(87 days)
The Reprocessed ClosureFast RF Catheter is intended for endovascular coagulation of blood vessels in patients with superficial vein reflux.
The Reprocessed ClosureFast RF Catheter relies on the delivery of radiofrequency (RF) energy via an intravascular catheter that delivers temperature controlled heat to a predetermined length of superficial vessel. The ClosureFast RF Catheter is a sterile device with a 7cm radiofrequency heating element, a molded handle with start/stop switch, a flexible shaft and an integrated instrument cable, and is designed for use with the ClosureRFG (RF) Generator.
This document is a 510(k) summary for a reprocessed medical device, the Reprocessed ClosureFast Radiofrequency (RF) Catheter, and does not describe a study involving AI or machine learning.
Therefore, I cannot provide information on the following points:
- A table of acceptance criteria and the reported device performance (no AI performance metrics are discussed).
- Sample size used for the test set and data provenance.
- Number of experts used to establish ground truth and their qualifications.
- Adjudication method for the test set.
- Multi-Reader Multi-Case (MRMC) comparative effectiveness study (not applicable).
- Standalone (algorithm only) performance (not applicable).
- Type of ground truth used.
- Sample size for the training set.
- How the ground truth for the training set was established.
However, I can extract information related to the device's performance data and acceptance criteria as described in this 510(k) summary:
1. Acceptance Criteria and Device Performance:
The document states that the Reprocessed ClosureFast RF Catheter was tested to demonstrate appropriate functional characteristics and that process validation testing was performed to validate the cleaning process in line with AAMI TIR 30:2011 requirements. The manufacturing process includes visual and validated functional testing of all products.
While specific numerical acceptance criteria (e.g., minimum tensile strength, maximum leakage current, etc.) for functional characteristics are not detailed in this summary, the implicit acceptance criterion is that the reprocessed device performs identically and safely to the new predicate device.
| Acceptance Criterion (Implicit) | Reported Device Performance |
|---|---|
| Appropriate functional characteristics (identical to predicate device in design and operation) | "Representative samples of Reprocessed ClosureFast RF Catheters were tested to demonstrate appropriate functional characteristics through bench top verification testing." The document also states, "The Reprocessed ClosureFast RF Catheters are identical to the predicate device in design, materials of construction, and intended use. There are no changes to the clinical applications, patient population, performance specifications, or method of operation." Additionally, "The manufacturing process includes...validated functional testing of all products produced prior to release." |
| Validated cleaning process | "Process validation testing was performed to validate the cleaning process in line with AAMI TIR 30:2011 requirements." |
| Visual integrity | "The manufacturing process includes visual...testing of all products produced prior to release." |
| Substantially Equivalent to Predicate Device | "Covidien concludes that the Reprocessed ClosureFast RF Catheter is safe, effective, and substantially equivalent to the predicate device, ClosureFast RF Catheters K061373..." |
2. Study Description:
The summary refers to bench top verification testing and process validation testing as the studies conducted. These are not clinical trials, but rather engineering and quality control studies.
- Bench Top Verification Testing: This type of testing involves evaluating the device's functional performance in a laboratory setting. It would assess physical and electrical properties relevant to the device's intended use.
- Process Validation Testing: Specifically mentions validating the cleaning process in line with AAMI TIR 30:2011 requirements. AAMI TIR 30:2011 provides guidance on processing reusable medical devices in health care settings.
Summary of what's provided in the document regarding the "study":
- Study Goal: Demonstrate appropriate functional characteristics and validate the cleaning process to ensure the reprocessed device is safe, effective, and substantially equivalent to the predicate device.
- Type of Study: Bench top verification testing and process validation testing.
- "Test Set" / Samples Used: "Representative samples of Reprocessed ClosureFast RF Catheters." (Specific number of samples not provided).
- Data Provenance: Not explicitly stated, but assumed to be internal laboratory testing by the manufacturer (Covidien) in Mansfield, MA, USA. This is an internal, prospective testing of reprocessed devices.
- Ground Truth: For the "functional characteristics," the ground truth is likely the performance specifications of the original predicate device (ClosureFast RF Catheter K061373). For the cleaning process, the ground truth is adherence to the standards outlined in AAMI TIR 30:2011.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains a logo and the word "COVIDIEN". The logo is a white cross on a black background. The word "COVIDIEN" is written in black, sans-serif font. The logo is on the left side of the image, and the word "COVIDIEN" is on the right side of the image.
K131614 510(K) SUMMARY
1. 510(k) Owner:
Applicant:
Telephone: Fax:
Covidien 15 Hampshire Street Mansfield, MA 02048 (508) 452 - 4135 (508) 452 - 4135
AUG 2 9 2013
Contact: Title: Date Prepared: Daniel Campion Associate Director Regulatory Affairs May 31, 2013
2. Device:
Trade Name: Classification Name: Regulation Number: Product Code(s): Classification:
Reprocessed ClosureFast Radiofrequency (RF) Catheter Electrosurgical Cutting & Coagulation, Reprocessed 21 CFR 878.4400 nui Class II
3. Predicate Device:
The Reprocessed ClosureFast RF Catheter is substantially equivalent in intended use and operation to the predicate ClosureFast RF Catheter (K061373).
4. Device Description:
The Reprocessed ClosureFast RF Catheter relies on the delivery of radiofrequency (RF) energy via an intravascular catheter that delivers temperature controlled heat to a predetermined length of superficial vessel. The ClosureFast RF Catheter is a sterile device with a 7cm radiofrequency heating element, a molded handle with start/stop switch, a flexible shaft and an integrated instrument cable, and is designed for use with the ClosureRFG (RF) Generator.
5. Intended Use:
The Reprocessed ClosureFast RF Catheters are intended for endovascular coagulation of blood vessels in patients with superficial vein reflux .
6. Technological Characteristics:
The Reprocessed ClosureFast RF Catheters are identical to the predicate device in design, materials of construction, and intended use. There are no changes to the clinical applications, patient population, performance specifications, or method of operation.
{1}------------------------------------------------
7. Performance Data:
1
Representative samples of Reprocessed ClosureFast RF Catheters were tested to demonstrate appropriate functional characteristics through bench top verification testing and process validation testing was performed to validate the cleaning process in line with AAMI TIR 30:2011 requirements. The manufacturing process includes visual and validated functional testing of all products produced prior to release.
8. Conclusion:
Covidien concludes that the Reprocessed ClosureFast RF Catheter is safe, effective, and substantially equivalent to the predicate device, ClosureFast RF Catheters K061373, as described in this premarket notification submission.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles an eagle or bird-like figure with three curved lines representing its wings or body.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Covidien Daniel Campion Associate Director Regulatory Affairs 15 Hampshire Street Mansfield, Massachusetts 02048
August 29, 2013
Re: K131614
Trade/Device Name: Reprocessed ClosureFast Radiofrequency (RF) Catheter Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: NUJ Dated: July 26, 2013 Reccived: July 31, 2013
Dear Mr. Campion:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRI-1 does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class 11 (Special Controls) or class 111 (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{3}------------------------------------------------
Page 2 - Daniel Campion
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOfTices/CDRH/CDRHOffices/ucm 115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
FOR
Sincerely yours,
Peter D. Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known): K131614
Device Name: Reprocessed ClosureFast RF Catheter
Indications for Use:
The Reprocessed ClosureFast RF Catheter is intended for endovascular coagulation of blood vessels in patients with superficial vein reflux.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Joshua C. Nipper -S
Page 1 of 1
§ 878.4400 Electrosurgical cutting and coagulation device and accessories.
(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.