K Number
K133743

Validate with FDA (Live)

Date Cleared
2014-04-07

(119 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Baxano Surgical MIS Pedicle Screw System is intended for posterior, non-cervical pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally-mature patients as an adjunct to fusion for the following indications: (1) Degenerative Disc Disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), (2) Spondylolisthesis, (3) Trauma (i.e. fracture or dislocation), (4) Spinal stenosis, (5) Curvatures (i.e. scoliosis, and/or lordosis), (6) Tumor, (7) Pseudoarthrosis and (8) Failed previous fusion.

Device Description

The Baxano Surgical MIS Pedicle Screw System is a multiple component system comprised of non-sterile, single-use implantable components fabricated from titanium alloy and cobalt chrome. When assembled, the components are implanted via a percutaneous, mini-open or open delivery to create a rigid structure to provide stabilization and promote spinal fusion. The system consists of an assortment of polyaxial cannulated pedicle screws, rods, cross connectors and locking caps, and associated instruments.

AI/ML Overview

This document describes the premarket notification (510(k)) for the Baxano Surgical MIS Pedicle Screw System, a medical device. The information provided focuses on the mechanical testing of the device to demonstrate its substantial equivalence to a predicate device, rather than a clinical study involving human readers or algorithm performance for image analysis.

Therefore, many of the requested categories related to clinical studies, human reader performance, ground truth establishment for images, and training sets for algorithms are not applicable to this submission.

Here is an analysis based on the provided document:

Acceptance Criteria and Device Performance (Mechanical Testing)

The acceptance criteria for the Baxano Surgical MIS Pedicle Screw System were established by applicable ASTM standards for spinal implant testing. The reported device performance demonstrated that it met or exceeded these requirements.

TestStandardAcceptance Criteria (Implied)Reported Device Performance
Static CompressionASTM F1717-13Met or exceeded requirements of ASTM F1717-13Met or exceeded requirements as established by test protocol and ASTM standards
Static TorsionASTM F1717-13Met or exceeded requirements of ASTM F1717-13Met or exceeded requirements as established by test protocol and ASTM standards
Dynamic CompressionASTM F1717-13Met or exceeded requirements of ASTM F1717-13Met or exceeded requirements as established by test protocol and ASTM standards
Axial PulloutASTM F543-07Met or exceeded requirements of ASTM F543-07Met or exceeded requirements as established by test protocol and ASTM standards
Torque to FailureASTM F543-07Met or exceeded requirements of ASTM F543-07Met or exceeded requirements as established by test protocol and ASTM standards
Axial GripASTM F1798-97Met or exceeded requirements of ASTM F1798-97Met or exceeded requirements as established by test protocol and ASTM standards
Torsional GripASTM F1798-97Met or exceeded requirements of ASTM F1798-97Met or exceeded requirements as established by test protocol and ASTM standards
Flexural GripASTM F1798-97Met or exceeded requirements of ASTM F1798-97Met or exceeded requirements as established by test protocol and ASTM standards

Study Information (Mechanical Testing)

This submission relies on non-clinical mechanical testing rather than a clinical study.

  1. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The document states that testing was performed on the "worst case subject device" and "components of the Baxano Surgical MIS Pedicle Screw System." It does not specify an exact numerical sample size for the mechanical tests, but implied that multiple components or configurations were tested to represent the worst-case scenario.
    • Data Provenance: The testing was conducted internally or by a contracted lab, as it refers to "test protocol." It is non-clinical, controlled laboratory testing.
  2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Not applicable. The "ground truth" for mechanical testing is defined by the performance standards (ASTM) and the physical properties of the materials and design, not by expert interpretation.
  3. Adjudication Method for the Test Set:

    • Not applicable. Adjudication methods like 2+1 or 3+1 are used for human expert disagreement in image interpretation or clinical diagnosis. Mechanical testing results are objective measurements against defined standards.
  4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No. This is a mechanical device submission, not an AI or image analysis device subject to MRMC studies.
  5. Standalone (Algorithm Only) Performance Study:

    • No. This is not an algorithm-based device.
  6. Type of Ground Truth Used:

    • The "ground truth" for this device's performance is the mechanical performance standards set by ASTM (e.g., F1717-13, F543-07, F1798-97).
  7. Sample Size for the Training Set:

    • Not applicable. There is no AI model or algorithm being trained for this medical device.
  8. How the Ground Truth for the Training Set was Established:

    • Not applicable. As there is no training set for an algorithm, there is no ground truth to establish for it.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for BAXANO SURGICAL. The logo is in black and white. The word "BAXANO" is in large, bold letters, and the word "SURGICAL" is in smaller letters below it. To the right of the word "BAXANO" is a symbol that looks like a circle with three lines coming out of it.

APR 0 7 2014

510(k) Summary Baxano Surgical MIS Pedicle Screw System

Submitter:Baxano Surgical, Inc.110 Horizon Drive, Ste 230Raleigh, NC 27615
Contact Person:Kristen AllenRegulatory Project Manager910-332-1733 (P)910-332-1701 (F)kristen.allen@baxsurg.com (e-mail)
Date Prepared:February 4, 2014
Trade Name:Baxano Surgical MIS Pedicle Screw System
Common Name:Spinal Pedicle Fixation Device
Device Product Codeand Classification:MNH, 888.3070, Class II, SpondylolisthesisSpinal Fixation Device SystemMNI, 888.3070, Class II, Pedicle ScrewSpinal SystemNKB, 888.3070, Class III, Pedicle Screw SpinalSystem, For Degenerative Disc Disease
Predicate Device:Life Spine® Conquest® Spinal System (K090320)CD Horizon® SOLERA™ Spinal System (K113174)

Device Description:

The Baxano Surgical MIS Pedicle Screw System is a multiple component system comprised of non-sterile, single-use implantable components fabricated from titanium alloy and cobalt chrome. When assembled, the components are implanted via a percutaneous, mini-open or open delivery to create a rigid structure to provide stabilization and promote spinal fusion. The system consists of an assortment of polyaxial cannulated pedicle screws, rods, cross connectors and locking caps, and associated instruments.

Indications and Intended use:

The Baxano Surgical MIS Pedicle Screw System is intended for posterior, non-cervical pedicle fixation of the spine to provide immobilization and stabilization of spinal segments in skeletally-mature patients as an adjunct to fusion for the following indications: (1) Degenerative Disc Disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), (2) Spondylolisthesis, (3) Trauma (i.e. fracture or dislocation), (4) Spinal stenosis, (5)

{1}------------------------------------------------

Curvatures (i.e. scoliosis, and/or lordosis), (6) Tumor, (7) Pseudoarthrosis and (8) Failed previous fusion.

Summary of Technological Characteristics:

The Baxano Surgical MIS Pedicle Screw System shares the same technological characteristics as the predicate device Life Spine Conquest® Spinal System, including design, materials, cleaning/sterilization process, range of sizes, performance characteristics and indication for use. The Baxano Surgical MIS Pedicle Screw System implant components are supplied non-sterile and are single-use. The Cannulated, selftapping pedicle screw implants are fabricated from Titanium alloy (Ti6Al4V ELI, ASTM F136) and Cobalt-chromium-molybdenum alloy (ASTM F1537). The subject device system also contains implantable rods (straight and lordosed), locking caps and cross connectors, which are fabricated from Titanium alloy (Ti6Al4V ELI, ASTM F136). The device provides correction and rigid stabilization of the spine during development of solid bone fusion following corrective spine surgery for a number of indications (listed above). The Baxano Surgical MIS Pedicle Screw System is designed to be implanted via a percutaneous, mini-open or open delivery method.

Summary of Performance Testing:

Non-clinical mechanical testing for the Baxano Surgical MIS Pedicle Screw System was performed on the worst case subject device in accordance with ASTM standards. Comparative testing was also performed including components of the Baxano Surgical MIS Pedicle Screw System and the predicate Conquest Pedicle Screw System.

TestStandard
Static CompressionASTM F1717-13
Static TorsionASTM F1717-13
Dynamic CompressionASTM F1717-13
Axial PulloutASTM F543-07
Torque to FailureASTM F543-07
Axial GripASTM F1798-97
Torsional GripASTM F1798-97
Flexural GripASTM F1798-97

For all test methods, the subject devices met or exceeded the requirements as established by the test protocol and applicable ASTM standards. The results demonstrated that the Subject device is substantially equivalent to the Predicate.

Substantial Equivalence:

Based on the comparison and performance testing analysis provided in this premarket notification submission, the Baxano Surgical MIS Pedicle Screw System has been shown to be substantially equivalent to the Life Spine® Spinal System Pedicle Screw System (K090320) in indications for use, design, materials used, and device functional scientific technology as the cleared predicate Conquest System.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002

April 7, 2014

Baxano Surgical, Incorporated Ms. Kristen Allen Regulatory Project Manager 301 Government Center Drive, Suite 100 Wilmington, North Carolina 28403

Re: K133743

Trade/Device Name: Baxano Surgical MIS Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNH, MNI Dated: February 4, 2014 Received: February 5, 2014

Dear Ms. Allen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

1

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Kristen Allen

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also. please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safetv/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Ronald P. Jean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K133743

Device Name

Baxano Surgical MIS Pedicle Screw System

Indications for Use (Describe)

The Baxano Surgical MIS Pedicle Screw System is intended for posterior, non-cervical pedicle fixation of the spine to provide immobilization and stabilization of spinal in skeletally-mature patients as an adjunct to fusion for the following indications: (1) Degenerative Dise Disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiogruphic studies), (2) Spondylolisthesis, (3) Trauma (i.e. fracture or dislocation), (4) Spinal stenosis, (5) Curvatures (i.e. scoliosis, kyphosis, and/or fordosis). (6) Tumor, (7) Pseudoarthrosis and (8) Failed previous fusion.

Type of Use (Select one or both, as applicable)

[ Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpant C)

PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)

买 -

Image /page/4/Picture/13 description: The image shows the name "James P. Bertram-S" in bold font on the top line. The second line shows the date and time as "2014.04.07 10:52:40 -04'00'". The date and time are also in bold font.

.. .. ... ... ................................................................................................................................................................

Form Approved: OMB No. 0910-0120

{5}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

N/A