K Number
K133606
Date Cleared
2014-04-01

(127 days)

Product Code
Regulation Number
888.3040
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The N8TIVE™ ACL Anatomic Reconstruction System is indicated for use in the fixation of soft tissue grafts to bone during anterior cruciate ligament reconstruction surgeries of the knee.

Device Description

The N8TIVE ACL Reconstruction System is designed for ACL repair and consists of the N8TIVE ACL Femoral and Tibial Implants, as well as various Class I ancillary devices, all of which may be marketed or used individually or in conjunction with one another. The N8TIVE ACL System provides fixation of a soft-tissue graft within the femoral and tibial tunnels to reconstruct the anterior cruciate ligament and is designed to create femoral and tibial tunnels that are "figure-eight" or oval shaped.

The N8TIVE ACL Femoral Implant consists of a polyether etherketone (PEEK) spacer, an ultra-high molecular weight polyethylene (UHMWPE) graft suspension loop, UHMWPE suture tape, and a titanium cortical button. The suspension loop is threaded through the titanium cortical button and the PEEK spacer and forms a symmetrical graft suspension construct. The suspension loop is tightened by pulling each "tail" of the suture to remove the slack and adjust the construct to match the tunnel length. The UHMWPE suture tape is provided to secure the graft to the spacer so that it remains seated during insertion into the bone tunnel.

The N8TIVE ACL Tibial Implant consists of a PEEK screw, a PEEK sheath, and an UHMWPE tether suture pre-threaded through the screw and through an optional titanium cortical button. The Tibial Implant is available in two sizes in order to accommodate specific patient anatomy: a 29 mm length screw with a 33 mm length sheath for smaller tibial anatomy and a 36 mm screw with a 40 mm sheath for larger tibial anatomy.

Both N8TIVE ACL Implants are supplied preloaded onto an inserter handle and are provided sterile. for single patient use only.

The N8TIVE ACL Implants may be marketed and/or used in conjunction with various Class I ancillary instruments. These instruments include both sterile/reusable instrumentation for use in the ACL reconstruction procedure.

AI/ML Overview

The provided text describes a 510(k) summary for the ArthroCare® Corporation N8TIVE™ ACL Anatomic Reconstruction System. This document focuses on demonstrating substantial equivalence to predicate devices based on non-clinical (bench) testing, rather than device performance against specific acceptance criteria the way a clinical study would.

Here’s a breakdown of the information requested, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state numerical acceptance criteria for device performance. Instead, it relies on comparative testing to predicate devices and conformance to design and performance specifications.

Acceptance Criteria (Implied)Reported Device Performance
Magnetic Resonance Imaging compatibilityTesting conducted; results confirm compatibility.
Mechanical performance (cyclic and static loading) comparable to predicate devicesComparative testing conducted in a simulated human bone substrate; "All test results confirm that the N8TIVE ACL Reconstruction System is substantially equivalent to the predicate devices."
Insertion strength in dense bone modelInsertion strength testing conducted in a dense bone model; results confirm that the device meets its design, performance, and safety specifications.
Cyclic evaluation of the interface between UHMWPE suture tape and graft analogCyclic evaluation conducted; results confirm that the device meets its design, performance, and safety specifications.
Conformance to design and performance specificationsDesign Verification testing conducted; "All test results confirm that the N8TIVE ACL Reconstruction System is substantially equivalent to the predicate devices. Additionally, the test results confirm that the devices meet their design, performance, and safety specifications."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify the exact sample sizes (e.g., number of devices, specimens) used for each bench test. It indicates that "bench testing was conducted."
  • Data Provenance: The data is from non-clinical bench testing. The country of origin is not specified but implicitly would be related to the submitter, ArthroCare Corporation, located in Austin, TX, USA. The data is not retrospective or prospective in a clinical sense, as it’s bench testing.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This question is not applicable to the provided document. The "ground truth" for this submission is based on engineering specifications, material science, and comparison to legally marketed predicate devices, not on expert clinical interpretation of data from a test set.

4. Adjudication Method for the Test Set

This question is not applicable. There was no clinical test set requiring adjudication in the context of this 510(k) submission.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. This 510(k) submission is for a medical device (ACL reconstruction system), not an AI-powered diagnostic or assistive technology. No MRMC study was conducted or mentioned.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This question is not applicable. This is not an algorithm or AI-based device.

7. The Type of Ground Truth Used

The "ground truth" for this 510(k) submission refers to:

  • Established material properties and standards for medical implants.
  • The performance characteristics of the predicate devices.
  • Engineering design and performance specifications.

The testing conducted aims to demonstrate that the N8TIVE ACL System meets these engineering and comparative standards.

8. The Sample Size for the Training Set

This question is not applicable. This is a medical device, not a machine learning algorithm requiring a training set.

9. How the Ground Truth for the Training Set was Established

This question is not applicable. As noted above, this is not a machine learning algorithm.

{0}------------------------------------------------

K133606 (page 1/4)

Image /page/0/Picture/1 description: The image shows the logo for ArthroCare. The logo consists of a stylized, circular graphic to the left of the text "ArthroCare". The graphic is made up of several curved lines that overlap and intersect to create a sense of movement and flow. The text "ArthroCare" is written in a bold, sans-serif font, with the "A" slightly larger than the other letters.

APR 0 1 2014

510(k) Summary

ArthroCare® Corporation N8TIVETM ACL Anatomic Reconstruction System

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

General Information

Submitter Name:ArthroCare Corporation
Address7000 West William Cannon DriveAustin, TX 78735
Contact Person:Laurie CartwrightSr. Manager, Regulatory AffairsPhone: 949-585-2465Fax: 512-895-1489
Date Prepared:31 March 2014

Device Name

Proprietary Name:N8TIVE™ ACL Anatomic Reconstruction SystemN8TIVE™ ACL Femoral ImplantN8TIVE™ ACL Tibial Implant
Common Name:Fastener, Fixation, Nondegradable, Soft Tissue
Classification Name:Smooth or Threaded Metallic Bone Fixation Fastener
Device Class:II
Product Code:MBI
CFR Section:888.3040

Predicate Devices

The predicate devices for the N8TIVE ACL Anatomic Reconstruction System are:

  • For the N8TIVE ACL Femoral Implant, the Biomet ToggleLoc Femoral Fixation Device cleared under K083070 and K130033.
  • . For the N8TIVE ACL Tibial Implant, the Arthrex Interference Screws cleared under K052607 and K062466.

The tables on the following page provide a comparison of the technological characteristics for the N8TIVE ACL Implants and the predicate devices.

{1}------------------------------------------------

:

.

Image /page/1/Picture/1 description: The image shows the logo for ArthroCare. The logo consists of a stylized graphic on the left and the word "ArthroCare" on the right. The graphic is made up of several curved lines that are arranged in a circular pattern. The word "ArthroCare" is written in a bold, sans-serif font.

Comparison of Device Characteristics:N8TIVE ACL Femoral Implant and Biomet ToggleLoc Femoral Fixation Device
CharacteristicsBiomet ToggleLocK083070, K130033N8TIVE ACL Femoral ImplantProposed Device
Intended UseFixation of Soft Tissue to BoneSame
How SuppliedSterileSame
Sterilization MethodEthylene OxideSame
Implant MaterialsSame
• Cortical ButtonTitanium(TI-6AL-4V Alloy)Same
• Suspension LoopUHMWPE/Polypropylene/PolyesterUHMWPE/Polyester
• Suture/Suture TapeNot ApplicableUHMWPE
• Graft SpacerNot ApplicablePEEK Optima®
Method of InsertionInserted into drilled hole/tunnelSame
Design TechnologyGraft Suspension Loop with Cortical ButtonGraft Spacer and Suspension Loop withCortical Button
Implant Size/Dimensions
• Cortical Button13.2 mm x 2.0 mm x 1.8 mm12.1 mm x 3.9 mm x 1.8 mm
• Suspension LoopAdjustableSame
• Graft SpacerNot Applicable23 mm x 6 mm x 6 mm
Comparison of Device Characteristics:N8TIVE ACL Tibial Implant and Arthrex Interference Screws
CharacteristicsArthrex Interference ScrewsK052607, K062466N8TIVE Tibial ImplantProposed Device
Intended UseFixation of Soft Tissue to BoneSame
How SuppliedSterileSame
Sterilization MethodEthylene OxideSame
Implant Materials
• Cortical ButtonNot ApplicableTitanium(TI-6AL-4V Alloy)
• ScrewPEEK OptimaSame
• SheathNot ApplicablePEEK Optima
• Tether SuturesNot ApplicableUHMWPE
Method of InsertionScrewed into drilled hole/tunnel usingreusable driverSame, using included inserter driver
Design TechnologyInterference ScrewSame
Implant Size/Dimensions
• Cortical ButtonNot Applicable18 mm x 11.5 mm x 6.1 mm
• ScrewRange of 23 to 35 mm length,6 to 12 mm diameterRange of 29 to 36 mm length,8 mm diameter
• SheathNot ApplicableRange of 33 to 40 mm length

:

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for ArthroCare. The logo consists of a stylized wave-like design on the left, followed by the word "ArthroCare" in a bold, sans-serif font. The wave-like design is made up of several curved lines that create a sense of motion and flow.

Description

The N8TIVE ACL Reconstruction System is designed for ACL repair and consists of the N8TIVE ACL Femoral and Tibial Implants, as well as various Class I ancillary devices, all of which may be marketed or used individually or in conjunction with one another. The N8TIVE ACL System provides fixation of a soft-tissue graft within the femoral and tibial tunnels to reconstruct the anterior cruciate ligament and is designed to create femoral and tibial tunnels that are "figure-eight" or oval shaped.

The N8TIVE ACL Femoral Implant consists of a polyether etherketone (PEEK) spacer, an ultra-high molecular weight polyethylene (UHMWPE) graft suspension loop, UHMWPE suture tape, and a titanium cortical button. The suspension loop is threaded through the titanium cortical button and the PEEK spacer and forms a symmetrical graft suspension construct. The suspension loop is tightened by pulling each "tail" of the suture to remove the slack and adjust the construct to match the tunnel length. The UHMWPE suture tape is provided to secure the graft to the spacer so that it remains seated during insertion into the bone tunnel.

The N8TIVE ACL Tibial Implant consists of a PEEK screw, a PEEK sheath, and an UHMWPE tether suture pre-threaded through the screw and through an optional titanium cortical button. The Tibial Implant is available in two sizes in order to accommodate specific patient anatomy: a 29 mm length screw with a 33 mm length sheath for smaller tibial anatomy and a 36 mm screw with a 40 mm sheath for larger tibial anatomy.

Both N8TIVE ACL Implants are supplied preloaded onto an inserter handle and are provided sterile. for single patient use only.

The N8TIVE ACL Implants may be marketed and/or used in conjunction with various Class I ancillary instruments. These instruments include both sterile/reusable instrumentation for use in the ACL reconstruction procedure.

Indications For Use

The N8TIVE™ ACL Anatomic Reconstruction System is indicated for use in the fixation of soft tissue grafts to bone during anterior cruciate ligament reconstruction surgeries of the knee.

The indications for use for the N8TIVE ACL System are comprised of a subset of the indications for use cleared for the predicate devices. While the N8TIVE ACL System is specifically designed and intended for use in anterior cruciate ligament reconstruction, the predicate devices may additionally be used for treatment in other anatomic locations. The differences between the predicate device indications for use and the N8TIVE ACL System indications for use do not affect the safety or effectiveness of the N8TIVE ACL System when used in accordance with its labeling.

Non-Clinical Data

Bench testing was conducted as follows:

  • Magnetic Resonance Imaging compatibility testing.
  • . Comparative testing of the proposed and predicate devices in which the proposed and predicate devices were inserted into a simulated human bone substrate and subjected to cyclic and static loading.
  • Insertion strength testing in a dense bone model.
  • Cyclic evaluation of the interface between the UHMWPE suture tape and a graft analog.

{3}------------------------------------------------

iv

Image /page/3/Picture/1 description: The image shows the logo for ArthroCare. The logo consists of a stylized wave-like design on the left, followed by the word "ArthroCare" in a bold, italicized font. A small trademark symbol is visible to the right of the word "Care".

  • 1 Design Verification testing to demonstrate conformance to design and performance specifications.
    All test results confirm that the N8TIVE ACL Reconstruction System is substantially equivalent to the predicate devices. Additionally, the test results confirm that the devices meet their design, performance, and safety specifications.

Clinical Data

No clinical or animal data are included in this submission.

Summary

All testing demonstrates that the N8TIVE ACL Anatomic System performs as intended and has acceptable mechanical properties when used in accordance with the device labeling.

As the intended use, operating principle, materials, and technological characteristics are comparable to the predicate devices, ArthroCare believes the N8TIVE ACL Anatomic Reconstruction System is substantially equivalent to the predicate devices. The minor differences between the ArthroCare devices and their predicate devices do not raise any new questions of safety or effectiveness.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three curved lines representing its wings and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue

Document Control Center - WO66-G609 Silver Spring MD 20993-0002

Public Health Service

April 1, 2014

Arthrocare® Corporation Mr. Mitchell Dhority Vice President, Regulatory Affairs 7000 West William Cannon Drive Building One Austin, Texas 78735

Re: K133606

Trade/Device Name: N8TIVE™ ACL Anatomic Reconstruction System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI, JDR Dated: March 19, 2014 Received: March 20, 2014

Dear Mr. Dhority:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{5}------------------------------------------------

Page 2 - Mr. Mitchell Dhority

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Lori A. Wiggins

  • for
    Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K133606

Device Name

The N8TIVE™ ACL Anatomic Reconstruction System

Indications for Use (Describe)

The N8TVE™ ACL Abatomic Reconstruction System is indicated for use in the fixation of soft tissue grafts to bone during antenor cruciate ligament reconstruction surgeries of the knee.

Type of Use (Select one or both, as applicable) 図 Prescription Use (Part 21 CFR 801 Subpart D) Over - The Counter Use (21 CFR 801 Subpart C) PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. PORTED TO CHARM CONTRACT FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR F Concurrence of Center for Devices and Radiological Heatth (CDRH) (Signature) Page (1/1) asev I Carre Devices Division O FORM FDA 3881 (1/14) Page -1 - of 2

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement on last page.

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.