(70 days)
The K7 Lumbar Spacers are indicated for intervertebral body fusion of the lumbar spine, from L2 to S1, in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and with autograft to facilitate fusion.
The K7 Lumbar Spacers are a collection of radiolucent interbody devices having variously shaped cross-sections. The superior and inferior surfaces are open with parallel serrations to facilitate implant stability. The implants are available in an assortment of height, length, width and anteroposterior angulation combinations to accommodate a variety of anatomic requirements.
The provided text describes the K7 Lumbar Spacers, a medical device for intervertebral body fusion. The study presented focuses on demonstrating substantial equivalence to predicate devices through mechanical testing, rather than a clinical study evaluating the device's performance in a human population.
Therefore, many of the requested categories related to clinical performance metrics, ground truth establishment, expert involvement, and reader studies are not applicable or cannot be extracted from this document, as the FDA 510(k) submission for this device type typically relies on demonstrating engineering and material equivalence for clearance.
Here's a breakdown of the information that can be extracted, and where the requested information is not applicable:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (from predicate devices / standards) | Reported Device Performance (K7 Lumbar Spacers) |
|---|---|
| Mechanical Testing: - Static Compression (according to ASTM F2077) - Dynamic Compression (according to ASTM F2077) | Performance shown to be substantially equivalent to predicate devices. |
| Subsidence Properties: - Evaluated according to ASTM F2267 | Performance shown to be substantially equivalent to predicate devices. |
| Technological Characteristics: - Intended Use - Basic Design (hollow column) - Material (PEEK polymer and tantalum) - Sizes (widths, lengths, heights within predicate range) | Same as predicate devices. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated as a "test set" in the context of clinical data. The mechanical testing was likely performed on a representative sample of K7 Lumbar Spacers of the "worst case" configuration. The specific number of units tested for static, dynamic compression, and subsidence is not provided.
- Data Provenance: The data is from mechanical testing of the device itself (K7 Lumbar Spacers) in a laboratory setting, not from human subjects. Therefore, country of origin or retrospective/prospective clinical data is not applicable.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- This is not applicable. The "ground truth" for this device's regulatory clearance is established by engineering standards (ASTM F2077, ASTM F2267) and comparison to predicate device characteristics, not by expert clinical consensus on a "test set" of patient data.
4. Adjudication Method for the Test Set
- This is not applicable, as there was no clinical "test set" involving human interpretation or adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, an MRMC comparative effectiveness study was not done. This device is a spinal implant, and its regulatory clearance here is based on mechanical performance and material equivalence, not on the interpretative performance of a diagnostic algorithm or image analysis by human readers.
6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study
- No, a standalone study in the context of AI algorithm performance was not done. This device is a physical implant, not an AI algorithm.
7. Type of Ground Truth Used
- For the mechanical performance, the "ground truth" is defined by engineering standards (ASTM F2077 for compression, ASTM F2267 for subsidence) and the established performance/characteristics of legally marketed predicate devices. The goal was to demonstrate that the K7 Lumbar Spacers met or were equivalent to these benchmarks.
8. Sample Size for the Training Set
- This is not applicable. There is no "training set" in the context of machine learning or AI as this is a physical medical device.
9. How the Ground Truth for the Training Set Was Established
- This is not applicable. See point 8.
{0}------------------------------------------------
:
.
| 510(k) Summary | ||
|---|---|---|
| Date: | 26 September 2013 | DEC - 9 2013 |
| Sponsor: | K7 LLC54 Moonrise WayHenderson, NV 89074Phone: 817.219.4441Facsimile: 817.326.5524 | |
| Contact Person: | Michael D. Smith, Manager | |
| Trade Names: | K7 Lumbar Spacers | |
| Device Classification | Class II | |
| Classification Name: | Intervertebral fusion device with bone graft, lumbar | |
| Regulation: | 888.3080 | |
| Device ProductCode: | MAX | |
| Device Description: | The K7 Lumbar Spacers are a collection of radiolucent interbodydevices having variously shaped cross-sections. The superior andinferior surfaces are open with parallel serrations to facilitate implantstability. The implants are available in an assortment of height,length, width and anteroposterior angulation combinations toaccommodate a variety of anatomic requirements. | |
| Intended Use: | The K7 Lumbar Spacers are indicated for intervertebral body fusionof the lumbar spine, from L2 to S1, in skeletally mature patients whohave had six months of non-operative treatment. The device isintended for use at either one level or two contiguous levels for thetreatment of degenerative disc disease (DDD) with up to Grade Ispondylolisthesis or retrolisthesis. DDD is defined as back pain ofdiscogenic origin with degeneration of the disc confirmed by historyand radiographic studies. The device is intended for use withsupplemental fixation and with autograft to facilitate fusion | |
| Materials: | K7 Lumbar Spacers are manufactured from polyetheretherketone(PEEK) per ASTM F2026 (VESTAKEEP® i4 R, Evonik PolymersTechnologies GmbH). Integral marker pins are manufactured fromtantalum according to ASTM F560. | |
| Predicate Devices: | Eminent Spine, Eminent Spine Interbody Fusion System (K090064)DePuy AcroMed, Lumbar I/F Cage® (P960025)DePuy Spine, Cougar LS Lateral Cage System (K090899, K110454)Choice Spine LP, ORIA Natura® (K073669)K2M, Aleutian® IBF System (K082698, K101302 and K110843)Stryker Spine, AVS® PEEK Spacers (K073470, K082014, K101051)Icotec, ETurn Spinal Implant (K100305) |
.
. . . .
.
.
. :
:
{1}------------------------------------------------
Performance Data: Mechanical testing of the worst case K7 Lumbar Spacer was performed according to ASTM F2077 and included static and dynamic compression. The subsidence properties were evaluated according to ASTM F2267. The mechanical test results demonstrate that the K7 Lumbar Spacers performance is substantially equivalent to the predicate devices. Technological The K7 Lumbar Spacers possess the same technological Characteristics: characteristics as the predicate devices. These include: intended use (as described above), . . basic design (hollow column), material (PEEK polymer and tantalum), and . sizes (widths, lengths and heights are within the range(s) . offered by the predicate systems).
Therefore the fundamental scientific technology of the K7 Lumbar Spacers is the same as previously cleared devices.
The K7 Lumbar Spacers possess the same intended use and technological characteristics as the predicate devices. Therefore the K7 Lumbar Spacer system is substantially equivalent for its intended use.
Conclusion:
Page 119
{2}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 9. 2013
K7. LLC % BackRoads Consulting, Incorporated Karen E. Warden. Ph.D. 8202 Sherman Road Chesterland, Ohio 44026
Re: K133126
Trade/Device Name: K7 Lumbar Spacers Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: September 26, 2013 Received: September 30, 2013
Dear Dr. Warden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Karen E. Warden, Ph.D.
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda,gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Ronalde Jean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Section 7 - Indications for Use Statement
510(k) Number:_K133126
Device Name: K7 Lumbar Spacers
Indications for Use:
The K7 Lumbar Spacers are indicated for intervertebral body fusion of the lumbar spine, from L2 to S1. in skeletally mature patients who have had six months of non-operative treatment. The device is intended for use at either one level or two contiguous levels for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The device is intended for use with supplemental fixation and with autograft to facilitate fusion.
Prescription Use X OR Over-the-Counter Use_ ________________________________________________________________________________________________________________________________________________________
(Per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anton E. Dmitriev, PhD Division of Orthopedic Devices
§ 888.3080 Intervertebral body fusion device.
(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.