The Medisystems Dual Lumen Needle is indicated for use as a vascular access device for dialysis procedures. The Medisystems Dual Lumen Buttonhole Needle is indicated fo use as a vascular access device for dialysis procedures using a constant-site or "buttonhole" method of needle insertion.
The proposed devices are to be used as a vascular access device for dialysis procedures using either a standard needle insertion technique (for Dual Lumen Needle) or a constant-site or "buttonhole" method of insertion (for Dual Lumen Buttonhole Needle).
This document is a 510(k) premarket notification for a medical device -- a dual lumen needle for dialysis. Based on the provided text, this is not an AI/ML medical device. Therefore, the questions related to AI/ML device performance criteria, training sets, expert review, MRMC studies, and ground truth establishment are not applicable to this document.
The document discusses the substantial equivalence of the proposed devices (Medisystems Dual Lumen Needle and Medisystems Dual Lumen Buttonhole Needle) to predicate devices already on the market (Medisystems Buttonhole Needles cleared under K990803 and Cordis Corp. Bionics Bi-Flo double lumen needle cleared under K801355).
The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to non-clinical bench testing for a physical medical device, not a software algorithm.
Here's the relevant information that can be extracted from the document:
1. A table of acceptance criteria and the reported device performance
The document broadly states:
Category | Acceptance Criteria | Reported Device Performance |
---|---|---|
Non-Clinical Test/Performance Testing - Bench | Predetermined acceptance criteria was met. | Information and data provided clearly describes the proposed devices and demonstrates that the devices are adequately designed for the labeled indications for use. Performance, verification and validation testing was conducted to characterize performance of the proposed devices. Results of this testing have documented that the proposed devices are substantially equivalent to the predicate devices and are suitable for the labeled indications for use. |
Note: The document itself does not provide the specific "predetermined acceptance criteria" or the detailed "results of this testing" for the bench tests. It only states that they were conducted and met.
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Sample Size: Not specified in the provided text. The document refers to "Performance, verification and validation testing," which implies a test set, but the size is not mentioned.
- Data Provenance: Not applicable in the context of this type of testing for a physical device. It would involve laboratory bench testing, not patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
- Not applicable. This is a physical medical device, not an AI/ML algorithm requiring expert annotation for ground truth.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not applicable. This relates to expert review of data for AI/ML, not bench testing of a physical device.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This is not an AI/ML device.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Not applicable. This is not an AI/ML device.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Not applicable. For a physical device, "ground truth" would relate to engineering specifications, material properties, and functional performance metrics established through bench testing, rather than medical "ground truth" like disease presence.
8. The sample size for the training set
- Not applicable. This is not an AI/ML device; there is no "training set" in the AI/ML sense.
9. How the ground truth for the training set was established
- Not applicable. Again, this is for AI/ML devices.
Summary based on the document:
The document serves as an FDA 510(k) clearance letter and summary for a physical medical device. It attests that "Performance, verification and validation testing was conducted to characterize performance of the proposed devices and the predetermined acceptance criteria was met." However, it does not provide the specifics of these tests, such as sample sizes, detailed criteria, or specific results, as this information would typically be contained in the full 510(k) submission, not the summary letter itself.
§ 876.5540 Blood access device and accessories.
(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.