K Number
K140606
Date Cleared
2014-05-12

(63 days)

Product Code
Regulation Number
892.1000
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of this PROCURE Array coil family is to provide high-quality images of the reproductive and urological anatomies in an easy-to-position, wearable, and very flexible design.

The coil is indicated for use by the order of a physician to be used as an accessory to a General Electric 1.5T or 3.0T magnetic resonance scanner for general human anatomy imaging as supported by the scanner. These images, when interpreted by a trained physician, may assist in medical diagnosis.

Device Description

The PROCURE™ Coil Family interfaces with a 1.5T or 3.0T 8-channel GE MRI scanner and provides high-quality images of the reproductive and urological anatomies in an easy-to-position, wearable, and very flexible design. This lightweight SemiFlex™ design facilitates effortless and accurate positioning similar to wearing a diaper and positions the multiple antenna elements as close as possible to the target anatomies regardless of patient size. The enclosure for the antenna set is made of flexible liquid impermeable, biocompatible materials, and the coil is accompanied by disposable liners (USP Class 6 PET) easily changed between patients, should scanning without clothing be desired. The two coils in the product family are identical in design, construction, materials, and operation, with the exception of resonant frequency. The designs and materials used for manufacture of both coils are identical to standard MRI coil technology that has existed for several years. The coils use similar blocking networks and impedance matching circuits, and they do not transmit energy into the patient, neither predicate nor current submission.

This phased array coil also accommodates popular biopsy systems, both delivered via an opening near the anus and/or vagina, neither included in the scope of this submission as they are independently operated and cleared by the FDA

AI/ML Overview

This device is an MRI coil, a non-AI medical device. The provided text is a 510(k) summary for a substantial equivalence determination, not a study evaluating acceptance criteria for an AI/ML powered device. As such, acceptance criteria, efficacy study, and information regarding ground truth and expert involvement are not applicable in the context of AI/ML evaluation.

The 510(k) summary details the "PROCURE™ Array Coil Family" and compares it to a predicate device (1.5T ScanMed® PV Array, K022395) to demonstrate substantial equivalence.

Here's an analysis based on the information provided, specifically addressing the non-AI nature:

1. Table of acceptance criteria and the reported device performance

For a non-AI device like an MRI coil, "acceptance criteria" for performance primarily revolve around meeting established safety standards and demonstrating image quality comparable to or better than a predicate device. The document does not present a formal table of quantitative acceptance criteria and reported device performance in the way one would for an AI/ML model's diagnostic accuracy. Instead, it relies on demonstrating equivalence through various engineering and scientific analyses.

The key performance metrics for an MRI coil typically involve Signal-to-Noise Ratio (SNR) and image uniformity.

Criteria CategoryDevice Performance / Justification
Safety and Biocompatibility
Biocompatibility"Both predicate and modified devices have a successful biocompatibility track record, as demonstrated by cytotoxicity testing and by their history of use in previously cleared devices." The nylon fabric used has been "successfully used for over 10 years in the predicate device with no reported biocompatibility issues."
Mechanical & Electrical Safety"Both modified and predicate devices have been tested for mechanical and electrical safety using IEC60601-1 3rd Edition."
Risk Management"The predicate and current submission have been subject to similar risk management studies,...and determined to be substantially equivalent." (Included: Blocking Analysis, Heat Testing, Risk management (including hazard analysis and FMEA))
Performance
Image Quality"SNR and uniformity analysis" was conducted. Clinical tests involved "Analyses in all 3 planes (sagittal, coronal and transverse) were run on the PROCURE Array product to show that the anatomies of the submitted and predicate coils have substantial equivalence". The device is intended to "provide high-quality images of the reproductive and urological anatomies." The comparison to the predicate device covering "very similar ranges of anatomy" suggests that the image quality is expected to be diagnostically equivalent.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document mentions "Clinical Tests" were performed with "Analyses in all 3 planes (sagittal, coronal and transverse) were run on the PROCURE Array product." However, it does not specify the sample size of these clinical tests (i.e., number of patients or imaging studies). The data provenance (country of origin, retrospective/prospective) is also not stated.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. For MRI coils, the "ground truth" is not established in the same way as for AI models requiring expert annotation. Image quality is assessed based on physical properties and visual interpretation by a "trained physician" for diagnosis, but there's no mention of a formal expert panel to establish ground truth for a test set in the context of an efficacy study as would be seen for an AI product.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. As this is not an AI/ML product requiring diagnostic accuracy adjudication, no such method is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted device, and no MRMC study or AI-related comparative effectiveness study was performed. The comparison is between two physical MRI coils.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a passive receive-only MRI coil, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable in the AI/ML sense. The "ground truth" for an MRI coil's performance typically refers to its ability to produce clinically usable images, which is assessed by "trained physicians" for "medical diagnosis." However, this is not a formal ground truth as required for validating an AI diagnostic algorithm.

8. The sample size for the training set

Not applicable. This is a hardware device, not an AI/ML model that requires training data.

9. How the ground truth for the training set was established

Not applicable, as there is no training set for an MRI coil.

{0}------------------------------------------------

MAY 12 2014

Image /page/0/Picture/1 description: The image shows the logo for ScanMed. The logo is in black and white, with the word "ScanMed" in a bold, sans-serif font. Below the word "ScanMed" is the tagline "The Image is Everything" in a smaller, serif font. The logo is simple and modern, and it conveys the company's focus on imaging.

Resonance Innovations, LLC 9840 S. 140" St., Suite 8 Omaha, NE 68138 (0) 402.934.2650 (F) 402.778.9699 www.scanmed.com

Section 8.510(k) Summary, K140606
Submitter's Name:Resonance Innovations LLC
Submitter's Address:9840 South 140th St., Suite 8Omaha, NE 68138
Submitter's Telephone:402-934-2650
Submitter's Contact:Randall Jones, President
Date 510(k) Summary prepared:April 21, 2014
Proprietary Name:PROCURE™ Array Coil Family
Common or Usual Name:MRI coil(s)
Classification Name:Coil, Magnetic Resonance, Specialty
Classification Code:MOS
Predicate Device:1.5T ScanMed® PV Array, K022395

Description of the Device

ﻴﺔ ﺍﻟ

The PROCURE™ Coil Family interfaces with a 1.5T or 3.0T 8-channel GE MRI scanner and provides high-quality images of the reproductive and urological anatomies in an easy-to-position, wearable, and very flexible design. This lightweight SemiFlex™ design facilitates effortless and accurate positioning similar to wearing a diaper and positions the multiple antenna elements as close as possible to the target anatomies regardless of patient size. The enclosure for the antenna set is made of flexible liquid impermeable, biocompatible materials, and the coil is accompanied by disposable liners (USP Class 6 PET) easily changed between patients, should scanning without clothing be desired. The two coils in the product family are identical in design, construction, materials, and operation, with the exception of resonant frequency. The designs and materials used for manufacture of both coils are identical to standard MRI coil technology that has existed for several years. The coils use similar blocking networks and impedance matching circuits, and they do not transmit energy into the patient, neither predicate nor current submission.

This phased array coil also accommodates popular biopsy systems, both delivered via an opening near the anus and/or vagina, neither included in the scope of this submission as they are independently operated and cleared by the FDA

40606
page 1 of 3

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for ScanMed. The text "ScanMed" is written in a bold, sans-serif font, with the words "The Image is Everything" written in a smaller font below. The logo is simple and modern, and it conveys the company's focus on medical imaging.

Resonance Innovations, LLC 9840 S. 140" St., Suite 8 Omaha, NE 68138 (0) 402.934.2650 (F) 402.778.9699 www.scanmed.com

Device Model NumberDevice Description
508GE15011.5T PROCURE Array Coil
508GE30013.0T PROCURE Array Coil

Intended Use

The intended use of this PROCURE Array coil family is to provide high-quality images of the reproductive and urological anatomies in an easy-to-position, wearable, and very flexible design.

Indications for Use

The coil is indicated for use by the order of a physician to be used as an accessory to a General Electric 1.5T or 3.0T magnetic resonance scanner for general human anatomy imaging as supported by the scanner. These images, when interpreted by a trained physician, may assist in medical diagnosis.

Technological Characteristics

The comparison between the predicate and the current submission is described, below.

    1. Design. This submission is for a dedicated coil that may give diagnostic quality images of the reproductive and urological anatomies. It is a multi-channel coil with all channels designed to work at once, receive-only, in conjunction with the system Body Coil. The predicate multi-channel, PV Array covers very similar ranges of anatomy, also working as receive-only to the Body Coil. Both predicate and modified coils have semi-flexible anterior and posterior sections, with an antenna section between the legs. The modified device is more form-fitted to allow the antenna geometry to physically closer to the patient's anatomy. All design principles employed are mature and well-known throughout the industry.
    1. Principles of operation: The scientific principles of operation (magnetic resonance) are identical between the predicate and modified devices. Theory of operation is very well understood throughout the industry.
    1. Materials. The same materials are used in the construction of the predicate and modified devices. All internal circuitry is encapsulated in flame retardant EVA foam, then completely covered with a nylon fabric or compressed EVA. This nylon fabric has been successfully used for over 10 years in the predicate device with no reported biocompatibility issues. Both modified and predicate devices have been tested for mechanical and electrical safety using IEC60601-1 3rd Edition.
    1. Chemical Composition. Both predicate and modified devices have a successful biocompatibility track record, as demonstrated by cytotoxicity testing and by their history of use in previously cleared devices.

{2}------------------------------------------------

K140606
Page 3 of 3

Resonance Innovations, LLC 9840 S. 140th St., Suite 8 Omaha, NE 68138 (0) 402.934.2650 (F) 402.778.9699 www.scanmed.com

    1. Energy Source. Both of these products are receive-only coils not generating their own power, but rather controlled by the MRI system as the energy source.

Non-Clinical Tests

The coils have similar dimension in the head area and have induced similar fields by the transmit coil (Body Coil) that stimulates them. The predicate and current submission have been subject to similar risk management studies, as listed below, and determined to be substantially equivalent.

    1. Blocking Analysis
    1. SNR and uniformity analysis
    1. Risk management (including hazard analysis and FMEA)
    1. Heat Testing
    1. Compliance testing to IEC60601-1 3rd Edition

Clinical Tests

Analyses in all 3 planes (sagittal, coronal and transverse) were run on the PROCURE Array product to show that the anatomies of the submitted and predicate coils have substantial equivalence; the predicate device images the thorax, abdomen, pelvis and hip regions, and the current submission is designed for imaging the reproductive and urological anatomies.

Substantial Equivalence Decision

As described in this summary, the modified device is substantially equivalent to the predicate device based on the analysis herein. The modified device raises no new concerns of safety or efficacy. Note that the predicate device was made by Medic, Inc., which is a historic company name. Resonance Innovations is under the same leadership, and for both the dba is ScanMed®

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the words "Public Health Service" in a bold, sans-serif font. The text is arranged on a single line. The words are clearly legible and the image appears to be a close-up of the text.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 12, 2014

Resonance Innovations, LLC % Mr. Randall Jones President & CEO 9840 South 140th Street OMAHA NE 68138-3693

Re: K140606

Trade/Device Name: 1.5T PROCURE™ Array Coil; 3.0T PROCURE™ Array Coil 2 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: March 28, 2014 Received: April 22, 2014

Dear Mr. Jones:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2-Mr. Jones

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Smh.7)

for

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K140606

Device Name PROCURE Array Family

Indications for Use (Describe)

The coil is indicated for use by the order of a physician to be used as an accessory to a General Electric 1.5T or 3.0T magnetic resonance scanner for general human anatomy imaging as supported by the scanner. These images, when interpreted by a trained physician. may assist in medical diagnosis.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)

Michael D. O'Hara

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.