K Number
K133045

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-04-10

(195 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Element™ Lite Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the fingertips, ventral palm, dorsal hand, upper arm, forearm, calf and thigh. The Element™ Lite Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared.

The Element™ Lite Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

The Element™ Lite Test Strips are for use with the Element™ Lite Meter to quantitatively measure glucose in fresh capillary whole blood. Fresh capillary whole blood samples may be drawn from the fingertips, ventral palm, dorsal hand, upper arm, forearm, calf and thigh.

The Element™ Lite Control Solutions are for use with the Element™ Lite Meter and Element™ Lite Test Strips to check that the meter and test strips are working together properly and the test is performing correctly.

Device Description

The Element™ Lite Blood Glucose Monitoring System consists of a meter, test strips and control solutions (level 1, level 2 and level 3), a lancing device and sterile lancets. This blood glucose test system is an in vitro diagnostic device designed for measuring the concentration of glucose in whole blood sample by means of an electrical current produced in the test strip and sent to the meter for measurement.

AI/ML Overview

The Element™ Lite Blood Glucose Monitoring System underwent performance testing in accordance with ISO 15197:2003 and other international standards.

Here's a breakdown of the acceptance criteria and study details, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document describes that "The device passed all of the tests based on predetermined Pass/Fail criteria." However, it does not explicitly list the specific acceptance criteria or the numerical performance results for the glucose measurements. It only states that clinical evaluation included method comparison, user performance, and alternative-site blood glucose measurement.

Therefore, the table can only reflect what is mentioned:

Acceptance Criteria CategoryReported Device Performance
Method ComparisonDevice passed, meeting predetermined Pass/Fail criteria.
User PerformanceDevice passed, meeting predetermined Pass/Fail criteria.
Alternative-Site MeasurementDevice passed, meeting predetermined Pass/Fail criteria.
Disinfectant CompatibilityNo change in performance or external materials after 1,098 cleaning/disinfection cycles with Cavi Wipes.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the sample size used for the test set. It mentions "Clinical evaluation included method comparison, user performance and alternative-site blood glucose measurement," but provides no details on the number of participants or samples.

Regarding data provenance, the study was conducted by Infopia Co., Ltd., which is based in the Republic of Korea. Therefore, the data likely originated from Republic of Korea. The study implicitly appears to be prospective as it involves clinical evaluation of the device's performance.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

The document does not provide information on the number of experts used or their qualifications to establish ground truth for the test set. For blood glucose meters, the "ground truth" is typically established by laboratory reference methods (e.g., YSI glucose analyzer), not expert consensus in the traditional sense of image interpretation.

4. Adjudication Method for the Test Set

The document does not provide information on any adjudication method used for the test set. Given the nature of blood glucose measurement, adjudication by multiple experts is not typically applicable in the same way as it might be for diagnostic image interpretation. The comparison would be against a reference method.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not reported. This type of study is more relevant for AI algorithms that assist human readers in tasks like interpreting medical images. The Element™ Lite Blood Glucose Monitoring System is a standalone diagnostic device for direct measurement by patients.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, a standalone performance evaluation was conducted. The "Element™ Lite Blood Glucose Monitoring System" itself is a standalone device. The performance data discussed (method comparison, user performance, alternative-site measurement) are assessments of the device's ability to accurately measure glucose independently. There is no indication of a human-in-the-loop component for the measurement process itself.

7. The Type of Ground Truth Used

While not explicitly stated, for a blood glucose monitoring system, the ground truth for performance evaluation is typically established using a laboratory reference method (e.g., a glucose analyzer like a YSI STAT PLUS Glucose & Lactate Analyzer or equivalent). This is implied by "method comparison" studies. It is not expert consensus, pathology, or outcomes data in this context.

8. The Sample Size for the Training Set

The document does not provide information on the sample size for a training set. As this is a direct measurement device rather than an AI/machine learning algorithm that requires extensive training data, a traditional "training set" in that sense might not exist or be explicitly identified in the same way. The device's performance is based on its chemical and electrical design, calibrated and validated against reference standards.

9. How the Ground Truth for the Training Set Was Established

Since the document does not mention a specific "training set" for an algorithm, it also does not provide information on how ground truth for such a set would have been established. Any internal calibration or design validation would involve comparing the device's readings against established reference standards.

{0}------------------------------------------------

K133045

510(k) Summary

APR 1 0 2014

This summary of 510(k)-safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: ___ 03/13/2014

1. Submission Applicant:

Infopia Co., Ltd. 891, Hogye-Dong, Dongan-Gu, Anyang, Kyunggi, Republic of Korea 431-080

Phone +82 31 460 0300 Fax +82 31 460 0401

2. Submission Correspondent:

Priscilla Chung LK Consulting Group USA, Inc. 2651 E Chapman Ave Ste 110, Fullerton, CA 92833 Phone: 714-202-5789 Fax: 714-409-3357 Email: info@lkconsultinggroup.com

3. Device: ·

  • Device's Trade Name:
    Element™ Lite Blood Glucose Monitoring System

■ Device's Common Name:

■ Device's Classification Name:

Blood Glucose Test System Quality control material (assayed and unassayed)

Glucose oxidase, glucose Single (specified) analyte controls Calculator/data processing module for clinical use

■ Classification Regulation:

  • Classification Product Code:
    21CFR 862.1345 21CFR 862.1660

CGA, NBW, JJX

Infopia Co., Ltd.

{1}------------------------------------------------

4. Predicate Device:

Element™ Blood Glucose Monitoring System (K113670) by Infopia Co., Ltd.

5. Description:

The Element™ Lite Blood Glucose Monitoring System consists of a meter, test strips and control solutions (level 1, level 2 and level 3), a lancing device and sterile lancets. This blood glucose test system is an in vitro diagnostic device designed for measuring the concentration of glucose in whole blood sample by means of an electrical current produced in the test strip and sent to the meter for measurement.

6. Indications for use:

The Element™ Lite Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the fingertips, ventral palm, dorsal hand, upper arm, forearm, calf and thigh. The Element™ Lite Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared.

The Element™ Lite Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

The Element™ Lite Test Strips are for use with the Element™ Lite Meter to quantitatively measure glucose in fresh capillary whole blood. Fresh capillary whole blood samples may be drawn from the fingertips, ventral palm, dorsal hand, upper arm, forearm, calf and thigh.

The Element™ Lite Control Solutions are for use with the Element™ Lite Meter and Element™ Lite Test Strips to check that the meter and test strips are working together properly and the test is performing correctly.

7. Technological Characteristics:

The Element™ Lite Blood Glucose Monitoring System has the same fundamental scientific technology as the predicate device and has similar performance specifications and features.

8. Performance Data:

The performance tests for the Element™ Lite Blood Glucose Monitoring System were performed in accordance with ISO 15197:2003 and some other international standards. Clinical evaluation included method comparison, user performance and alternative-site blood glucose measurement. Non-clinical performance evaluations were conducted to establish the performance, functionality and reliability characteristics of the Element™ Lite Blood Glucose Monitoring System. The device passed all of the tests based on predetermined Pass/Fail criteria.

.

{2}------------------------------------------------

Disinfectant Cavi Wipes with the EPA registration number of 46781-8 has been validated demonstrating complete inactivation of live virus of use with the meter and the reusable lancing device. There was also no change in performance or in the external materials of the meter and the lancing device after 1,098 cleaning/disinfection cycles designed to simulate 3 years of device use.

9. Conclusion:

Infopia Co., Ltd. concludes that the Element™ Lite Blood Glucose Monitoring System is safe and effective and also substantially equivalent to the predicate device, Element™ Blood Glucose Monitoring System (K113670).

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K133045

Device Name

Element™ Lite Blood Glucose Monitoring System

Indications for Use (Describe)

The Element™ Lite Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the fingetips, ventral palm, dorsal hand, upper arm, calf and thigh. The Element™ Lite Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared.

The Element™ Lite Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. It should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

The Element™ Lite Test Strips are for use with the Element™ Lite Meter to quantitatively measure glucose in fresh capillary whole blood. Fresh capillary whole blood samples may be drawn from the fingertips, ventral palm, dorsal hand thigh.

The Element™ Lite Control Solutions are for use with the Element™ Lite Test Strips to check that the meter and test strips are working together properly and the test is performing correctly.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/3/Picture/16 description: The image shows the text "Stayce Beck" in a stylized font. The letters "ayce" have a layered, outlined effect, creating a three-dimensional appearance. The rest of the letters are in a bold, sans-serif font.

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

{4}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 10, 2014

INFOPIA CO., LTD. C/O PRISCILLA CHUNG 2651 E. CHAPMAN AVE STE. 110 FULLERTON CA 92831

Re: K133045

Trade/Device Name: Element™ Lite Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW, CGA, JJX Dated: September 24, 2013 Received: September 27, 2013

Dear Ms. Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{6}------------------------------------------------

Page 2-Ms. Priscilla Chung

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.lda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/McdicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Courtney H. Lias -S

Courtney H. Lias. Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known) K133045

Device Name

Element™ Lite Blood Glucose Monitoring System

Indications for Use (Describe)

The Element™ Lite Blood Glucose Monitoring System is intended for the quantitative measurement of glucose in fresh capillary whole blood from the fingerips, ventral pain, dorearn, calf and thigh. The Element™ Life Blood Glucose Monitoring System is intended to be used by a single patient and should not be shared.

The Element™ Lite Blood Glucose Monitoring System is intended the body (in viro diagnostic use) by people with diabeles at home as an aid to monitor the effectiveness of diabetes control. It should not be used for screening of diabeles or for neonatal use. Alternative should be done only during steady-state times (when glucose is not changing rapidly).

The Element™ Lite Test Strips are for use with the Element™ Lite Meter to quantitatively measure glacese in fresh capillary whole blood. Fresh capillary whole blood samples may be fingertips, ventral palm, dorsal hand, upper arm, forcarn, calf and thigh.

The Element™ Lite Control Solutions are for use with the Element™ Lite Test Strips to check (trat the meter and test strips are working together properly and the test is performing correctly.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

. FOR FDA USE ONLY

currence of Center for Devices and Radiological Health (CDRH) (Signature)

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.