K Number
K132462
Date Cleared
2013-10-28

(82 days)

Product Code
Regulation Number
862.1770
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The S TEST Reagent Cartridge Blood Urea Nitrogen (BUN) is intended for the quantitative measurement of BUN in serum, lithium heparin plasma, K3 EDTA plasma, and sodium citrate plasma on the Hitachi Clinical Analyzer E40. The test system is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. BUN measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.

The S TEST Reagent Cartridge Creatinine (CRE) is intended for the quantitative measurement of creatinine in serum, lithium heparin plasma, K3 EDTA plasma, and sodium citrate plasma on the Hitachi Clinical Analyzer E40. The test system is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Device Description

The Hitachi Clinical Analyzer is an automatic, bench-top, wet chemistry system intended for use in clinical laboratories or physician office laboratories. The instrument consists of a desktop analyzer unit, an operations screen that prompts the user for operation input and displays data, a printer, and a unit cover. The analyzer unit includes a single probe, an incubation rotor, carousels for sample cups and reagent cartridges, and a multi-wavelength photometer. The single-use reagent cartridges may be placed in any configuration on the carousel, allowing the user to develop any test panel where the reagent cartridges are available.

The S TEST reagent cartridges are made of plastic and include two small reservoirs capable of holding two separate reagents (R1 and R2), separated by a reaction cell/photometric cuvette. The cartridges also include a dot code label that contains all chemistry parameters, calibration factors, and other production-related information, c.g., expiration dating, dimensions of the reagent cartridges are: 13.5 mm (W) × 28 mm (D) × 20.2 mm (H).

System operation: After the sample cup is placed into the carousel, the analyzer pipettes the sample, pipettes the reagent, and mixes (stirs) the sample and reagent together. After the sample and reagent react in the incubator bath, the analyzer measures the absorbance of the sample, and based on the absorbance of the reactions. it calculates the concentration of analyte in the sample. The test system can measure analytes in scrum or plasma and results are available in approximately 15 minutes per test. This submission is for reagent cartridge test systems for glucose.

AI/ML Overview

The provided text describes the performance of the Hitachi S TEST Reagent Cartridge Blood Urea Nitrogen (BUN) and S TEST Reagent Cartridge Creatinine (CRE) for use with the Hitachi Clinical Analyzer E40. These are in vitro diagnostic devices, not AI/ML-driven as commonly understood in medical imaging or other AI applications. Therefore, many of the requested categories (e.g., number of experts, adjudication method, MRMC study, training set ground truth) are not applicable to this type of device.

Below is a summary of the acceptance criteria and reported device performance based on the provided document, adapted for an in vitro diagnostic device regulatory submission.

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance Criteria (Implicit from Predicate/Study Design)Reported Device Performance (BUN)Reported Device Performance (CRE)
Analytical Sensitivity (LoD)Comparable to predicate or suitable for intended use0.8 mg/dL0.1 mg/dL
Linearity/Reportable RangeConsistent linear correlation across dynamic rangeLinear between 0.9 mg/dL and 110 mg/dL. Reportable range: 1.5 mg/dL to 80 mg/dL.Linear between 0.1 mg/dL and 31.3 mg/dL. Reportable range: 0.1 mg/dL to 25 mg/dL.
In-house Precision (%CV)Low %CVs indicating good reproducibilityLevel 1: 5.0% (Total), Level 2: 2.7% (Total), Level 3: 2.3% (Total)Level 1: 8.5% (Total), Level 2: 3.4% (Total), Level 3: 2.9% (Total), Level 4: 1.4% (Total)
InterferenceNo significant interference at specified levels (within 10% of neat)Hemoglobin: no interference up to 1,000 mg/dL.Unconjugated bilirubin: no interference up to 50 mg/dL.Lipemia (Intralipid®): no interference up to 1,000 mg/dL. Ascorbic acid: no interference up to 50 mg/dL.Hemoglobin: no interference up to 250 mg/dL.Unconjugated bilirubin: no interference up to 25 mg/dL.Lipemia (Intralipid®): no interference up to 1,000 mg/dL. Ascorbic acid: no interference up to 25 mg/dL.
Method Comparison (Regression)High correlation (r-value close to 1), slope close to 1, intercept close to 0 compared to a standard lab systemn=162, r=0.997, Slope=0.96 (95% CI: 0.95 to 0.97), y-intercept=-0.27 (95% CI: -0.64 to 0.10)n=100, r=0.999, Slope=0.99 (95% CI: 0.98 to 1.00), y-intercept=-0.13 (95% CI: -0.18 to -0.07)
Matrices Comparisons (Plasma vs. Serum Regression)High correlation (r-value close to 1), slope close to 1, intercept close to 0 compared to serumHeparinized: r=0.999, Slope=1.01 (-0.56 int.)EDTA: r=0.999, Slope=1.01 (-0.61 int.)Na Citrate: r=0.998, Slope=0.99 (-0.98 int.)Heparinized: r=0.999, Slope=0.99 (-0.02 int.)EDTA: r=0.999, Slope=1.01 (-0.06 int.)Na Citrate: r=0.999, Slope=1.00 (-0.05 int.)
External Site Precision (%CV)Demonstrates acceptable reproducibility in a POL settingSite 1: 0.9-1.8% Total CVSite 2: 1.2-3.6% Total CVSite 3: 0.6-2.0% Total CV (across levels A, B, C)Site 1: 2.1-6.8% Total CVSite 2: 0.0-3.8% Total CVSite 3: 4.4-6.7% Total CV (across levels A, B, C)
External Site Method Comparison (Regression)High correlation (r-value close to 1), slope close to 1, intercept close to 0 compared to a reference methodSite 1: n=75, r=0.999, $y = 0.98x - 0.23$Site 2: n=74, r=0.999, $y = 0.94x - 0.24$Site 3: n=73, r=0.999, $y = 0.95x - 0.05$Site 1: n=45, r=0.999, $y = 0.97x - 0.06$Site 2: n=46, r=0.999, $y = 0.98x - 0.09$Site 3: n=47, r=0.999, $y = 0.96x - 0.04$

2. Sample Size Used for the Test Set and Data Provenance

  • Analytical Sensitivity (LoD): No specific sample size for a test set is provided, as this is typically determined by testing multiple replicates of low-concentration samples.
  • Linearity/Reportable Range: Not specified as a distinct "test set" sample size beyond the creation of concentration series.
  • 20-day In-house Precision:
    • BUN: n=80 per level (3 levels), so 240 measurements in total.
    • CRE: n=80 per level (4 levels), so 320 measurements in total.
  • Interference Testing: Two serum pools per analyte, spiked with various concentrations of interfering substances. Specific number of total samples not explicitly stated but implies multiple measurements for each.
  • Method Comparison (Internal):
    • BUN: 162 clinical specimens.
    • CRE: 100 clinical specimens.
  • Matrices Comparisons:
    • BUN: 36 matched serum/plasma samples.
    • CRE: 39 matched serum/plasma samples.
  • External Site Precision: Each of 3 sites tested 3 blinded serum samples (low, middle, high levels) 6 times a day for 5 days. For each level, n=30 replicates per site.
  • External Site Method Comparison:
    • BUN: Approximately 75 serum specimens at each of 3 sites (total ~225).
    • CRE: Approximately 45 serum specimens at each of 3 sites (total ~135).

Data Provenance: The studies were performed "in-house" (Hitachi Chemical Diagnostics, Inc.) and at "three external POL-type sites" (Physician Office Laboratory). The clinical specimens are implied to be human serum or plasma. No country of origin is explicitly stated for the clinical data, but the company is in the USA. The studies are prospective in the sense that they were designed and executed to evaluate the device performance for this 510(k) submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This is not applicable to this type of in vitro diagnostic device. The "ground truth" for chemical analyzers is established through reference methods and calibrated samples, not expert consensus in the way a radiologist would interpret an image. The comparative methods used in the method comparison studies serve as the reference.

4. Adjudication Method for the Test Set

Not applicable. This is a quantitative chemical assay. Discrepancies are resolved through re-testing or investigation of analytical issues, not by expert adjudication of interpretations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

Not applicable. This device is an automated chemical analyzer, not an AI-assisted diagnostic imaging tool that involves human readers interpreting cases.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the device operates in a standalone (algorithm only) manner. It is an automated instrument that performs the assay and calculates results based on the detected absorbance. Humans are involved in operating the instrument, loading samples, and interpreting the numerical output, but the measurement and calculation itself is automated. The performance metrics listed (precision, linearity, method comparison, etc.) reflect this standalone analytical performance.

7. The Type of Ground Truth Used

The ground truth for evaluating the device's performance is established by:

  • Reference materials/calibrators: Used for linearity, detection limits, and precision studies.
  • Comparative methods/Standard laboratory systems: The results from the Hitachi system are compared against established, legally marketed systems (e.g., Roche Cobas c systems) which serve as the reference or "ground truth" for method comparison and accuracy studies. These are not "expert consensus, pathology, or outcomes data" in the traditional sense, but rather established analytical methods.

8. The Sample Size for the Training Set

Not applicable. This device is a wet chemistry system, not an AI/ML model that requires a training set in that context. The "training" of such a system involves chemical reagent formulation and instrument calibration, not data-driven model training.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As noted above, this is not an AI/ML model with a data-driven training set.

{0}------------------------------------------------

HITACHI
Inspire the Next

OCT 2 8 2013

510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is K132462.

807.92 (a)(1): Name: Hitachi Chemical Diagnostics Address: c/o Erika Ammirati 575 Shirlynn Court Los Altos, CA 94022

Phone: FAX: Contact:

(650) 949-2768 (650) 949-5347

Erika Ammirati, Consultant to Hitachi Chemical Diagnostics, Inc.

807.92 (a)(2): Device name- trade name and common name, and classification

Trade name:

S TEST Reagent Cartridge Blood Urea Nitrogen (BUN) S TEST Reagent Cartridge Creatinine (CRE)

Common Name: Routine chemistry analyzer for blood urea nitrogen (BUN) Routine chemistry analyzer for creatinine (CRE)

Classification: Class II, 21 CFR § 862.1770- BUN test system (BUN) Class II. 21 CFR § 862.1225- creatinine test system (CRE)

807.92 (a)(3): Identification of the legally marketed predicate devices

Cobas c systems UREAL (Roche Diagnostics, Inc., Indianapolis, IN)- K100853 Cobas c systems CREJ2 (Roche Diagnostics, Inc., Indianapolis, IN)- K100853

807.92 (a)(4): Device Description

The Hitachi Clinical Analyzer is an automatic, bench-top, wet chemistry system intended for use in clinical laboratories or physician office laboratories. The instrument consists of a desktop analyzer unit, an operations screen that prompts the user for operation input and displays data, a printer, and a unit cover. The analyzer unit includes a single probe, an incubation rotor, carousels for sample cups and reagent cartridges, and a multi-wavelength photometer. The single-use reagent cartridges may be placed in any configuration on the carousel, allowing the user to develop any test panel where the reagent cartridges are available.

The S TEST reagent cartridges are made of plastic and include two small reservoirs capable of holding two separate reagents (R1 and R2), separated by a reaction cell/photometric cuvette. The cartridges also include a dot code label that contains all chemistry parameters.

Page 1 of 8

O Hitachi Chemical Diagnostics, Inc. 630 Clyde Court, Mountain Viaw, CA 94043-2239 Jiel: 800-233-6278

.hcdiagnostics.com

{1}------------------------------------------------

calibration factors, and other production-related information, c.g., expiration dating, dimensions of the reagent cartridges are: 13.5 mm (W) × 28 mm (D) × 20.2 mm (H).

System operation: After the sample cup is placed into the carousel, the analyzer pipettes the sample, pipettes the reagent, and mixes (stirs) the sample and reagent together. After the sample and reagent react in the incubator bath, the analyzer measures the absorbance of the sample, and based on the absorbance of the reactions. it calculates the concentration of analyte in the sample. The test system can measure analytes in scrum or plasma and results are available in approximately 15 minutes per test. This submission is for reagent cartridge test systems for glucose.

Chemistry reactions (BUN):

Urea is degraded by urcase into ammonia. When glutamate dehydrogenase (GLD) reacts with ammonia and alfa-ketoglutaric acid to produce glutamic acid. NADPH is converted into NADP with a decrease of absorbance at 340 mm. The concentration of urea nitrogen can be determined by measuring the amount of change in absorbance.

Image /page/1/Figure/5 description: This image shows two chemical reactions. In the first reaction, Urea is converted to NH3 and CO2 by Urease. In the second reaction, NH3, Alfa-ketoglutaric acid, and NADPH are converted to Glutamic acid, H2O, and NADP by GLD.

Chemistry reactions (CRE):

As the first reaction, creatine in samples is decomposed into water and oxygen by the action of creatinase, sarcosine oxidase, and catalase. Subsequently, as the second reaction, creatinine in samples is converted into creatine by the action of creatininase, and then sarcosine is formed by creatinase. Then, the quinone pigment is formed by oxidation condensation between N.N-Bis (4-suffobuty)-3-methylaniline disodium salt (TODB) and 4aminoantipyrine in the presence of peroxidase (POD) and hydrogen peroxide is by sarcosine oxidase. The concentration of creatinine is determined by measuring the absorbance of the resulting quinone pigment (purple-red).

The first reaction:

Image /page/1/Figure/9 description: This image shows three chemical reactions. In the first reaction, Creatine and H2O react in the presence of Creatinase to produce Sarcosine and Urea. In the second reaction, Sarcosine, O2, and H2O react in the presence of Sarcosine oxidase to produce Glycine, HCHO, and H2O2. In the third reaction, 2H2O2 reacts in the presence of Catalase to produce O2 and 2H2O.

Page 2 of 8

@ Hitachi Chemical Diagnostics, Inc. 630 Clyde Court, Mountain View, CA 94043-2239 Tol: B00 233 6278 Fax: 650 969 2745

The first of the management of the comments of the research of the many of the many of the commend of the reasonal property of the reasonal property of the reasonal property
www.hcdiaonostics.com
ile={Melanosal (.i. ) "Invelseads, , all "informacing of the comment of the of encomment of the one of the comments of the comments of the comments of the comments of the fir

{2}------------------------------------------------

The second reaction:

Image /page/2/Figure/2 description: This image shows a series of chemical reactions. In the first reaction, Creatinine and H2O react with Creatininase to produce Creatine. In the second reaction, Creatine and H2O react with Creatinase to produce Sarcosine and Urea. In the third reaction, Sarcosine, O2, and H2O react with Sarcosine oxidase to produce Glycine, HCHO, and H2O2. In the fourth reaction, H2O2, TODB, and 4-aminoantipyrine react with POD to produce Quinone pigment (purple-red).

807.92 (a)(5): Intended Use (BUN and CRE)

The S TEST Reagent Cartridge Blood Urea Nitrogen (BUN) is intended for the quantitative measurement of BUN in serum, lithium heparin plasma, K3 EDTA plasma, and sodium citrate plasma on the Hitachi Clinical Analyzer E40. The test svstem is intended for use in clinical laboratories or physician office laboratories. For in viro diagnostic use only. BUN measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.

The S TEST reagent cartridge Creatinine (CRE) is intended for the quantitative measurement of creatinine in serum, lithium heparin plasma, K3 EDTA plasma, and sodium citrate plasma on the Hitachi Clinical Analyzer E40. The test system is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.

Page 3 of 8

O Hitachi Chemical Diagnostics, Inc.

630 Clyde Court, Mountain View, CA 94043-2239 Tel: 800 233 6278 Fax: 650 969 2745

развительно при при при при при странии при страниции в при сединания в подательно подательности политические поведниками по политические поведниками
www.hcdiaonostics.com
WE !!(1994) 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 - 1992 -

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Hitachi. The word "HITACHI" is in bold, black letters at the top of the image. Below the company name is the slogan "Inspire the Next" in a smaller font. The background is white.

807.92 (a)(6): Technological Similarities and Differences to the Predicate

CharacteristicHitachi S TEST SystemsPREDICATE(S)
Instrument PlatformHitachi Clinical AnalyzerRoche cobas 8000 - K100853
BUN Test SystemK number- pendingRoche K-number- K100853
Device Class. Regulation CodeClass II, 21 CFR 862.1770Same
Classification Product CodeCDNSame
Intended UseQuantitative measurement of BUNSame
Testing EnvironmentPhysician office or clinical labClinical lab
Test PrincipleKinetic test (UV rate) with ureaseand glutamate dehydrogenaseKinetic test with urease andglutamate dehydrogenase
Specimen TypeHuman serum or plasmaHuman serum, plasma, or urine
Reportable Range1.5 to 80.0 mg/dL1.4 to 112 mg/dL
Detection Wavelength340/546 nm (main/sub)700/340 nm (sub/main)
Detection Limit0.8 mg/dL1.4 mg/dL
Linearity0.9 to 110 mg/dL1.4 to 112 mg/dL.
Precision%CVs range from 2.3% to 5.0%%CVs range from 0.9% to 1.3%(from product labeling)
CRE Test SystemK number- pendingRoche K-number- K100853
Device Class. Regulation CodeClass II, 21 CFR 862.1225Same
Classification Product CodeCGXSame
Intended UseQuantitative measurement of CRESame
Testing EnvironmentPhysician office or clinical labClinical lab- cobas
Test PrincipleEnzymatic with creatinase andformation of quinone pigmentKinetic colorimetric assay based onJaffe method
Specimen TypeHuman serum or plasmaHuman serum, plasma, or urine
Reportable Range0.1 to 25.0 mg/dL.0.2 to 24.9 mg/dL
Detection Wavelength546/700 nm (main/sub)570/505 nm (sub/main)
Detection Limit0.1 mg/dL.0.2 mg/dL.
Linearity0.1 to 31.3 mg/dL0.2 to 24.9 mg/dL.
Precision%CVs range from 1.4% to 8.5%%CVs range from 1.1% to 5.0%(from product labeling)

The following chart describes similarities and differences between the two test systems.

807.92 (b)(1): Brief Description of Nonclinical Data

A series of studies were performed that evaluated the following nonclinical performance characteristics for glucose: analytical sensitivity (limits of detection), linearity, 20-day inhouse precision, interference testing, in-house method comparisons, and matrices comparison between serum and various plasma options.

Page 4 of 8

ത്ര Hitachi Chemical Diagnostics, Inc.

630 Clyde Court, Mountain Vlow, CA 94043-2239 Tel: 800 233 6278 Fax: 650 989 2745

A But the summer state of the capital and the support of the secondcomments of the many of the consistence and the consistence and the consistence and consistence and consistence of the consistence of the consistence of the consistence of thwww.hcdiaonostics.com
Particle Collection Comments of Children Comments on Minimal Proventure and Childer Comments of Child Childer Collection Comments of Child Childer Collection Comments of Comm

{4}------------------------------------------------

Analytical Sensitivity (Limits of Detection)

The sudy followed CLSI EP17-A2. The sensitivity/limit of detection (LoD) for BUN was calculated to be 0.8 mg/dL, and the LoD for CRE was calculated to be 0.1 mg/dL.

Linearity/Reportable Range

The study followed CLSI EP-6A. The BUN test is linear between 0.9 mg/dL and 110 mg/dL. and the CRE test is linear between 0.1 mg/dL and 31.3 mg/dL. The reportable range for BUN is 1.5 mg/dL to 80 mg/dL, and the reportable range for CRE is 0.1 mg/dL to 25 mg/dL.

20-dav In-house Precision

The studies followed CLSI EP5-A2, where three (BUN) or four (CRE) levels of samples were each tested four times a day for 20 days. The results were as follows:

Mean (mg/dL)Within-Run %CVTotal %CV
BUNLevel 17.243.05.0
n= 80 per levelLevel 214.071.62.7
Level 349.051.22.3
CRELevel 10.596.88.5
n= 80 per levelLevel 21.752.33.4
Level 36.551.22.9
Level 420.340.91.4

Precision Summary:

Interference Testing

The studies followed CLSI EP7-A2. Two serum pools per analyte were spiked with increasingly high concentrations of substances. The data demonstrated that the BUN and CRE systems were not affected by high levels of the following substances at the levels noted. Non interference was defined as spiked samples quantitating within 10% of the neat samples,

BUN (serum pools: ~ 12 mg/dL and 30 mg/dL)

  • Hemoglobin: no interference up to 1,000 mg/dL .
  • Unconjugated bilirubin: no interference up to 50 mg/dL .
  • . Lipemia (Intralipid ®): no interference up to 1,000 mg/dL
  • Ascorbic acid: no interference up to 50 mg/dL .

CRE (serum pools: ~1.5 mg/dL and 5.7 mg/dL)

  • Hemoglobin: no interference up to 250 mg/dL .
  • Unconjugated bilirubin no interference up to 25 mg/dL .
  • Lipemia (Intralipid ®): no interference up to 1,000 mg/dL .
  • Ascorbic acid: no interference up to 25 mg/dL .

Page 5 of 8

Hitachi Chemical Diagnostics, Inc.

630 Clyde Court, Mountain Vlow, CA 94043-2239 Tel: 800 233 6278

Comments of the comments of the more and the many of the results of the research and the server and one of the comments of the server of the server of the comments of the sen
www.hcdiaanostics.com
Ity-world-in-formality-in-frame-station-result-mail-an-mail-a-mail-a-mail-a-mail-a-mail-a-mail-a-mail-a-mail-a-mail-a-mail-a-mail-a-

{5}------------------------------------------------

Method Comparison - BUN

A total of 162 clinical specimens spanning the dynamic range (2.4 to 78.5 mg/dL), were assayed in singleton and in a blinded fashion by both the Hitachi E40 system (y-axis) and a standard laboratory system (x-axis). The comparative data were analyzed by linear regression and are shown below. (CI = confidence interval)

BUN Regression Statistics:

nrSlope(95% CI)y-intercept(95% CI)X meanY mean
1620.9970.96(0.95 to 0.97)-0.27(-0.64 to 0.10)27.2 mg/dL25.8 mg/dL

Method Comparison - CRE

A total of 100 clinical specimens spanning the dynamic range (0.5 to 24.7 me/dL) were assayed in singleton and in a blinded fashion by both the Hitachi E40 system (v-axis) and a standard laboratory system (x-axis). The comparative data were analyzed by lincar regression and are shown below. (CI = confidence interval)

CRE Regression Statistics:

0Stope(95% CI)y-intercept(95% Cl)X meanY mean
1000.9990.99(0.98 to 1.00)-0.13(-0.18 to -0.07)4.26 mg/dl.4.09 mg/dL

Matrices Comvarisons- BUN

A study was performed to validate the use of three plasma types as an alternative to serum for S TEST Reagent Cartridge BUN. The plasma types were lithium heparin, KS EDTA, and sodium citrate. Thirty-six (36) matched serum/plasma samples that spanned the dynamic range (2.4 to 75.3 mg/dL) were assayed in singleton and the results were compared using linear regression (plasma = y-axis, each type). The performance characteristics were as follows.

BUN: n = 36. range (serum) = 2.4 to 75.3 mg/dL

Heparinized PlasmaEDTA PlasmaNa Citrate Plasma
Slope (95% CIs)1.01 (1.00 to 1.03)1.01(1.00 to 1.03)0.99 (0.97 to 1.01)
v-intercept (95% CIs)-0.56 (-0.84 to -0.28)-0.61 (-0.94 to -0.27)-0.98 (-1.42 to -0.54)
r0.9990.9990.998

Matrices Comparisons- CRE

A study was performed to validate the use of three plasma tvpes as an alternative to serum for S TEST Reagent Cartridge CRE. The plasma types were lithium heparin, K3 EDTA, and sodium citrate. Thirty-nine (39) matched serum/plasma samples that spanned the dynamic range (0.1 to 24.5 mg/dL) were assayed in singleton and the results were compared using linear regression (plasma = y-axis, each type). The performance characteristics were as follows.

Page 6 of 8

Hitachi Chemical Diagnostics, Inc

630 Clyde Court, Mountain View, CA 94043-2239 Tol: 800 233 6278 Fax: 650 969 2745

Comments of the controlled on the controlled on the month of the month of the province of the many of the comments of the consistence in
www.hcdiaanostics.com ========================================================================================================================================================
(1) == == == == ==============================================================================================================================================================of the program and and and services and services and services and a grand a management of the specifical and any a management of the management of the management of the manag

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Hitachi, a Japanese multinational conglomerate. The logo consists of the word "HITACHI" in bold, uppercase letters. Below the company name is the slogan "Inspire the Next" in a smaller font size. The logo is simple and modern, and it is likely used in a variety of contexts, such as on the company's website, in its marketing materials, and on its products.

CRE: n = 39, range (serum) = 0.1 to 24.5 mg/dL
Heparinized PlasmaEDTA PlasmaNa Citrate Plasma
Slope (95% Cls)0.99 (0.98 to 1.00)1.01 (1.00 to 1.02)1.00 (0.99 to 1.02)
y-intercept (95% CIs)-0.02 (-0.08 to 0.05)-0.06 (-0.15 to 0.03)-0.05 (-0.14 to 0.05)
r0.9990.9990.999

807.92 (b)(2): Brief Description of Clinical Data

Studies for precision and method comparison (accuracy) were performed at three external POL-type sites to evaluate the Hitachi E40 Clinical Analyzer with S TEST Reagent Cartridges for BUN and CRE in one of its targeted intended use environments, the physician's office laboratory.

For the external site precision study. cach site received three blinded serum samples (the Precision Panel, labeled A. B. and C) that were chosen to represent low, middle, and high concentrations of BUN or CRE. Each sample was assayed six times per day for five days, reporting 30 results per level. Precision estimates for total precision were as follows:

Site #SampleMean(mg/dL)Within-run PrecisionTotal Precision
SD (mg/dL)%CVSD (mg/dL)%CV
1A12.040.131.10.221.8
2A10.920.262.30.393.6
3A11.680.242.00.232.0
1B46.940.370.80.691.5
2B45.400.481.10.651.4
3B46.400.481.00.661.4
1C75.680.710.90.690.9
2C74.290.600.80.851.2
3C75.230.410.50.470.6

BUN (mg/dL) n = 30 replicates per sample per site

n = 30 replicates per sample per site
Site #SampleMean(mg/dL)Within-run PrecisionTotal Precision
SD (mg/dL)%CVSD (mg/dL)%CV
1A0.580.035.00.046.8
2A0.600.000.00.000.0
3A0.510.036.70.036.7
1B1.800.031.80.073.8
2B1.800.031.80.073.8
3B1.610.032.00.074.4
1C6.520.050.70.132.1
2C6.380.050.80.081.2
3C5.950.071.20.315.2

CRE (mg/dL)

Page 7 of 8

O Hitachl Chemical Diagnostics, Inc.

630 Clyde Court, Mountain View, CA 94043-2239 Tel: 800 233 6278 Fax: 650 969 2745

Company of the control of the status and the first to the first to the first and theсобразования продажения продавания продавания по приведельные в принамента в глада на при выпанияthe contract and the control control and the program and the program and the comments of the comments of the comments of the comments of the comments of the contribution of t
www.hcdiaonostics.com
1 · 1479 · Pent meet meet and the manufacture and the production of a provincial in a provincial of a provincial of a program and service of a program a program a program a p

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Hitachi. The logo consists of the word "HITACHI" in bold, black letters. Below the company name is the slogan "Inspire the Next" in a smaller font, also in black. The logo is simple and modern.

For the BUN external method comparison studies, a series of approximately 75 serum specimens with BUN values ranging from 5.7 mg/dL to 73.6 mg/dL, were assayed on the Hitachi Clinical Analyzer E40 at three sites using S TEST Reagent Cartridge BUN (v) and a comparative method as the reference method (x). The same study design was used for the S TEST Reagent CRE, using approximately 45 serum specimens ranging from 0.6 mg/dL to 24.1 mg/dL CRE. Linear regression analyses (least squares) vielded the following results.

Site #nRange(mg/dL)RegressionEquation"r"CI*SlopeCI*Intercept
1756.0 to 73.6$y = 0.98x - 0.23$0.9990.96 to 0.99-0.59 to 0.13
2745.7 to 69.3$y = 0.94x - 0.24$0.9990.93 to 0.95-0.50 to 0.01
3735.7 to 70.8$y = 0.95x - 0.05$0.9990.93 to 0.96-0.41 to 0.32

POL ACCURACY DATA SUMMARY- BUN (mg/dL)

*95% Confidence Interval

SilenKange(mg/dl.)RegressionEquationﺗﺄﺳﻴﺴﺎﺕ ﺳﻨﺔ 1992 ﻓﻲ ﺇﺳﺒﺎﻧﻴﺎ، ﻭﺃﻧﺎ ﻳﻜﻮﻥ ﻓﻲ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤCl *SlopeCl*Intercept
ને રે0.6 to 23.5v = ().97x - 0.060.9990.96 to 0.98-(). 13 to 0.02
में में में में में में में में में में में में में में में में में में में में में में में में में में में में में में में में में में में में में में में में में में में म0.6 to 24.1v = 0.98x - 0.090.9990.96 to 0.99-0.19 to 0.02
470.6 to 22.8v= 0.96x - 0.040.9990.95 to 0.96-0.10 to 0.02
. A comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the contribution of the contri

POL ACCURACY DATA SUMMARY- CRE (mg/dL)

*95% Confidence Interval

807.92 (b)(3): Conclusions from Nonclinical and Clinical Testing

Nonclinical and clinical testing was performed for the S TEST Reagent Cartridge BUN and CRE. The test system was shown to be safe and effective for its intended use.

Chitachi Chemical Dlagnostics, Inc.

630 Clyde Court, Mountain Vlew, CA 94043-2239 Tol: 800 233 6278 Fax: 650 969 2745

- Le de de la mini de la mini de man de man de mana de mana de mana de mana de mana de mana de mana de mana mana mana mana mana mana mana mana mana mana mana mana mana mana mW. Ann a localified to Manhanting agallanting again the Philipping and the man has
www.hcdiagnostics.com
Comments of the comments of the comments of the program and the program and the research and one of the contraction of the consistence of contraction of consequence of conseq

{8}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/8/Picture/1 description: The image shows a logo for the Department of Health & Human Services USA. The logo features the department's seal, which includes the department's name in a circular arrangement around the perimeter. In the center of the seal is a stylized representation of three lines, possibly symbolizing people or services.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 28, 2013

Hitachi Chemical Diagnostics, Inc. c/o Erika Ammirati Ammirati Regulatory Consulting 575 Shirlynn Court LOS ALTOS CA 94022

Re: K132462

Trade/Device Name: S TEST Reagent Cartridge Blood Urea Nitrogen (BUN) S TEST Reagent Cartridge Creatinine (CRE) Regulation Number: 21 CFR 862.1770 Regulation Name: Urea nitrogen test system Regulatory Class: II Product Code: CDN, CGX Dated: August 6, 2013 Received: August 12, 2013

Dear Ms. Ammirati:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{9}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm

Sincerely yours,

Carol||G, Benson -S$_{for}$

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

INDICATIONS FOR USE

510(k) Number (if Known): K132462

Device Name:

S TEST Reagent Cartridge Blood Urea Nitrogen (BUN) S TEST Reagent Cartridge Creatinine (CRE)

Indications for Use:

The S TEST Reagent Cartridge Blood Urea Nitrogen (BUN) is intended for the quantitative measurement of BUN in serum, lithium heparin plasma, K3 EDTA plasma, and sodium citrate plasma on the Hitachi Clinical Analyzer E40. The test system is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. BUN measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.

The S TEST Reagent Cartridge Creatinine (CRE) is intended for the quantitative measurement of creatinine in serum, lithium heparin plasma, K3 EDTA plasma, and sodium citrate plasma on the Hitachi Clinical Analyzer E40. The test system is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis. and as a calculation basis for measuring other urine analytes.

Prescription Use (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use . (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE: CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Ruth A. Chesler -S

Division Sign-Off Office of In Vitro Diagnostics and Radiological Health (OIR)

510(k)_k132462

Page 1 of 1

§ 862.1770 Urea nitrogen test system.

(a)
Identification. A urea nitrogen test system is a device intended to measure urea nitrogen (an end-product of nitrogen metabolism) in whole blood, serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases.(b)
Classification. Class II.