K Number
K132201

Validate with FDA (Live)

Date Cleared
2013-11-27

(134 days)

Product Code
Regulation Number
880.6250
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or fingers to prevent contamination between patient and examiner.(21 CFR 880.6250)

Device Description

Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Examination Vinyl Glove, 80LYZ, and meets all requirements of ASTM Standard D5250-06.

AI/ML Overview

The provided 510(k) summary is for the Synmax Synthetic Examination Vinyl Gloves, Powder Free, Blue. As such, the "device" in this context refers to these gloves. This document describes the non-clinical tests performed to demonstrate substantial equivalence to a predicate device, rather than a clinical study of an AI algorithm. Therefore, many of the requested categories (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, training set) are not applicable to this type of medical device submission.

Here's the information extracted from the document, focusing on the acceptance criteria and the study (non-clinical testing) that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally based on the ASTM-D-5250-06 standard for examination vinyl gloves, and FDA requirements for pinholes and biocompatibility. The reported device performance is compared to the predicate device, which implicitly meets these standards.

Acceptance Criteria (Standard)Reported Device Performance (Synmax Synthetic Examination Vinyl Gloves)
ASTM-D-5250-06 ComplianceConforms fully to ASTM-D-5250-06 standard
Overall Pinhole AQL 2.5 (FDA 1000 ml. Water Fill Test, Inspection Level I)Meets these requirements (AQL 2.5, Inspection Level I)
Primary Skin IrritationNo primary skin irritant reactions
Skin Sensitization (Allergic Contact Dermatitis)No sensitization reactions
Powder Content (for "powder-free" claim)Contains no more than 2 mg powder per glove
Physical Dimensions (for Medium Size, based on comparison to predicate, implied to meet standards)
Overall LengthAverage over 230mm
Width of PalmAverage 95mm
Palm ThicknessAverage 0.095 mm
Finger ThicknessAverage 0.090 mm
Material Properties (Before Aging, based on comparison to predicate, implied to meet standards)
Average Tensile Strength (Mpa)16.9
Average Ultimate Elongations (%)550%
Material Properties (After Aging, based on comparison to predicate, implied to meet standards)
Average Tensile Strength (Mpa)14.4
Average Ultimate Elongations (%)500%

2. Sample size used for the test set and the data provenance

  • Sample Size: Not explicitly stated as a single "test set" sample size. For physical and dimensions testing, the inspection levels are S-2, AQL 2.5. For the FDA 1000 ml. Water Fill Test (pinhole), the inspection level is I, AQL 2.5. These are industry standard sampling plans, but the specific number of gloves sampled is not provided in this summary.
  • Data Provenance: The tests were conducted by Zibo Intco Medical Products, Co. Ltd. (the manufacturer) in China. The data is prospective, generated specifically for this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is not an AI/diagnostic device that requires expert ground truth establishment in the traditional sense. The "ground truth" for glove performance is determined by established engineering and material science standards (ASTM, FDA water fill test).

4. Adjudication method for the test set

Not applicable. This is not an AI/diagnostic device. Performance is measured against objective standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is not an AI algorithm.

7. The type of ground truth used

The "ground truth" for the device's performance is based on:

  • Industry Standards: ASTM D5250-06 for physical properties and dimensions.
  • Regulatory Requirements: FDA 1000 ml. Water Fill Test for pinhole integrity.
  • Biocompatibility Testing: Primary Skin Irritation and Skin Sensitization tests (in vivo tests on biological systems, typically animals or human volunteers for patches, but the document doesn't specify for this case, only "with results showing no primary skin irritant or sensitization reactions").
  • Chemical Analysis: Measurement of powder content (no more than 2 mg/glove).

8. The sample size for the training set

Not applicable. This is not an AI device that requires a training set.

9. How the ground truth for the training set was established

Not applicable.

{0}------------------------------------------------

510(K) SUMMARY

This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(K) number is: K132201

    1. Submitter's Identification: Zibo Intoo Medical Products, Co. Ltd. No. 18, Qingtian Road, Linzi District Oilu Chemical Industry Park Zibo City, Shandong Province China

Contact Person

John Zhao Tel: 909-548-4828, Fax: 909-548-4808 NOV 2 7 2013

Date summary prepared: Nov. 4, 2013

    1. Name of the Device:
      Synmax Synthetic Examination Vinyl Gloves, Powder Free, Blue
    1. Common Name: Synmax Synthetic Examination Vinyl Gloves, Powder Free, Blue

4. Predicate Device Information:

Shijiazhuang Eversharp Plastic Products Co., Ltd. Synthetic Powder Free (Yellow) Vinyl Examination Gloves (K011882)

    1. Device Description:
      Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Examination Vinyl Glove, 80LYZ, and meets all requirements of ASTM Standard D5250-06.
    1. Intended Use:
      A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or fingers to prevent contamination between patient and examiner.(21 CFR 880.6250)

7. Comparison to Predicate Devices:

Zibo Intco Medical Products, Co. Ltd.'s Synmax Synthetic Examination Vinyl Gloves, Powder Free, Blue is substantially equivalent in safety and effectiveness

{1}------------------------------------------------

to Shijiazhuang Eversharp Plastic Products Co., Ltd. Synthetic Powder Free (Yellow) Vinyl Examination Gloves (K011882). Please see table 7-2 for comparison details.

8. Discussion of Non-Clinical Test Performed for Determination of Substantial Equivalence are as Follows:

The standards used for Zibo Intco Medical Products, Co. Ltd. glove production are based on ASTM-D-5250-06. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2. AOL 2.5.

The FDA 1000 ml. Water Fill Test was also conducted with samplings of AOL 2.5, Inspection Level I, meeting these requirements. Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.

There are no special labeling claims and we do not claim our gloves as hypoallergenic is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove).

9. Sterilization

There is no specific device for non-sterile examination gloves. Hand hygiene by rubbing with an alcohol-based hand rub or by washing with soap and water should be performed when appropriate.

10. Discussion of Clinical Tests Performed:

Not Applicable - There is no hypoallergenic Claim.

11. Conclusions:

Zibo Intco Medical Products, Co. Ltd. Synmax Synthetic Examination Vinyl Gloves, Powder Free, Blue, conform fully to ASTM-D-5250-06 standard as well as applicable 21 CFR references, and, meets pinhole FDA requirements. biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.

{2}------------------------------------------------

Page 3 of 4

Proposed DevicePredicate Device (K051662)
DescriptionZibo Intco Medical Products, Co. Ltd.Synmax Synthetic Examination VinylGloves, Powder Free, BlueShijiazhuang Eversharp PlasticProducts Co., Ltd. SyntheticPowder Free (Yellow) VinylExamination Gloves(K011882)
Indication for UseA patient examination glove is adisposable device intended for medicalpurposes that is worn upon theexaminer's hands or finger to preventcontamination between patient andexaminer.Substantially equivalent
Labeling: Instructionfor useA garment covering the hand and wristarea. That is a disposable device whichis worn upon the examiner's 'hands orfingers to prevent contaminationbetween patient and examiner.Substantially equivalent
Labeling: Labels onthe cartonLabels include: Product name; color:"single use only" size, piece count, lotnumber,distributor name, andmanufacturer address.Substantially equivalent
Device MaterialsPoly Vinyl ChloridePolyurethaneDiisononyl Phthalate (DINP)Substantially equivalent
Before Aging:TensileStrength(Mpa) andUltimate ElongationsAverage Tensile Strength (Mpa): 16.9Average Ultimate Elongations: 550%Substantially equivalent
After Aging: TensileStrength(Mpa) andUltimate ElongationsAverage Tensile Strength (Mpa): 14.4Average Ultimate Elongations: 500%Substantially equivalent
Overall Length onMedium SizeAverage over 230mmSubstantially equivalent
Width of Palm onMedium SizeAverage 95mmSubstantially equivalent
Palm ThicknessAverage 0.095 mmSubstantially equivalent
Figure ThicknessAverage 0.090 mmSubstantially equivalent

Table 7-2. Side-by-Side Comparison of Intended Use, Design, Material, Physical,

{3}------------------------------------------------

.

.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo of the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 27, 2013

Zibo Intco Medical Products Company, Limited C/O Mr. John Zhao Official Correspondent Basic Medical Industries, Inc. 12390 East End Avenue CHINO CA 91710

Re: K132201

Trade/Device Name: Synthetic Examination Vinyl Gloves, Powder Free, Blue Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LYZ Dated: October 28, 2013 Received: October 31, 2013

Dear Mr. Zhao:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Zhao

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Image /page/5/Picture/8 description: The image contains the text "Erin With". The text is in a bold, sans-serif font. The word "Erin" is clearly legible, while the word "With" is partially obscured by a decorative design.

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Indications for Use

Form Approvad: OMB No. 0910-0120 Expiration Date: Docember 31, 2013 See PRA Statement on last page.

610(k) Number (if known) K 132201

Davice Name

Synroax Synthetis Examination Vinyl Oloves, Powder Free, Blue

Indications for Use (Describe)

A putiat coaznination between patant and comments of the more the is worn upon the exemitor's hands of fingers
to provent countinction between pation and examiner.(21 CFR 800

Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Elizabeth F. ClaverieS
2013.11.26 01:19:15 -05'00'

FORM FDA 3881 (9/13)

Page 1 of 2

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.