(100 days)
The NuVasive CoRoent Single Tab System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft to facilitate fusion.
The CoRoent Single Tab System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to L5, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The devices are to be used in patients who have had at least six months of non-operative treatment. The system is intended to be used with supplemental internal spinal fixation systems (e.g., pedicle or facet screws) that are cleared by the FDA for use in the lumbar spine in addition to the integrated screw.
The NuVasive CoRoent Single Tab System is manufactured from PEEK-Optima® LT-1 conforming to ASTM F2026, MP35N conforming to ASTM F562, and titanium alloy conforming to ASTM F136 and ISO 5832-3. The implants are available in a variety of different shapes and sizes to suit the individual pathology and anatomical conditions of the patient.
The provided document is a 510(k) summary for the NuVasive CoRoent Single Tab System, which is an intervertebral body fusion device. This type of regulatory submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving efficacy through clinical studies or setting acceptance criteria for device performance based on a new study.
Therefore, the document does not contain the information requested in your prompt regarding acceptance criteria and a study proving the device meets those criteria, as well as details about sample sizes, ground truth establishment, expert adjudication, or MRMC comparative effectiveness studies.
Here's why and what information is available:
1. A table of acceptance criteria and the reported device performance:
- Not provided. The 510(k) pathway for this device relies on demonstrating substantial equivalence through bench testing, not on establishing and meeting new clinical performance acceptance criteria.
- The document states: "The results demonstrate that the subject CoRoent Single Tab System presents no new worst-case for performance testing, and the subject device was therefore found to be substantially equivalent to the predicate."
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
- Not applicable/Not provided. No clinical test sets or patient data were used for this 510(k) submission. All testing was bench-based.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/Not provided. No human expert ground truth was established as there were no clinical studies.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable/Not provided. No clinical test set.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable/Not provided. This is a physical medical device (intervertebral body fusion system) and does not involve AI, human readers, or image interpretation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable/Not provided. This device is a physical implant, not an algorithm. The document explicitly states: "This device does not contain software or electrical equipment."
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable/Not provided. Ground truth, in the clinical sense, was not established as there were no clinical studies. For bench testing, the "ground truth" is typically the established engineering and materials standards (e.g., ASTM standards).
8. The sample size for the training set:
- Not applicable/Not provided. No training set was used as this device does not involve machine learning or AI.
9. How the ground truth for the training set was established:
- Not applicable/Not provided. No training set was used.
Summary of available "Performance Data" from the document:
The device underwent bench testing to demonstrate substantial equivalence to predicate devices. The specific tests performed were:
- Static and dynamic axial compression and compression shear per ASTM F2077.
- Wear debris testing per ASTM F2077, ASTM F1714, and ASTM F1877.
The document explicitly states: "No animal or clinical studies were conducted."
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a stylized letter N on the left, followed by the word "NUVASIVE" in a sans-serif font. Below the word "NUVASIVE" is the tagline "Speed of Innovation" in a smaller font. The logo is simple and modern, and the tagline emphasizes the company's focus on innovation.
Traditional 510(k) Premarket Notification CoRoent Single Tab System
510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR $807.92, the following summary of information is provided:
A. Submitted by:
Jeremy Markovich Senior Specialist, Regulatory Affairs NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 909-1800
SEP 20 2013
Date Prepared: September 19, 2013
B. Device Name
| Trade or Proprietary Name: | CoRoent Single Tab System |
|---|---|
| Common or Usual Name: | Intervertebral Body Fusion Device |
| Classification Name: | Spinal Intervertebral Body Fixation orthosis |
| Device Class: | Class II |
|---|---|
| Classification: | 21 CFR § 888.3080 |
| Product Code: | OVD, MAX |
C. Predicate Devices
The subject CoRoent Single TabSystem is substantially equivalent to the following predicate devices: NuVasive CoRoent System (K071795), NuVasive Sage Lateral Plate System (K121815), and Globus InterContinental Plate-Spacer (K103382).
D. Device Description
The NuVasive CoRoent Single Tab System is manufactured from PEEK-Optima® LT-1 conforming to ASTM F2026, MP35N conforming to ASTM F562, and titanium alloy conforming to ASTM F136 and ISO 5832-3. The implants are available in a variety of different shapes and sizes to suit the individual pathology and anatomical conditions of the patient.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a stylized eye-like shape on the left, followed by the word "NUVASIVE" in a simple, sans-serif font. Below the company name, there is a tagline that reads "Speed of Innovation" in a smaller font size.
Intended Use E.
The NuVasive CoRoent Single Tab System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft to facilitate fusion.
The CoRoent Single-Tab System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to L5, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The lumbar devices are to be used in patients who have had at least six months of non-operative treatment. The system is intended to be used with supplemental internal spinal fixation systems (e.g., pedicle or facet screws) that are cleared by the FDA for use in the lumbar spine in addition to the integrated screw.
F. Technological Characteristics
As was established in this submission. the subject CoRoent Single Tab System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, and function. This device does not contain software or electrical equipment.
G. Performance Data
Bench testing was performed to demonstrate that the subject CoRoent Single Tab System is substantially equivalent to other predicate devices. The following testing was performed:
- Static and dynamic axial compression and compression shear per ASTM F2077 .
- Wear debris testing per ASTM F2077, ASTM F1714 and ASTM F1877 .
The results demonstrate that the subject CoRoent Single Tab System presents no new worstcase for performance testing, and the subject device was therefore found to be substantially equivalent to the predicate. No animal or clinical studies were conducted.
H. Conclusions
Based on the indications for use, technological characteristics, and comparison to predicate devices. the subject CoRoent Single Tab System has been shown to be substantially equivalent to legally marketed predicate devices for its intended use.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure with three flowing lines above it.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002
September 20, 2013
NuVasive, Incorporated Mr. Jeremy Markovich Senior Specialist Regulatory Affairs 7475 Lusk Boulevard San Diego. California 92121
Re: K131723
Trade/Device Name: NuVasive * CoRent * Single Tab System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: OVD, MAX Dated: June 21, 2013 Received: June 24, 2013
Dear Mr. Markovich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice. Iabeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 - Mr. Jeremy Markovich
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mark N. Melkerson -S
Mark Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number (if known):__K131723
Device Name: NuVasive® CoRoent® Single Tab System
Indications For Use:
The NuVasive CoRoent Single Tab System is indicated for intervertebral body fusion of the spine in skeletally mature patients. The System is designed for use with autogenous bone graft to facilitate fusion.
The CoRoent Single Tab System is intended for use at either one level or two contiguous levels in the lumbar spine, from L2 to L5, for the treatment of degenerative disc disease (DDD) with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. The devices are to be used in patients who have had at least six months of non-operative treatment. The system is intended to be used with supplemental internal spinal fixation systems (e.g., pedicle or facet screws) that are cleared by the FDA for use in the lumbar spine in addition to the integrated screw.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anton E. Dmitriev, PhD Division of Orthopedic Devices
N/A