(87 days)
The Presto Breast Biopsy Device is intended for diagnostic sampling of breast biopsy procedures. It is to be used for diagnostic purposes only and is not intended for therapeutic uses.
The Presto Breast Biopsy Device is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.
The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
The Presto Breast Biopsy Device is a sterile, single-use percutaneous biopsy device. The working end of the device includes a stainless steel coring cannula with a razor edge and a stationary coil located within the coring cannula. The handle of the device contains an actuation button, a sample collection chamber, a drive mechanism for rotating the coring cannula and a DC power jack for a 12V input. A reusable, medical grade AC/DC power supply provides 12V to the disposable device. Depressing the button on the handle rotates the coring cannula - allowing the operator to core and transport tissue samples to the collection chamber. The device is used with a coaxial introducer. Actuating the partoff button mechanically adjusts the distal end of the coring cannula between a coring & partoff configuration.
Here's an analysis of the acceptance criteria and study information for the Presto Breast Biopsy Device, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
Test ID | Risk | Acceptance Criteria | Reported Device Performance (Result) |
---|---|---|---|
1 | Cannula buckles during use | Tube must withstand >15lbf | Pass |
2 | Tissue not cored/transported | Mass of 5 samples > 0.13g | Pass |
3 | Weld breaks | Weld must withstand >15lbf | Pass |
4 | Partoff tab breaks | Tab must last >30 cycles | Pass |
5 | Cytotoxicity (biocompatibility) | None of the cell cultures exposed to the test sample shall show greater than mild reactivity defined as: ≤ 50% of cells round, devoid of intracytoplasmic granules; no extensive cell lysis; and ≤ 50% growth inhibition present. Positive and negative control samples must demonstrate test system suitability. | Pass (discrete granules, no lysis, no reduction of growth) |
6 | Sensitization (biocompatibility) | The material will be considered acceptable if it has an overall grade of "0" (no visible change) or "1" (discrete or patchy erythema). The reagent blank must show no sensitization reaction. | Pass (grade "0", no sensitization reaction) |
7 | Irritation or Intracutaneous reactivity (biocompatibility) | The sample will be considered a non-irritant if the difference between the test extracts and the corresponding control mean score is 1.0 or less. | Pass (overall mean difference (test article - reagent control) was 0.0 for all extracts) |
8 | Systemic Toxicity (biocompatibility) | The sample will be considered non-toxic if all of the following conditions are met: a) none of the test extract animals exhibit a significantly greater reaction than the corresponding control animals; b) no more than one animal dies; c) no more than 1 animal displays abnormal behavior such as convulsions or prostration, and c) no more than 2 animals display a body weight loss > 2 grams. | Pass (no mortality, morbidity or weight loss observed) |
9 | Hemocompatibility (biocompatibility) | The test article shall have a hemolytic index below 2% (nonhemolytic). Positive and negative control samples must demonstrate test system suitability (the negative control must have a blank corrected % hemolysis value 30 times | Pass |
13 | Introducer does not engage w/ device | Device must operate with introducer attached | Pass |
14 | Partoff tab actuated prematurely | Mass of 5 samples > 0.13g | Pass |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state a specific "test set" in the context of a clinical study or human performance evaluation. The performance data presented are from non-clinical studies.
-
Sample sizes mentioned for non-clinical tests:
- Test #2 (Tissue not cored/transported): "Mass of 5 samples > 0.13g"
- Test #8 (Systemic Toxicity): "none of the test extract animals exhibit a significantly greater reaction than the corresponding control animals; b) no more than one animal dies; c) no more than 1 animal displays abnormal behavior... and c) no more than 2 animals display a body weight loss > 2 grams."
- Test #10 (Pyrogenicity): "each of the three animals does not experience a temperature rise ≥ 0.5°C"
- Test #14 (Partoff tab actuated prematurely): "Mass of 5 samples > 0.13g"
-
Data Provenance: The studies are described as "non-clinical performance data" and include "ex-vivo device performance," "tensile strength & fatigue," "biocompatibility," and "simulated use testing." This indicates the data is from laboratory and bench testing, rather than retrospective or prospective human data from a specific country of origin.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
Not applicable. The reported studies are non-clinical (bench testing, simulated use, biocompatibility). There is no mention of a "test set" requiring expert-established ground truth in a clinical context.
4. Adjudication Method for the Test Set
Not applicable. As the studies are non-clinical and do not involve human readers or interpretation of medical images/data, no adjudication method is mentioned or required.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The document focuses solely on non-clinical performance data and a substantial equivalence comparison to predicate devices, not on human reader performance with or without AI assistance.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Study was Done
No, a standalone algorithm performance study was not mentioned. The device described is a physical medical instrument (a breast biopsy device), not an artificial intelligence algorithm. The device's performance is evaluated based on its mechanical function and material properties.
7. The Type of Ground Truth Used
For the non-clinical studies described, the "ground truth" is established by:
- Engineering specifications and measurements: For mechanical tests like compression, weld force, and cycling tests (e.g., tube must withstand >15lbf, mass of samples > 0.13g, tab must last >30 cycles).
- Standardized biological/chemical assays: For biocompatibility tests, following ISO 10993 standards (e.g., cell reactivity, sensitization, irritation scores, hemolytic index, temperature rises in animal models).
This is not a clinical "ground truth" derived from expert consensus, pathology, or outcomes data, as it is a non-clinical device evaluation.
8. The Sample Size for the Training Set
Not applicable. This document describes the evaluation of a physical medical device (biopsy device), not an AI algorithm that would require a training set.
9. How the Ground Truth for the Training Set was Established
Not applicable. As there is no AI algorithm or training set discussed, there is no mention of how ground truth was established for such a set.
§ 876.1075 Gastroenterology-urology biopsy instrument.
(a)
Identification. A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification. (1) Class II (performance standards).(2) Class I for the biopsy forceps cover and the non-electric biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.