(224 days)
The Cogent Med-LIF is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These devices are intended for intervertebral body fusion, and are intended to be used with supplemental fixation instrumentation which has been cleared by the FDA for use in the lumbar spine.
The Cogent Med-LIF is an interbody fusion device intended to stabilize the spinal segment to promote fusion. The Cogent Med-LIF consists of two PEEK spacers with axial voids to contain bone graft material, a titanium linkage that connects the PEEK spacers, angular anti-migration teeth, and tantalum x-ray markers. The Cogent Med-LIF is available in various sizes to accommodate varying patient anatomy. The implants come in two lordotic options, 0° (parallel) and 6° (lordotic). The parallel implants are available in heights ranging from 8mm to 16mm and the lordotic implants are available in heights ranging from 10mm to 16mm. All implants are 10mm wide and are available in three lengths; 20, 25 and 30mm. The Cogent Med-LIF implants may be inserted via an open or minimally invasive approach and may be placed as a single implant or as two units bilaterally in the same intervertebral space. The Cogent Med-LIF implants are non-sterile and are to be sterilized by the end user.
The provided document describes the Cogent Med-LIF, an intervertebral body fusion device, and its performance testing for a 510(k) submission. It does not describe a study involving AI, software or an analytical device. Therefore, it is impossible to answer the following sections, as they are not presented in the document:
- Acceptance criteria and reported device performance related to a diagnostic or analytical AI/software device.
- Sample size used for the test set and data provenance.
- Number of experts used to establish ground truth for the test set and their qualifications.
- Adjudication method for the test set.
- Multi-reader multi-case (MRMC) comparative effectiveness study, effect size.
- Standalone (algorithm only) performance.
- Type of ground truth used (expert consensus, pathology, outcomes data, etc.) for an AI/software device.
- Sample size for the training set.
- How ground truth for the training set was established.
However, I can extract information regarding the physical device's performance testing and its overall conclusion of substantial equivalence.
1. Table of Acceptance Criteria and Reported Device Performance
For the physical device, Cogent Med-LIF, the "acceptance criteria" are implied by compliance with ASTM standards and demonstrating performance comparable to legally marketed predicates. The "reported device performance" is a general statement that the device met these criteria.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Mechanical Testing: | |
| - Static Axial Compression (ASTM F2077) | No new risks indicated; substantial equivalence to predicate. |
| - Dynamic Axial Compression (ASTM F2077) | No new risks indicated; substantial equivalence to predicate. |
| - Static Shear Compression (ASTM F2077) | No new risks indicated; substantial equivalence to predicate. |
| - Dynamic Shear Compression (ASTM F2077) | No new risks indicated; substantial equivalence to predicate. |
| - Subsidence (ASTM F2267) | No new risks indicated; substantial equivalence to predicate. |
| - Expulsion Testing (ASTM Draft Standard F04.25.02.02) | No new risks indicated; substantial equivalence to predicate. |
| Functional/Clinical (Cadaveric Study): | |
| - Adequacy of autogenous bone graft delivery compared to predicate | Cadaveric study confirmed adequate delivery compared to a legally marketed predicate. |
Summary of the Study:
The document describes bench testing and a cadaveric implantation study for the Cogent Med-LIF device.
- Bench Tests: These tests focused on the mechanical properties of the device, adhering to established ASTM standards (F2077, F2267, and a draft expulsion standard). The specific acceptance criteria for these tests would be the performance benchmarks defined by these ASTM standards or comparative performance to the predicate devices.
- Cadaveric Implantation Study: This study's purpose was to confirm that autogenous bone graft could be adequately delivered using the Cogent Med-LIF, specifically when compared to a legally marketed predicate device.
Overall Conclusion:
The manufacturer concluded that "mechanical and non-clinical testing of the Cogent Med-LIF device indicated no new risks and demonstrated substantial equivalence in performance compared to a legally marketed predicate." This statement serves as the final "reported device performance" and meets the overarching "acceptance criteria" for a 510(k) submission, which is to demonstrate substantial equivalence to a predicate device.
{0}------------------------------------------------
APR 1 5 2014
SUBMITTER:
Submitted Bv:
| Company Name: | Cogent Spine |
|---|---|
| Address: | 101 N. Acacia, Suite 106Solana Beach, CA 92075 |
| Telephone: | (858) 774-7891 |
CONTACT PERSON: Jude Paganelli
DATE PREPARED: April 15, 2014
TRADE NAME: Cogent Med-LIF
COMMON NAME: Intervertebral Body Fusion Device
CLASSIFICATION NAME: Intervertebral Body Fusion Device (21 CFR 888.3080)
PRODUCT CODE: MAX
SUBSTANTIALLY EQUIVALENT TO:
The Cogent Med-LIF is substantially equivalent to the predicates in all facets including: indications, technology, method of use and performance. The predicate devices for the Cogent Med-LIF system are:
Eisertech PLIF Cage (K113478) Stryker Spine AVS PL PEEK Spacers (K073470) Custom Spine Pathway Avid (K090566) Sapphire Medical Group A-Wedge Anterior Interbody System (K121693)
DESCRIPTION of the DEVICE:
The Cogent Med-LIF is an interbody fusion device intended to stabilize the spinal segment to promote fusion. The Cogent Med-LIF consists of two PEEK spacers with axial voids to contain bone graft material, a titanium linkage that connects the PEEK spacers, angular anti-migration teeth, and tantalum x-ray markers.
The Cogent Med-LIF is available in various sizes to accommodate varying patient anatomy. The implants come in two lordotic options, 0° (parallel)
510(K) Submission:
{1}------------------------------------------------
510(k) Summary-Continued
and 6° (lordotic). The parallel implants are available in heights ranging from 8mm to 16mm and the lordotic implants are available in heights ranging from 10mm to 16mm. All implants are 10mm wide and are available in three lengths; 20, 25 and 30mm.
The Cogent Med-LIF implants may be inserted via an open or minimally invasive approach and may be placed as a single implant or as two units bilaterally in the same intervertebral space.
The Cogent Med-LIF implants are non-sterile and are to be sterilized by the end user.
MATERIALS:
The Cogent Med-LIF is manufactured from polyetheretherketone (PEEK) as per ASTM F2026 and contains titanium alloy (Ti-6Al-4V) per ASTM F1472 and tantalum per ASTM F560.
INDICATIONS FOR USE:
The Cogent Med-LIF is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These devices are intended for intervertebral body fusion, and are intended to be used with supplemental fixation instrumentation which has been cleared by the FDA for use in the lumbar spine.
PERFORMANCE TESTING:
Cogent Spine conducted the following bench tests:
ASTM F2077
- . Static Axial Compression
- Dynamic Axial Compression ●
- Static Shear Compression .
- Dynamic Shear Compression .
ASTM F2267
· Subsidence
Expulsion Testing per ASTM Draft Standard F04.25.02.02
{2}------------------------------------------------
510(k) Summary-Continued
Additionally, Cogent Spine conducted a cadaveric implantation study to confirm autogenous bone graft can be adequately delivered when compared to a legally marketed predicate.
In summary, mechanical and non-clinical testing of the Cogent Med-LIF device indicated no new risks and demonstrated substantial equivalence in performance compared to a legally marketed predicate.
CONCLUSIONS:
The Cogent Med-LIF device has shown to be substantially equivalent to legally marketed predicates based on indications for use, technological characteristics, performance testing and comparison to predicate devices.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 15, 2014
Cogent Spine LLC % Mr. Jude Paganelli Cor Medical Ventures LLC 101 North Acacia Avenue, Suite 106 Solana Beach, California 92075
Re: K132738
Trade/Device Name: Cogent Med-LIF Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: February 12, 2013 Received: March 21, 2014
Dear Mr. Paganelli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical CI K Fall 607); adoling (21 OFF 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Jude Paganelli
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Ronald P. Jean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K132738
Device Name Cogent Med-LIF
Indications for Use (Describe)
The Cogent Med-LIF is indicated for use with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous levels from L2 to S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved levels. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment. These devices are intenertebral body fusion, and are intended to be used with supplemental fixation instrumentation which has been cleared by the FDA for use in the lumbar spine.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Anton E. Dmitriev, PhD
.
Division of Orthopedic Devices
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
N/A