K Number
K132454

Validate with FDA (Live)

Date Cleared
2013-10-04

(59 days)

Product Code
Regulation Number
870.5900
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VITAHEAT™ Patient Warming System is intended to prevent or treat hypothermia and to provide warmth to patients. The VITAHEAT™ Patient Warming System should be used in circumstances in which patients may not maintain a state of normothermia.

The System is intended primarily for use in hospitals and surgical centers including without limitation operating, recovery and emergency rooms and on medical/surgical floors.

Device Description

The VITAHEAT™ Patient Warming System consists of a Controller, a reusable Warming Mattress, single use disposable covers and accessories. The Controller is powered by a rechargeable 18V lithium ion battery and a separate charging unit is provided with the device. Each System also includes a second rechargeable 18V lithium ion battery to allow for uninterrupted operation. Typical battery life is 3-5 hours maximum on a fully charged battery depending on the usage. Other accessories provided with the device include a non-heating Thigh Pad that may be used by the patient for thigh support while lying on the mattress and an extension cord to connect the Controller to the Warming Mattress. The specific mechanism of action/heating is detailed further in Section 12.3.

To operate the VITAHEAT™ Patient Warming System, the Controller is connected to the reusable Warming Mattress using the extension cord provided with the device. The Controller may be placed on hospital bed rail, stretcher rail or on patient bed. The reusable Warming Mattress is then placed into a pocket on a single use disposable sheet cover also provided with the device. The patient then lies on top of the covered Warming Mattress for safe warmth. After the patient is in position on the covered Warming Mattress, it is recommended that the VITAHEAT™ Thigh Pad be placed under the patient's thighs to provide thigh support.

The VITAHEAT™ Patient Warming System is designed to be controlled by either the patient or the clinician. The output temperature range of the device is 35°C - 38.5°C ± 1.5°C: temperatures are adjusted to depending on the power setting selected by the user. From the time the device is turned ON, it takes 8-10 minutes for the Warming Mattress to reach the maximum temperature of 38.5℃ ± 1.5℃ when the highest power setting is selected. When the 'LOWER' button is pressed, the desired temperature setpoint is lowered and correspondingly when the HIGHER button is pressed, the setpoint is raised gradually. A different set of LED lights on the Controller indicate the battery life. If the temperature of the Warming Mattress exceeds a maximum safety limit of 41ºC, an automatic system cutoff initiates and the Warming Mattress is discontinued and the unit becomes inoperable.

The VITAHEAT™ device contains a controllable thermal warming component comprised of a series of connected lines printed with conductive ink on a substrate.

AI/ML Overview

Acceptance Criteria and Study for VitaHEAT™ Patient Warming System

The provided text describes a 510(k) premarket notification for the VitaHEAT™ Patient Warming System. This document focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing. It does not include information about clinical studies with human participants for effectiveness or safety as would be the case for a standalone algorithm or a multi-reader multi-case study.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in a quantified, pass/fail format for each test. Instead, it lists the types of non-clinical tests performed and implies that the device met the "predetermined specifications" for each. The reported performance is generally qualitative, stating the device is able to achieve or shuts off at certain parameters.

Acceptance Criteria (Implied)Reported Device Performance
Cleaning Effectiveness: Patient-contacting components are adequately cleaned.Cleaning method validation of patient contacting components supplied was performed. (Assumed successful based on overall conclusion of substantial equivalence).
Packaging Integrity: Packaging protects the device during transport.Drop testing, simulated shipping, and overall packaging validation were performed. (Assumed successful).
Biocompatibility: Device materials are not harmful to patients.Cytotoxicity, Irritation, and Sensitization testing were performed. (Assumed successful, aligned with ISO 10993-1).
Software Functionality: Software operates as intended.A complete testing of the software used in the device was performed. (Assumed successful).
Electromagnetic Compatibility & Electrical Safety: Device meets relevant standards.Electrical safety testing per relevant standards (IEC 60601-1, IEC 60601-2-35) was conducted, considering the battery-operated nature. (Assumed successful). The device is also considered safer due to battery operation compared to AC mains.
Temperature Performance: Achieves and maintains specified temperature range.Testing showed the device is able to achieve the temperature range provided in labeling (35°C - 38.5°C ± 1.5°C) and maintain a heating set point once steady state is reached. It takes 8-10 minutes to reach maximum temperature (38.5°C ± 1.5°C) on highest power.
High Temperature Cutoff: Device shuts off at a safe maximum temperature.Testing showed the VITAHEAT™ device shuts off at an acceptable predetermined cutoff temperature of 41°C.
Pressure & Burns Safety: Device does not cause thermal injury under pressure.Testing showed that even when pressure is applied to any portion of the device (surrogate for pressure points), the surface temperature does not exceed the threshold temperature that could cause thermal injury to patients. (Assumed successful).
General Functional, Mechanical, Dimensional Performance: Device meets specifications.General functional, mechanical, dimensional, and other performance testing against predetermined specifications was conducted. (Assumed successful).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document does not specify exact sample sizes for each test mentioned (e.g., how many units were drop-tested, how many samples for biocompatibility, etc.). It generally refers to "multiple tests" or "testing was conducted."
  • Data Provenance: The data is generated from non-clinical laboratory testing performed by the submitter (VitaHEAT Medical, LLC) or their contractors, as part of the device design, development, and validation process. The country of origin for this data is not explicitly stated but can be inferred as the USA, where the company and contact information are located. The data is prospective in the sense that it was specifically generated for this 510(k) submission to demonstrate compliance with standards and equivalence.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This section is not applicable as the described study is a series of non-clinical, engineering, and performance tests rather than a study requiring expert readers or interpretation of results against a "ground truth" established by experts (e.g., image-based diagnostic systems). The "ground truth" for these tests is defined by the relevant engineering standards (e.g., IEC 60601-1, IEC 60601-2-35) and predetermined device specifications.

4. Adjudication Method for the Test Set

This section is not applicable for the same reasons as point 3. There is no human interpretation or adjudication process described for these non-clinical performance and safety tests.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is typically conducted for diagnostic devices (e.g., AI in radiology) where human readers interpret medical images or data. The VitaHEAT™ Patient Warming System is a patient warming device, and its evaluation focuses on its physical and electrical safety and performance, not on diagnostic accuracy or human interpretation.

6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was Done

No, a "standalone" study in the artificial intelligence/algorithm sense was not done. The VitaHEAT™ Patient Warming System is a physical medical device with associated embedded software and controls, not a standalone AI algorithm. The software testing mentioned is to ensure the device's operational software functions correctly, not to evaluate an AI's performance.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

The "ground truth" for the non-clinical tests conducted was primarily based on:

  • Engineering and Safety Standards: Compliance with recognized international standards such as IEC 60601-1 and IEC 60601-2-35.
  • Predetermined Device Specifications: Internal design specifications for temperature range, cutoff temperature, material properties, etc.
  • Biocompatibility Standards: ISO 10993-1 for biological evaluation of medical devices.
  • Physical Measurements and Observations: Direct measurement of temperatures, forces, observation of packaging integrity, etc.

8. The Sample Size for the Training Set

This section is not applicable. The VitaHEAT™ Patient Warming System is not an AI/machine learning model where "training sets" are used. The device's operation is based on established engineering principles and pre-programmed controls, not on learning from data.

9. How the Ground Truth for the Training Set was Established

This section is not applicable for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text "K132454" at the top. Below the text is a black square with three curved lines inside. Underneath the square is the text "VITA HEAT" in bold letters.

6. 510(k) Summary per 807.92(a)(1)

Date Prepared: July 31, 2013

Submitter's Information

Owner/Operator: James Gaffney VitaHEAT Medical, LLC 1585 Beverly Drive #121 Aurora, Illinois 60502 Phone: 800-850-8350 Fax: 630-340-3336

OCT 04 2013

Submission Contact Information

Name:Jack Slovick
Title:Quality and Regulatory
Phone:(763) 639-0238
Email:jack_slovick@yahoo.com

Proposed Device

Common Name: Classification name: Class: Product Code: Regulatory Class:

VITAHEATTM Patient Warming System Thermal Regulating System Class II /21CFR 870.5900 DWJ Class II (Two)

Indications for Use

The VITAHEAT™ Patient Warming System is intended to prevent or treat hypothermia and to provide warmth to patients. The VITAHEAT™ Patient Warming System should be used in circumstances in which patients may not maintain a state of normothermia.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the text 'K132454' at the top. Below that is a black rectangle with three white, curved lines inside, resembling heat waves. Underneath the rectangle, the words 'VITA HEAT' are printed in a bold font.

The System is intended primarily for use in hospitals and surgical centers including without limitation operating, recovery and emergency rooms and on medical/surgical floors.

Description of Device

The VITAHEAT™ Patient Warming System consists of a Controller, a reusable Warming Mattress, single use disposable covers and accessories. The Controller is powered by a rechargeable 18V lithium ion battery and a separate charging unit is provided with the device. Each System also includes a second rechargeable 18V lithium ion battery to allow for uninterrupted operation. Typical battery life is 3-5 hours maximum on a fully charged battery depending on the usage. Other accessories provided with the device include a non-heating Thigh Pad that may be used by the patient for thigh support while lying on the mattress and an extension cord to connect the Controller to the Warming Mattress. The specific mechanism of action/heating is detailed further in Section 12.3.

To operate the VITAHEAT™ Patient Warming System, the Controller is connected to the reusable Warming Mattress using the extension cord provided with the device. The Controller may be placed on hospital bed rail, stretcher rail or on patient bed. The reusable Warming Mattress is then placed into a pocket on a single use disposable sheet cover also provided with the device. The patient then lies on top of the covered Warming Mattress for safe warmth. After the patient is in position on the covered Warming Mattress, it is recommended that the VITAHEAT™ Thigh Pad be placed under the patient's thighs to provide thigh support.

The VITAHEAT™ Patient Warming System is designed to be controlled by either the patient or the clinician. The output temperature range of the device is 35°C - 38.5°C ± 1.5°C: temperatures are adjusted to depending on the power setting selected by the user. From the time the device is turned ON, it takes 8-10 minutes for the Warming Mattress to reach the maximum temperature of 38.5℃ ± 1.5℃ when the highest power setting is selected. When the 'LOWER' button is pressed, the desired temperature setpoint is lowered and correspondingly when the HIGHER button is pressed, the setpoint is raised gradually. A different set of LED lights on the Controller indicate the battery life. If the temperature of the Warming Mattress exceeds a maximum safety limit of 41ºC, an automatic system cutoff initiates and the Warming Mattress is discontinued and the unit becomes inoperable.

Below is a picture of the complete VITAHEAT™ Patient Warming System:

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the text "K132454" at the top. Below that is a logo with three wavy lines. Under the logo, the words "VITA HEAT" are written in a bold font, with the word "medical" written below in a smaller font.

Image /page/2/Picture/1 description: The image shows a Vita Heat medical heating pad with a control unit and power supply. The heating pad is white and has the Vita Heat logo printed on it. The control unit is connected to the heating pad with a black cord. The power supply is a small, rectangular box.

The VITAHEAT™ device contains a controllable thermal warming component comprised of a series of connected lines printed with conductive ink on a substrate. The components of this device are given in the table below:

S.No.ComponentQuantitySuppliedUse
1Controller1Device used to accurately control thetemperature of the Warming Mattress peruser selected levels.
2Warming Mattress1Using the energy supplied by the Controllerto provide warmth to the patient
3Non-heating Thigh pad1To provide thigh support to the patient ashe/she lies down on the Warming Mattress
418V rechargeablelithium ion battery2Batteries are used to power the Controllerthat supplies electrical energy to theWarming Mattress. Additional battery isprovided as a backup once the originalbattery is drained after 3-5 hours ofcontinuous use.
5Battery Charging Unit1To charge the 2 rechargeable lithium ionbatteries provided with the device

Table 6A: Product Components

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text "K132454" in a simple, sans-serif font. The text is black against a white background. The letters and numbers are evenly spaced and clearly legible.

Image /page/3/Picture/1 description: The image shows a logo for a company called "Vita Heat". The logo consists of three curved lines above the company name. The lines are white and the background is black. The company name is in bold, black letters.

6Single use disposable covers24To cover the Warming Mattress when it is used to supply warmth to a patient as a layer between the Mattress and the patient
7Detachable extension cord1To supply the energy from the Controller to the Warming Mattress
8Battery Charger Power Supply1To connect the Battery Charger to A/C mains
9Instructions for Use booklet1To provide instructions for use, safety features, system precautions and warnings

Table 6B below gives the details on the device component properties

Table 6B: Component Properties

S.No.ComponentPatientContact(Yes/No)Provided Sterile orNon-SterileSingle Use orReusable
1ControllerNoNon-SterileReusable
2Warming MattressNo*Non-SterileReusable
3Non-heating Thigh padNoNon-SterileReusable
418V rechargeable lithiumion batteryNoNon-SterileReusable
5Battery Charging UnitNoNon-SterileReusable
6Single use disposablecoversYesNon-SterileSingle Use
7Detachable extension cordNoNon-SterileReusable
8Battery Charger PowerSupplyNoNon-SterileReusable
9Instructions for UsebookletNoNon-SterileReusable

*Since the disposable cover provides a layer between the patient and the Warming Mattress, the Warming Mattress does not directly come into contact with the patient; however blood, soils, etc. could leak through the cover and touch the mattress, thus, a cleaning method validation is discussed in Section 15 of this 510(k).

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the text "K132454" at the top. Below the text is a logo with three curved lines inside of a rectangle. Underneath the logo is the text "VITA HEAT medical".

Predicate Device

Hot Dog Patient Warming Mattress System by Augustine Biomedical & Design LLC, (K092807).

Summary of Technical Characteristics

The VITAHEAT™ Patient Warming System was compared to the predicate device in areas of components, packaging and compatibility with material, design, and size range. The VitaHEAT device is considered to be substantially equivalent to the predicate device Hot Dog Patient Warming Mattress System by Augustine Biomedical & Design LLC, (K092807).

Per 21 CFR Part 807.92(a)(5), the following shows where the VitaHEAT device is similar and different in terms of technological characteristics. A comparison between the new and predicate device shows that the technological characteristics and indications for use are equivalent.

Summary of Non-Clinical Testing/Statement of Equivalence

The VITAHEAT™ Patient Warming System was designed, developed, tested and validated according to written procedures. Multiple tests concerning product functionality, biocompatibility, packaging and sterilization have been performed to ensure that the VITAHEAT™ device is as safe and as effective as the predicate device. Specific testing included the following:

  • . Cleaning: A cleaning method validation of patient contacting components supplied
  • Packaging: Drop testing, simulated shipping and overall packaging validation
  • . Biocompatibility: Cytotoxicity, Irritation and Sensitization testing
  • . Software: A complete testing of the software used in the device
  • . Electromagnetic Compatibility and Electrical Safety: Keeping in mind that the VITAHEAT™ device is battery operated, electrical safety testing per relevant standards was conducted

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the text "K132454" at the top. Below the text is a black box with three white, curved lines inside. Underneath the black box is the text "VITA HEAT" in a larger, bold font.

  • . Temperature Performance Testing: Testing was conducted to show that the device is able to achieve the temperature range that is provided in the labeling and is able to maintain a heating set point once steady state is reached
  • . High Temperature Cutoff Testing: Testing to show that the VITAHEAT™ device shuts off at an acceptable predetermined cutoff temperature of 41℃
  • . Pressure & Burns Safety Testing: Testing was conducted to show that even when pressure is applied to any portion of the device, a surrogate for pressure points of the human body such as elbows, head, buttocks etc, the temperature on the surface of the . device does not exceed the threshold temperature that could cause thermal injury to patients.
  • General functional, mechanical, dimensional and other performance testing against predetermined specifications

The system is designed to meet the following performance standards:

  • . IEC 60601-1: 1998 + A1:1991 + A2:1995 Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance, edition 2.
  • . IEC 60601-2-35 Particular requirements for the safety of Disposable Covers, pad and mattresses intended for heating in medical use, edition 1.

Conclusion

The VITAHEAT™ device has the same indications for use, similar scientific fundamental technology, materials and packaging as the predicate device. The VITAHEAT™ device has very similar technological characteristics to the predicate and both have the same general shape, size and identical principles of use. Both devices contain a Patient Warming Mattress and a Temperature Controller that provides energy to the Warming Mattress and controls the heat it provides to the patient. Both devices are provided non-sterile and contain no components that are meant to be sterilized. Both devices have identical patient positioning and nearly identical device positioning requirements. Both have the same set of single use and re-usable components. Both devices have nearly identical patient contacting components and use a similar patient barrier, each of which has been tested for biocompatibility safety as per ISO 10993-1. Both devices have identical environmental conditions for storage, transportation and use.

Although there are certain differences between the VITAHEAT™ device and the predicate, these differences do not raise any new concerns in terms of safety and efficacy. For example, in terms of power source requirements, the predicate device requires an A/C mains power source whereas

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the text "K132454" at the top. Below the text is a logo with three flame-like shapes inside of a square. Underneath the logo is the text "VITA HEAT" in bold letters. The word "medical" is located below "VITA HEAT".

the VITAHEAT™ device is battery operated, making the VITAHEAT™ device safer to operate in this aspect.

The data and information provided in this submission supports a substantial equivalence determination and, therefore, 510(k) premarket notification clearance of the VITAHEATTM system. The minor technological differences such as shape, size, and compatibility have been evaluated through multiple verification and validation activities as well as critical analyses. Results of these evaluations demonstrate that these differences do not affect the ability of the VITAHEAT™ to achieve the indications for use. The VITAHEAT™ system does not raise any new questions regarding safety or effectiveness of the device, compared to the predicate Hot Dog device and is therefore substantially equivalent.

Traditional 510(k) Application VitaHEAT™ Patient Warming System VitaHEAT™ Medical, LLC Page 7 of 7

b

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle or other bird with its wings spread.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 4, 2013

VitaHEAT Medical, LLC c/o Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, MN 55313

Re: K132454

Trade/Device Name: VitaHEAT Patient Warming System Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulating System Regulatory Class: Class II Product Code: DWJ Dated: August 9, 2013 Received: August 12. 2013

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{8}------------------------------------------------

Page 2 - Mr. Mark Job

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

M.A. Willhemen

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Vita Heat. The logo consists of a square with three wavy lines inside, above the words "Vita HEAT" in a sans-serif font. Below that is another word in a smaller font, but it is illegible.

5. Indications for Use Statement

510(k) Number (if known): K132454 (To be assigned)

Device Name: VITAHEATTM Patient Warming System

Indications for Use: The VITAHEAT™ Patient Warming System is intended to prevent or treat hypothermia and to provide warmth to patients. The VITAHI:AT™ Patient Warming System should be used in circumstances in which patients may not maintain a state of normothermia.

The System is intended primarily for use in hospitals and surgical centers including without limitation operating, recovery and emergency rooms and on medical/surgical floors.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH. Office of Device Evaluation (ODE)

Page 1 of ا__ا

M.A. Willemann

Traditional 510(k) Application VITAHEAT™ Patient Warming System VITAHEAT™ Medical, LLC Page 1 of 1

§ 870.5900 Thermal regulating system.

(a)
Identification. A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.(b)
Classification. Class II (performance standards).