K Number
K140479

Validate with FDA (Live)

Date Cleared
2014-07-28

(152 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NuVasive CoRoent XL-F System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft to facilitate fusion.

The CoRoent XL-F System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to L5, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar devices are to be used in patients who have had at least six months of non-operative treatment. The system is intended to be used with supplemental internal spinal fixation systems (e.g., pedicle or facet screws) that are cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

Device Description

The NuVasive CoRoent XL-F System is manufactured from PEEK-Optima® LT-1 conforming to ASTM F2026, MP35N conforming to ASTM F562, and titanium alloy conforming to ASTM F136 and ISO 5832-3. The implants are available in a variety of different shapes and sizes to suit the individual pathology and anatomical conditions of the patient.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the NuVasive CoRoent XL-F System, which is an intervertebral body fusion device. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving that the device meets specific acceptance criteria through a study with performance metrics in the typical sense of a novel medical AI/software device.

Therefore, the requested information elements related to performance criteria, sample sizes, ground truth establishment, expert involvement, and MRMC studies are not applicable to this document. The submission relies on non-clinical performance data (engineering analyses and cadaveric study) to demonstrate safety and effectiveness relative to existing devices.

Here's a breakdown of what can be extracted and why other elements are not applicable:

1. A table of acceptance criteria and the reported device performance

  • Not Applicable in the traditional sense. For a 510(k) for an intervertebral body fusion device, "acceptance criteria" are typically related to meeting established performance standards for mechanical properties and biocompatibility, as demonstrated through engineering analyses and material testing, rather than clinical performance metrics (like sensitivity, specificity, accuracy) for an AI/software device. The document states that the objective was to demonstrate substantial equivalence to predicate devices.
  • Reported Device Performance (based on non-clinical studies):
    • Axial Compression Finite Element Analysis: "Results demonstrate that the subject CoRoent XL-F System presents no new worst-case for performance testing."
    • Compression Shear Finite Element Analysis: "Results demonstrate that the subject CoRoent XL-F System presents no new worst-case for performance testing."
    • Wear Debris Analysis: No specific performance reported, but implied to be acceptable for substantial equivalence.
    • Subsidence Analysis: No specific performance reported, but implied to be acceptable for substantial equivalence.
    • Clinical Literature Analysis: Used for comparison, not direct performance measurement of the device itself.
    • Cadaveric Study: "Did not identify any new risks associated with the subject device."

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not Applicable. No test set in the context of an AI/software device performance study was used. The studies conducted were engineering analyses (Finite Element Analysis) and a cadaveric study.
    • For the cadaveric study, the sample size is not specified in the provided text.
    • Data provenance for FEA is theoretical modeling; for the cadaveric study, the source/provenance is not detailed.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not Applicable. This device is a physical implant, not an AI/software device that requires human expert annotation for ground truth establishment. Engineering analyses and cadaveric studies are evaluated by engineers and medical professionals specialized in those fields, but not in the context of "ground truth" for a test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not Applicable. There was no "test set" in the context of a diagnostic or predictive performance study, and therefore no adjudication method as described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is a physical intervertebral body fusion device, not an AI or imaging diagnostic tool. Therefore, an MRMC study is irrelevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This device does not contain software or electrical equipment, as explicitly stated in section F. It is a physical implant.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not Applicable in the typical sense of AI/software device. For this physical device, the "ground truth" for its safety and effectiveness is established through adherence to engineering standards (e.g., ASTM F2026, F562, F136, ISO 5832-3 for material composition), and the findings of the finite element analyses and cadaveric study which confirm it performs similarly to predicate devices without introducing new risks. Clinical literature analysis also contributes to supporting the established safety and effectiveness of similar devices.

8. The sample size for the training set

  • Not Applicable. This device does not involve a "training set" in the context of machine learning.

9. How the ground truth for the training set was established

  • Not Applicable. As there is no training set for machine learning involved with this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf-like shape on the left, followed by the word "NUVASIVE" in a sans-serif font. Below the word "NUVASIVE" is the tagline "Speed of Innovation" in a smaller font.

Traditional 510(k) Premarket Notification NuVasive® CoRoent® XL-F System

510(k) Summary

In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR $807.92, the following summary of information is provided:

Submitted by: A.

Cynthia Adams Regulatorv Affairs Associate NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 909-1800

Date Prepared: July 23, 2014

B. Device Name

NuVasive® CoRoent® XL-F System Trade or Proprietary Name: Common or Usual Name: Intervertebral Body Fusion Device Spinal Intervertebral Body Fixation orthosis Classification Name:

Device Class:Class II
Classification:21 CFR § 888.3080
Product Code:OVD

C. Predicate Devices

The subject NuVasive CoRoent XL-F System is substantially equivalent to the following predicate devices: NuVasive Single Tab System (K131723), NuVasive Brigade Hyperlordotic System (K123045), and Globus Medical InterContinental Plate-Spacer (K103382).

D. Device Description

The NuVasive CoRoent XL-F System is manufactured from PEEK-Optima® LT-1 conforming to ASTM F2026, MP35N conforming to ASTM F562, and titanium alloy conforming to ASTM F136 and ISO 5832-3. The implants are available in a variety of different shapes and sizes to suit the individual pathology and anatomical conditions of the patient.

{1}------------------------------------------------

E. Intended Use

The NuVasive CoRoent XL-F System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft to facilitate fusion.

The CoRoent XL-F System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to L5, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar devices are to be used in patients who have had at least six months of non-operative treatment. The system is intended to be used with supplemental internal spinal fixation systems (e.g., pedicle or facet screws) that are cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

Technological Characteristics F.

As was established in this submission, the subject CoRoent XL-F System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, and function. This device does not contain software or electrical equipment.

G. Performance Data

The following activities were performed to demonstrate that the subject CoRoent XL-F System is substantially equivalent to other predicate devices for the following:

  • Axial Compression Finite Element Analysis .
  • Compression Shear Finite Element Analysis .
  • Wear Debris Analysis .
  • Subsidence Analysis .
  • Clinical literature analysis .
  • Cadaveric study .

The results demonstrate that the subject CoRoent XL-F System presents no new worst-case for performance testing, and a cadaveric study did not identify any new risks associated with the subject device. The subject device was therefore found to be substantially equivalent to its predicate devices. No clinical studies were conducted.

H. Conclusions

Based on the indications for use, technological characteristics, and comparison to predicate devices, the subject CoRoent XL-F System has been shown to be substantially equivalent to legally marketed predicate devices.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 28, 2014

NuVasive, Incorporated Ms. Cynthia Adams Regulatory Affairs Associate 7475 Lusk Boulevard San Diego, California 92121

Rc: K140479

Trade/Device Name: NuVasive® CoRoent® XL-F System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD Dated: June 17, 2014 Received: June 18, 2014

Dear Ms. Adams:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Cynthia Adams

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Ronald P. Jean -S for

  • Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No, 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) KI40479

Device Name NuVasive® CoRoent® XL-F System

Indications for Use (Describe)

The NuVasive CoRoent XL-F System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft to facilitate fusion.

The CoRoent XL-F System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to L5, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar devices are to be used in patients who have had at least six months of non operative treatment. The used with supplemental internal spinal fixation systems (e.g., pedicle or facet screws) that are cleared by the FDA for use in the lumbar spine in addition to the integrated screws.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY . . . . . . . . . . . .. 11.15

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Katherine D. Kavlock, PhD

Division of Orthopedic Devices

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.1

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

N/A