K Number
K140611
Manufacturer
Date Cleared
2014-05-21

(72 days)

Product Code
Regulation Number
888.3060
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Trinica® Anterior Lumbar Plate System is indicated for use via the lateral or anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach, below the bifurcation of the great vessels. This system is indicated in the treatment of lumbar or lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), pseudoarthrosis, spondylolisthesis, scoliosis, lordotic deformities of the spine, spinal stenosis, or a failed previous fusion.

Device Description

The Trinica® Anterior Lumbar Plate (ALP) System is a temporary supplemental fixation device consisting of a variety of shapes and sizes of plates and screws. The Trinica Anterior Lumbar Plate System is used as an implant for the correction and stabilization of the spine. This system provides temporary stabilization and augments the development of a solid spinal fusion. Additionally, this system provides the surgeon with the ability to supplement an interbody device with anterior plate fixation. The Trinica Anterior Lumbar Plate System components can be locked into a variety of configurations and each construct may be customized to individual cases. The plates are low profile and anatomically designed to provide optimal fit from either anterior or anterior-lateral approach. This system also features anti-migration locking caps to help secure the fixation screws. All Trinica Anterior Lumbar Plate System implant components are made from titanium alloy (Ti-6AI-4V ELI).

The Trinica Anterior Lumbar Plate System is supplied with the instrumentation necessary for implantation of the system. The Trinica" Anterior Lumbar Plate System is for single use only.

AI/ML Overview

The provided documentation focuses on a 510(k) summary for the Zimmer Spine Trinica® Anterior Lumbar Plate System, which is a medical device for spinal fixation. This type of submission typically demonstrates substantial equivalence to a predicate device rather than presenting a performance study with acceptance criteria in the way a diagnostic AI/ML device would.

Therefore, the requested information regarding acceptance criteria, study details (sample size, data provenance, expert adjudication, MRMC study, standalone performance), and ground truth establishment is not applicable to this document. The document describes a sterilization validation as the primary "performance testing" conducted.

Here's a breakdown of the applicable information:

1. A table of acceptance criteria and the reported device performance:

Acceptance CriteriaReported Device Performance
Device meets design controls for sterilization.Sterilization testing of the components/instruments contained in the subject Trinica ALPS tray was conducted. The test results demonstrated the modified device met the requirements of ISO 17665 and AAMI TIR12.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

  • Not Applicable. The submission does not detail a "test set" for a performance study in the context of diagnostic accuracy. The performance testing was related to sterilization validation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

  • Not Applicable. Ground truth, in the context of expert review, is not relevant to sterilization validation.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not Applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This device is a spinal implant, not an AI-powered diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Not Applicable. This device is a spinal implant.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

  • Not Applicable. For sterilization validation, the "ground truth" would be the successful killing of microorganisms, which is assessed through specific microbiological testing methods and standards (ISO 17665, AAMI TIR12).

8. The sample size for the training set:

  • Not Applicable. There is no "training set" for this type of medical device submission.

9. How the ground truth for the training set was established:

  • Not Applicable.

{0}------------------------------------------------

Page 1 of 2 K140611

Zimmer Spine 510(k) - Trinica ALP

510(k) Summary

MAY 2 1 20 %

510(k) SUMMARY

Trinica® Anterior Lumbar Plate System

March 07, 2014 Date of Summary Preparation: Submitter: Zimmer Spine, Inc. 7375 Bush Lake Road · Minneapolis, MN 55439 USA Establishment Registration Number: 2184052 (Minneapolis) Donna M. Semlak Company Contact (Primary): Senior Requlatory Affairs Specialist Email: Donna.Semlak@zimmer.com Office: 952.857.5643 Email Fax: 952.857.5843 Trade Name(s): Trinica® Anterior Lumbar Plate System Device Name (Common Name): Spinal Intervertebral Body Fixation Orthosis Device Classification: Class II Requlation Number and 21 CFR § 888.3060 / KWQ Product Code(s): Predicate Devices: Trinica® Anterior Lumbar Plate System (K061353)

General Device Description:

The Trinica® Anterior Lumbar Plate (ALP) System is a temporary supplemental fixation device consisting of a variety of shapes and sizes of plates and screws. The Trinica Anterior Lumbar Plate System is used as an implant for the correction and stabilization of the spine. This system provides temporary stabilization and augments the development of a solid spinal fusion. Additionally, this system provides the surgeon with the ability to supplement an interbody device with anterior plate fixation. The Trinica Anterior Lumbar Plate System components can be locked into a variety of configurations and each construct may be customized to individual cases. The plates are low profile and anatomically designed to provide optimal fit from either anterior or anterior-lateral approach. This system also features anti-migration locking caps to help secure the fixation screws. All Trinica Anterior Lumbar Plate System implant components are made from titanium alloy (Ti-6AI-4V ELI).

The Trinica Anterior Lumbar Plate System is supplied with the instrumentation necessary for implantation of the system. The Trinica" Anterior Lumbar Plate System is for single use only.

{1}------------------------------------------------

510(k) Summary

Indications for Use:

The Trinical Anterior Lumbar Plate System is indicated for use via the lateral or anterolateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach, below the bifurcation of the great vessels. This system is indicated in the treatment of lumbar or lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), pseudoarthrosis, spondylolisthesis, scoliosis, lordotic deformities of the spine, spinal stenosis, or a failed previous fusion.

Summary of Technological Characteristics

The technological characteristics remain the same for the subject Trinica® Anterior Lumbar Plate System as the currently marketed (predicate) Trinica® Anterior Lumbar Plate System, which received FDA clearance K061353. There are no changes to the implants (plates and/or screws) and instrumentation within this submission is only addressing a sterilization tray bracket modification.

All the technology characteristics remain the same system's intended use, same mechanical and functional scientific technology; same materials and the same substantially equivalent performance characteristics.

The Trinical Anterior Lumbar Plate System provides temporary supplemental fixation devices consisting of a variety of shapes and sizes of plates and screws: intended to provide anterior lumbar temporary stabilization and the development of a solid spinal fusion per the indications for use, as stated in the Section above.

Trinica® Anterior Lumbar Plate System includes: plates, bone screws with locking caps, and instrumentation necessary to implant this specific system. The Trinica Anterior Lumbar Plate System is for single use only.

Summary of Performance Testing

Sterilization testing of the components/instruments contained in the subject Trinica ALPS tray were assessed and tested appropriately to design controls; i.e. sterilization validation. The test results demonstrated the modified device met the requirements of ISO 17665 and AAMI TIR12.

Substantial Equivalence

Zimmer Spine considers the subject Trinica Anterior Lumbar Plate System to be substantially equivalent to the currently marketed Trinica Anterior Lumbar Plate System, K061353, as there are no changes to: the intended use, mechanical and functional performance, functional scientific technology, the implants (plates or screw) and/or instrumentation.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/12 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, rendered in black.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 21, 2014

Zimmer Spine, Incorporated Ms. Donna M. Semlak Senior Regulatory Affairs Specialist 7375 Bush Lake Road Minneapolis, Minnesota 55439

K140611 Re:

Trade/Device Name: Trinica® Anterior Lumbar Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: March 7, 2014 Received: March 10, 2014

Dear Ms. Semlak:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Donna M. Semlak

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDcvices/Safetv/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Ronald P. Jean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K140611

Device Name

Trinica® Anterior Lumbar Plate System

Indications for Use (Describe)

The Trinica® Anterior Lumbar Plate System is indicated for use via the lateral surgical approach above the bifurcation of the great vessels or via the anterior surgical approach, below the great vessels. This system is indicated in the treatment of lumbosacral (L)-SI) spine instability as a result of fracture (including dislocation and subluxation), tumor, degencrative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondyfolisthesis, scoliosis. lordotic deformities of the spine, spinal stenosis, or a failed previous fusion.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Ronald P. Jean -S 2014.05.21 09:48:15 -04'00'

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

§ 888.3060 Spinal intervertebral body fixation orthosis.

(a)
Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.(b)
Classification. Class II.