K Number
K132633
Date Cleared
2014-08-14

(357 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OKmeter Direct Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper-arm, calf, thigh, or palm. The OKmeter Direct Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

The OKmeter Direct Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The OKmeter Direct Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

The OKmeter Direct Test Strips are for use with the OKmeter Direct Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper-arm, calf, thigh, or palm.

Device Description

Based on an electrochemical biosensor technology and the principle of capillary action, Okmeter Direct Blood Glucose Monitoring System only needs a small amount of blood. Capillary action at the end of the test strip draws the blood into the action chamber and your blood glucose result is precisely and is displayed in 6 seconds.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the OKmeter Direct Blood Glucose Monitoring System, as extracted from the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for glucose meters are typically defined by regulatory bodies like the FDA, often referring to standards like ISO 15197. While the document doesn't explicitly state "acceptance criteria" numbers as a separate table, it presents performance data in a way that implies these are the criteria against which the device is being tested for accuracy. The core criteria revolve around the percentage of results falling within specific deviation ranges from a reference method (YSI 2300).

The performance data is presented separately for two user groups: "Home Users" (lay users) and "Healthcare Professionals."

For Glucose Concentration < 75 mg/dL (4.2 mmol/L):

Acceptance Criteria (Deviation from Reference)Lay Users Performance (Range: Fingertip to Thigh)Healthcare Professionals Performance (Range: Fingertip to Thigh)
Within ±5 mg/dL (±0.28 mmol/L)67% - 86%67% - 81%
Within ±10 mg/dL (±0.56 mmol/L)95% - 100%95% - 100%
Within ±15 mg/dL (±0.83 mmol/L)100%100%

For Glucose Concentration ≥ 75 mg/dL (4.2 mmol/L):

Acceptance Criteria (Deviation from Reference)Lay Users Performance (Range: Fingertip to Thigh)Healthcare Professionals Performance (Range: Fingertip to Thigh)
Within ±5%58% - 73%59% - 74%
Within ±10%84% - 100%85% - 100%
Within ±15%96% - 99%96% - 97%
Within ±20%100%100%

Summary of Reported Performance (Across all alternative sites and user types, presenting the minimum reported compliance):

Glucose Concentration RangeAccuracy (Within ±5 mg/dL or ±5%)Accuracy (Within ±10 mg/dL or ±10%)Accuracy (Within ±15 mg/dL or ±15%)Accuracy (Within ±20% for ≥ 75 mg/dL)
< 75 mg/dL (4.2 mmol/L)67% (Forearm/Upper arm/Calf)95% (Palm/Forearm/Upper arm/Calf)100%Not applicable
≥ 75 mg/dL (4.2 mmol/L)58% (Calf)84% (Thigh)96% (Palm/Forearm/Upper arm/Calf/Thigh)100%

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 135 patients
  • Data Provenance: The document does not explicitly state the country of origin. The manufacturer is "OK Biotech Co., Ltd." based in "Hsin Chu City, Taiwan, ROC." It is reasonable to infer the study might have been conducted in Taiwan, but this is not definitively stated.
  • Retrospective or Prospective: The document does not specify whether the data was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • The ground truth was established by a YSI 2300 Analyzer. This is a laboratory reference instrument, not a human expert. Therefore, the concept of "number of experts" and their qualifications for establishing ground truth does not directly apply in the human sense.

4. Adjudication Method for the Test Set

  • Since the ground truth was established by an objective laboratory instrument (YSI 2300 Analyzer), no human adjudication method (like 2+1, 3+1) was required. The device's readings were directly compared to the YSI 2300 readings.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • The document describes a study comparing the device's performance to a reference standard (YSI 2300) when used by both lay users and healthcare professionals. While different user groups ("readers") are involved, this is not a traditional MRMC comparative effectiveness study designed to measure the improvement of human readers with AI vs. without AI assistance. It's a performance study of the device itself under different user scenarios.
  • Therefore, an "effect size of how much human readers improve with AI vs without AI assistance" is not applicable in this context.

6. Standalone (Algorithm Only) Performance

  • Yes, a standalone performance study was done. The entire study presented quantifies the accuracy of the OKmeter Direct Blood Glucose Monitoring System (the algorithm/device) against a reference standard (YSI 2300 Analyzer). The metrics (e.g., % within ±5 mg/dL or ±5%) are direct measurements of the device's accuracy.

7. Type of Ground Truth Used

  • The ground truth used was established by a laboratory reference method: the YSI 2300 Analyzer. This is a highly accurate instrument used for glucose measurement in clinical settings.

8. Sample Size for the Training Set

  • The document does not provide information regarding a training set sample size. The data presented is for the evaluation of the final device rather than its development.

9. How the Ground Truth for the Training Set Was Established

  • Since information on a training set is not provided, how its ground truth was established is not available in this document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Auqust 14, 2014

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

OK BIOTECH CO., LTD. JEN KE-MIN 1F, NO. 87, SEC. 2, GONGDAOWU ROAD HSIN CHU CITY 30070 TAIWAN

Re: K132633

Trade/Device Name: OKmeter Direct Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW. CGA Dated: July 17, 2014 Received: July 15, 2014

Dear Dr. Jen. Ke-min:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K132633

Device Name

Okmeter Direct Blood Glucose Monitoring System

Indications for Use (Describe)

The OKmeter Direct Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper-arm, calf, thigh, or palm. The OKmeter Direct Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

The OKmeter Direct Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The OKmeter Direct Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

The OKmeter Direct Test Strips are for use with the OKmeter Direct Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, calf, thigh, or palm.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows a logo with the letter K inside of a circle. Next to the K is the word "biotech" in a bold, sans-serif font. Below the word "biotech" is some text in another language.

5. 510(K) Summary of Safety and Effectiveness

(Per 21 CFR 807.92)

5.1. General Information Establishment

The assigned 510(k) number is :K132633
Manufacturer:OK Biotech Co., Ltd.
Address:1F, No. 87, Sec. 2, Gongdaowu Road, Hsin Chu City,30070, Taiwan, ROC
Owner Number:9090860
Contact Person:Dr. Jen, Ke-Min E-mail: ceirs.jen@msa.hinet.net886-3-5208829 (Tel); 886-3-5209783 (Fax)
Address:No.58, Fu Chiun Street, Hsin Chu City, 30067, Taiwan, ROC
Date Prepared:July 26, 2013
Device
Proprietary Name:Okmeter Direct Blood Glucose Monitoring System
Common Name:Blood Glucose Monitoring System
● Classification Name:SYSTEM, TEST, BLOOD GLUCOSEOVER THE COUNTER, Class II
● Product Code:NBW

5.2. Safety and Effectiveness Information

  • Predicate Device: Claim of Substantial Equivalence (SE) is made to Okmeter Match Blood Glucose Monitoring System (K090609)
    • Device Description: Based on an electrochemical biosensor technology and the principle of capillary action, Okmeter Direct Blood Glucose Monitoring System only needs a small amount of blood. Capillary action at the end of the test strip draws the blood into the action chamber and your blood glucose result is precisely and displayed in 6 seconds.

● Intended Use:

The OKmeter Direct Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper-arm, calf, thigh, or palm. The OKmeter Direct Blood Glucose Monitoring System is intended to be used by a single person and should not be shared.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo with the letter 'K' inside of an oval shape. To the right of the 'K' is the word 'biotech' in bold, sans-serif font. Below the word 'biotech' is a line of text in a different language, possibly Chinese or Japanese.

E-mail:service@okbiotech.com

The OKmeter Direct Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The OKmeter Direct Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).

The OKmeter Direct Test Strips are for use with the OKmeter Direct Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper-arm, calf, thigh, or palm.

● Substantial Equivalence (SE)

A claim of substantial equivalence is made to Okmeter Match Blood Glucose Monitoring System (K090609). Both of them have the same working principle and technologies including sample volume, measuring time, detecting range, HCT range, calibration method, and memory data number. The major differences for the two devices are meter dimension, weight; especially the predicate device using two colors of the test strips and has some audible features but it is not for use by the visually impaired. B Besides, the subject device and predicate device is same intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. Thus the differences are due to the feature design aspects, not relating to the safety or effectiveness aspects. They are substantially equivalent.

● Synopsis of Test Methods and Results

Pre-clinical and clinical data are employed upon submission of this 510(K) premarket notification according to the Guidance Document for In Vitro Diagnostic Test System; Guidance for Industry and FDA document provided by CDRH/ FDA.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo with the letter K inside of an oval shape. Next to the K is the word "biotech" in bold font. Below the word biotech is a line of Chinese characters.

For Home Users

The table below was based on a study done with 135 patients to see how well the OKmeter Direct Blood Glucose Monitoring System compared to the YSI 2300.

  • Finger capillary whole blood
    Image /page/5/Figure/6 description: The image is a scatter plot titled "Lay users vs. YSI 2300". The x-axis is labeled "YSI2300 (mg/dL)" and ranges from 0 to 720. The y-axis is labeled "Meter (mg/dL)" and ranges from 0 to 720. The plot shows a positive correlation between the two variables, with a regression equation of y = 0.9820x + 1.2484 and an R-squared value of 0.9866.

Results for glucose concentration<75 mg/dL(4.2 mmol/L)

Within±5 mg/dLWithin±10 mg/dLWithin±15 mg/dL
(Within±0.28 mmol/L)(Within±0.56 mmol/L)(Within±0.83 mmol/L)
15/21 (71%)21/21 (100%)21/21 (100%)

Results for glucose concentration ≥ 75 mg/dL (4.2 mmol/L)

Within±5%Within±10%Within±15%Within±20%
83/114 (73%)100/114 (88%)111/114 (97%)114/114 (100%)
OKmeter Direct Blood Glucose Monitoring System vs. YSI 2300 Analyzer
nSlopeInterceptR2
1350.9821.24840.9866

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Kbiotech. The logo features a stylized letter K enclosed in an oval shape, followed by the word "biotech" in a bold, sans-serif font. Below the word "biotech" is a line of text in a different language, possibly Chinese or Japanese.

Image /page/6/Figure/3 description: The image is a scatter plot titled "Lay users vs. YSI 2300". The x-axis is labeled "YSI 2300 (mg/dL)" and ranges from 0 to 720. The y-axis is labeled "Meter (mg/dL)" and ranges from 0 to 720. The scatter plot shows a positive correlation between the two variables, and the equation of the line of best fit is y = 0.997x - 0.421, with an R-squared value of 0.987.

Palm capillary whole blood 0

Results for glucose concentration<75 mg/dL(4.2 mmol/L)

Within±5 mg/dL(Within±0.28 mmol/L)Within±10 mg/dL(Within±0.56 mmol/L)Within±15 mg/dL(Within±0.83 mmol/L)
15/21 (71%)20/21 (95%)21/21 (100%)

Results for glucose concentration ≥ 75 mg/dL (4.2 mmol/L)

Within±5%Within±10%Within±15%Within±20%
74/114 (65%)102/114 (89%)111/114 (97%)114/114 (100%)
OKmeter Direct Blood Glucose Monitoring System vs. YSI 2300 Analyzer
nSlopeInterceptR2
1350.9972-0.42120.9874

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Kbiotech. The logo features a stylized letter K enclosed in an oval shape. Below the K is the word "biotech" in a bold, sans-serif font. Underneath the word "biotech" is some Chinese text.

Image /page/7/Figure/3 description: The image is a scatter plot titled "Lay users vs. YSI 2300". The x-axis is labeled "YSI2300 (mg/dL)" and ranges from 0 to 720. The y-axis is labeled "Meter (mg/dL)" and ranges from 0 to 720. The scatter plot shows a positive correlation between the two variables, and the equation of the regression line is y = 0.977x + 1.668 with an R-squared value of 0.981.

  • Forearm capillary whole blood 0
    Results for glucose concentration<75 mg/dL(4.2 mmol/L)
Within±5 mg/dL(Within±0.28 mmol/L)Within±10 mg/dL(Within±0.56 mmol/L)Within±15 mg/dL(Within±0.83 mmol/L)
15/21 (71%)20/21 (95%)21/21 (100%)

Results for glucose concentration ≥ 75 mg/dL (4.2 mmol/L)

Within±5%Within±10%Within±15%Within±20%
64/114 (56%)99/114 (87%)113/114 (99%)114/114 (100%)
OKmeter Direct Blood Glucose Monitoring System vs. YSI 2300 Analyzer
nSlopeInterceptR2
1350.97781.66890.9817

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows a logo for Kbiotech. The logo features a stylized letter "K" enclosed in a circular shape. Below the "K" is the word "biotech" in a bold, sans-serif font. Underneath the word "biotech" is some text in a different language, possibly Chinese.

Image /page/8/Figure/3 description: The image is a scatter plot titled "Lay users vs. YSI 2300". The x-axis is labeled "YSI 2300 (mg/dL)" and ranges from 0 to 720. The y-axis is labeled "Meter (mg/dL)" and ranges from 0 to 720. The scatter plot shows a positive correlation between the two variables, and the equation of the line of best fit is y = 1.008x - 1.377 with an R-squared value of 0.980.

  • Upper arm capillary whole blood
    Results for glucose concentration<75 mg/dL(4.2 mmol/L)
Within±5 mg/dL(Within±0.28 mmol/L)Within±10 mg/dL(Within±0.56 mmol/L)Within±15 mg/dL(Within±0.83 mmol/L)
14/21 (67%)20/21 (95%)21/21 (100%)

Results for glucose concentration ≥ 75 mg/dL (4.2 mmol/L)

Within±5%Within±10%Within±15%Within±20%
75/114 (66%)102/114 (89%)111/114 (97%)114/114 (100%)
OKmeter Direct Blood Glucose Monitoring System vs. YSI 2300 Analyzer
nSlopeInterceptR2
1351.0089-1.3770.9805

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Kbiotech. The logo consists of a stylized letter K inside of a circle, with the word "biotech" written in a bold, sans-serif font to the right of the K. Below the word "biotech" is some text in a different language, possibly Chinese. The logo is simple and modern, and the use of the letter K and the word "biotech" suggests that the company is involved in the biotechnology industry.

Image /page/9/Figure/3 description: The image is a scatter plot titled "Lay users vs. YSI 2300". The x-axis is labeled "YSI 2300 (mg/dL)" and ranges from 0 to 720. The y-axis is labeled "Meter (mg/dL)" and ranges from 0 to 720. The scatter plot shows a positive correlation between the two variables, and the equation of the line of best fit is y = 0.988x + 1.366, with an R-squared value of 0.979.

Calf capillary whole blood ●

Results for glucose concentration<75 mg/dL(4.2 mmol/L)

Within±5 mg/dL(Within±0.28 mmol/L)Within±10 mg/dL(Within±0.56 mmol/L)Within±15 mg/dL(Within±0.83 mmol/L)
14/21 (67%)20/21 (95%)21/21 (100%)

Results for glucose concentration ≥ 75 mg/dL (4.2 mmol/L)

Within±5%Within±10%Within±15%Within±20%
66/114 (58%)98/114(86%)112/114 (98%)114/114 (100%)
OKmeter Direct Blood Glucose Monitoring System vs. YSI 2300 Analyzer
nSlopeInterceptR2
1350.98841.36620.9794

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Kbiotech. The logo features a stylized letter K enclosed in an oval shape, followed by the text "biotech" in a bold, sans-serif font. Below the English text, there is Chinese text, which is likely the company's name in Chinese characters. The logo is simple and modern, with a focus on the company's name and brand identity.

Image /page/10/Figure/3 description: The image is a scatter plot titled "Lay users vs. YSI 2300". The x-axis is labeled "YSI 2300 (mg/dL)" and ranges from 0 to 720, while the y-axis is labeled "Meter (mg/dL)" and also ranges from 0 to 720. The plot shows a positive correlation between the two variables, with data points clustered around a regression line represented by the equation y = 1.019x - 3.188 and an R-squared value of 0.986.

Thigh capillary whole blood ●

Results for glucose concentration<75 mg/dL(4.2 mmol/L)

Within±5 mg/dL(Within±0.28 mmol/L)Within±10 mg/dL(Within±0.56 mmol/L)Within±15 mg/dL(Within±0.83 mmol/L)
18/21 (86%)21/21 (100%)21/21 (100%)

Results for glucose concentration ≥ 75 mg/dL (4.2 mmol/L)

Within±5%Within±10%Within±15%Within±20%
72/114 (63%)96/114 (84%)110/114 (96%)114/114 (100%)
OKmeter Direct Blood Glucose Monitoring System vs. YSI 2300 Analyzer
nSlopeInterceptR2
1351.0191-3.18860.986

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image contains a logo with the letter K inside a circular design. Next to the logo is the word "Kbiotech" in a bold, sans-serif font. Below the word "Kbiotech" is some text in a smaller font size, which appears to be in Chinese. The text is arranged in a horizontal line.

For Healthcare Professionals

The table below was based on a study done with 135 patients to see howwell the OKmeter Direct Blood Glucose Monitoring System compared to theYSI 2300.

  • Finger capillary whole blood
    Image /page/11/Figure/6 description: This image is a scatter plot comparing a healthcare professional's meter readings to YSI 2300 readings, both measuring in mg/dL. The x-axis represents YSI 2300 readings, while the y-axis represents the meter readings. A regression line is plotted through the data points, with the equation y = 1.000x - 0.603 and an R-squared value of 0.988, indicating a strong positive correlation between the two sets of readings.

Results for glucose concentration<75 mg/dL(4.2 mmol/L)

Within±5 mg/dL(Within±0.28 mmol/L)Within±10 mg/dL(Within±0.56 mmol/L)Within±15 mg/dL(Within±0.83 mmol/L)
17/21 (81%)21/21 (100%)21/21 (100%)

Results for glucose concentration ≥ 75 mg/dL (4.2 mmol/L)

Within±5%Within±10%Within±15%Within±20%
84/114 (74%)100/114 (88%)111/114 (97%)114/114 (100%)
OKmeter Direct Blood Glucose Monitoring System vs. YSI 2300 Analyzer
nSlopeInterceptR2
1351.0001-0.60330.9884

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image contains a logo with the letter K inside a circular design. Next to the logo is the word "biotech" in bold, followed by Chinese characters. The logo appears to represent a company or organization in the biotechnology field, with the Chinese characters likely representing the company's name or a related term in Chinese.

  • Palm capillary whole blood 0
    Image /page/12/Figure/4 description: The image is a scatter plot comparing a healthcare professional's measurements against YSI 2300 measurements. The x-axis represents YSI 2300 in mg/dL, ranging from 0 to 720, while the y-axis represents the meter readings in mg/dL, also ranging from 0 to 720. A regression line is plotted through the data points, with the equation y = 0.99x + 1.757 and an R-squared value of 0.985, indicating a strong positive correlation between the two measurement methods.

Results for glucose concentration<75 mg/dL(4.2 mmol/L)

Within±5 mg/dL(Within±0.28 mmol/L)Within±10 mg/dL(Within±0.56 mmol/L)Within±15 mg/dL(Within±0.83 mmol/L)
14/21 (67%)21/21 (100%)21/21 (100%)

Results for glucose concentration ≥ 75 mg/dL (4.2 mmol/L)

Within±5%Within±10%Within±15%Within±20%
77/114 (68%)99/114 (87%)110/114 (96%)114/114 (100%)
OKmeter Direct Blood Glucose Monitoring System vs. YSI 2300 Analyzer
SlopeIntercept
1350.991.75720.9859

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image contains a logo with the letter K inside a circular design. Next to the logo is the word "biotech" in bold, followed by Chinese characters. The logo appears to represent a company or organization in the biotechnology field, with the Chinese characters likely representing the company's name or a related term in Chinese.

  • Forearm capillary whole blood ●
    Image /page/13/Figure/4 description: The image is a scatter plot comparing measurements from a healthcare professional versus a YSI 2300 device. The x-axis represents the YSI 2300 measurements in mg/dL, ranging from 0 to 720. The y-axis represents the healthcare professional's measurements in mg/dL, also ranging from 0 to 720. A regression line is plotted through the data points, with the equation y = 0.996x - 0.920 and an R-squared value of 0.984, indicating a strong positive correlation between the two measurement methods.

Results for glucose concentration<75 mg/dL(4.2 mmol/L)

Within±5 mg/dL(Within±0.28 mmol/L)Within±10 mg/dL(Within±0.56 mmol/L)Within±15 mg/dL(Within±0.83 mmol/L)
14/21 (67%)20/21 (95%)21/21 (100%)

Results for glucose concentration ≥ 75 mg/dL (4.2 mmol/L)

Within±5%Within±10%Within±15%Within±20%
67/114 (59%)97/114 (85%)109/114 (96%)114/114 (100%)
OKmeter Direct Blood Glucose Monitoring System vs. YSI 2300 Analyzer
nSlopeInterceptR2
1350.9962-0.92020.9843

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image contains a logo with the letter K inside a circular design. Next to the logo is the word "biotech" in bold, followed by Chinese characters. The logo appears to represent a company or organization in the biotechnology field, with the Chinese characters likely representing the company's name or a related term in Chinese.

  • Upper arm capillary whole blood ●
    Image /page/14/Figure/4 description: The image is a scatter plot comparing a healthcare professional's meter readings against YSI 2300 readings, both measuring in mg/dL. The x-axis represents YSI 2300 readings, ranging from 0 to 720 mg/dL, while the y-axis represents the meter readings, also ranging from 0 to 720 mg/dL. A regression line is fitted to the data, with the equation y = 1.011x - 1.241 and an R-squared value of 0.980, indicating a strong positive correlation between the two sets of readings.

Results for glucose concentration<75 mg/dL(4.2 mmol/L)

Within±5 mg/dL(Within±0.28 mmol/L)Within±10 mg/dL(Within±0.56 mmol/L)Within±15 mg/dL(Within±0.83 mmol/L)
15/21 (71%)20/21 (95%)21/21 (100%)

Results for glucose concentration ≥ 75 mg/dL (4.2 mmol/L)

Within±5%Within±10%Within±15%Within±20%
69/114 (61%)100/114 (88%)109/114 (96%)114/114 (100%)
OKmeter Direct Blood Glucose Monitoring System vs. YSI 2300 Analyzer
nSlopeInterceptR2
1351.0111-1.24140.9808

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image contains a logo with the letter K inside a circular design. Next to the logo is the word "biotech" in bold, followed by Chinese characters. The logo appears to represent a company or organization in the biotechnology field, with the Chinese characters likely representing the company's name or a related term in Chinese.

  • Calf capillary whole blood ●
    Image /page/15/Figure/4 description: The image is a scatter plot comparing two different methods of measuring glucose levels: a healthcare professional's meter and a YSI 2300 analyzer. The x-axis represents the glucose levels measured by the YSI 2300 in mg/dL, while the y-axis represents the glucose levels measured by the healthcare professional's meter in mg/dL. The scatter plot shows a strong positive correlation between the two methods, with most of the data points clustered around a regression line. The regression equation is y = 0.967x + 3.901, and the R-squared value is 0.982, indicating a good fit of the regression line to the data.

Results for glucose concentration<75 mg/dL(4.2 mmol/L)

Within±5 mg/dLWithin±10 mg/dLWithin±15 mg/dL
(Within±0.28 mmol/L)(Within±0.56 mmol/L)(Within±0.83 mmol/L)
16/21 (76%)20/21 (95%)21/21 (100%)

Results for glucose concentration ≥ 75 mg/dL (4.2 mmol/L)

Within±5%Within±10%Within±15%Within±20%
67/114 (59%)99/114 (87%)109/114 (96%)114/114 (100%)
OKmeter Direct Blood Glucose Monitoring System vs. YSI 2300 Analyzer
nSlopeInterceptR2
1350.9673.90170.9823

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image contains a logo with the text "Kbiotech" in bold, stylized font. Above the text is a circular graphic element with the letter "K" inside. Below the text is a line of smaller text in a different language, possibly Chinese or Japanese. The logo appears to be for a company named Kbiotech.

  • Thigh capillary whole blood ●
    Image /page/16/Figure/4 description: The image is a scatter plot comparing two different methods of measuring glucose levels: a healthcare professional's meter and a YSI 2300 analyzer. The x-axis represents the glucose levels measured by the YSI 2300 in mg/dL, while the y-axis represents the glucose levels measured by the meter in mg/dL. The scatter plot shows a strong positive correlation between the two methods, with most of the data points clustered around a line. The equation of the line is y = 1.009x - 1.186, and the R-squared value is 0.985, indicating a very good fit.

Results for glucose concentration<75 mg/dL(4.2 mmol/L)

Within±5 mg/dLWithin±10 mg/dLWithin±15 mg/dL
(Within±0.28 mmol/L)(Within±0.56 mmol/L)(Within±0.83 mmol/L)
14/21 (67%)20/21 (95%)21/21 (100%)

Results for glucose concentration ≥ 75 mg/dL (4.2 mmol/L)

Within±5%Within±10%Within±15%Within±20%
74/114 (65%)97/114 (85%)109/114 (96%)114/114 (100%)
OKmeter Direct Blood Glucose Monitoring System vs. YSI 2300 Analyzer
nSlopeInterceptR2
1351.0095-1.18610.9856

{17}------------------------------------------------

Image /page/17/Picture/0 description: The image shows the logo for "Kbiotech". The logo consists of a stylized letter "K" with a ring around it, followed by the word "biotech" in a bold, sans-serif font. Below the word "biotech" is some Chinese text. The logo is black and white.

Comparison Table

ItemOkmeter Match Blood GlucoseMonitoring System( K090609 )Okmeter Direct Blood GlucoseMonitoring System( K132633 )
Intend UseThe Okmeter Match Blood GlucoseMonitoring System is intended for use inthe quantitative measurement of glucose infresh capillary whole blood from the fingerand the following alternative sites: thepalm, the forearm, the upper-arm, the calfand the thigh. It is intended for use byhealthcare professionals and people withdiabetes mellitus at home as an aid inmonitoring the effectiveness of diabetescontrol program. It is not intended for thediagnosis of or screening for diabetesmellitus, and is not intended for use onneonates. The alternative site testing in thissystem can be used only during steady-stateblood glucose conditions. The meter hassome audible features but it is not for useby the visually impaired.The OKmeter Direct Blood GlucoseMonitoring System is intended to be usedfor the quantitative measurement of glucose(sugar) in fresh capillary whole bloodsamples drawn from the fingertips, forearm,upper-arm, calf, thigh, or palm. TheOKmeter Direct Blood Glucose MonitoringSystem is intended to be used by a singleperson and should not be shared.The OKmeter Direct Blood GlucoseMonitoring System is intended forself-testing outside the body (in vitrodiagnostic use) by people with diabetes athome as an aid to monitor the effectivenessof diabetes control. The OKmeter DirectBlood Glucose Monitoring System shouldnot be used for the diagnosis of or screeningof diabetes or for neonatal use. Alternativesite testing should be done only duringsteady-state times (when glucose is notchanging rapidly).The OKmeter Direct Test Strips are for usewith the OKmeter Direct Blood GlucoseMeter to quantitatively measure glucose(sugar) in fresh capillary whole bloodsamples drawn from the fingertips, forearm,upper-arm, calf, thigh, or palm.
TestPrincipleElectrochemical biosensor with carbonelectrodessame
SpecimenTypecapillary whole blood, alternative sitessame
SampleVolume$0.7 \mu L$same
MeasuringTime6 secsame
CalibrationMethodControl Solutionsame
OperatingTemperature10 to 40 °Csame
StripStorageTemperature4 to 40 °Csame

Similarities:

{18}------------------------------------------------

Image /page/18/Picture/0 description: The image shows a logo with the letter 'K' inside an oval shape. Next to the logo is the word "biotech" in a stylized font. Below the logo and the word is some text in a different language, possibly Chinese or Japanese.

Differences:

ItemOkmeter Match Blood GlucoseMonitoring System( K090609 )Okmeter Direct Blood GlucoseMonitoring System( K132633 )
Test StripOKmeter voice blood glucose test strip Aand strip B(identical structure design for different color,white for strip A and Green for strip B)OKmeter Direct blood glucose test stripB
MeterDimension105(L) x 55(W) x 18(H)unit: mm95(L) x 45(W) x 18.4(H)
MeterWeight82 gwith battery:70g
DetectingRange25 ~ 580 mg/dL20 ~ 600 mg/dL
HCTRange20 ~ 55 %20 ~ 60 %
SpeakingFunctionYes(some audible features but it is not for use bythe visually impaired)No
MemoryStorage450 test results180 test results

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.