K Number
K133727
Date Cleared
2014-01-16

(41 days)

Product Code
Regulation Number
888.3040
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Q-Fix Suture Anchor is intended to be used for soft tissue to bone fixation for:

Shoulder: Bankart lesion repair; SLAP lesion repair; acromio-clavicular repair; capsular shift/capsulolabral reconstruction; deltoid repair; rotator cuff tear repair; biceps tenodesis
Foot & Ankle: Medial/Lateral repair and reconstruction; midfoot and forefoot repair; Hallux valgus reconstruction; Metatarsal ligament/tendon repair or reconstruction; Achilles tendon repair
Elbow: Ulnar or radial collateral ligament reconstruction; lateral epicondylitis repair; biceps tendon reattachment
Knee: Extra-capsular repair: medial collateral ligament (MCL), lateral collateral ligament (LCL) and posterior oblique ligament; Iliotibial band tenodesis (IBT); patellar tendon repair; vastus medialis obliquus advancement (VMO); joint capsule closure
Hand & Wrist: Collateral ligament repair; Scapholunate ligament reconstruction; Tendon transfers in phalanx; Volar plate reconstruction
Hip: Acetabular labral repair

Device Description

The purpose of this subnission is to seek clearance for modifications and additions to the accessory instrumentation used with the previously cleared Q-Fix Suture Anchor System (K132513). There are no changes to the previously cleared indications for use, materials, design, technology, method of anchor insertion and or tissue attachment.

The Q-Fix Suture Anchor System (Q-Fix) is a bone anchor with inserter handle designed for use in arthroscopic and orthopedic procedures.

The Q-Fix consists of two primary parts: a bone anchor inserter, which is preloaded with the anchor. The anchor inserter is a disposable tool.

The entire product is packaged in a tray with a Tyvek® Iid, and the finished product is sterilized by ethylene oxide. Both the anchor and inserter are designed for single use only.

The O-Fix Suture Anchor System consists of the bone anchor and associated instruments for implanting the bone anchor.

AI/ML Overview

The Q-Fix™ Suture Anchor System is a medical device designed for soft tissue to bone fixation. The device underwent mechanical testing to demonstrate its safety and effectiveness.

1. Acceptance Criteria and Reported Device Performance:

The acceptance criteria are not explicitly quantifiable performance metrics in the provided text, but rather a demonstration of substantial equivalence to the predicate device and adherence to design, performance, and safety specifications based on mechanical testing.

Acceptance CriteriaReported Device Performance
Meets design, performance, and safety specifications"The test results demonstrate that the Q-Fix meets its design, performance, and safety specifications."
Performs as intended"Based on the test results, the proposed device performs as intended..."
Mechanical properties are substantially equivalent to the predicate device (ArthroCare Q-Fix™ Suture Anchor: K132513)"...mechanical properties are substantially equivalent to the predicate devices when used in accordance with labeling." Also, "All testing demonstrates that the Q-Fix performs as intended and has acceptable mechanical properties..."
Testing performed in accordance with FDA Guidance Document, Testing Bone Anchors, April 1996"Mechanical testing was performed in accordance with the requirements of the FDA Guidance Document, Testing Bone Anchor Devices, April 1996."

2. Sample Size and Data Provenance for Test Set:

  • Sample Size: Not explicitly stated. The text mentions "Bench testing was performed on both the proposed and predicate devices," implying multiple units were tested, but the exact number is not provided.
  • Data Provenance: The testing was "in vitro," involving insertion of anchors into a "simulated human bone substrate." This indicates the data is from laboratory bench testing, rather than from human or animal subjects. The origin (country) is not specified beyond "ArthroCare Corporation" which is located in Austin, Texas, USA. The testing is prospective as it was conducted specifically for this submission.

3. Number of Experts and Qualifications for Ground Truth:

Not applicable. This device is a mechanical implant, and its performance was assessed through bench testing, not through expert interpretation of data or images. Therefore, clinical experts were not required to establish ground truth for the device's mechanical performance.

4. Adjudication Method:

Not applicable, as no expert adjudication was involved for this type of mechanical bench testing.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No MRMC comparative effectiveness study was done. This type of study is typically used for diagnostic or screening devices where human readers interpret medical images or data. The Q-Fix™ Suture Anchor System is a surgical implant, and its performance is evaluated mechanically rather than through human interpretation.

6. Standalone (Algorithm Only) Performance:

Not applicable. The Q-Fix™ Suture Anchor System is a medical implant, not an AI algorithm. Its performance is inherent to its physical design and mechanical properties, not an algorithm.

7. Type of Ground Truth Used:

The ground truth for evaluating the device's performance was established through mechanical testing against predefined specifications and comparison to the predicate device. This involved "static and cyclic fatigue testing" in a "simulated human bone substrate."

8. Sample Size for Training Set:

Not applicable. As a medical implant, there is no "training set" in the context of machine learning or AI. The design and manufacturing process would involve internal testing and validation, but not a training set as understood in AI/ML.

9. How Ground Truth for Training Set was Established:

Not applicable, as there is no training set. The "ground truth" for the device's design and manufacturing would be based on engineering principles, material science, and regulatory standards.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for ArthroCare. The logo consists of a stylized wave-like design on the left and the word "ArthroCare" on the right. The wave-like design is made up of several curved lines, and the word "ArthroCare" is in a sans-serif font.

K133727

JAN 16 2014

510(k) Summary

ArthroCare® Corporation Q-Fix™ Suture Anchor System

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

General Information

Submitter Name:ArthroCare Corporation
Address7000 West William Cannon DriveAustin, TX 78735
Contact Person:Laura KasperowiczSr. Manager, Regulatory AffairsPhone: 949-585-2406Fax: 949-585-2401
Date Prepared:December 5, 2013

Device Name

Proprietary Name:Q-Fix™ Suture Anchor System
Common Name:Bone Anchor
Classification Name:Smooth or threaded metallic bone fixation fastener
Device Class:Class II
Product Code:MBI
CFR Section:21 CFR 888.3040

Predicate Device

ArthroCare Q-Fix™ Suture Anchor: K132513 (cleared September 19, 2013)

Description

The purpose of this subnission is to seek clearance for modifications and additions to the accessory instrumentation used with the previously cleared Q-Fix Suture Anchor System (K132513). There are no changes to the previously cleared indications for use, materials, design, technology, method of anchor insertion and or tissue attachment.

The Q-Fix Suture Anchor System (Q-Fix) is a bone anchor with inserter handle designed for use in arthroscopic and orthopedic procedures.

The Q-Fix consists of two primary parts: a bone anchor inserter, which is preloaded with the anchor. The anchor inserter is a disposable tool.

{1}------------------------------------------------

K133727

Image /page/1/Picture/1 description: The image shows the logo for ArthroCare. The logo consists of a stylized wave-like design on the left and the word "ArthroCare" on the right. The wave-like design is made up of several curved lines that are arranged in a circular pattern. The word "ArthroCare" is written in a bold, sans-serif font.

it

The entire product is packaged in a tray with a Tyvek® Iid, and the finished product is sterilized by ethylene oxide. Both the anchor and inserter are designed for single use only.

The O-Fix Suture Anchor System consists of the bone anchor and associated instruments for implanting the bone anchor. In accordance with the ArthroCare Product Development Process, testing was performed to demonstrate the proposed device is substantially equivalent to the predicate device. Mechanical testing was performed in accordance with the requirements of the FDA Guidance Document, Testing Bone Anchor Devices, April 1996.

Intended Use/Indications For Use

The O-Fix Suture Anchor is intended to be used for soft tissue to bone fixation for:

Shoulder:Bankart lesion repair; SLAP lesion repair; acromio-clavicular repair; capsularshift/capsulolabral reconstruction; deltoid repair; rotator cuff tear repair; bicepstenodesis
Foot & Ankle:Medial/Lateral repair and reconstruction; midfoot and forefoot repair; Halluxvalgus reconstruction; Metatarsal ligament/tendon repair or reconstruction;Achilles tendon repair
Elbow:Ulnar or radial collateral ligament reconstruction; lateral epicondylitis repair;biceps tendon reattachment
Knee:Extra-capsular repair: medial collateral ligament (MCL), lateral collateral ligament(LCL) and posterior oblique ligament; Iliotibial band tenodesis (IBT); patellartendon repair; vastus medialis obliquus advancement (VMO); joint capsule closure
Hand & Wrist:Collateral ligament repair; Scapholunate ligament reconstruction; Tendon transfersin phalanx; Volar plate reconstruction
Hip:Acetabular labral repair

Non-Clinical Data

Bench testing was performed on both the proposed and predicate devices in accordance with the FDA Guidance Document, Testing Bone Anchors, April 1996. This in vitro testing involved insertion of the anchors into a simulated human bone substrate followed by both static and cyclic fatigue testing.

The test results demonstrate that the Q-Fix meets its design, performance, and safety specifications. Based on the test results, the proposed device performs as intended and mechanical properties are substantially equivalent to the predicate devices when used in accordance with labeling.

Clinical Data

No clinical or animal data are included in this submission.

Summary

All testing demonstrates that the Q-Fix performs as intended and has acceptable mechanical properties when used in accordance with its labeling.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring. MD 20993-0002

January 16, 2014

ArthroCare Corporation Mr. Mitchell Dhority Vice President, Regulatory Affairs 7000 West William Cannon Drive, Building I Austin. Texas 78735

Re: K133727

Trade/Device Name: Q-Fix™ Suture Anchor System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: December 18, 2013 Received: December 19, 2013

Dear Mr. Dhority:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Mitchell Dhority

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable. the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Also. please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safetv/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometries/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.

Sincerely yours,

Ronald® Jean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K133727

Device Name: Q-Fix™ Suture Anchor System

Indications for Use:

:

The Q-Fix Suture Anchor is intended to be used for soft tissue to bone fixation for:

Shoulder:Bankart lesion repair; SLAP lesion repair; acromio-clavicular repair; capsularshift/capsulolabral reconstruction; deltoid repair; rotator cuff tear repair; bicepstenodesis
Foot & Ankle:Medial/Lateral repair and reconstruction; midfoot and forefoot repair; Halluxvalgus reconstruction; Metatarsal ligament/tendon repair or reconstruction;Achilles tendon repair
Elbow:Ulnar or radial collateral ligament reconstruction; lateral epicondylitis repair;biceps tendon reattachment
Knee:Extra-capsular repair: medial collateral ligament (MCL), lateral collateral ligament(LCL) and posterior oblique ligament; Iliotibial band tenodesis (IBT); patellartendon repair; vastus medialis obliquus advancement (VMO); joint capsule closure
Hand & Wrist:Collateral ligament repair; Scapholunate ligament reconstruction; Tendon transfersin phalanx; Volar plate reconstruction
Hip:Acetabular labral repair

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth司班rank -S

Division of Orthopedic Devices

Page 1 of ____________________________________________________________________________________________________________________________________________________________________

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.