K Number
K131759
Date Cleared
2013-08-08

(52 days)

Product Code
Regulation Number
888.3030
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Metallic Locking Compression Bone Plates and Screws System can be used for adult patients with age above 21 as indicated for fixation of fractures, including ulna, radius, humerus, femur and tibia.

Device Description

Metallic Locking Compression Bone Plates and Screws System contains (1) locking compression plates (LCPs) with various specifications, (2) two kinds of screws with various specifications and (3) various specific instruments. Locking Compression Plates that made of Titanium. They have combi-holes. the threaded hole sections on the plates for locking screws provides ability to create fixed-angle constructs; the un-threaded hole sections for cortex screws allows utilization of familiar AO plating techniques. The limited-contact design of LCPs reduces plate-to-bone contact, thus limiting vascular trauma.The screws are available in locking screws and cortex screws. There are various instruments specific to the proposed device intend for completing the surgery.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Metallic Locking Compression Bone Plates and Screws System, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Standard)Reported Device Performance (Test)
ASTM F382-99 (Reapproved 2008)Static and Dynamic Performance tested per ASTM F382
ASTM F543-07Torsional, Driving Torque, and Pull out strength tested per ASTM F543

2. Sample Size Used for the Test Set and Data Provenance

The provided document describes non-clinical tests performed to demonstrate that the device met design specifications and was substantially equivalent to predicate devices. It does not mention a traditional "test set" in the context of clinical data or a sample size of patients. Instead, the tests were conducted on the device components themselves.

  • Sample Size for Test Set: Not applicable in the context of clinical data. The tests were performed on the device components.
  • Data Provenance: The tests were conducted by the manufacturer, Tianjin Walkman Biomaterial Co., Ltd, located in Tianjin, P.R. China. The data would be considered "non-clinical" and laboratory-based, not patient-derived.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable as there was no clinical test set requiring expert ground truth establishment. The performance was assessed against established engineering standards (ASTM F382 and F543).

4. Adjudication Method for the Test Set

This information is not applicable as there was no clinical test set requiring adjudication. The device performance was compared directly to the specifications outlined in the ASTM standards.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study was not done. The submission explicitly states: "No clinical study is included in this submission."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

This question is not applicable as the device is a physical medical implant (bone plates and screws), not a software algorithm or AI-based diagnostic tool. Therefore, "standalone" performance in the context of an algorithm or AI is not relevant. The performance tests done were solely on the physical device components.

7. The Type of Ground Truth Used

The ground truth used for evaluating the device's performance was the specifications and requirements outlined in the relevant ASTM standards:

  • ASTM F382-99 (Reapproved 2008) for metallic bone plates.
  • ASTM F543-07 for metallic medical bone screws.

These standards define the acceptable mechanical properties and performance characteristics for such devices.

8. The Sample Size for the Training Set

This information is not applicable. As this is a submission for a physical medical device (bone plates and screws) and not a software/AI algorithm, there is no "training set" in the computational sense. The device's design and manufacturing are based on established engineering principles and materials science, not on machine learning from a dataset.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reason as point 8. There is no concept of a "training set" or "ground truth for a training set" for this type of device submission.

{0}------------------------------------------------

AUG 8 2013 K131759

Project #: M0152013Ad Section 3 510k Summary Premarket Notification 510(k) Submission

Section 3 510(k) Summary

This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.

The assigned 510(k) Number: J

  • 3.1 Date of Submission
    06/07/2013

  • 3.2 Sponsor Identification
    Tianjin Walkman Biomaterial Co., Ltd No. 19, Technology Road, Tianjin Tianyu Science and Technology Garden Jinghai East, Tianjin, P.R. China 301609

Establishment Registration Number: Not yet registered

Contact Person: Ms. Fengmei Ren Position: Management Representative Tel: +86-22-68660780 Fax: +86-22-68660776 Email: wm-rfm@126.com

  • 3.3 Submission Correspondent
    Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net

{1}------------------------------------------------

K131759

Project #: M0152013Ad Section 3 510k Summary Premarket Notification 510(k) Submission

3.4 Proposed Devices Identification

Proposed Device Name: Metallic Locking Bone Plates and Metallic Locking Bone Screws

Regulatory Information of Plates: Classification Name: Plate, Fixation, Bone Common Name: Bone Plates Class: Class II Product Code: HRS Regulation Number: 21 CFR 888.3030 Review Panel: Orthopedic

Regulatory Information of Screws: Classification Name: Screw, Fixation, Bone Common Name: Bone Screws Class: Class II Product Code: HWC Regulation Number: 21 CFR 888.3040 Review Panel: Orthopedic

Intended Use Statement:

Metallic Locking Compression Bone Plates and Screws System can be used for adult patients with age above 21 as indicated for fixation of fractures, including ulna, radius, humerus, femur and tibia.

3.5 Device Description

Metallic Locking Compression Bone Plates and Screws System contains (1) locking compression plates (LCPs) with various specifications, (2) two kinds of screws with various specifications and (3) various specific instruments. Locking Compression Plates that made of Titanium. They have combi-holes. the threaded hole sections on the plates for locking screws provides ability to create fixed-angle constructs; the un-threaded hole sections for cortex screws allows utilization of familiar AO plating techniques. The limited-contact design of LCPs reduces plate-to-bone contact, thus limiting vascular trauma.The screws are available in locking screws and cortex screws. There are various instruments specific to the proposed device intend for completing the surgery.

{2}------------------------------------------------

Premarket Notification 510(k) Submission Section 3 510k Summary

Project #: M01 52013Ad

3.6 Predicate Device Identification

Predicate Device 1

.

510(k) Number K101400

Predicate Device Name Locking Compression Plate

Manufacturer Changzhou Orthmed Medical Instrument Co., Ltd

Predicate Device 2

. 510(k) Number K100721

Predicate Device Name . Locking Bone Screw

Manufacturer Changzhou Orthmed Medical Instrument Co., Ltd

Predicate Device 3

510(k) Number K073159

Predicate Device Name Trauson Bone Screw

Manufacturer Trauson (Jiangsu) Medical Instrument Co., Ltd

{3}------------------------------------------------

3.7 Non-Clinical Test Conclusion

Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

  • ASTM F382-99 (Reapproved 2008), Standard Specification and Test Method for Metallic Bone a) Plates.
  • ASTM F543-07, Standard Specification and Test Methods for Metallic Medical Bone Screws. b)
  • 3.8 Clinical Testing Conclusion

No clinical study is included in this submission.

  • 3.9 Substantially Equivalent (SE) Conclusion
    The following table compares the DEVICE to the predicate device with respect to intended use, technological characteristics and principles of operation, etc.
ItemProposed DevicePredicate Device1Predicate Device2Predicate Device3
Product CodeHRSHWCHRS//HWC/HWC
Regulation Number21 CFR 888.303021 CFR 888.304021 CFR 888.3030//21 CFR 888.3040/21 CFR 888.3040
Intended UseMetallic LockingCompression BonePlate and ScrewSystem is intended foradult patients with ageabove 21 as indicatedfor fixation offractures, includingulna, radius, humerus,femur and tibia.Locking CompressionPlate can be used foradult patients with ageabove 21 as indicatedfor fixation offractures, includingulna, radius, humerus,femur and tibia.Locking Bone Screwis indicated for bonereconstruction,osteotomy,arthrodesis, jointfusion, fracture repair,and fracture fixationof bones appropriatefor the size of thedevice.Trauson Bone Screwis indicated for bonereconstruction,osteotomy.arthrodesis, jointfusion, fracture repair,and fracture fixationof bones appropriatefor the size of thedevice.
MaterialPlate: TitaniumCortex Screw:Titanium alloyLocking Screw:Plate: Titanium////Locking Screw:/Cortex Screw:Stainless Steel/

Table 3-1 Comparison of Technology Characteristics

{4}------------------------------------------------

Titanium alloyTitanium alloy
How suppliedNon-SterileNon-SterileNon-SterileNon-Sterile
Single UseYesYesYesYes
SterileSubject to steamsterilized prior to use.Subject to steamsterilized prior to use.Subject to steamsterilized prior to use.Subject to steamsterilized prior to use.
PerformanceStatic and DynamicPerformance testedper ASTM F382Torsional, DrivingTorque and Pull outstrength tested perASTM F543.Static and DynamicPerformance testedper ASTM F382Torsional, DrivingTorque and Pull outstrength tested perASTM F543.Torsional, DrivingTorque and Pull outstrength tested perASTM F543.

Differences in intended use, material and performance between the proposed and predicate device have been discussed and address. The proposed device is determined to be Substantially Equivalent (SE) to the predicate devices, in respect of safety and effectiveness.

{5}------------------------------------------------

Image /page/5/Picture/12 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all capital letters and is smaller than the caduceus symbol.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 8, 2013

Tianiin Walkman Biomaterial Company, Limited % Mid-Link Consulting Company, Limited Ms. Diana Hong General Manager PO Box 120-119 Shanghai, 200120, China

Re: K131759

Trade/Device Name: Metallic Locking Compression Bone Plates and Screws System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HRS, HWC Dated: July 19, 2013 Received: July 22, 2013

Dear Ms. Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{6}------------------------------------------------

Page 2 - Ms. Diana Hong

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincercly yours.

Erin I. Keith

For

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Section 2 Indications for Use

Section 2 Indications for Use

K131759 510(k) Number: Device Name: Metallic Locking Compression Bone Plates and Screws System

Indications for Use:

Mctallic Locking Compression Bone Plates and Screws System can be used for adult patients with age above 21 as indicated for fixation of fractures, including ulna, radius, humerus, femur and tibia.

�PRESCRIPTION USE (Part 21 CFR 801 Subpart D)

OR

[]over-the-counter use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page I of I

Elizabeth L. Frank -S

Division of Orthopedic Devices

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.