K Number
K132585
Date Cleared
2014-06-04

(292 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Zibone one-piece ceramic dental implants are indicated for implantation into the upper or lower jaw to replace missing teeth. They are indicated for (delayed or immediate) loading once primary stability has been achieved.

Device Description

Zibone ® Ceramic (Zirconia) Dental Implants are threaded; root-form dental implants intended for use in the upper and/or lower jaw to support prosthetic devices. The implants are manufactured from yttrium stabilized zirconium oxide bioceramics material conforming to ISO 13356-2008: Implants for surgery-ceramic materials based on yttria-stablized tetragonal ziconia (Y-TZP). The implant has a one-piece design (both implant and abutment), which simplifies the surgical procedures.

The implants are single use devices and are delivered in sterile condition have been sterilized with moisture heat. Zibone ceramic dental implants are also subjected to a sandblasting process that is used to impart the surface of the devices to enhance the osseointegration process.

Zibone dental implants are offered in three diameters (3.6mm, 4mm and 5mm) and 5 insertion lengths (8mm, 10mm, 11.5mm, 13mm and 14.5mm) for different teeth. The general rule is to select largest and longest implant size that the indication permits.

AI/ML Overview

The provided text is a 510(k) summary for the Zibone Ceramic Dental Implant System. It primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study with specific acceptance criteria and detailed device performance metrics in the way one might find for an AI/CADe device.

Therefore, many of the requested sections about acceptance criteria, study design, and performance metrics (especially those pertaining to AI/CADe systems like sensitivity, specificity, reader studies, etc.) are not applicable to this type of submission.

Here's a breakdown based on the information provided:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Bench Testing)Reported Device Performance
Mechanical Performance:
Fatigue Test to ISO 14801Reviewed and supports performance
Material Performance:
Surface AnalysisReviewed and supports performance
Biocompatibility TestingReviewed and supports performance
Sterilization & Shelf Life:
Sterilization ValidationReviewed and supports performance
Shelf Life TestingReviewed and supports performance
Absence of new questions of safety or effectiveness associated with identified differences compared to predicate device.Bench testing demonstrates any differences do not raise any new questions of safety or effectiveness.

Note: The document states "The following were reviewed to support the performance of Zibone Summary of 8. Ceramic Dental Implant System: Fatigue test to ISO 14801, Nonclinical Testing: Sterilization validation, Shelf life testing, Surface analysis & Biocompatibility testing." It does not provide specific numerical acceptance criteria or performance values for each of these tests, but concludes they support performance and substantial equivalence.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not applicable as no clinical studies with test sets are reported for this 510(k) submission. The testing conducted was bench testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable as no clinical studies requiring ground truth by experts are reported.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable as no clinical studies requiring adjudication are reported.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This information is not applicable as no MRMC study or AI components are described. This device is a physical dental implant, not an AI/CADe system.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable as this is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

This information is not applicable as no clinical studies requiring ground truth are reported. Performance was evaluated through non-clinical (bench) testing against established standards (e.g., ISO 14801) and material properties.

8. The sample size for the training set

This information is not applicable as no training set (for an algorithm) is mentioned or relevant to this physical device submission.

9. How the ground truth for the training set was established

This information is not applicable as no training set or ground truth in that context is mentioned or relevant.

{0}------------------------------------------------

K132585

510(k) SUMMARY

JUN 0 4 2014

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR §807.92

Date of preparation: June 4, 2014

1. Submitter's Name:Coho Technology Co., Ltd.
Address:No. 21, Dafeng St., Luju Township, Taoyan County, TAIWAN
Phone:+886-3-3112203 Ex.: 512
Fax:+886-3-3125626
Contact:Jacky Hsieh / Vice General Manager
2. Device Name :
Trade Name:Zibone Ceramic Dental Implant System
Common Name:Endosseous dental implant
Classification nameImplant, Endosseous, Root-Form
3. Device Class:Zibone Ceramic Dental Implant System has been classified as
Regulatory Class:II
Product Code:DZE
Panel :Dental
Regulation Number:21CFR 872.3640
4. Predicate Device:The predicate device is the
• Z-Systems AG - Z-Look3 Dental Implant System -(K062542) marketed byZ-Look3 dental implant system
5. Indications for Use"Zibone one-piece ceramic dental implants are indicated forimplantation into the upper or lower jaw to replace missing teeth.They are indicated for (delayed or immediate) loading onceprimary stability has been achieved."
6. Device Description:Zibone ® Ceramic (Zirconia) Dental Implants are threaded; root-formdental implants intended for use in the upper and/or lower jaw tosupport prosthetic devices. The implants are manufactured fromyttrium stabilized zirconium oxide bioceramics material

{1}------------------------------------------------

conforming to ISO 13356-2008: Implants for surgery-ceramic materials based on yttria-stablized tetragonal ziconia (Y-TZP). The implant has a one-piece design (both implant and abutment), which simplifies the surgical procedures.

The implants are single use devices and are delivered in sterile condition have been sterilized with moisture heat. Zibone ceramic dental implants are also subjected to a sandblasting process that is used to impart the surface of the devices to enhance the osseointegration process.

Zibone dental implants are offered in three diameters (3.6mm, 4mm and 5mm) and 5 insertion lengths (8mm, 10mm, 11.5mm, 13mm and 14.5mm) for different teeth. The general rule is to select largest and longest implant size that the indication permits.

ItemSubject DevicePredicate Device
510(k) NumberK062542
FDA regulation no.872.3640872.3640
Device NameZibone ceramic dental implant systemZ-Look3 dental implant system
ManufacturerCoho Tech Co., Ltd.Z-Systems AG
Intended UseTo be surgically place in the upper or lower jaw to provide support for dental prosthetics and restore chewing functions.Same
Device ConfigurationScrew TypeOne-Piece; without the need to attach an abutmentScrew TypeOne-Piece; without the need to attach an abutment
Device DimensionDiameter: 3.6, 4.0, 5.0mmInsertion length: 8, 10, 11.5mm, 13mm and 14.5mmDiameter: 3.6, 4.0, 5.0mmInsertion length: 10mm, 11.5mm, 13mm and 14mm
Material CompositionZirconium OxideZirconium Oxide
Surface TreatmentSandblastSandblast
Sterilization MethodMoisture HeatMoisture Heat

7. Substantial Equivalence:

{2}------------------------------------------------

Summary of substantial Equivalence Comparison

The Zibone Dental Implant System has the same device characteristics as the predicate device -- Z-lock3 dental implant system (K062542), intended use, material, design, surface treatment, sterilization method. Even slight differences in design characteristics and dimension do not affect the submission of the device. Therefore, we state that Zibone Ceramic Dental Implant System is substantially equivalent to predicate device.

  • The following were reviewed to support the performance of Zibone Summary of 8. Ceramic Dental Implant System: Fatigue test to ISO 14801, Nonclinical Testing: Sterilization validation, Shelf life testing, Surface analysis & Biocompatibility testing.
  • No Clinical studies are submitted. 9. Summary of Clinical Testing:
  • The Zibone Ceramic Dental Implant System has the same Conclusions: 10. indication for use , the same basic designs and materials, the same surface treatment and the same sterilization method as the Z-Look3 Dental Implant System (K062542) marketed by. Z-Systems AG. Moreover, bench testing contained in this submission demonstrates that any differences do not raise any new questions of safety or effectiveness. Thus, the Zibone Ceramic Dental Implant System is substantially equivalent to the predicate device.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-Gib09 Silver Spring, MD 20993-0002

June 4, 2014

Coho Technology Company Limited C/O Jennifer Reich Senior Consultant Harvest Consulting Corporation 2904 N. Boldt Drive Flagstaff, AZ 86001

Re: K132585

Trade/Device Name: Zibone Ceramic Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseus Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: April 24, 2014 Received: May 5, 2014

Dear Ms. Reich:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Oosinene fore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warrantics. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it n your device to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Reich

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mary's Runner -S

Erin 1. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

003_Indications for Use Statement (C)

ﺮ ﻣﺼﺮ

{6}------------------------------------------------

Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Zibone Ceramic Dental Implant System COHO TECHNOLOGY CO., LTD.

Indications for use:

Zibone one-piece ceramic dental implants are indicated for implantation into the upper or lower jaw to replace missing teeth. They are indicated for (delayed or immediate) loading once primary stability has been achieved.

V Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sheena A. Greena 2014.06.04 13:23

Page 1 of 1

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.