(143 days)
The Saphena Medical OnePass Endoscopic Vessel Harvesting System is in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. It is indicated for cutting tissue and controlling bleeding through coagulation, and for patients requiring blunt dissection of tissue including dissection of blood vessels, dissection of blood vessels of the extremities, dissection of ducts and other structures in the extra peritoneal or subcutaneous extremity and thoracic space. Extremity procedures include tissue dissection along the saphenous vein for use in coronary artery bypass grafting and peripheral artery bypass or radial artery for use in coronary artery bypass grafting. Thorascopic procedures include exposure and dissection of structures external to the parietal pleura, including nerves, blood vessels, and other tissues of the chest wall.
The OnePass™ Endoscopic Vessel Harvesting System attaches to existing laparoscopic endoscopes (e.g. 7mm) to provide a means to endoscopically harvest vessels for use in arterial bypass graft procedures. Endoscopic Vessel Harvesting is a procedure that has been in use for approximately 15 years. The OnePass™ device differs in that it provides for single device entry into the body cavity without the need for additional instruments. Also, the distal end effector design allows for excellent visibility and quick thermal ligation of vessel trunks while minimizing trauma to the target vessel.
The provided text describes a 510(k) submission for the OnePass™ Endoscopic Vessel Harvesting System. The focus of the submission is to demonstrate substantial equivalence to a predicate device, not to establish performance against specific acceptance criteria through a standalone clinical study. Therefore, direct answers to many of the requested points are not available in this document.
Here's an analysis based on the information provided:
1. A table of acceptance criteria and the reported device performance:
The document focuses on demonstrating substantial equivalence to an existing predicate device rather than fulfilling specific, quantifiable acceptance criteria for a novel device performance claim. The reported device performance is defined by its ability to be "as safe and effective and performs at least as safely and effectively as the predicate device."
| Acceptance Criteria (Inferred from Substantial Equivalence Claim) | Reported Device Performance |
|---|---|
| Safe and Effective as Predicate Device | "Performance testing demonstrates that the device is as safe and effective and performs at least as safely and effectively as the predicate device stated in this summary." |
| Functional Specification Met | "Non-clinical testing of the subject device has shown that it meets its functional specification..." |
| Performs According to Intended Use | "...and performs according to the prescribed intended use." |
| Equivalent Intended Use to Predicate | "As described in this submission, the intended use of the OnePass™ Endoscopic Vessel Harvesting System is equal to the predicate device." |
| Similar Technological Characteristics to Predicate | "Further, the subject and predicate device have similar technological characteristics." |
| Equivalent in terms of Intended Use, Principles of Operations, Technology, Design, Materials and Performance to Predicate | "The OnePass™ Endoscopic Vessel Harvesting System is equivalent to the predicate device in terms of intended use, principles of operations, technology, design, materials and performance." |
| No new issues of safety or effectiveness raised by noted differences | "The noted difference between the OnePass™ Endoscopic Vessel Harvesting and predicate device does not raise new issues of safety or effectiveness." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
The document describes non-clinical performance data, which includes:
- Design Verification Testing (dimensional verification, tensile/compressive strength, durability)
- Compatibility Testing with standard 7mm scope
- Electrical Safety and Electromagnetic Testing
- Packaging and Shipping Testing
- Sterility Testing
- Biocompatibility Testing
- Simulated animal testing use evaluation
These tests are typically laboratory-based and do not involve human subjects or clinical data in the traditional sense of a "test set" like in an AI/diagnostic algorithm study. Therefore, information on sample size for a "test set" (human cases) and data provenance (country, retrospective/prospective) is not applicable in this context. The "simulated animal testing use evaluation" implies some form of animal study, but no details on sample size or specific outcomes are provided beyond its inclusion in the non-clinical testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Since this is a submission for a surgical device demonstrating substantial equivalence via non-clinical performance and animal testing, there is no mention of experts used to establish ground truth for a clinical test set, as such a set was not used in the described studies.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
As no clinical test set with human data and expert ground truth establishment is described, an adjudication method is not applicable and not mentioned.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This device is an electrosurgical cutting and coagulation device, not an AI or diagnostic imaging device. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and not described.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
This is a physical surgical device, not an algorithm. Therefore, "standalone algorithm performance" is not applicable and not described.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
For the non-clinical and simulated animal testing, the "ground truth" would be established through engineering specifications, validated test methods, and potentially pathological examination or physiological endpoints in the animal models, rather than expert consensus on human data or pathology from real clinical cases. The specific types of "ground truth" for each non-clinical test (e.g., tensile strength values, sterility test results, biocompatibility standards compliance) are implied but not explicitly detailed as "ground truth types" in the provided summary.
8. The sample size for the training set:
As this is a physical device and not an AI algorithm, the concept of a "training set" in the context of data for model development is not applicable and not mentioned. The device's design and manufacturing processes are based on engineering principles and regulatory standards, not statistical training data.
9. How the ground truth for the training set was established:
Since there is no "training set" as defined for an AI algorithm, this question is not applicable. The "ground truth" for the device's design and functionality would be established through engineering design specifications, material properties, and manufacturing standards.
{0}------------------------------------------------
JUL 3 1 2014
K140603 Section 005
510(k) Summary (21 CFR § 807.92(c))
| Submitter: | Saphena Medical, Inc.375 West StreetWest Bridgewater, MA 02379 |
|---|---|
| Contact: | Cheryl BlakeRA / QA, Saphena Medical ConsultantTelephone: 949.285.3517Email: cherylblake@cox.net |
| Date Summary Prepared: | January 30, 2014 |
| Device Trade Name: | OnePass™ Endoscopic Vessel Harvesting System |
| Common Name: | Electrosurgical cutting and coagulation device and accessories |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
| Product Code: | GEI |
| Class II | 21 CFR 878.4400 |
| Predicate Devices: | Maquet Cardiovascular, LLC Vasoview 6 Pro EndoscopicHarvesting System (K091733)Decision Date: August 28, 2009 |
Device Description:
The OnePass™ Endoscopic Vessel Harvesting Systemattaches to existing laparoscopic endoscopes (e.g. 7mm) to provide a means to endoscopically harvest vessels for use in arterial bypass graft procedures. Endoscopic Vessel Harvesting is a procedure that has been in use for approximately 15 years. The OnePass™ device differs in that it provides for single device entry into the body cavity without the need for additional instruments. Also, the distal end effector design allows for excellent visibility and quick thermal ligation of vessel trunks while minimizing trauma to the target vessel.
Intended Use:
The Saphena Medical OnePass Endoscopic Vessel Harvesting System is indicated for use in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. It is indicated for cutting tissue and controlling bleeding through coagulation, and for patients requiring blunt dissue including dissection of blood vessels, dissection of blood vessels of the extremities, dissection of ducts and other structures in the extra peritoneal or subcutaneous extremity and thoracic
{1}------------------------------------------------
space. Extremity procedures include tissue dissection along the saphenous vein for use in coronary artery bypass grafting and peripheral artery bypass or radial artery for use in coronary artery bypass grafting. Thorascopic procedures include exposure and dissection of structures external to the parietal pleura, including nerves, blood vessels, and other tissues of the chest wall.
Non-Clinical Performance Data:
The OnePass™ device performance characteristics were evaluated in the following studies:
- Design Verification Testing (including dimensional verification, tensile/compressive . strength and durability)
- Compatibility Testing with standard 7mm scope .
- . Compatibility Electrical Safety and Electromagnetic Testing
- Packaging and Shipping Testing .
- Sterility Testing .
- Biocompatibility Testing .
- Simulated animal testing use evaluation .
Performance testing demonstrates that the device is as safe and effective and performs at least as safely and effectively as the predicate device stated in this summary.
Statement of Equivalence:
The OnePass™ Endoscopic Vessel Harvesting System is substantially equivalent to the Maquet Cardiovascular, LLC VasoView 6 Pro Endoscopic Harvesting System (K091733) decision date of August 8, 2009. As described in this submission, the intended use of the OnePass™ Endoscopic Vessel Harvesting System is equal to the predicate device. Further, the subject and predicate device have similar technological characteristics. The OnePass™ Endoscopic Vessel Harvesting System is equivalent to the predicate device in terms of intended use, principles of operations, technology, design, materials and performance. The noted difference between the OnePass™ Endoscopic Vessel Harvesting and predicate device does not raise new issues of safety or effectiveness.Laboratory studies have been performed to assess the safety and effectiveness of the new characteristics. The testing performed in support of this 510(k) submission was conducted in accordance with accepted scientific methods, including compliance with ISO 60601-1 30 addition (Electrical Safety Testing for Medical Devices). Guidance documents and standards were followed as appropriate.
Summary:
Non-clinical testing of the subject device has shown that it meets its functional specification and performs according to the prescribed intended use. The OnePass™ Endoscopic Vessel Harvesting System is substantially equivalent to the predicate device in terms of its intended use, performance characteristics and product labeling. Based on the product technical information provided in this premarket notification, the OnePass™ Endoscopic Vessel Harvesting System has been shown to be substantially equivalent to the currently marketed predicate device.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle-like symbol with three curved lines representing wings or feathers. The symbol is positioned to the right of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA".
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 31, 2014
Saphena Medical Incorporated Ms. Cheryl Blake Regulatory Affairs/Quality Assurance. Saphena Medical Consultant 375 West Street West Bridgewater, Massachusetts 02379
Re: K140603
Trade/Device Name: OnePass Endoscopic Vessel Harvesting System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class H Product Code: GEI Dated: July 8, 2014 Received: July 9, 2014
Dear Ms. Blake:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract fiability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 - Ms. Cheryl Blake
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K140603
Device Name OnePass Endoscopic Vessel Harvesting System
Indications for Use (Describe)
The Saphena Medical OnePass Endoscopic Vessel Harvesting System is in minimally invasive surgery allowing access for vessel harvesting, and is primarily indicated for patients undergoing endoscopic surgery for arterial bypass. It is indicated for cutting tissue and controlling bleeding through coagulation, and for patients requiring blunt dissection of tissue including dissection of blood vessels, dissection of blood vessels of the extremities, dissection of ducts and other structures in the extra peritoneal or subcutaneous extremity and thoracic space. Extremity procedures include tissue dissection along the saphenous vein for use in coronary artery bypass grafting and peripheral artery bypass or radial artery for use in coronary artery bypass grafting. Thorascopic procedures include exposure and dissection of structures external to the parietal pleura, including nerves, blood vessels, and other tissues of the chest wall.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the The to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
§ 878.4400 Electrosurgical cutting and coagulation device and accessories.
(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.