLINITEK Novus system can analyze multiple sample pipette aspirates a sample from the sample tube and dispenses sample onto each test pad.
Calibration
The CLINITEK Novus system uses four liquid calibration solutions to perform a calibration reflectance check. The calibrators have a fixed value with no lot specific values. A calibration must be performed when a new lot of reagent cassette is loaded onto the system or a new cassette of the same lot is loaded and the calibration is older than 24 hours.
The calibrators are identical to those cleared in K932674.
Quality Control
Recommendations for testing quality control are in the labeling. Routinely test at least 2 levels (negative/low and positive) of an appropriate commercially available control to confirm performance of the CLINITEK Novus urinalysis cassette. Controls may also be randomly placed in each batch of samples tested. Each laboratory should establish its own qoals for acceptable standards of performance, and should question handling and testing procedures if these standards are not met.
The CLINITEK Atlas controls, cleared in K932674, may be used with the CLINITEK Novus automated urine chemistry analyzer.
Other Supportive Instrument Performance Characteristics Data Not Covered in the "Performance Characteristics" Section above:
None
11.0 Conclusion
The results of these studies demonstrate that the CLINITEK Novus automated urine chemistry analyzer and CLINITEK Novus 10 urinalysis cassette is similar to the predicate in both Technological Characteristics and Intended Use. The data presented is a summary of external clinical evaluation, internal laboratory evaluation, and software development information. The CLINITEK Novus performance was shown to be substantially equivalent to the predicate urinalysis device.
§ 862.2900 Automated urinalysis system.
(a)
Identification. An automated urinalysis system is a device intended to measure certain of the physical properties and chemical constituents of urine by procedures that duplicate manual urinalysis systems. This device is used in conjunction with certain materials to measure a variety of urinary analytes.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.