K Number
K132656

Validate with FDA (Live)

Date Cleared
2014-03-24

(210 days)

Product Code
Regulation Number
888.3045
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

PRO-DENSE® resultant paste is intended for use as a bone graft substitute to be injected or digitally packed into open bone voids/gaps that are not intrinsic to the stability of bony structure of the skeletal system (i.e., the extremities and pelvis) to cure in situ. These open bone voids may be the result of benign bone cysts and turnors (in adults and pediatric patients ≥ 6 years old), surgically created osseous defects or osseous defects created from traumatic injury to the bone. The paste provides a bone graft substitute that resorbs and is replaced with bone during the healing process.

The PRO-DENSE® paste cured in situ provides an open void/gap filler that can augment provisional hardware (e.g. K. Wires) to help support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process.

PRO-DENSE® is provided sterile for single use only.

For the PRO-DENSE® Core Decompression Procedure Kit:

The PRO-DENSE® Core Decompression Procedure Kit, consisting of a bone void filler and manual surgical instruments, is intended to be used during core decompression procedures. The bone void filler component resorbs and is replaced with bone during the healing process. The bone void filler included in the PRO-DENSE® Core Decompression Procedure Kit is not intended to be used as a load-bearing device.

Device Description

PRODENSE® is indicated as a bone graft substitute to be injected and/or digitally packed into open bone voids/gaps that are not intrinsic to the structural stability of the skeletal system and cure in-situ. It is supplied in separate powder and liquid vials along with the instruments for mixing it into a paste and delivering it to the defect site. The triphasic resorption of PRODENSE® results in a scaffold that is osteoconductive allowing tissue infiltration and must eventually be degraded through osteoclastic action as bone remodels within the scaffold. The clinical use of calcium sulfate, calcium phosphate, and composites thereof as a bone void filler has been well established through many peer reviewed publications.

PRO-DENSE® is a calcium sulfate - calcium phosphate composite bone graft substitute consisting of a powder component and an aqueous mixing solution. When the two components are mixed according to directions, an injectable paste forms and subsequently hardens via hydration reactions.

AI/ML Overview

Here's an analysis of the provided text regarding the PRODENSE® Bone Graft Substitute, focusing on acceptance criteria and supporting studies:

  1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state quantitative (e.g., numerical thresholds for strength, resorption rate) acceptance criteria in the way one might expect for a typical medical device performance study, nor does it present a direct alongside comparison of such criteria with the device's reported performance. Instead, the document focuses on substantial equivalence to predicate devices. The "performance" assessment is based on demonstrating that the modified PRODENSE® is equivalent to its original form and established predicate devices.

    The key "acceptance criteria" appear to be:

    • Equivalence of Characteristics: The modified device's characteristics, especially set-time, should be equivalent or adequately managed compared to the original product.
    • No New Questions of Safety or Effectiveness: The design changes should not introduce new safety or effectiveness concerns.
    • Adequate Risk Mitigation: All risks associated with the product line must remain adequately mitigated.
    Acceptance CriterionReported Device Performance
    Equivalence of Characteristics (specifically set-time)"The components of PRODENSE® were slightly adjusted to decrease its set-time. The characteristics of this PRODENSE® modification were analyzed according to design control testing and found to be equivalent to the original product." (This implies the new set-time is acceptable).
    No New Questions of Safety or Effectiveness"The design characteristics of the subject system do not raise any new types of questions of safety or effectiveness."
    Adequate Risk Mitigation"a risk analysis indicated all risks associated with the PRODENSE® product line remain adequately mitigated."
    Indications for Use unchanged as a result of modification"The indications for use of the subject device were not changed as a result of the modification..."
    Fundamental scientific technology/mechanism of action remains identical"...and the fundamental scientific technology of the device, including the bone void filler's mechanism of action, remain identical to the original product."
    Performance at least as well as predicate systems"From the evidence submitted in this 510(k), the subject devices can be expected to perform at least as well as the predicate systems." (This is the overarching conclusion of a substantial equivalence submission, effectively serving as the ultimate acceptance criterion for this type of submission).
  2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify a distinct "test set" in the context of human subjects or clinical data. The evaluation was primarily based on design control testing for the modified device. No sample size for a test set (e.g., number of patients or bone voids) is provided, as the submission did not involve new clinical studies for this specific modification. The provenance of any underlying data from the original PRODENSE® or predicate devices is also not mentioned, as the focus is on the equivalence of the modified product. The study is non-clinical in nature, focusing on engineering and material characteristics.

  3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable. This was a non-clinical submission based on design control testing and comparison of material properties, not on interpretation of clinical images or patient outcomes requiring expert ground truth in that sense. The FDA regulatory review process itself involves experts (e.g., Mark N. Melkerson, Director, Division of Orthopedic Devices, and Laurence D. Coyne, Division Sign-Off) but they are evaluating the submission, not establishing ground truth for a test set within the submission.

  4. Adjudication Method:

    Not applicable. There was no clinical test set requiring adjudication.

  5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, a MRMC comparative effectiveness study was not done. The submission explicitly states "Substantial Equivalence - Clinical Evidence N/A". The evaluation was based on non-clinical design control testing.

  6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance):

    Yes, in a conceptual sense, the "standalone" performance was assessed through design control testing. This testing evaluated the material properties and characteristics of the PRODENSE® modification itself, independent of human interaction beyond the mixing and application steps as intended. The "set-time" was a key characteristic evaluated. This isn't an algorithm, but the concept of evaluating the device's intrinsic properties applies.

  7. Type of Ground Truth Used:

    The "ground truth" for the non-clinical evaluation was based on engineering and material specifications of the original PRODENSE® device and the established expectations for bone void fillers as outlined in FDA's special control guidance for resorbable calcium salt bone void filler devices. The comparison was to the original product's known characteristics and regulatory standards, rather than pathology, outcomes data, or expert consensus in a clinical trial setting.

  8. Sample Size for the Training Set:

    Not applicable. This device is a bone graft substitute, not an AI/ML algorithm that requires a training set. The term "training set" doesn't apply to the evaluation described in this 510(k) submission.

  9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there was no training set in the context of an AI/ML algorithm.

{0}------------------------------------------------

MAR 2 4 2014 Create Motion."

K132656

510(K) SUMMARY

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92(c), this information serves as a Summary of Safety and Effectiveness for the use of the PRODENSE® Bone Graft Substitute.

1. Submitted By:Wright Medical Technology, Inc.5677 Airline RoadArlington, TN 38002
Date:August 23, 2013
Contact Person:Leslie Fitch, PhDRegulatory Affairs Specialist IIPhone: (901) 867-4120Fax: (901) 867-4190
2. Proprietary Name:PRO-DENSE® Bone Graft Substitute
Common Name:Bone Void Filler
Classification Name and Reference:21 CFR 888.3045 - Class II
Device Product Code, Device Panel:MQV: Orthopedic
3. Predicate Devices:K113871 – PRODENSE® Bone Graft SubstituteK072597 – PRODENSE® Core DecompressionProcedure Kit

Device Description 4.

PRODENSE® is indicated as a bone graft substitute to be injected and/or digitally packed into open bone voids/gaps that are not intrinsic to the structural stability of the skeletal system and cure in-situ. It is supplied in separate powder and liquid vials along with the instruments for mixing it into a paste and delivering it to the defect site. The triphasic resorption of PRODENSE® results in a scaffold that is osteoconductive allowing tissue infiltration and must eventually be degraded through osteoclastic action as bone remodels within the scaffold. The clinical use of calcium sulfate, calcium phosphate, and composites thereof as a bone void filler has been well established through many peer reviewed publications.

Page E2-1

{1}------------------------------------------------

Create Motion."

Image /page/1/Picture/1 description: The image shows the text "K132656 - Page 2 of 3" at the top. Below that is the word "WRIGHT." in large, bold letters. Underneath the word is a design that looks like three horizontal lines that are thicker on the ends and thinner in the middle.

5. Materials

PRO-DENSE® is a calcium sulfate - calcium phosphate composite bone graft substitute consisting of a powder component and an aqueous mixing solution. When the two components are mixed according to directions, an injectable paste forms and subsequently hardens via hydration reactions. The benefits of this composite include:

  • Calcium Sulfate .
    • Primary osteoconductive filler 트
    • . Resorbs first primarily through simple dissolution to allow early vascular infiltration
    • . Excellent clinical history
  • Calcium Phosphate ●
    • Osteoclastic resorption I
    • 로 Secondary porous scaffold that is resorbed after primary filler
    • . TCP granules are resorbed in the third and final phase

6. Intended Use

PRO-DENSE® resultant paste is intended for use as a bone graft substitute to be injected or digitally packed into open bone voids/gaps that are not intrinsic to the stability of bony structure of the skeletal system (i.e., the extremities and pelvis) to cure in situ. These open bone voids may be the result of benign bone cysts and tumors (in adults and pediatric patients > 6 years old), surgically created osseous defects or osseous defects created from traumatic injury to the bone. The paste provides a bone graft substitute that resorbs and is replaced with bone during the healing process.

The PRO-DENSE® paste cured in situ provides an open void/gap filler that can augment provisional hardware (e.g. K. Wires) to help support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process.

PRO-DENSE® is provided sterile for single use only.

PRODENSE® Core Decompression Procedure Kit

The PRO-DENSE® Core Decompression Procedure Kit, consisting of a bone void filler and manual surgical instruments, is intended to be used during core decompression procedures. The bone void filler component resorbs and is replaced with bone during the healing process. The bone void filler included in the PRO-DENSE® Core Decompression Procedure Kit is not intended to be used as a load-bearing device.

7. Technological Characteristics Comparison

The components of PRODENSE® were slightly adjusted to decrease its set-time. The characteristics of this PRODENSE® modification were analyzed according to design control testing and found to be equivalent to the original product. The indications for use of the subject device were not changed as a result of the modification and the fundamental scientific technology of the device, including the bone void filler's mechanism of action, remain identical to the original product.

Page E2-2

{2}------------------------------------------------

Create Motion."

Image /page/2/Picture/1 description: The image shows the text "K132656 - Page 3 of 3" at the top, followed by the word "WRIGHT." in large, bold letters. Below the word "WRIGHT." is a decorative, stylized line. The text at the top indicates that this is the third page of a document with the identifier K132656.

Substantial Equivalence - Non-Clinical Evidence 8.

Design control testing was conducted in accordance with the FDA special control guidance for resorbable calcium salt bone void filler devices. The results of this analysis determined that set-time was the only significantly altered characteristic of the PRODENSE® modification relative to the original product and a risk analysis indicated all risks associated with the PRODENSE® product line remain adequately mitigated.

9. Substantial Equivalence - Clinical Evidence

N/A

10. Substantial Equivalence - Conclusions

The design characteristics of the subject system do not raise any new types of questions of safety or effectiveness. From the evidence submitted in this 510(k), the subject devices can be expected to perform at least as well as the predicate systems.

www.wmt.com

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an abstract caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring. MD 20993-0002

March 24, 2014

Wright Medical Technology, Incorporated Leslie Fitch, Ph.D. Regulatory Affairs Specialist II 5677 Airline Road Arlington, Tennessee 38002

Re: K132656

Trade/Device Name: PRO-DENSE® Bone Graft Substitute Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: February 7, 2014 Received: February 20, 2014

Dear Dr. Fitch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Leslie Fitch. Ph.D.

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to -

http://www.fda.gov/MedicalDevices/Safetv/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Ronald P. Jean -S for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K132656 - Page 1 of 1

INDICATIONS FOR USE STATEMENT

510(k) Number: K132656

PRO-DENSE® Bone Graft Substitute Device Name:

Indications For Use:

PRO-DENSE® resultant paste is intended for use as a bone graft substitute to be injected or digitally packed into open bone voids/gaps that are not intrinsic to the stability of bony structure of the skeletal system (i.e., the extremities and pelvis) to cure in situ. These open bone voids may be the result of benign bone cysts and turnors (in adults and pediatric patients ≥ 6 years old), surgically created osseous defects or osseous defects created from traumatic injury to the bone. The paste provides a bone graft substitute that resorbs and is replaced with bone during the healing process.

The PRO-DENSE® paste cured in situ provides an open void/gap filler that can augment provisional hardware (e.g. K. Wires) to help support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process.

PRO-DENSE® is provided sterile for single use only.

For the PRO-DENSE® Core Decompression Procedure Kit:

The PRO-DENSE® Core Decompression Procedure Kit, consisting of a bone void filler and manual surgical instruments, is intended to be used during core decompression procedures. The bone void filler component resorbs and is replaced with bone during the healing process. The bone void filler included in the PRO-DENSE® Core Decompression Procedure Kit is not intended to be used as a load-bearing device.

Prescription Use (Per21 CFR 801.109) OR

Over-The Counter Use (Optional Format 1-2-96)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Laurence D. Coyne -S

(Division Sign-OM) Division of Orthopedic Devices 510(k) Number: K132656

§ 888.3045 Resorbable calcium salt bone void filler device.

(a)
Identification. A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.