K Number
K133549

Validate with FDA (Live)

Date Cleared
2014-08-27

(281 days)

Product Code
Regulation Number
880.2800
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SteriTec Biological Test Pack with Instant Readout Integrator is a single-use device designed specifically for testing of 3 or 4 minutes at 132°C (270°F) and 3 minutes at 135°C (275°F) in pre-vacuum steam sterilizers. The test pack consists of a selfcontained biological indicator containing Geobacillus stearothermophilus inside a small package of porous and nonporous materials, simulating the biological indicator 16 towel test packs defined by AAMI ST79. When used as directed, the Chemical Indicator Record Card gives visible indication that sterilizing conditions were met.

Device Description

The test pack consists of a selfcontained biological indicator containing Geobacillus stearothermophilus inside a small package of porous and nonporous materials, simulating the biological indicator 16 towel test packs defined by AAMI ST79. When used as directed, the Chemical Indicator Record Card gives visible indication that sterilizing conditions were met.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device called the "SteriTec Biological Test Pack with Instant Readout Integrator, Model No.: B1 111." This device is a biological indicator designed for testing pre-vacuum steam sterilizers.

However, the document is a regulatory clearance letter, not a scientific study report. It states that the device is "substantially equivalent" to legally marketed predicate devices.

Therefore, the detailed information requested regarding acceptance criteria, device performance, study design, sample sizes, ground truth establishment, expert involvement, and comparative effectiveness studies (MRMC) is not available within this document. This letter does not provide the results of any clinical or performance studies.

To answer your questions comprehensively, information from the actual device validation study would be needed, which is not included in this FDA 510(k) clearance letter.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three profiles facing right, symbolizing service to the people. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle emblem.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 27, 2014

Steritec Products Manufacturing, Company Incorporated. Jonathan Rutigliano 74 Inverness Drive East Englewood, CO 80112 US

Re: K133549

Trade/Device Name: Steritec biological test pack with instant readout integrator, Model No.: BI 111 Regulation Number: 21 CFR 880.2800 Regulation Name: Biological Test Pack Regulatory Class: II Product Code: FRC Dated: July 31, 2014 Received: August 1, 2014

Dear Jonathan Rutigliano:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Erin I. Keith -S

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known): _ K133549

Device Name: SteriTec Biological Test Pack with Instant Readout Integrator, ModelNo.: Bl 111

Indications For Use:

SteriTec Biological Test Pack with Instant Readout Integrator is a single-use device designed specifically for testing of 3 or 4 minutes at 132°C (270°F) and 3 minutes at 135°C (275°F) in pre-vacuum steam sterilizers. The test pack consists of a selfcontained biological indicator containing Geobacillus stearothermophilus inside a small package of porous and nonporous materials, simulating the biological indicator 16 towel test packs defined by AAMI ST79. When used as directed, the Chemical Indicator Record Card gives visible indication that sterilizing conditions were met.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).